Signal sensitivity & linearity<br />

The OCTOPLUS system detects fluorescent signals from minimally labeled proteins (one<br />

fluorophore per protein) in the lower nanogram range. Even in this range, the signal<br />

linearity remains at an ideal level (R 2 = 0.996 [fig.3.]).<br />

1.000 0.500 0.250 0.125 0.062 0.031 0.016 0.008<br />

Protein load (μg)<br />

Signal intensity (AU)<br />

100<br />

80<br />

60<br />

R² = 0.996<br />

40<br />

20<br />

0<br />

0.0 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.0<br />

Amount of protein (µg)<br />

Fig. 3. Serial dilution of BSA minimally labeled with T-Rex 330. Proteins were separated by<br />

SDS-PAGE and scanned with OCTOPLUS. Signal intensities were analyzed by LabImage 1D L340<br />

(Kapelan GmbH, Germany).<br />