D I A G N O S T I C S<br />
SPECIMEN<br />
Serum/plasma<br />
Collect serum using standard sampling tubes.<br />
Heparin, or EDTA-plasma<br />
Stability: 5 days at +2°C to +8°C<br />
6 moths at –20°C<br />
Urine<br />
Collect urine without using preservatives.<br />
Stability: Assay urinary uric acid as soon as possible. Do not refrigerate.<br />
Centrifuge samples containing precipitate before performing the assay.<br />
TESTING PROCEDURE<br />
Materials provided<br />
• Working solutions as described above<br />
Additional materials required<br />
• Calibrators and controls as indicated below<br />
• 0.9% NaCl<br />
Manual procedure:<br />
Wavelength: Hg 546 nm, (490 - 550)<br />
Temperature:<br />
+25 / +30 / +37°C<br />
Cuvette:<br />
1 cm light path<br />
Zero adjustment:<br />
against reagent blank one, reagent blank per series only<br />
Blank Standard Sample<br />
Calibrator --- 20 µl ---<br />
Sample --- --- 20 µl<br />
R1 1000 µl 1000 µl 1000 µl<br />
Mix, incubate for 5 min. at +37°C or 10 min at +20°C or +25°C. Read absorbance of the sample against reagent<br />
blank within 30 min.<br />
Calculation:<br />
By standard:<br />
ΔA sample<br />
x standard conc. = <strong>Uric</strong> acid conc.<br />
ΔA standard<br />
Standard concentration: 6 mg/dl (356.9 μmol/l)<br />
CALIBRATION FREQUENCY<br />
Two-point calibration is recommended<br />
• after lot change<br />
• as required following quality control procedures<br />
LIMITATIONS - INTERFERENCE<br />
Criterion: Recovery within ±10% of initial values<br />
Icterus: No significant interference up to an index I of 12 (approximate conjugated and unconjugated bilirubin<br />
concentration 12 mg/dl)<br />
Hemolysis: No significant interference up to an index H of 50 (approximate hemoglobin concentration: 50 mg/dl).<br />
Lipemia (Intralipid): Triglycerides may interfere. There is poor correlation between turbidity and triglycerides<br />
concentration.<br />
Elevated levels of ascorbic acid produce false low values.<br />
<strong>Uric</strong>ase reacts specifically with uric acid. Other purine derivatives can inhibit the uric acid reaction.<br />