<strong>IAAF</strong> <strong>Calibration</strong> <strong>and</strong> <strong>Testing</strong> <strong>Manual</strong><br />
Graduation Test Intervals:<br />
Every 1m for initial <strong>and</strong> subsequent calibrations.<br />
Recalibration Intervals:<br />
Every four years or after damage / repair of the bar.<br />
Limitation on Use:<br />
The calibrated steel tape should be used for checking the accuracy of other<br />
working measuring bars.<br />
Vernier Callipers <strong>and</strong> Micrometers<br />
Applicable St<strong>and</strong>ards:<br />
ISO 6906:1984 Vernier calipers reading to 0.02 mm or<br />
BS 887:2008 Precision Vernier calipers. Requirements <strong>and</strong> test methods or<br />
AS 1984-1997: Vernier Callipers (metric series) or<br />
JIS B 7507:1993 Vernier, dial <strong>and</strong> digital callipers or<br />
DIN 862 (1988-12) Vernier calipers requirements <strong>and</strong> testing.<br />
ISO 3611:1978 Micrometer callipers for external measurement or<br />
AS 2102-1989 Micrometer callipers for external measurement or<br />
BS EN ISO 3611 Micrometer callipers for external measurement or<br />
BS 870:2008 Micrometer callipers for external measurement or<br />
JIS B 7502:1994 Micrometer callipers for external measurement or<br />
DIN 863-1 (1994-04) Micrometer callipers for external measurement.<br />
Test Temperature:<br />
The calibration test shall be undertaken at a temperature selected by the<br />
laboratory within the range 20 to 23 degrees Celsius <strong>and</strong> stable to ±1.0 degree<br />
Celsius during the test.<br />
Required 95% Confidence Limit Uncertainty of a Single Measurement:<br />
±0.05mm<br />
Reading<br />
To the nearest 0.1mm.<br />
Recalibration Intervals:<br />
Every four years or after damage / repair.<br />
Electronic Distance Measuring<br />
Applicable St<strong>and</strong>ards:<br />
Australia: Regulation 80 of the National St<strong>and</strong>ard Measurement Act 1960<br />
Required 95% Confidence Limit Uncertainty of a Single Measurement:<br />
±(6mm + 4ppm.)<br />