B-Scan Option for the epOch XT<br />

Our advanced digital ultrasonic flaw<br />

detector, the Panametrics-NDT<br />

EPOCH XT, offers a fully encoded B-scan<br />

option that displays scanned thickness<br />

measurements of internally corroded<br />

pipe or tank walls as easy-to-read crosssectional<br />

images. This time-saving B-scan<br />

option works with multiple crawlers and<br />

encoders providing important transducer<br />

location information of distance traveled.<br />

Complete with thickness markers the<br />

B-scan option provides readings at<br />

specific scan locations and the ability to<br />

automatically save A-scans for minimum<br />

thickness data or when triggered by alarm<br />

conditions.<br />

EPOCH XT B-Scan and Floating Gate<br />

B-<strong>SCAN</strong> <strong>OPTION</strong><br />

Flaw Detectors<br />

Features<br />

■<br />

■<br />

■<br />

Bi-directional, uni-directional, and<br />

manual scan modes<br />

Ability to simultaneously view<br />

A-scan and B-scan<br />

Scanning resolution up to 0.020 in.<br />

(0.5 mm)<br />

■ Fast scanning rate (variable based<br />

upon operator selectable resolution)<br />

of 20 in./sec. (500 mm/sec) with a<br />

resolution of 0.020 in. (0.5 mm)<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

Automatic audible alarm if scan speed<br />

exceeds limit<br />

Operates in all test modes including<br />

Time Varied Gain (TVG) and Floating Gate<br />

Floating Gate improves precision and<br />

consistency of thickness readings<br />

Ability to save a single screenshot or an<br />

entire B-scan<br />

Internal storage of up to 150,000<br />

B-scan thickness readings per scan<br />

Ability to scroll and review internally<br />

stored B-scan data<br />

■ Audible and visual Minimum Depth Alarm<br />

for thickness below an operator defined<br />

level<br />

■<br />

Minimum Thickness continuously<br />

captured and updated on screen<br />

920-145-EN


powerful Software and Reporting capabilities<br />

Color-coded Scan Views<br />

The EPOCH XT B<strong>SCAN</strong> software features color-coded<br />

scan views for striking visual feedback. Triggering a<br />

Minimum-Depth Alarm changes the scan color from<br />

green to red, immediately alerting the operator of a<br />

critically thin area of material. By selecting the Color<br />

Gradient view, the B-scan image gradually shifts from<br />

red to green over the full thickness of the inspection<br />

window, allowing the operator to visualize thickness<br />

variations across the entire inspection range!<br />

B-scan with Floating Gate<br />

GageView ® Pro Optional Interface Program<br />

GageView Pro is <strong>Olympus</strong> NDT’s PC interface program, compatible with<br />

all EPOCH 4 series, EPOCH XT and EPOCH LTC flaw detectors. GageView<br />

Pro allows the operator to download and review saved B-scan data from the<br />

EPOCH XT. GageView Pro features a detailed B-scan Review window that<br />

allows the user to review: full compressed B-scan, uncompressed B-scan<br />

segments, stored associated A-scan data, minimum and maximum thickness<br />

values, averages, and other important scan statistics, and much more.<br />

GageView Pro also allows the user to either print a B-scan report, including<br />

images, statistics, and calibration data, or export B-scan data to a spreadsheet<br />

program for further review and integration in other reports or databases.<br />

B-scan Print View<br />

www.olympusNDT.com<br />

Alarm Color Change<br />

Color Gradient View<br />

Includes Floating Gate Software Option<br />

Included with the B-scan software is <strong>Olympus</strong> NDT’s Floating Gate software option.<br />

This option allows the operator to “float” Gate 1 and/or Gate 2 at a selected height<br />

compared to a gated echo (-1 dB to -14 dB). This feature is intended to allow more<br />

consistent, precise readings, and is especially useful while acquiring B-scan data in<br />

Edge Detect mode.<br />

B-scan Review Window<br />

Data Export<br />

Image Export


B-Scan Options Specifications*<br />

B-Scan Standard Kit<br />

EPXT-B<strong>SCAN</strong>-KIT-XX<br />

Includes:<br />

EPXT-B<strong>SCAN</strong>: B-scan operating software<br />

for the EPOCH XT<br />

EP4/ENC: B-scan Position Encoder<br />

EPXT-ENC-CBL-XX: Position Encoder Cable<br />

with specified cable length. (Cable lengths<br />

available 10, 25, 50, and 100 feet)<br />

Notes:<br />

• Panametrics-NDT D790-SM, 5.0 MHz,<br />

0.434 in. (11.0 mm) diameter, transducer<br />

sold separately<br />

• Panametrics-NDT BCMD-316-XXB or<br />

L1CMD-316-XXB transducer cables also<br />

sold separately<br />

• Transducer cables available in 5, 10, 25,<br />

50, and 100 feet lengths<br />

• B-scan option can be added to existing<br />

EPOCH XT units remotely if the instrument<br />

is equipped with the Hardware I/O<br />

option.<br />

www.olympusNDT.com<br />

info@olympusNDT.com<br />

Optional Accessories<br />

EP4/ENC-PH: Probe holder assembly<br />

EP4/MW: One pair of magnetic hold down<br />

wheels<br />

EP4/EPB: Extension pole bracket<br />

EP4/WC: Replaceable wear cap<br />

EP4/EWC: Replaceable extended wear cap<br />

GAGEVIEWPRO-KIT-USB: Interface<br />

Program on CD-ROM, includes USB<br />

interface cable<br />

EP4/ENC-PH<br />

EP4/WC<br />

or<br />

EP4/EWC<br />

<strong>Olympus</strong> NDT<br />

48 Woerd Avenue • Waltham, MA 02453 • USA<br />

Tel.: (1) 781-419-3900 • Fax: (1) 781-419-3980<br />

12569 Gulf Freeway • Houston, TX 77034 • USA<br />

Tel.: (1) 281-922-9300 • Fax: (1) 952-487-8877<br />

<strong>Olympus</strong> NDT u.K. lTD.<br />

12 Nightingale Close • Rotherham, South Yorkshire S60 2AB • UK<br />

<strong>Olympus</strong> siNgapOre pTe. lTD.<br />

491B River Valley Road 12-01/04, Valley Point Office Tower, 248373 • Singapore<br />

<strong>Olympus</strong> ausTralia pTy. lTD.<br />

PO Box 985 • Mount Waverley, VIC 3149 • Australia<br />

EP4/MW<br />

Copyright © 2008 by <strong>Olympus</strong> NDT. All brands are trademarks or registered trademarks of their respective owners.<br />

*All specifications are subject to change without notice.<br />

Encoder Specifications<br />

Resolution: 50 pulses per inch (0.020 in.<br />

per reading) 2 pulses per mm (0.5 mm<br />

per reading)<br />

Outputs: Standard quadrature position<br />

outputs, TTL/CMOS compatible<br />

Power: Single +5 V supply provided by<br />

EPOCH XT<br />

Connector: LEMO ® , four conductor<br />

Water Fitting: Bronze barbed fitting<br />

requires 5/32 in. (4.0 mm) ID tubing<br />

Pipe Diameter: Minimum 3.0 in. (76.2 mm)<br />

Dimensions: (L x W x H)<br />

4.41 in. x 2.75 in. x 2.70 in.<br />

(111 mm x 69 mm x 68 mm)<br />

Weight: 0.93 lb (425 g)<br />

Construction: Rugged, sealed cast<br />

aluminum<br />

Note: Other commercially available<br />

encoders or crawlers can be used with the<br />

EPOCH XT B-scan option.<br />

EP4/EPB