772 Westenberg et al.<br />
Figure 4. Bl<strong>and</strong>-Altman analysis for studying the differences between the MV-flow volume [determined with the 1-dir MV-flow<br />
method (A) <strong>and</strong> the 3-dir MV-flow method (B), respectively] <strong>and</strong> the AO flow. The 1-dir MV flow shows a statistically significant<br />
overestimation compared to the AO flow. The 3-dir MV flow is not statistically significantly different. The dashed lines represent the<br />
mean differences, the dotted line determines the 95% confidence intervals.<br />
methods <strong>and</strong> the AO-flow volume is tested with<br />
Pearson correlation coefficients. The correlation<br />
between the 1-dir MV-flow volume <strong>and</strong> the AO-flow<br />
volume is not statistically significant (r P =0.15,<br />
p=0.68), whereas the correlation between the 3-dir<br />
MV-flow volume <strong>and</strong> the AO-flow volume is statistically<br />
significant: (r P =0.92, p