<strong>Thermo</strong> <strong>Scientific</strong><br />
<strong>Multi</strong>-<strong>Stream</strong> <strong>Analyzer</strong> (<strong>MSA</strong>)<br />
XRF Elemental Slurry <strong>Analyzer</strong><br />
The <strong>Thermo</strong> <strong>Scientific</strong> <strong>Multi</strong>-<strong>Stream</strong><br />
<strong>Analyzer</strong> (<strong>MSA</strong>) provides accurate, fast,<br />
cost-effective elemental analysis of up<br />
to 12 slurry streams. The information<br />
provided by the <strong>MSA</strong> allows plant<br />
operators to follow process trends in<br />
real time, making manual or automatic<br />
changes to the process possible.<br />
This <strong>Analyzer</strong> enables improvements<br />
in product quality and metal recovery.<br />
Product Specifications<br />
Optimizes<br />
• Reagent dosage<br />
• Flotation air<br />
• Water balanced flows<br />
• Feed tonnage<br />
Flotation Circuit Monitoring<br />
• Tailings losses<br />
• Concentrate grades<br />
• Key impurity levels<br />
• Head grade variations<br />
Measures and Reports<br />
• Up to eight elements (Ca and higher)<br />
simultaneously and percent solids<br />
• Up to 12 streams<br />
• Period averages: rolling, hourly, shift, day<br />
• Trends and historical data logging<br />
• Metal recoveries<br />
The <strong>Thermo</strong> <strong>Scientific</strong> <strong>Multi</strong>-<strong>Stream</strong> <strong>Analyzer</strong><br />
(<strong>MSA</strong>) provides simultaneous analysis of up<br />
to 20 elements and percent solids. Elements<br />
in the periodic table from Ca and higher can<br />
be analyzed using the X-ray Fluorescence<br />
(XRF) technique, which is a proven and<br />
robust technology for plant environments.<br />
The <strong>MSA</strong> can measure from 12 to 18<br />
streams that are kept completely separate<br />
in their own analysis tanks, moving the<br />
probe from zone to zone. This overall design<br />
philosophy provides the following benefits:<br />
• No additional multiplexer or<br />
de-multiplexer is required.<br />
• Low slurry head loss of less than 1,000 mm<br />
across the <strong>MSA</strong>, which combined with a<br />
sensible location, dramatically reduces<br />
the need for sample pumps.<br />
• The standard 300 mm wide analysis tank<br />
consists of a de-aeration zone, followed<br />
by an agitated analysis zone. It can handle<br />
flows up to approximately 12 m 3 /hr.<br />
Larger width tanks up to 600 mm are<br />
available, which means the full flow of<br />
some process streams can be analyzed,<br />
eliminating the need for primary samplers<br />
and providing enhanced accuracy.<br />
• Assay updates are fast (seven minutes<br />
total cycle for six streams).<br />
• Cross cut metallurgical samplers are<br />
located at the outlet of each analysis tank<br />
for calibration and process control samples.<br />
When used with full flow or a representative<br />
primary sampler (e.g., the <strong>Thermo</strong> <strong>Scientific</strong><br />
SamStat), the <strong>MSA</strong> can collect shift<br />
composite samples for metallurgical<br />
accounting, providing cost savings.
Specifications<br />
<strong>Thermo</strong> <strong>Scientific</strong> <strong>Multi</strong>-<strong>Stream</strong> <strong>Analyzer</strong> (<strong>MSA</strong>)<br />
Analysis Performance<br />
Number of <strong>Stream</strong>s 3-12<br />
Analysis Time per <strong>Stream</strong><br />
Typically 1 minute<br />
Sample Flow Rate<br />
Nominally 12 m 3 /hr for standard tank, from 1m 3 /hr to 50 m 3 /hr<br />
Detection System<br />
Si (Li) solid state detector, LN 2 -cooled. Typical resolution 170-180 eV.<br />
Accuracy<br />
All elements with atomic number above 20 (i.e., above Ca in the periodic table)<br />
Assay Range (% element by weight) Relative Error 1<br />
0.05 to 0.2 4%-6%<br />
0.2 to 1.0 3%-5%<br />
1 to 10 2%-4%<br />
10 to 80 1%-2%<br />
1 Guide only; dependent on particle size, mineralogy and matrix variations<br />
Utilities Required<br />
Electrical Power<br />
Factory selectable 380/415/440/460 Volts AC ±10%, 3-phase, 48-62 Hz ±2 Hz (3 wire plus earth)<br />
Maximum power consumption: 3-stream, 1.8 kW; 6-stream, 2.9 kW; 9-stream, 3.8 kW—with standard 250W stirrer motors<br />
Air<br />
Instrument quality air (clean and dry to 0.1 microns with dew point