Comet F - catalogue sheet
Comet F - catalogue sheet
Comet F - catalogue sheet
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
EMKOMETER<br />
Ledeč nad Sázavou tel: +420 569 721 622, +420 569 721 549,<br />
+420 569 720 539, e-mail: emko_ledec@iol.cz<br />
COMET F2<br />
INDUCTIVE FLOWMETER<br />
COMET F2 - CONTROL UNIT<br />
It is at all points certified unit of inductive flowmeters CPMET F2, with wide range of additional functions, that give to<br />
user comfortable service.<br />
Important properties:<br />
- High precision of flow measuring in both directions.<br />
- Easy option of mode and parameters (by means of 5 buttons or from PC).<br />
- Backup clock of real time.<br />
- Possibility of periodical clearing of sensor’s electrodes.<br />
- Possibility of saving 1.000 data samples.<br />
- Liquid batching.<br />
- Option of dead zone and damping of impression.<br />
- Two independent multifunctional outputs.<br />
- Frequency output, analogue output (up to 800Ω).<br />
- Calibration methods adaptable to test rooms.<br />
In case of set meter (metrology seal), the User cannot change either the calibration constants or cumulated values.<br />
Detail information you will find in manuals: COMET F2 – Installation Manual, COMET F2 – Reference Manual.<br />
Voltage 230 V~ (+10%,-15%), 50Hz (±2%)<br />
Power supply<br />
12 VA max<br />
Shielding IP 66<br />
Operational temperature<br />
- 5º C ÷ + 45º C (protect against direct sunshine)<br />
Warehousing temperature - 20º C ÷ + 80º C (relative humidity max 85%)<br />
Measuring error<br />
≤ 0,5% from measured value in range 5 ÷ 100% Q max<br />
Indication<br />
LCD display 2 x 16 characters<br />
Manual control<br />
five buttons<br />
Input<br />
batch start<br />
(Galvanic separation)<br />
I v = 2 ÷ 10 mA (diode)<br />
Outputs<br />
2 independent multifunctional<br />
(Galvanic separation)<br />
impulses of flowed quantity,<br />
one- or two-levels comparison,<br />
batch start and stop<br />
(transistor 30V / 20mA max.; closing or opening contact<br />
relay (120V~ / 0,8A max.)<br />
active current<br />
( 0 ÷ 20mA or 4 ÷ 20mA up to ≤ 800Ω)<br />
frequency<br />
2 Hz ÷ 20 kHz at 30V/5mA max;<br />
max 50kHz at 30V/1mA max.<br />
(transistor)<br />
supply 24V ss , unstable
EMKOMETER<br />
Serial ports<br />
(Galvanic separation)<br />
Case material<br />
Weight<br />
Ledeč nad Sázavou tel: +420 569 721 622, +420 569 721 549,<br />
+420 569 720 539, e-mail: emko_ledec@iol.cz<br />
(R i = 470Ω; C = 470MF)<br />
current loops<br />
TxD output ( 30V ss / 20mA max, transistor)<br />
RxD input ( I i = 2 ÷ 10mA, diode)<br />
RS232<br />
connector Canon 9<br />
another data bus<br />
by request<br />
ABS<br />
2 kg<br />
F.OUT<br />
FLOW<br />
DAMP<br />
ANALOG<br />
OUTPUT<br />
FLOW SENSORS<br />
Conductivity of measured liquid ≥ 5µS.cm -1<br />
Shielding<br />
IP 65, IP67<br />
coat<br />
technical rubber, PTFE<br />
PN<br />
0,6 MPa ÷ 2,5 MPa<br />
Electrodes<br />
by default steel 17.246, hasteloy, titanium, and platinum<br />
Flanges<br />
flat, screwing, without flanges<br />
Cable length 5 m standard (max. 30m)<br />
In the table are mentioned standard parameters of sensors in flanges design. The delivery of different dimensions and designs<br />
is possible.<br />
DN/PN Qmax[l/s] Qmax[m3/h] A[mm] B[mm] Weight[kg]<br />
25/16 0,5 - 5 1,8 - 18 150/200 175 4<br />
32/16 0,8 - 8 2,9 - 29 150/200 195 5<br />
40/16 1,2 - 12,5 4,5 - 45 150/200 205 7<br />
50/16 2 - 20 7 - 70 200 220 9<br />
65/16 3 - 30 12 - 120 200 240 10<br />
80/16 5 - 50 18 - 180 200 255 17<br />
100/16 8 - 80 28 - 280 200 270 19<br />
125/16 12 - 120 44 - 440 300 305 25<br />
150/16 18 - 180 63 - 630 300 330 30<br />
200/16 31 - 310 115 - 1150 350 390 40<br />
250/10 50 - 500 180 - 1800 350 460 55<br />
300/10 70 - 700 255 - 2550 500 520 90<br />
350/10 96 - 960 350 - 3500 500 550 120<br />
400/10 125 - 1250 450 - 4500 600 615 150<br />
500/10 196 - 1960 700 - 7000 600 720 220<br />
600/10 280 - 2800 1000 - 10000 600 830 300<br />
800/10 500 - 5000 1800 - 18000 800 1060 550<br />
A - construction length; B – maximal height.<br />
B<br />
A