Receiver Modules RX - Radiometer Physics Gmbh
Receiver Modules RX - Radiometer Physics Gmbh
Receiver Modules RX - Radiometer Physics Gmbh
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Receiver</strong> <strong>Modules</strong> <strong>RX</strong><br />
50-500 GHz<br />
active frequency down converter / receiver<br />
Product Data Sheet – April 2011 Iss 1 Rev 1<br />
General Description<br />
RPG offers full band waveguide active<br />
frequency down converters / receivers<br />
covering the frequency range 50 – 500<br />
GHz. Standard versions are optimized for<br />
broadband operation; higher dynamics<br />
versions with reduced bandwidth can be<br />
customized.<br />
All receiver modules are equipped with<br />
UG-387 waveguide flange and delivered<br />
with testport adapters and power supplies.<br />
Matched feed horn antennas are part of<br />
RPG’s product portfolio.<br />
Key Features<br />
• full band WR15 to WR2.2<br />
• high dynamic range<br />
• no mechanical tuners<br />
• passive cooling<br />
Order Code: <strong>RX</strong>-yy<br />
option<br />
yy<br />
-A<br />
description<br />
upper frequency of desired frequency band,<br />
e.g. yy=110 for WR10<br />
select option –A for additional tunable<br />
waveguide attenuator (WTA-series)<br />
-40<br />
dynamics [dB]<br />
-60<br />
-80<br />
-100<br />
-120<br />
<strong>RX</strong>-75<br />
<strong>RX</strong>-90<br />
<strong>RX</strong>-110<br />
<strong>RX</strong>-140<br />
<strong>RX</strong>-170<br />
<strong>RX</strong>-220<br />
<strong>RX</strong>-325<br />
<strong>RX</strong>-500<br />
-140<br />
-160<br />
50 75 100 150 200 250 300 400 500<br />
frequency [GHz]<br />
<strong>Radiometer</strong> <strong>Physics</strong> GmbH, Birkenmaarstr. 10, D - 53340 Meckenheim, Germany<br />
Tel.: +49 (0) 2225 99981 – 0; Fax: +49 (0) 2225 999811 – 99; www.radiometer-physics.de; info@radiometer-physics.de<br />
All data and specification are subject to change without notice! ©<strong>Radiometer</strong> <strong>Physics</strong> GmbH 2010
<strong>Receiver</strong> <strong>Modules</strong> <strong>RX</strong><br />
50-500 GHz<br />
active frequency down converter / receiver<br />
Product Data Sheet – April 2011 Iss 1 Rev 1<br />
Technical specifications <strong>RX</strong> modules*<br />
specification<br />
unit<br />
<strong>RX</strong>-75<br />
<strong>RX</strong>-90<br />
<strong>RX</strong>-110<br />
<strong>RX</strong>-140<br />
<strong>RX</strong>-170<br />
<strong>RX</strong>-220<br />
<strong>RX</strong>-325<br />
<strong>RX</strong>-500<br />
frequency range<br />
GHz<br />
50 – 75<br />
60 – 90<br />
75 – 110<br />
90 – 140<br />
110 – 170<br />
140 – 220<br />
220 – 325<br />
325 – 500<br />
waveguide band<br />
WR15<br />
WR12<br />
WR10<br />
WR8<br />
WR6.5<br />
WR5.1<br />
WR3.4<br />
WR2.2<br />
minimal dynamic<br />
range +<br />
dB<br />
120<br />
125<br />
125<br />
120<br />
115<br />
110<br />
95<br />
80<br />
typical dynamic<br />
range +<br />
dB<br />
140<br />
140<br />
140<br />
130<br />
130<br />
125<br />
115<br />
100<br />
1dB compression<br />
point<br />
dBm<br />
-5<br />
-5<br />
-5<br />
-5<br />
-5<br />
-5<br />
-5<br />
-5<br />
input match<br />
dB<br />
-15<br />
-15<br />
-15<br />
-15<br />
-15<br />
-15<br />
-15<br />
-15<br />
input connector<br />
UG-387<br />
UG-387<br />
UG-387<br />
UG-387<br />
UG-387<br />
UG-387<br />
UG-387<br />
UG-387<br />
LO frequency<br />
range<br />
GHz<br />
8.3 - 12.5<br />
10 - 15<br />
12.5 - 18.4<br />
11.2 - 17.5<br />
6.8 - 10.7<br />
11.6 - 18.4<br />
13.7 – 20.4<br />
13.5 - 20.9<br />
LO multiplication<br />
factor<br />
6<br />
6<br />
6<br />
8<br />
16<br />
12<br />
16<br />
24<br />
LO input power<br />
dBm<br />
+7<br />
+7<br />
+7<br />
+7<br />
+7<br />
+7<br />
+7<br />
+7<br />
LO input connector<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
IF frequency range<br />
MHz<br />
5 – 1000* 5 – 1000*<br />
5 – 1000*<br />
5 – 1000*<br />
5 – 1000*<br />
5 – 1000*<br />
5 – 1000*<br />
5 – 1000*<br />
IF frequency power<br />
level<br />
dBm<br />
-15 †<br />
-15 †<br />
-15 †<br />
-15 †<br />
-15 †<br />
-15 †<br />
-15 †<br />
-15 †<br />
IF output connector<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
SMA(f)<br />
supply voltage<br />
V DC<br />
+9<br />
+9<br />
+9<br />
+9<br />
+9<br />
+9<br />
+9<br />
+9<br />
supply current<br />
mA<br />
300<br />
350<br />
300<br />
300<br />
300<br />
300<br />
300<br />
350<br />
power supply<br />
voltage<br />
V AC<br />
100 – 240 100 – 240<br />
100 – 240<br />
100 – 240<br />
100 – 240<br />
100 – 240<br />
100 – 240<br />
100 – 240<br />
* Contact RPG for <strong>RX</strong> modules with modified specifications<br />
† input signal power level -15 dBm<br />
+ resolution bandwidth 10Hz<br />
Absolute maximum ratings<br />
LO input power<br />
supply voltage<br />
case temperature<br />
dBm<br />
V DC<br />
°C<br />
+10<br />
+12<br />
+85<br />
<strong>Radiometer</strong> <strong>Physics</strong> GmbH, Birkenmaarstr. 10, D - 53340 Meckenheim, Germany<br />
Tel.: +49 (0) 2225 99981 – 0; Fax: +49 (0) 2225 999811 – 99; www.radiometer-physics.de; info@radiometer-physics.de<br />
All data and specification are subject to change without notice! ©<strong>Radiometer</strong> <strong>Physics</strong> GmbH 2010
1 2<br />
3<br />
4 5<br />
6 7 8<br />
A<br />
A<br />
35,0 130,0<br />
Mounting Interface<br />
4x M6<br />
B<br />
58<br />
RF Interface<br />
UG387 anti cocking precision flange<br />
Label / Interface<br />
16 12<br />
46,5<br />
65<br />
30<br />
72,0<br />
B<br />
27 146<br />
2x<br />
3,00 H7<br />
C<br />
C<br />
D<br />
D<br />
26<br />
E<br />
F<br />
2<br />
180<br />
~ 192<br />
Toleranzen / TOLERANCES:<br />
X = 0,2 mm<br />
X.X = 0,05 mm<br />
X.XX = 0,02 mm<br />
X.XXX = 0,005 mm<br />
Gezeichn. / DRAWN<br />
Geprüft / CHK'D<br />
Freigegeben / APPV'D<br />
Gefertigt / MFG<br />
NAME<br />
Abeling<br />
<strong>Radiometer</strong> <strong>Physics</strong> GmbH<br />
Birkenmaarstraße 10<br />
53340 Meckenheim<br />
Germany<br />
Unterschrift / SIGN.<br />
DATE<br />
30.3.2011<br />
Eigenschaften / PROPERTIES:<br />
Material / MATERIAL:<br />
Gewicht / WEIGHT:<br />
Oberfläche / SURFACE:<br />
Scharfe Kanten<br />
fasen und<br />
brechen<br />
DEBUR AND<br />
BREAK SHARP<br />
EDGES<br />
---<br />
---<br />
---<br />
--- g<br />
--- mm²<br />
UNLESS OTHERWISE SPECIFIED<br />
DIMENSIONS ARE IN MILLIMETERS<br />
Änderungen siehe / CHANGES SEE<br />
---<br />
Benennung<br />
TITLE<br />
Version / Überarbeitung<br />
VERSION / REVISION<br />
Ersatz für / REPLACEMENT FOR<br />
<strong>Receiver</strong><br />
<strong>RX</strong>-xxx<br />
Outline Drawing<br />
1B<br />
1 2 3 4<br />
Qualitätskontrolle / Q.A<br />
Zeichnung nicht skalieren<br />
DO NOT SCALE DRAWING<br />
Oberfläche vergolden /<br />
SURFACE GILDING:<br />
--- mm²<br />
Zeichnungsnr.<br />
DRAWING NO.<br />
Bestellcode / Order Code: --- ---<br />
Maßstab / SCALE: ---<br />
OUT-0000-950-100-00-1B<br />
SHEET 1 OF 3<br />
A3