Binding Complex on Capture Strip<br />

1) Wash capture strips on the platewasher using<br />

the “3X Wash” program.<br />

2) Using an 8-channel pipetter, add 30 µL of<br />

each reaction to a well of the capture strips.<br />

30 µL<br />

3) Place the rack of capture strips back on the<br />

plateshaker and shake for 30 minutes.<br />

Prepare Capture Strips for PCR amplification<br />

1) Wash capture strips on the platewasher using<br />

the “5x Wash” program. The series of washes<br />

and soaks takes 15 minutes.<br />

2) Thaw primer probe (white cap) and add 40 µL<br />

primer/probe, 160 µL of water (DNase, RNase<br />

free) and 200 µL of Universal Master Mix per<br />

capture strip used to a glass test tube. Mix<br />

gently.<br />

+ Master + H 2 O<br />

Mix<br />

3) Add the primer/probe + Master Mix solution<br />

to a pipetting trough, and using an 8-channel<br />

pipetter, add 40 µL to each well of the capture<br />

strips.<br />

40 µL<br />

4) Seal the wells with an adhesive optical cover,<br />

cover with 2 compression pads and insert the<br />

strips into the thermocylcer of the PCR<br />

instrument.