Chapter Four: Operation<br />

U How To Optimize Long Cable Operation<br />

How To Optimize Long Cable Operation<br />

When long transducer cables are used in a system, frequently it is necessary to adjust the system<br />

gain to optimize system performance which might be degraded due to the cable length. When the<br />

simulator is installed on the end of the transducer cable, the system gain is adjusted by using the<br />

procedures described in How To Perform a System Check, <strong>and</strong> How To Recalibrate the<br />

Electronics Unit, page 36. Note that on occasion, the Electronics Unit span adjust may have to be<br />

adjusted if the full scale adjustment is insufficient to set full scale.<br />