2100 Series Portable Turbidimeter<br />
<strong>Laboratory</strong> Turbidimeters<br />
Bringing laboratory-level performance on-site.<br />
Lab-Quality Results in a Portable Unit<br />
The 2100P Portable Turbidimeter performs at a level normally<br />
associated with laboratory instruments. Its microprocessor-controlled<br />
touch-key functions and patented Ratio Optical System bring<br />
greater accuracy, sensitivity, and reliability to field testing. The twodetector<br />
optical system compensates for color in the sample, light<br />
fluctuation, and stray light, allowing analysts to achieve laboratorylevel<br />
results with a wide range of samples, even under difficult<br />
on-site conditions.<br />
Direct Digital Readout of Results<br />
The 2100P provides direct digital readout in nephelometric<br />
turbidity units (NTUs), completely eliminating the uncertainty of<br />
analog indicators and the need for calculations or interpolation<br />
of calibration charts.<br />
• Selectable signal averaging mode compensates for fluctuations in readings<br />
caused by movement of large particles in the light path<br />
• Pre-programmed calibration procedure, with microprocessor-controlled<br />
adjustment of calibration curve. No potentiometers to adjust<br />
• Electronic zeroing: compensates for electronic and optical offsets; no manual<br />
adjustments are required<br />
• Built-in diagnostics<br />
accessible with<br />
one keystroke<br />
• Range: 0 to 1000 NTUs<br />
• Direct digital readout<br />
in NTU<br />
• Meets or exceeds EPA<br />
method 180.1 criteria<br />
• Two-year warranty<br />
Ideal for Regulatory Reporting<br />
With a measurement range of 0 to 1000 NTUs and resolution to 0.01<br />
NTU, the 2100P is ideal for regulatory monitoring, process control<br />
applications, and field studies. Choose from three manual range<br />
modes: 0 to 9.99, 0 to 99.9, or 0 to 1000 NTU. Or select automatic<br />
range mode for samples that vary widely in turbidity.<br />
Ready for <strong>Field</strong> Use<br />
The 2100P comes field-ready with four AA alkaline batteries<br />
(optional battery eliminator available), all necessary apparatus, a set<br />
of primary and secondary standards, and a rugged carrying case.<br />
The instrument itself weighs less than 0.45 kg (~1 lb). The complete<br />
analytical package, including case, apparatus, standards, and<br />
instructions, weighs just under 2.7 kg (6 lb).<br />
For more information, call to request<br />
Literature #1619, or visit www.hach.com<br />
Specifications*<br />
Applications<br />
<strong>Field</strong> and laboratory monitoring<br />
and regulatory reporting of drinking<br />
water, wastewater, and treated<br />
(low-turbidity) process water.<br />
Principle of Operation<br />
Nephelometric Ratio<br />
Measurement Modes<br />
NTU<br />
Ranges<br />
0-1000 NTU in automatic range<br />
mode or manual range selection of<br />
0-9.99, 0-99.9 and 0-1000 NTU<br />
Accuracy †<br />
±2% of reading or ±1 least<br />
significant digit from 0-500 NTU;<br />
±3% of reading between<br />
500-1000 NTU<br />
Resolution<br />
0.01 on lowest range<br />
Repeatability †<br />
±1% of reading, or 0.01 NTU,<br />
whichever is greater (with Gelex<br />
standards)<br />
Response Time<br />
6 sec. for full step change (signal<br />
avg. off) in constant reading mode<br />
Light Source<br />
Tungsten lamp.<br />
Temperature Ranges<br />
Operating (inst. only)<br />
0 to 50°C (32 to 122°F)<br />
Storage (inst. only)<br />
-40 to 60°C (-40 to 140ºF)<br />
Sample<br />
0 to 95°C (32 to 203ºF)<br />
Sample Cells<br />
60.0-mm high x 25-mm diam.<br />
(2.36” high x 1” diam.)<br />
Sample Volume Required<br />
15 mL (0.5 oz)<br />
*Subject to change without notice.<br />
† Reference conditions: 23 ±2°C,<br />
50% RH ± 10% RH Noncondensing,<br />
115/230 Vac ± 17%, 50/60 Hz.<br />
Each Model 2100P Portable Turbidimeter is supplied with six sample<br />
cells, set of four standard ampules of