24.12.2014 Views

MQP – 014 EQA HANDBOOK EDITION 6 A participant ... - Thistle QA

MQP – 014 EQA HANDBOOK EDITION 6 A participant ... - Thistle QA

MQP – 014 EQA HANDBOOK EDITION 6 A participant ... - Thistle QA

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

P.O. Box 131375, Bryanston, 2074<br />

Ground Floor, Block 5<br />

Bryanston Gate, 170 Curzon Road<br />

Bryanston, Johannesburg, South Africa<br />

804 Flatrock, Buiten Street, Cape Town, 8001<br />

www.thistle.co.za<br />

Tel: +27 (011) 463 3260<br />

Fax: +27 (011) 463 3036<br />

Fax to Email: + 27 (0) 86-557-2232<br />

e-­‐mail : service@thistle.co.za<br />

<strong>MQP</strong>-­‐<strong>014</strong><br />

Edition 10<br />

Where do I start investigating problems<br />

This is never simple. We always suggest, however, checking the simple things<br />

first.<br />

1. Have we got the correct registration details for you Each report shows<br />

the categories or instruments for which you are registered. If this is not<br />

correct, tell us. Do not change anything unless you have taken this simple<br />

step.<br />

2. Have you made a reconstitution or transcription or mixing error Is it<br />

possible that your reconstitution procedure was not followed Could you<br />

have a problem with water quality Are you sure you sent us the <strong>E<strong>QA</strong></strong><br />

sample results – could there have been a transcription error Finally, for<br />

haematology, did you follow the mixing instructions Again, check these<br />

stages before you make any changes.<br />

3. Is the error you see a once-off Do not over-react to a single error on<br />

<strong>E<strong>QA</strong></strong>. Start your investigations, and certainly monitor your procedures,<br />

but do not begin altering calibration factors – not yet!<br />

4. What does your Internal Quality Control (IQC) look like If you have<br />

imprecision on both <strong>E<strong>QA</strong></strong> and IQC, investigate and resolve. If on one<br />

only, investigate and resolve. BUT, if your IQC looks fine, without a bias,<br />

and you have a bias on your <strong>E<strong>QA</strong></strong>, give us a call. We need to discuss the<br />

way in which you integrate your information from your IQC (for<br />

monitoring precision) and your <strong>E<strong>QA</strong></strong> (for assessing accuracy).<br />

5. If this information above is not clear, or does not answer your problem,<br />

either look on our web page, or give us a call. We are here to help.<br />

Page 11 of 12

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!