Ametek 900 ADA PDF
Ametek 900 ADA PDF
Ametek 900 ADA PDF
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
PERFORMANCE SPECIFICATIONS<br />
Methodology: Multiple wavelength, high resolution,<br />
non-dispersive UV<br />
Species Minimum Maximum<br />
Measured Full Scale Full Scale<br />
H 2 S 5000 ppm 10%<br />
SO 2 2500 ppm 10%<br />
CS 2 5000 ppm 10%<br />
COS 5000 ppm 10%<br />
Sv 50 ppm 500 ppm<br />
Accuracy 1 : Better than 1% full scale<br />
Reproducibility 1 : Better than 0.5% full scale<br />
Linearity 1 : Better than 1% full scale<br />
Response Time: Typically less than 30 s to T 90 (excl. sample<br />
system)<br />
Number of Gases: Up to 5 simultaneously (refer to AMETEK for<br />
possible combinations)<br />
Power: 120 VAC +10%, 47-63 Hz or 220 VAC +10%, 47-63 Hz<br />
600 W (for analyzer only)<br />
Approvals and Certifications<br />
NEC/CEC Class I, Division 2, Groups C & D<br />
CENELEC EEx pd IIB T3<br />
GOST 1ExpdIIBT3 (certification pending)<br />
Complies with all relevant European Directives<br />
GOST Pattern Approval<br />
Typical Sample Flow: 3 to 5 1/min (0.1 to 0.2 SCFM)<br />
Sample Transport: Air aspiration<br />
Temperature Control: Independent control of four zones<br />
(oven, sample line, probe, vent line)<br />
Pressure and Temperature Compensation: Standard<br />
Ambient Temperature: 41 O F to 122 O F (5 O C to 50 O C)<br />
Instrument Air: Minimum 30 psig, 1 CFM; instrument quality air<br />
Physical Dimensions: 61 x 44 x 12 in. (1554 x 1118 x 305 mm)<br />
Weight: Approximately 250 to 350 lbs. (115 to 160 kg)<br />
Communications:<br />
Analog 4 x 4 to 20 mA self powered<br />
Digital One RS-422 with Modbus protocol<br />
Relays 3 independent sets of SPDT relays alarm conditions<br />
1. NOTE: Stated specifications apply to SO 2 and H 2 S only. Please consult factory for<br />
specifications on other species.<br />
IN<br />
(MM)<br />
ALLOW MIN.<br />
10.0 (254.0)<br />
OVEN DOOR<br />
OUTLINE<br />
2.94<br />
(74.7)<br />
44.0<br />
(1117.6)<br />
42.25<br />
(1073.2)<br />
SAMPLE LINE<br />
CONNECTION<br />
3/8 IN.<br />
SWAGELOK<br />
Ø 1.76 (44.6)<br />
(2 PLACES)<br />
VENT LINE<br />
CONNECTION<br />
1/2 IN.<br />
SWAGELOK<br />
27.12<br />
(688.8)<br />
ANALYZER<br />
OVEN<br />
58.99<br />
(1498)<br />
SAMPLE<br />
INLET<br />
ELECTRONICS<br />
ENCLOSURE<br />
DOOR OUTLINE<br />
ELECTRONICS<br />
ENCLOSURE<br />
BLANK PLATE<br />
RECOMMENDED<br />
ENTRY LOCATION<br />
FOR PROBE HTR/<br />
RTD CONDUIT<br />
NOTE:<br />
SAMPLE INLET/VENT<br />
HOLE LOCATIONS ARE<br />
BASED ON THE<br />
RECOMMENDED<br />
MOUNTING HEIGHT<br />
SHOWN ON THE<br />
FRONT VIEW.<br />
55.05<br />
(1398)<br />
SAMPLE<br />
VENT<br />
ALLOW MIN.<br />
27.8 (706.0)<br />
DOOR SWING<br />
41.44<br />
(1052.7)<br />
RECOMMENDED<br />
HEIGHT<br />
27.12<br />
(688.8)<br />
3.99<br />
(101.3)<br />
SAMPLE/VENT<br />
LINE & PROBE<br />
TERMINATION BOX<br />
8 x 6 x 3.5 IN.<br />
3/4 IN.<br />
CONDUIT (A)<br />
3/4 IN.<br />
CONDUIT (B)<br />
USE 3/8 x 1-1/2 IN.<br />
STUDS<br />
(6 PLACES)<br />
FLOOR<br />
FLOOR<br />
One of a family of innovative process analyzer solutions from AMETEK Process Instruments. Specifications subject to change without notice.<br />
PROCESS INSTRUMENTS<br />
CAN<strong>ADA</strong> CHINA FRANCE GERMANY USA - Delaware USA - Texas<br />
2876 Sunridge Way N.E. Rm #B5, 16F Harvest Bldg. 3 Avenue Des Coudriers Postfach 2165 455 Corporate Blvd 9750 Whithorn Drive<br />
Calgary, AB T1Y 7H9 #585 Long Hua Xi Road Z.A. De L’Observatoire D-40644 Meerbusch OR Newark, DE 19702 Houston, TX 77095<br />
Ph. 403-235-8400 Shanghai, 200232 78180 Montigny Le Bretonneux Rudolf-Diesel-Str. 16 Ph. 302-456-4400 Ph. 281-463-2820<br />
Fax 403-248-3550 Ph. 86 21 64 284 067 Ph. 33 1 30 64 89 70 D-40670 Meerbusch Fax 302-456-4444 Fax 281-463-2701<br />
Fax 86 21 64 875 329 Fax 33 1 30 64 89 79 Ph. 49 21 59 91 36 0<br />
Fax 49 21 59 91 3680<br />
www.ametekpi.com<br />
© 2001, by AMETEK, Inc. All rights reserved.<br />
901<strong>PDF</strong> (080116)