<strong>Chlorine</strong> <strong>Controller</strong> <strong>MV</strong> <strong>5060</strong> (<strong>CAN</strong>), <strong>MV</strong> <strong>5060</strong>-F<br />
page 1/3 02/12<br />
<strong>Sensortechnik</strong> <strong>Meinsberg</strong> GmbH<br />
Quality System certified to DIN EN ISO 9001<br />
Kurt-Schwabe-Straße 6<br />
D-04720 Ziegra-Knobelsdorf/Germany<br />
<strong>MV</strong> 5000<br />
CAL CONF LOG INFO<br />
SENSORTECHNIK MEINSBERG GMBH<br />
<strong>MV</strong> <strong>5060</strong>, <strong>MV</strong> <strong>5060</strong> <strong>CAN</strong> - wall mounting enclosure<br />
Dimensions: mm<br />
<strong>MV</strong> 5000-F<br />
CAL CONF LOG INFO<br />
<strong>MV</strong> <strong>5060</strong>-F – front panel mounting enclosure<br />
Cut out acc. DIN 43700: 138 x 138<br />
Dimensions: mm<br />
The controller series <strong>MV</strong> <strong>5060</strong> features reliable processing of chlorine, chlorine dioxide, ozone and temperature.<br />
User-friendly installation & operation as well as high function especially controller functions, data logger<br />
and field bus networking allow universal application.<br />
Internet: www.meinsberg.de<br />
Tel.: +49 (0) 34327 623-0<br />
Fax: +49 (0) 34327 623-79
<strong>Chlorine</strong> <strong>Controller</strong> <strong>MV</strong> <strong>5060</strong> (<strong>CAN</strong>), <strong>MV</strong> <strong>5060</strong>-F<br />
page 2/3 02/12<br />
Specifications<br />
Measuring ranges / 0.00 ... 10.00 mg/l (free adjustable); -10 ... 130 °C / 0.1 °C<br />
Resolution (display)<br />
Configuration directly by 5 keys and display (plain text menu structure) or<br />
with PC-configuration software program "DinModule"<br />
Display graphic OLED, 128 x 64 pixel, self-luminous<br />
Output signals 2 x 0(4) … 20 mA or 0 … 5 V, isolated, resolution 12 bit<br />
RS-232 (protocol on request)<br />
Current output signal load ≤ 500 Ω, accuracy ≤ 0.2%<br />
Voltage output signal input resistance ≥ 2 kΩ, accuracy ≤ 0.2%<br />
<strong>CAN</strong> Bus networking via <strong>CAN</strong> bus up to 16 devices (slaves) with KM3000 (master)<br />
(only <strong>MV</strong>50XX<strong>CAN</strong>)<br />
<strong>Controller</strong> function bidirectional PID controller with pulse length (PWM)- or analogue<br />
(not <strong>MV</strong>50XX<strong>CAN</strong>) controller<br />
Relay output 2 x changeover contact max. 250 V AC, 5A<br />
Power supply 100 … 240 V AC, about 4 VA or 18 … 36 V DC, about 1.5 VA<br />
Ambient temperature -10 … 55 °C<br />
EMC acc. EN 61326 class B<br />
Enclosure <strong>MV</strong>50XX(<strong>CAN</strong>): aluminium case for wall mounting, IP 65 (NEMA 4X)<br />
<strong>MV</strong>50XX-F: Noryl (plastic) enclosure for front panel mounting;<br />
front side: IP 63, with transparent front cover: IP 65, rear side: IP 20<br />
Dimensions <strong>MV</strong>50XX(<strong>CAN</strong>): 160 x 130 x 70 mm (W x H x D)<br />
<strong>MV</strong>50XX-F: 144 x 144 x 70 mm (W x H x D)<br />
Electrodes / Sensors sensors with 2-wire output especially amperometric sensors for measurement<br />
of disinfectants (chlorine, chlorine dioxide, ozone); temperature sensor<br />
Pt 1000<br />
Electrical connections mountable screw-terminals for wires cross section 0.2 … 2.5 mm²;<br />
3-pins socket for RS-232 cable connector; optional rail track bus-connector<br />
<strong>Sensortechnik</strong> <strong>Meinsberg</strong> GmbH<br />
Quality System certified to DIN EN ISO 9001<br />
Kurt-Schwabe-Straße 6<br />
D-04720 Ziegra-Knobelsdorf/Germany<br />
Internet: www.meinsberg.de<br />
Tel.: +49 (0) 34327 623-0<br />
Fax: +49 (0) 34327 623-79
<strong>Chlorine</strong> <strong>Controller</strong> <strong>MV</strong> <strong>5060</strong> (<strong>CAN</strong>), <strong>MV</strong> <strong>5060</strong>-F<br />
page 3/3 02/12<br />
Accessories<br />
Interface <strong>MV</strong> USB Interface cable 1.8 m for USB connection (3 pin stereo jack / USB connector)<br />
Interface <strong>MV</strong> Interface cable 1.8 m for RS-232 interface (3 pin stereo jack / D-SUB9 connector)<br />
Adapter <strong>MV</strong> 5000 Adapter for Interface <strong>MV</strong> 5000 USB for internal installation (only for<br />
<strong>MV</strong>50XX)<br />
Interface <strong>MV</strong> 5000 USB Interface cable 1.8 m for USB connection (Adapter <strong>MV</strong> 5000 is required, only<br />
for <strong>MV</strong>50XX)<br />
Front cover <strong>MV</strong> 5000-F transparent cover for front side, gum, IP 65 (only for <strong>MV</strong>50XX-F)<br />
DinModule PC-configuration software program "DinModule"; CD-ROM; free download<br />
from http://download.meinsberg.de/<br />
<strong>Sensortechnik</strong> <strong>Meinsberg</strong> GmbH<br />
Quality System certified to DIN EN ISO 9001<br />
Kurt-Schwabe-Straße 6<br />
D-04720 Ziegra-Knobelsdorf/Germany<br />
Internet: www.meinsberg.de<br />
Tel.: +49 (0) 34327 623-0<br />
Fax: +49 (0) 34327 623-79