14.08.2013 Views

pH Transmitter MV 4010 - Sensortechnik Meinsberg

pH Transmitter MV 4010 - Sensortechnik Meinsberg

pH Transmitter MV 4010 - Sensortechnik Meinsberg

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>pH</strong> <strong>Transmitter</strong> <strong>MV</strong> <strong>4010</strong><br />

page 1/2 02/12<br />

99<br />

22,5<br />

Dimensions: mm<br />

Specifications<br />

<strong>Sensortechnik</strong> <strong>Meinsberg</strong> GmbH<br />

Quality System certified to DIN EN ISO 9001<br />

Kurt-Schwabe-Straße 6<br />

D-04720 Ziegra-Knobelsdorf/Germany<br />

114<br />

Internet: www.meinsberg.de<br />

Tel.: +49 (0) 34327 623-0<br />

Fax: +49 (0) 34327 623-79<br />

The <strong>pH</strong> <strong>Transmitter</strong> <strong>MV</strong> <strong>4010</strong> realises<br />

reliable processing of <strong>pH</strong> and temperature<br />

in connection with a <strong>pH</strong> combination<br />

electrode (with integrated temperature<br />

sensor). A small display (OLED) for<br />

measuring value and calibration instructions<br />

facilitates control and service. The<br />

transmitter features programmable calibration<br />

functions, automatic temperature<br />

compensation, easy installation and programming<br />

by means of a PC interface.<br />

Input signal terminals:<br />

A1 Guard A5 Pt 1000 (1)<br />

A2 Measuring electrode<br />

A6 Pt 1000 (1)<br />

A3 Guard A7 Pt 1000 (2)<br />

A4 Reference<br />

electrode<br />

A8 Pt 1000 (2)<br />

Power supply and output signal terminals:<br />

B1 Power supply AC/DC B5 output 1 referred to GND<br />

B2 Power supply GND B6 output 0 referred to GND<br />

B3 Power supply GND B7 Relay output closing contact<br />

B4 Power supply AC/DC B8 Relay output closing contact<br />

Measuring ranges <strong>pH</strong> -2.00 … 16.00; -2000 … 2000 mV; -10 …130 °C<br />

Resolution (Display) 0.01 <strong>pH</strong>; 1 mV; 0.1 °C<br />

Configuration with PC configuration software program "DinModule"<br />

Scaling and determining of the analogue output signals<br />

Calibration menu, determination calibration points (two-points calibration)<br />

Temperature compensation<br />

Buttons functions MODE: Changing into calibration mode<br />

C1, C2: Calibration values (pre-adjustment C1: buffer solution <strong>pH</strong> 6.87 (25<br />

°C); C2: buffer solution <strong>pH</strong> 4,01 (25 °C))<br />

Indication multi-coloured LED (stable value signalisation)<br />

LED yellow: Calibration LED green: Measuring LED red: Calibration error<br />

Display graphic OLED, 48 x 64 pixel, back-lighted<br />

display for 3 values (value 1, value 2 and temperature value) and clear<br />

calibration instructions<br />

Output signals 2 x 0(4) … 20 mA or 0 … 5 V, isolated, resolution 12 bit<br />

RS-232 (protocol on request)<br />

Relay output closing contact max. 125 V AC, 60 V DC; 30 VA


<strong>pH</strong> <strong>Transmitter</strong> <strong>MV</strong> <strong>4010</strong><br />

page 2/2 02/12<br />

Power supply 15 ... 24 V AC/DC, about 1.5 VA<br />

Ambient temperature 0 ... 50 °C<br />

EMC acc. EN 61326 class B<br />

Enclosure plastic case for mounting on rail tracks DIN EN 50022-35<br />

protection degree IP 40<br />

Dimensions 22.5 x 99 x 114.5 mm (W x H x D)<br />

Electrodes / Sensors <strong>pH</strong> combination electrode; <strong>pH</strong> combination electrode with integrated temperature;<br />

separated measuring and reference electrode; temperature sensor<br />

Pt 1000 (special version for Pt 100)<br />

Electrical connections mountable screw-terminals for wires cross section 0.2 … 2.5 mm²;<br />

3-pins socket for RS-232 cable connector; optional rail track bus-connector<br />

Accessories<br />

Interface <strong>MV</strong> USB Interface cable 1.8 m for USB connection (3 pin stereo jack / USB connector)<br />

Interface <strong>MV</strong> Interface cable 1.8 m for RS-232 interface (3 pin stereo jack / D-SUB9 connector)<br />

DinModule PC-configuration software program "DinModule"; CD-ROM; free download<br />

from http://download.meinsberg.de/<br />

<strong>Sensortechnik</strong> <strong>Meinsberg</strong> GmbH<br />

Quality System certified to DIN EN ISO 9001<br />

Kurt-Schwabe-Straße 6<br />

D-04720 Ziegra-Knobelsdorf/Germany<br />

Internet: www.meinsberg.de<br />

Tel.: +49 (0) 34327 623-0<br />

Fax: +49 (0) 34327 623-79