Individual Wind Gauge <strong>Calibration</strong><br />

Each wind gauge on sale shall be supplied with a calibration certificate stating<br />

the uncertainty of calibration for a test over a 10 second interval at a velocity<br />

of + 2.0m/s. Such calibration certificates may be provided by the manufacturer<br />

if their test facility has been accredited by the national accreditation authority.<br />

Recalibration<br />

Each wind gauge shall be tested over a 10 second interval at a velocity of<br />

+2.0m/s every four years. Simple test procedures that federations may<br />

consider are given hereunder.<br />

Comparison <strong>Testing</strong><br />

A wind gauge that has a current traceable calibration certificate may be used<br />

as a reference measure for other wind gauges. The wind gauges to be<br />

checked <strong>and</strong> the reference wind gauge should be set up in an open field when<br />

there are constant light winds with the measurement directions parallel to<br />

each other <strong>and</strong> the general wind direction.. At least five comparison measures<br />

of wind speed in the range of 1 to 3 metres per second but preferably as close<br />

to 2 metres per second as possible over a 10 second interval should be taken.<br />

Provided the average difference between the wind readings for the wind<br />

gauge being checked compared with the wind speeds of the reference wind<br />

gauge is not more than ±0.1metres per second then the wind gauge being<br />

tested may be certified as being in calibration. An appropriate sticker should<br />

be attached to the wind gauge indicating the date of test, the tester’s name<br />

<strong>and</strong> the reference wind gauge’s identification number. The test results should<br />

be filed by the Technical Manager for future reference.<br />