<strong>Model</strong> <strong>2010</strong> Specifications<br />
______________________________________________________<br />
<strong>Dilution</strong>s System<br />
Input <strong>Dilution</strong> <strong>Gas</strong>es:<br />
1 Standard<br />
Input Source <strong>Gas</strong>es:<br />
2 Standard, 1 Purge<br />
Additional Inputs:<br />
4 Additional<br />
<strong>Dilution</strong> Mass Flow Controller: 0-10 SLPM<br />
Optional Ranges:<br />
0 thru 20 SLPM<br />
Source Mass Flow Controller: 0-100 SCCM<br />
Optional Ranges:<br />
0-1 SLPM<br />
2 nd Source Mass Flow Controller: 0 thru 1 SLPM<br />
Flow Accuracy:<br />
< + 1% Full Scale<br />
Flow Repeatability:<br />
< + 0.15% Full Scale<br />
Linearity:<br />
< + 0.5% Full Scale<br />
Input Pressure:<br />
12-35 psig<br />
Output Manifold:<br />
4 Outputs Standard<br />
Optional Outputs:<br />
4 Additional<br />
Optional Internal Ozone Generator<br />
Output:<br />
0.05-1.5 ppm @ 5 SLPM<br />
Accuracy:<br />
+ 2% of Set Point or<br />
+ 3 ppb @ 5 SLPM<br />
Nominal Flow:<br />
100 SCCM, + 1 SCCM<br />
UV Lamp Temperature:<br />
50 deg. C, + 0.1 deg. C<br />
<strong>Calibrator</strong> Interface<br />
Operation:<br />
Calibration Definitions:<br />
Calibration Types:<br />
<strong>Gas</strong> Definitions:<br />
Auto Calibrations:<br />
Digital Inputs/Outputs:<br />
Optional I/O:<br />
Communications:<br />
<strong>Calibrator</strong> System<br />
Operating Temperature:<br />
Dimensions:<br />
Average Weight:<br />
Input Voltage:<br />
Manually through 42-button<br />
membrane keypad, Keyboard,<br />
serial ports<br />
20 factory-defined calibration<br />
sequences with up to 20 points per<br />
sequences<br />
<strong>Gas</strong> <strong>Dilution</strong>, Ozone, GPT<br />
1 Diluent <strong>Gas</strong>es, 20 Source <strong>Gas</strong><br />
20 timer driven cal routines that<br />
perform user-defined calibration<br />
sequences on a 7 day calendar of<br />
events<br />
8 status I/O bits for control of<br />
calibrator functions by contact<br />
closures or TTL logic and output<br />
bits for monitoring calibrator<br />
functions<br />
Additional 16 status I/O bits for a<br />
total of 24 I/O bits<br />
RS232<br />
5 deg. C to 40 deg. C<br />
6.2” (15.6 cm) H x 14.3” (36.4 cm)<br />
W x 12.4” (31.6 cm) D<br />
19 lbs. (8.6 kg)<br />
98-264 VAC, 100 VA, 50/60 Hz<br />
operation<br />
<strong>Sabio</strong> <strong>Instruments</strong>, Inc. • 103A Halmar Cove • P.O. Box 5059 • Georgetown, TX 78627-5059 • USA<br />
Phone: (512) 869-0544 • Fax: (512) 869-0993 • Email: sales@sabio.com