30.01.2015 Views

Calibration Check Procedure - NanoDrop

Calibration Check Procedure - NanoDrop

Calibration Check Procedure - NanoDrop

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

T030 INSTRUCTIONS <strong>NanoDrop</strong> 8000<br />

<strong>Calibration</strong> <strong>Check</strong> Best Practices<br />

The Thermo Scientific <strong>NanoDrop</strong> 8000 <strong>Calibration</strong> <strong>Check</strong> diagnostic utilizes five repeat measurements to verify the pathlength<br />

calibration. It is important to pay careful attention to technique when performing this procedure, as a single outlier caused by<br />

user error can result in a failed calibration check on one or more pedestals.<br />

• Position the instrument at an angle that will allow for optimal use of the pipette guide.<br />

• Ensure the <strong>NanoDrop</strong> 8000 is not situated near an air vent or an exhaust fan from a<br />

nearby instrument.<br />

• Do not open the CF-1 ampoule until ready to perform the calibration check.<br />

• Aliquot 60 uL of CF-1 into each of the 8 tubes in the PCR strip. A volume of 60 ul in each PCR tube is required<br />

to reduce the effects of evaporation and potential fluid concentration.<br />

• Use a small-volume (0.5-10 uL) multi-channel pipettor. If using an electronic pipettor do not draw up a large volumes<br />

to dispense multiple aliquots. The CF-1 reserve stored in the tips may begin to concentrate, affecting the quality of<br />

results. Draw and dispense individual 1.5 uL aliquots for each measurement cycle.<br />

• Ensure that adequate volumes of CF-1 are being pipetted onto the center of each pedestal. If tips are not<br />

seated properly on each channel of a multi-channel pipettor, they may not properly draw up, or dispense an adequate volume<br />

onto each pedestal. Columns that are too thin, or shifted off to one side may not completely cover the fiber-optic center of<br />

each pedestal.<br />

Pre-<strong>Calibration</strong> <strong>Check</strong> Cleaning Instructions<br />

Wiping the sample from both the upper and lower measurement surfaces upon completion of each sample measurement is<br />

usually sufficient to prevent sample carryover and avoid residue buildup. However, reagents and buffers may contain surfactants<br />

that will “un-condition” the measurement surfaces, therefore it is a good practice to use the <strong>NanoDrop</strong> Pedestal Reconditioning<br />

Compound (PR-1) as a rapid means of reconditioning the pedestals prior to performing the calibration check.<br />

1. Open the vial containing PR-1 and use the applicator provided in the kit to remove a pin-head sized amount of the<br />

compound.<br />

2. Apply a very thin, even layer of PR-1 to the surface of the upper and lower pedestals.<br />

3. Wait 30 seconds for the PR-1 to dry.<br />

4. Fold a clean, dry laboratory wipe into quarters and remove the PR-1 by aggressively rubbing<br />

the surface of the upper and lower measurement surfaces until all compound residue is removed and an<br />

aliquot of water “beads up” on the surface.<br />

Note: The residue on the lab wipe may appear black. This is normal.<br />

Rev 6/09<br />

Thermo Fisher Scientific - <strong>NanoDrop</strong> products Wilmington, Delaware USA Technical support: info@nanodrop.com 302-479-7707 www.nanodrop.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!