<strong>Corning</strong> ® <strong>HYPERStack</strong> Cell Culture Vessel<br />

Closed System Protocol Guide<br />

<strong>HYPERStack</strong> Vessel Set-Up<br />

G<br />

H<br />

J<br />

B<br />

F<br />

E<br />

D<br />

C<br />

B<br />

A<br />

A <strong>HYPERStack</strong> Vessel manifolds<br />

B Liquid handling tubing<br />

C Liquid handling tubing clamp<br />

D Y connector with sampling port<br />

E Chase tube clamp<br />

F Chase tubing<br />

G 0.2 µm chase tube filter<br />

H 0.2 µm air filter<br />

J Female MPC with plug<br />

A<br />

<strong>Corning</strong> <strong>HYPERStack</strong> vessels can be connected to tubing via tube welding<br />

or by using the pre-assembled, multi-purpose connects (MPCs).<br />

Seeding the <strong>HYPERStack</strong> Vessel<br />

1 2 3<br />

Tube weld serum and media bags, mix thoroughly.<br />

1: Clamp off ~300 mL of media in the top of the bag. Fill syringe with cell inoculum.<br />

Inject into bottom portion of media bag.<br />

Attach inoculated media bag to <strong>HYPERStack</strong> vessel’s 3/8" I.D. liquid handling tubing.<br />

2: Close liquid handling tubing clamp.<br />

3: Close chase tubing clamp.


Filling the <strong>HYPERStack</strong> Vessel<br />

1 2 3 4 5<br />

1: Place <strong>HYPERStack</strong> vessel into fill position at a 10° angle and position media bag at<br />

same level as the <strong>HYPERStack</strong> vessel.<br />

2: Open liquid handling tubing clamp.<br />

3: Open media bag tubing clamp. Fluid will enter the <strong>HYPERStack</strong> vessel.<br />

4: Raise media bag to increase flow into the <strong>HYPERStack</strong> vessel, emptying the inoculated<br />

portion of the media bag.<br />

5: Remove clamp on top of media bag; finish filling <strong>HYPERStack</strong> vessel to the fill line.<br />

To reduce rate of flow and prevent overfilling, lower media bag during final fill.<br />

Isolate Liquid into <strong>HYPERStack</strong> Layers<br />

1 2 3<br />

4<br />

✕<br />

5 6 7<br />

1: Place <strong>HYPERStack</strong> vessel into “isolation” position and lower media bag below<br />

the height of the <strong>HYPERStack</strong> vessel.<br />

2: Hold the chase filter upright and open the chase tube clamp. Allow excess media<br />

to empty back into the media bag.<br />

3: Close the media bag tubing clamp.<br />

4: Open the liquid handling tubing clamp.<br />

5: Return <strong>HYPERStack</strong> vessel to starting position.<br />

6: Close chase tube clamp.<br />

7: Close liquid handling tube clamp.<br />

Place <strong>HYPERStack</strong> vessel into incubator. (Option: Empty media bag can be<br />

removed or remain attached to the <strong>HYPERStack</strong> vessel.)<br />

Keep the vent filter upright to prevent wetting out the filter.


Empty Spent Media from <strong>HYPERStack</strong> Vessel<br />

1 2 3 4<br />

5 6 7 8<br />

1: Remove <strong>HYPERStack</strong> vessel from incubator; keep vent tubing upright.<br />

2: Place <strong>HYPERStack</strong> vessel in “fill” position at a 10° angle.<br />

Lower media bag below height of <strong>HYPERStack</strong> vessel.<br />

3: Open the media bag tubing clamp.<br />

4: Open the vessel’s liquid handling tubing clamp. Fluid will enter empty media bag.<br />

5: Place <strong>HYPERStack</strong> vessel in final empty position and let all media drain.<br />

6: Lift tubing to chase remaining fluid into the media bag.<br />

7: Close liquid handling tubing clamp.<br />

8: Close media bag tubing clamp. Remove spent media bag.<br />

Keep the vent filter upright to prevent wetting out the filter.<br />

<strong>Corning</strong> <strong>HYPERStack</strong> Cell Culture<br />

Vessel Product Information<br />

◗ Growth area per layer. . . . . 500 cm 2<br />

◗ Media per layer. . . . . . . . . . . . 100 mL<br />

◗ Media per cm 2 . . . . . . . . . . . . . 0.2 mL<br />

◗ Stackette internal height . . 2.0 mm<br />

◗ Triple bagged<br />

◗ <strong>Corning</strong> CellBIND® Surface treated<br />

◗ Sterile<br />

3 Sizes Available:<br />

HyperStack-12 Layer . . . 6,000 cm 2<br />

HyperStack-36 Layer. . 18,000 cm 2<br />

HyperStack-120 Layer 60,000 cm 2<br />

Suggested Reagent Volumes:<br />

Cell dissociation solution:<br />

HyperStack-12 Layer . . 200 mL<br />

HyperStack-36 Layer . . . 600 mL<br />

Complete media:<br />

HyperStack-12 Layer. . . . . . 1.3 L<br />

HyperStack-36 Layer. . . . . . 3.9 L<br />

Cell Harvest from <strong>HYPERStack</strong> Vessel<br />

1 2 3 4 5 6<br />

1: Connect cell dissociation and quench bags together to form harvest bag assembly.<br />

2: Attach to <strong>HYPERStack</strong> vessel.<br />

3: Raise harvest bag assembly above height of <strong>HYPERStack</strong> vessel and place it into<br />

“fill” position at a 10° angle.<br />

4: Open cell dissociation solution tube clamp.<br />

5: Open harvest bag assembly tube clamp. Fluid will enter the <strong>HYPERStack</strong> vessel.<br />

6: Close liquid handling tube clamp.<br />

(continued on back)


Cell Harvest from <strong>HYPERStack</strong> Vessel (continued)<br />

1 2 3 4 5 6<br />

1: Close cell dissociation tube clamp.<br />

2: Return <strong>HYPERStack</strong> vessel to the starting position; rock the vessel to distribute cell<br />

dissociation solution.<br />

3: After incubating the <strong>HYPERStack</strong> vessel for the required time for cell release, place<br />

<strong>HYPERStack</strong> vessel in “fill” position at a 10° angle.<br />

4: Open quench bag tube clamp.<br />

5: Open liquid handling tube clamp. Quench solution will enter the <strong>HYPERStack</strong> vessel.<br />

6: Close liquid handling tube clamp. Distribute the quench solution to completely<br />

deactivate the cell dissociation solution.<br />

1 2 3 4 5<br />

1: Place harvest bag assembly lower than <strong>HYPERStack</strong> vessel.<br />

2: Place the vessel in the empty position.<br />

3: Open liquid handling tube clamp.<br />

4: Transfer cell solution into empty quench bag.<br />

5: Disconnect harvest bag assembly from <strong>HYPERStack</strong> vessel. Cell suspension is<br />

ready for processing.<br />

Ordering Information<br />

For replacement filters, contact Sartorius Stedim North America Inc.<br />

Tel.: +1.631.254.4249 or E-mail: techsupport@sartorius-stedim.com.<br />

Cat. No. Description Qty/Pack<br />

17805----------BVN Nonsterile, Midisart BV, individually packed 100<br />

17805----------BVQ Nonsterile, Midisart BV, individually packed 500<br />

17805----------E (NPE) Sterile, Midisart 2000, individually packed 12<br />

17805----------G (NPG) Sterile, Midisart 2000, individually packed 25<br />

For additional product or technical information, please visit<br />

www.corning.com/lifesciences or call 1.800.492.1110.<br />

Outside the United States, please call 1.978.442.2200.<br />

<strong>Corning</strong> is a registered trademark of <strong>Corning</strong> <strong>Incorporated</strong>, <strong>Corning</strong>, NY.<br />

<strong>HYPERStack</strong> is a trademark of <strong>Corning</strong> <strong>Incorporated</strong>, <strong>Corning</strong>, NY.<br />

<strong>Corning</strong> <strong>Incorporated</strong>, One Riverfront Plaza, <strong>Corning</strong>, NY 14831-0001<br />

<strong>Corning</strong> <strong>Incorporated</strong><br />

Life Sciences<br />

Tower 2, 4th Floor<br />

900 Chelmsford St.<br />

Lowell, MA 01851<br />

t 800.492.1110<br />

t 978.442.2200<br />

f 978.442.2476<br />

www.corning.com/lifesciences<br />

© 2011 <strong>Corning</strong> <strong>Incorporated</strong> Printed in U.S.A. 4/11 POD CLS-AN-171 REV3