12.1 Specialist Patient Equipment<br />

In the clean decontamination room the storage trolley for the<br />

endoscopes was dusty. In the dirty decontamination room, the second<br />

sink used in the double sink manual cleaning process of the scopes did<br />

not have a fill line.<br />

In relation to effective cleaning, the HEIG inspectors noted that in the<br />

manual wash area the unit dispensing detergent was faulty and an<br />

incorrect ratio of detergent to water was being used. Detergents<br />

should be diluted as per the manufactures guidelines to be effective.<br />

12.2 Policies Procedures and Training<br />

This is a new open facility and while some policies, for example the<br />

decontamination of endoscopy equipment and stand operational<br />

procedures are in place. Others such as, procedures for regular audits<br />

and the hospitals own infection prevention and control policy have still<br />

to receive approval. Staff currently use the Regional Infection<br />

Prevention and Control guidance on line. A training and validation<br />

programme has just been put in place.<br />

Recommendations<br />

19. The environment of the endoscopy suite should be clean and<br />

free of clutter.<br />

<strong>20</strong>. The decontamination room should be fit for purpose and<br />

comply with relevant standards.<br />

21. Improvements required in the decontamination process<br />

should be addressed and work commenced to comply with<br />

CFPP.<br />

22. All recommendations as outlined by the HEIG inspectors in<br />

their key findings report should be addressed.<br />