Model 2010D Gas Dilution Calibrator - Sabio Instruments
Model 2010D Gas Dilution Calibrator - Sabio Instruments
Model 2010D Gas Dilution Calibrator - Sabio Instruments
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Gas</strong> <strong>Dilution</strong> <strong>Calibrator</strong><br />
<strong>Model</strong> <strong>2010D</strong> <strong>Gas</strong> <strong>Dilution</strong><br />
<strong>Calibrator</strong><br />
Users can easily perform gas dilution,<br />
ozone, and GPT calibration with<br />
<strong>Sabio</strong> Instrument’s intelligent portable<br />
<strong>Model</strong> 2010 gas dilution calibrator.<br />
The <strong>Model</strong> 2010 was designed to be<br />
small in size and weight for users<br />
who require true portability.<br />
Enhanced with the latest technology,<br />
<strong>Sabio</strong>’s <strong>Model</strong> 2010 calibration<br />
system is an uncomplicated and<br />
effective way to precisely control the<br />
dilution of gas standards for<br />
calibration of ambient air and source<br />
monitoring analyzers.<br />
The <strong>Model</strong> 2010 calibrator makes use<br />
of embedded microprocessor<br />
technology to enhance the accuracy<br />
and control of the featured calibration<br />
system. The unit can be operated<br />
manually, automatically, or semiautomatically<br />
by remote control to<br />
conduct calibrations. The calibrator<br />
has the capability to produce and<br />
store in memory 20 calibration<br />
sequences, with up to 20 levels of<br />
source/dilution in each sequence.<br />
Users can employ either the front<br />
panel membrane keypad, external<br />
keyboard, or serial port to input<br />
calibration parameters.<br />
The pneumatic system for the <strong>Model</strong><br />
2010 utilizes Mass Flow Controllers<br />
to precisely meter gas streams and<br />
implements multiple types of curve<br />
correction algorithms to linearize<br />
calibrations.<br />
The <strong>Model</strong> 2010 design exceeds the<br />
United States EPA calibration method<br />
requirements. <strong>Dilution</strong> components<br />
are calibrated with standards and test<br />
equipment traceable to the National<br />
Institute of Standards and<br />
Technology (NIST).<br />
Standard Features<br />
Pneumatics<br />
Diluent Mass Flow Controller, 0-10 SLM<br />
Source Mass Flow Controller, 0-100 SCCM<br />
2 Calibration <strong>Gas</strong> Input Ports and 1 Purge Port<br />
4 Output Ports and 1 Vent Port<br />
1 Diluent <strong>Gas</strong> Input Port<br />
Electronics<br />
Bright Active Matrix Color Display (shown with monochrome<br />
EL display)<br />
42-Button Membrane Keypad<br />
8 Bits Digital Input/Output<br />
2 Serial Ports for Data Communications<br />
External PC Keyboard Input Port<br />
Centronics Parallel Printer Port<br />
Air Source Control Port<br />
Electrical<br />
Standard 98-264 VAC, 150-300 VA, 50/60 Hz Operation<br />
Optional Features<br />
Alternate Diluent Mass Flow Controller, Ranges 0-20 SLM<br />
Alternate Source Mass Flow Controller, Ranges 0-1 SLM<br />
Second Source Mass Flow Controller, Ranges 0-1 SLM<br />
Additional Calibration <strong>Gas</strong> Input Ports<br />
Permeation Port<br />
Internal Ozone Generator<br />
Ozone Generator UV Optical Servo Control Loop<br />
Side mounted carrying straps<br />
High-impact transport case<br />
Ethernet TCP/IP Capability<br />
2548 Shell Road Units A-D Georgetown, TX 78628 USA<br />
Phone: (512) 869-0544 Fax: (512) 869-0993 Email: sales@sabio.com
<strong>Model</strong> 2010 Specifications<br />
______________________________________________________<br />
<strong>Dilution</strong>s System<br />
Input <strong>Dilution</strong> <strong>Gas</strong>es:<br />
1 Standard<br />
Input Source <strong>Gas</strong>es:<br />
2 Standard, 1 Purge<br />
Additional Inputs:<br />
2 Additional<br />
<strong>Dilution</strong> Mass Flow Controller: 0-10 SLPM<br />
Optional Ranges:<br />
0 thru 20 SLPM<br />
Source Mass Flow Controller: 0-100 SCCM<br />
Optional Ranges:<br />
0-1 SLPM<br />
2 nd Source Mass Flow Controller: 0 thru 1 SLPM<br />
Flow Accuracy:<br />
< + 1% Full Scale<br />
Flow Repeatability:<br />
< + 0.15% Full Scale<br />
Linearity:<br />
< + 0.5% Full Scale<br />
Input Pressure:<br />
20-35 psig<br />
Output Manifold:<br />
4 Outputs Standard<br />
Optional Internal Ozone Generator<br />
Output:<br />
0.05-1.5 ppm @ 5 SLPM<br />
Accuracy:<br />
+ 2% of Set Point or<br />
+ 3 ppb @ 5 SLPM<br />
Nominal Flow:<br />
100 SCCM, + 1 SCCM<br />
UV Lamp Temperature:<br />
50 deg. C, + 0.1 deg. C<br />
<strong>Calibrator</strong> Interface<br />
Operation:<br />
Calibration Definitions:<br />
Calibration Types:<br />
<strong>Gas</strong> Definitions:<br />
Auto Calibrations:<br />
Digital Inputs/Outputs:<br />
Communications:<br />
<strong>Calibrator</strong> System<br />
Operating Temperature:<br />
Dimensions:<br />
Average Weight:<br />
Input Voltage:<br />
Manually through 42-button<br />
membrane keypad, Keyboard,<br />
serial ports<br />
20 factory-defined calibration<br />
sequences with up to 20 points per<br />
sequences<br />
<strong>Gas</strong> <strong>Dilution</strong>, Ozone, GPT<br />
1 Diluent <strong>Gas</strong>es, 20 Source <strong>Gas</strong><br />
20 timer driven cal routines that<br />
perform user-defined calibration<br />
sequences on a 7 day calendar of<br />
events<br />
8 status I/O bits for control of<br />
calibrator functions by contact<br />
closures or TTL logic and output<br />
bits for monitoring calibrator<br />
functions<br />
RS232<br />
5 deg. C to 40 deg. C<br />
6.2” (15.6 cm) H x 14.3” (36.4 cm)<br />
W x 12.4” (31.6 cm) D<br />
19 lbs. (8.6 kg)<br />
98-264 VAC, 100 VA, 50/60 Hz<br />
operation<br />
2548 Shell Road Units A-D Georgetown, TX 78628 USA<br />
Phone: (512) 869-0544 Fax: (512) 869-0993 Email: sales@sabio.com