6<br />

INSTALLATION DRAWING<br />

1<br />

2 3 4 5 6 7 8<br />

A<br />

Lead wires not to scale<br />

Wires are 26 AWG and min 24” (610mm) long<br />

2x 1/4”-28 UNF<br />

Flat bottom port<br />

Depth of port = .25” min. (6.4mm)<br />

A<br />

B<br />

Ø1.0”<br />

(Ø25.4mm)<br />

.49”<br />

(12.4mm)<br />

Outlet<br />

.72”<br />

(18.3mm)<br />

B<br />

C<br />

D<br />

E<br />

F<br />

Set screw<br />

position<br />

variable<br />

to +.25” (6mm)<br />

(2.5”)<br />

(63.5mm)<br />

.84”<br />

(21.3mm)<br />

.688”<br />

(17.5mm)<br />

2x #4-40 UNF<br />

Mounting holes<br />

Inlet<br />

1.4”<br />

(35.6mm)<br />

<strong>Micro</strong>-Pump<br />

120SP series<br />

C<br />

D<br />

E<br />

F<br />

1 2 3 4 5 6 7 8<br />

SPECIFICATIONS<br />

120SP Fluid Data<br />

Dispense Volume (μl) 20 30 40 50 60<br />

Set-point accuracy +/- 10% +/- 10% +/- 10% +/- 10% +/- 10%<br />

Repeatability +/- 5% +/- 5% +/- 5% +/- 5% +/- 5%<br />

Max flow rate (μl/min) 2400 3600 4800 6000 7200<br />

Internal vol (μl) 105 105 105 105 105<br />

Voltage<br />

Power<br />

@70˚F (21˚C)<br />

120SP Electrical Data<br />

Current<br />

@70˚F (21˚C)<br />

Effective continuous<br />

power @ max cycle rate<br />

12 VDC 4.0 Watts 0.32 amps 1.2 Watts<br />

24 VDC 4.0 Watts 0.16 amps 1.2 Watts<br />

120SP Cycle Rates<br />

Fixed “on” time Min “off” time Max cycle rate<br />

150 msec 350 msec 2.0 Hz<br />

Recommended tubing for 120SP<br />

Inlet & outlet, 1/32” (0.80mm) ID, hardwall<br />

tubing, PART NO. 008T16-080<br />

120SP <strong>Micro</strong>-<strong>Pumps</strong> can be cycled at up to 2 Hz. To maintain<br />

pumping precision the voltage “on” time should remain fixed -<br />

the pumping rate can be changed by increasing the “off” time.<br />

Voltage (v)<br />

a = 150 msec (fixed)<br />

b = 350 msec (minimum, adjustable)<br />

a<br />

b<br />

Time (msec)<br />