03.03.2015 Views

Getting started with the confocal microscope Leica TCS SP5 Matrix ...

Getting started with the confocal microscope Leica TCS SP5 Matrix ...

Getting started with the confocal microscope Leica TCS SP5 Matrix ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

<strong>Getting</strong> <strong>started</strong> <strong>with</strong> <strong>the</strong> <strong>confocal</strong> <strong>microscope</strong> <strong>Leica</strong> <strong>TCS</strong> <strong>SP5</strong> <strong>Matrix</strong>, HCS A<br />

TABLE OF CONTENTS<br />

Switching on <strong>the</strong> <strong>microscope</strong> and <strong>the</strong> scanner………………………………………...……………………2<br />

Opening <strong>the</strong> scanning software LAS AF……………………………………………………...……….…2-3<br />

Activating <strong>the</strong> lasers……...…………………………………………………………………………………3<br />

Using <strong>the</strong> <strong>microscope</strong>…………………………………………….………………………………………4-5<br />

Field of view vs. scan area………………………………………………………………………………….5<br />

Arrangements for scanning<br />

Choosing <strong>the</strong> Beam Path Settings………….……………………………………………………………….6<br />

Settings for basic scanning……………………………………….……………………………………….6-7<br />

Start scanning……………………………………………………………………………………………….7<br />

Modifying emission detection range……………………………………………………………..…………8<br />

Setting <strong>the</strong> dynamic range………………………………………………………….……………………….9<br />

Averaging optical plane………………………………………….…………………………………………9<br />

Bidirectional scan……………………...…………………………………………………………………..10<br />

Zooming………………………………...…………………………………………………………………11<br />

Saving images………………………………………………………...…………………………………...11<br />

<strong>Leica</strong> <strong>TCS</strong> <strong>SP5</strong> shutdown procedure……………………………..……………………………………11-12<br />

Troubleshooting<br />

Nothing but blackness in <strong>microscope</strong> eyepieces…………………………………………………………....5<br />

No scanned image on screen…..……………………………………………………………………………7<br />

Page 1 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

<strong>Getting</strong> <strong>started</strong> <strong>with</strong> <strong>the</strong> HCS <strong>confocal</strong> <strong>microscope</strong> <strong>Leica</strong> <strong>TCS</strong> <strong>SP5</strong> <strong>Matrix</strong><br />

Located in <strong>the</strong> room 2209, at <strong>the</strong> Department of Biology, Åbo Akademi University, Tykistökatu 6 A, FI-20520 Turku,<br />

BioCity building, 2 nd floor.<br />

With this guide you should be able to turn on <strong>the</strong> <strong>confocal</strong> <strong>microscope</strong>, scan fixed samples, and close <strong>the</strong> system after use. In<br />

case you consider live-cell imaging, please contact responsible personnel well in advance. Note that gas- and ion-lasers prefer<br />

longer running times compared to continual switching on – switching off. It is better for <strong>the</strong> lifetime of <strong>the</strong> lasers that <strong>the</strong>y run at<br />

least one hour after switching on. Do not switch off <strong>the</strong> laser key, in case <strong>the</strong> next user has a reservation during <strong>the</strong> next three<br />

hours after you. It is your duty to check <strong>the</strong> electrical reservation book before you shut down <strong>the</strong> system.<br />

Switching on <strong>the</strong> <strong>microscope</strong> and <strong>the</strong> scanner<br />

First, tilt carefully <strong>the</strong> illumination arm of <strong>the</strong> inverted <strong>microscope</strong> backwards (fig. 1.) to prevent any collision between <strong>the</strong><br />

condenser and <strong>the</strong> moving motorized stage in initializing phase.<br />

Then, press <strong>the</strong> left green button “PC, Microscope” (fig. 2.) under <strong>the</strong> right corner of <strong>the</strong> table. You will here sound of <strong>the</strong><br />

<strong>microscope</strong> as <strong>the</strong> motorized stage starts to initialize, and <strong>the</strong> stage moves in pre-defined area.<br />

Next, instantly after <strong>the</strong> first button, press <strong>the</strong> middle green button “Scanner Power”. The scanner starts to hum.<br />

Finally, press on <strong>the</strong> right green button “Laser Power”, and turn <strong>the</strong> Laser Emission key to “On-I”-position; <strong>the</strong> yellow indicator<br />

lamp will be lit (fig. 2). At this point you have switched all <strong>the</strong> lasers only to standby-mode.<br />

Opening <strong>the</strong> scanning software LAS AF<br />

The operating system Windows XP opens automatically, after you have switched on <strong>the</strong> left green button “PC, Microscope”.<br />

Choose <strong>TCS</strong>-User, no password, in case you don’t have your own account.<br />

Double-click <strong>the</strong> icon LAS AF on <strong>the</strong> desktop. A message “Initializing LAS AF, Please wait …” appears on <strong>the</strong> left screen.<br />

In case you accidentally clicked <strong>the</strong> <strong>Leica</strong> AF Hardware Configurator, let this software open, until you shut it down again.<br />

Thereafter start <strong>the</strong> LAS AF scanning software. Both softwares can not be active concurrently.<br />

The next window to open is <strong>Leica</strong> Application Suite window (fig. 3), where you<br />

simply click <strong>the</strong> option OK. In case you want to check <strong>the</strong>m, default options<br />

Machine (not Simulator<strong>SP5</strong>) and DMI6000 (not DM-Manual 6) should be<br />

selected in Configuration and Microscope Stand options, respectively.<br />

Page 2 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

The next window “Microscope Stand” opens in less than a minute (fig. 4).<br />

Click “Yes”, in case you will use <strong>the</strong> LAS AF MATRIX M3 application, or<br />

<strong>the</strong> Tile Scan option. Click “No”, if you’re going to use conventional<br />

scanning (like in most cases).<br />

The scanning software LAS AF v.2.5.1, build 6757 opens <strong>with</strong> scanning<br />

options on <strong>the</strong> left screen, and <strong>the</strong> scanned image to be seen on <strong>the</strong> right<br />

screen. The Control Panel <strong>with</strong> six rotating knobs is behind <strong>the</strong> keyboard<br />

(fig. 5).<br />

Activating <strong>the</strong> lasers<br />

The first step is to activate <strong>the</strong> laser(s). Click “Configuration”-tab (fig. 13), and click “Laser” in Hardware Configuration<br />

window (fig. 14). The view <strong>with</strong> currently available lasers opens. Fill in checkboxes of <strong>the</strong> laser(s) to be used: Ar, HeNe 543 or<br />

HeNe 633. In case of Ar-laser, you will here <strong>the</strong> air-cooling start after you filled <strong>the</strong> checkbox. Raise <strong>the</strong> excitation power to<br />

20% <strong>with</strong> <strong>the</strong> slider. Return to scanning options by clicking <strong>the</strong> “Acquire”-tab (fig. 13).<br />

Page 3 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Using <strong>the</strong> <strong>microscope</strong><br />

The <strong>microscope</strong> stand is an inverted <strong>Leica</strong> DMI6000B. Place a <strong>microscope</strong> slide on <strong>the</strong> specimen stage (fig. 6.). Here is assumed<br />

that you are using a fixed sample. Place <strong>the</strong> slide on <strong>the</strong> shallow notch, cover glass facing down. In case you’re using an oil<br />

immersion objective, add a small drop of oil on <strong>the</strong> cover glass, and turn <strong>the</strong> specimen round before placing it over <strong>the</strong> objective.<br />

In case you used gloves when handling immersion oil, you are not allowed to use same gloves for operating <strong>the</strong> <strong>microscope</strong> or<br />

using <strong>the</strong> keyboard. Use <strong>the</strong> specimen holders to better keep <strong>the</strong> sample in place. Turn carefully <strong>the</strong> illuminator arm of <strong>the</strong><br />

<strong>microscope</strong> (fig. 1) back to place.<br />

ATTENTION! Read this revised paragraph thoroughly.<br />

Switch on <strong>the</strong> external light source (fig. 7).It stands on <strong>the</strong> table between <strong>the</strong> <strong>microscope</strong><br />

and <strong>the</strong> left screen. When you switch it on, both green (Power) and yellow (Shutter)<br />

indicator lamps will lit. Please, do not touch <strong>the</strong> intensity knob.<br />

NOTE! The lamp must be switched on at least one hour from ignition. It must cool down<br />

at least 20 minutes from switch off. Do not switch <strong>the</strong> lamp off after your session in case<br />

<strong>the</strong>re are o<strong>the</strong>r reservations after you.<br />

When scanning a time-lapse series lasting hours, switch <strong>the</strong> lamp off, because even <strong>the</strong><br />

transmitted image is scanned <strong>with</strong> laser excitation.<br />

Caution! The cover of <strong>the</strong> lamp housing gets warm or even hot during prolonged use.<br />

Never place anything on top of it!<br />

The light source itself is a metal halide bulb.<br />

Choose <strong>the</strong> light microscopy method to use. Choose fluorescent cube I3 or N2.1 for fluorescence microscopy of “green” and<br />

“red” samples, respectively. The cube I3 has a band-pass filter of 450-490nm, a dichromatic mirror of value 510, and a long-pass<br />

suppression filter of 515nm. For <strong>the</strong> cube N2.1 <strong>the</strong>se values are 515-560nm, 580, and 590nm. The buttons for selecting cubes are<br />

facing you above <strong>the</strong> LCD, on <strong>the</strong> front of <strong>the</strong> <strong>microscope</strong> (fig. 8).<br />

Page 4 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

In case you need transmitted light, press <strong>the</strong> lowest button “TL/IL” on <strong>the</strong> left side of <strong>microscope</strong> stand (fig. 9). To open <strong>the</strong> light<br />

path completely you must also press <strong>the</strong> button “Shutter” in front of <strong>the</strong> <strong>microscope</strong>, (fig. 10), in case you chose a fluorescent<br />

mode.<br />

Focus <strong>the</strong> image in <strong>microscope</strong> <strong>with</strong> motorized and manual fine focusing (fig. 11). Z ↑ raises <strong>the</strong> objective towards <strong>the</strong> specimen,<br />

and should stop to 0-position, when set right. Z ↓ lowers <strong>the</strong> objective. In case you want to reset <strong>the</strong> 0-position where <strong>the</strong><br />

motorized upwards focusing stops, find <strong>the</strong> focus, and press simultaneously <strong>the</strong> middle (Set) and <strong>the</strong> top (Z ↑) buttons more than<br />

a second, and repeat this procedure. Use <strong>the</strong> hand<br />

wheel for manual fine focusing.<br />

For moving <strong>the</strong> stage in X- and Y-directions, use<br />

<strong>the</strong> SmartMove remote control module (fig. 12).<br />

The preciseness (fastness) of movement can also be<br />

defined to ei<strong>the</strong>r “fast” or “precise” by pushing<br />

ei<strong>the</strong>r one of <strong>the</strong> black buttons on <strong>the</strong> left side. For<br />

focusing, z-movement, you can use ei<strong>the</strong>r <strong>the</strong><br />

focusing knob (z) of <strong>the</strong> SmartMove device, or <strong>the</strong><br />

manual fine focusing hand wheel of <strong>the</strong><br />

<strong>microscope</strong>.You can choose between Z-fine and Z-<br />

coarse <strong>with</strong> <strong>the</strong> single top right button of <strong>the</strong> SmartMove module, and <strong>the</strong> chosen option is visible on <strong>the</strong> right low corner of <strong>the</strong><br />

LCD of <strong>the</strong> <strong>microscope</strong>.<br />

Nothing but blackness in <strong>microscope</strong> eyepieces – troubleshooting:<br />

-is <strong>the</strong> external light source for fluorescence mode (fig. 7) switched on?<br />

-do you have <strong>the</strong> right filtercube (fig. 8) in use for your fluorescent sample?<br />

-is <strong>the</strong> shutter (fig. 10) on in case of fluorescence mode?<br />

-have you pressed <strong>the</strong> TL/IL –button (fig. 9) in case you need transmitted light?<br />

-is <strong>the</strong> sample in focus?<br />

-is <strong>the</strong> sample facing cover glass down, and sits nicely in <strong>the</strong> shallow groove?<br />

Field of view vs. scan area<br />

The scanned area is always a smaller square or rectangular in <strong>the</strong> middle of <strong>the</strong> field of view (table 1.).<br />

Table 1. Field of view and maximal scanning area <strong>with</strong> available objectives in <strong>TCS</strong> <strong>SP5</strong>.<br />

10x dry 20x dry 40x oil 63x oil<br />

ø 2.50 mm ø 1.25 mm ø 0.625 mm ø 0.397 mm<br />

1.55 mm x 1.55 mm 775 µm x 775 µm 390 µm x 390 µm 250 µm x 250 µm<br />

Find a presentable area in your sample and centre it in <strong>the</strong> field of view. This is, however, not critical as <strong>the</strong> sample can be<br />

moved also while scanning. Note! Do not choose <strong>the</strong> best area of <strong>the</strong> sample first, because some preparing scanning is required<br />

to achieve suitable scanning conditions for <strong>the</strong> first scan. In this way <strong>the</strong> best area will not get any “pre-bleaching” before <strong>the</strong><br />

actual scanning.<br />

Page 5 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Arrangements for scanning<br />

Choosing <strong>the</strong> Beam Path Settings<br />

In <strong>the</strong> “Beam Path Window” you modify <strong>the</strong> light path suitable for your sample,<br />

from excitation laser all <strong>the</strong> way to <strong>the</strong> detector. Choose <strong>the</strong> correct dye<br />

(identical or nearest match) from a list of “<strong>Leica</strong> Settings”, in “Load/Save single<br />

setting” window (fig. 15). Here we prepare settings for Alexa 488 dye.<br />

Therefore, choose FITC (<strong>the</strong> nearest match) in <strong>the</strong> list of dyes (fig. 16). After<br />

clicking FITC, <strong>the</strong> excitation line 488 is activated <strong>with</strong> certain percentage of<br />

laser output, certain beam splitter is chosen, as well as certain emission<br />

detection range is chosen. These are pre-defined manufacturer’s settings, which<br />

can be fur<strong>the</strong>r modified, but are good to start <strong>with</strong>.<br />

It is advisable to choose first some saved (factory or user) setting to start <strong>with</strong>.<br />

Settings for basic scanning<br />

In <strong>the</strong> Acquisition-window on <strong>the</strong> left side of <strong>the</strong> screen are <strong>the</strong> most settings for<br />

scanning. Small, one row windows can be opened by clicking <strong>the</strong> triangle in right end<br />

(fig. 17).<br />

In <strong>the</strong> top window “Acquisition Mode”, <strong>the</strong> acquisition mode can be changed from xyz<br />

to some o<strong>the</strong>r, if needed. The defined mode is xyz, when <strong>the</strong> scanning software is<br />

opened.<br />

Scan format, speed and area can be chosen in <strong>the</strong> next window. Predefined settings are<br />

format 512 pixels x 512 pixels <strong>with</strong> scanning speed 400 Hz and maximal scanning area<br />

(i.e. zoom factor 1, <strong>with</strong>out any zoom in). Note - <strong>the</strong> maximal scanning area gets<br />

smaller when you change to larger magnification objective (table 1.).<br />

Image size and pixel size <strong>with</strong> current settings are also visible here.<br />

Page 6 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

There are 13 different predefined scan formats, most of <strong>the</strong>m square, but some also rectangular (fig. 18). In case you choose <strong>the</strong> option<br />

“More”, a new window opens for setting <strong>the</strong> x- and y-format <strong>with</strong> sliders. The single file sizes become quickly fairly large in formats<br />

larger than 1024 pixels x 1024 pixels, where <strong>the</strong> file size is roughly 1Mb.<br />

There are seven predefined scan speed options ranging from 10 Hz to 1400 Hz (fig. 19). In option “More”, you can type <strong>the</strong><br />

wanted speed between those values. Remember, that scan speed 600 Hz is <strong>the</strong> fastest speed, which still can use <strong>the</strong> maximal<br />

scanning area. In case you choose speed 700 Hz, <strong>the</strong> zoom factor will grow to 1.4, and <strong>the</strong> scanning area will be somewhat<br />

smaller. With maximum speed 1400 Hz <strong>the</strong> zoom factor will be 6.0, when scanning area is clearly smaller than maximal scan<br />

area. This is normal, technical “speed limit”, which prevents overheating of critical parts of <strong>the</strong> scanner.<br />

Start scanning<br />

Click <strong>the</strong> “Live” button on <strong>the</strong> lower left corner of <strong>the</strong> left screen. You will hear mechanical sounds from <strong>the</strong> <strong>microscope</strong> stand,<br />

as mode is changed from viewing to scanning. You will also here a “click” sound from behind <strong>the</strong> table, where <strong>the</strong> system<br />

interlock opens to let laser beam through <strong>the</strong> scanhead into <strong>the</strong> sample. The scanned image also appears on <strong>the</strong> right screen.<br />

No scanned image on screen – troubleshooting:<br />

-are <strong>the</strong> correct excitation lasers activated (figs 13 & 14)?<br />

-have you chosen <strong>the</strong> correct dye in “Load/Save single setting” window (figs 15 & 16)?<br />

-is <strong>the</strong> sample in focus (fig. 11)?<br />

Use <strong>the</strong> rightmost knob of <strong>the</strong> Control Panel (fig. 5) to move in z direction, and <strong>the</strong> SmartMove remote control module (fig. 12)<br />

to move in x and y directions. In scanning it is advisable to use <strong>the</strong> panelbox knob ra<strong>the</strong>r than SmarMove module for z direction,<br />

as it is far more precise. The first will move <strong>the</strong> specimen stage, but <strong>the</strong> latter will move <strong>the</strong> objective.<br />

Now you have a scanned image of your sample on <strong>the</strong> screen, and <strong>the</strong> scanning continues as long as you click <strong>the</strong> stop button.<br />

This is where <strong>the</strong> work only starts. Next step is to set optimal laser excitation power and emission detection range for <strong>the</strong> sample.<br />

Preferred laser excitation power is such, that specific staining is bright, but not saturated. In a saturated image area you will not<br />

be able to detect any details, but it is homogenously all too bright. Fur<strong>the</strong>rmore, in most cases, it is not a bad idea to show also<br />

some unspecific areas in <strong>the</strong> image.<br />

Page 7 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Note! The image on <strong>the</strong> screen is rotated 90º to left in contrast to view in <strong>microscope</strong>. Type <strong>the</strong> value 90 in <strong>the</strong> numerical<br />

window on <strong>the</strong> right side of <strong>the</strong> slider “Rotation” (fig. 17), in case you want to have <strong>the</strong> image orientation on <strong>the</strong> screen to be<br />

identical <strong>with</strong> <strong>microscope</strong> view.<br />

Modifying emission detection range<br />

To collect effectively <strong>the</strong> specific emission signal to detectors, you can modify <strong>the</strong> emission detecting range, in case you are not<br />

satisfied <strong>with</strong> predefined settings. For our fixed sample stained <strong>with</strong> Alexa 488, we first select <strong>Leica</strong> / Alexa 488 dye in a list<br />

under <strong>the</strong> active detector, in this case PMT 1, i.e. Photomultiplier tube (fig. 20). An emission curve for <strong>the</strong> chosen dye appears<br />

(fig. 21). Note! This emission curve does not affect scanning in any way, as scanning can be also performed <strong>with</strong>out it, when<br />

“None” is chosen in <strong>the</strong> list. It is intended simply to help you in setting <strong>the</strong> emission detection ranges, especially <strong>with</strong> two or<br />

more active channels in use simultaneously. The bar representing emission collecting range for <strong>the</strong> active detector is connected<br />

to it <strong>with</strong> a ”shade”.<br />

You can affect <strong>the</strong> emission detection range bar while scanning, and see immediately possible change in scanned image on <strong>the</strong><br />

screen. The emission bar can be broadened or narrowed in both ends (fig. 22), or moved <strong>with</strong>out affecting <strong>the</strong> broadness by<br />

dragging it in <strong>the</strong> middle (fig. 23). Fur<strong>the</strong>rmore, emission range can be set numerically, after you double click <strong>the</strong> emission bar<br />

(fig. 24).<br />

In case you have a bright staining, try also to reduce emission range in <strong>the</strong> right end to fit under <strong>the</strong> top of <strong>the</strong> emission curve,<br />

like in figure 21. This also reduces unspecific staining to be collected in detector, and thus betters visibility of <strong>the</strong> specific signal<br />

in <strong>the</strong> final image. You can choose <strong>the</strong> colour of <strong>the</strong> final images by clicking <strong>the</strong> colour bar next to word PMT (figs. 20 & 21),<br />

which opens a new window. Remember, however, that <strong>confocal</strong> images are always pseudo-coloured!<br />

Page 8 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Setting <strong>the</strong> dynamic range<br />

To use <strong>the</strong> whole dynamic range of <strong>the</strong> detector (PMT), click<br />

<strong>the</strong> Quick LUT-button on <strong>the</strong> left side of <strong>the</strong> scanned image in<br />

<strong>the</strong> right screen (fig. 25). There are six rotating knobs on <strong>the</strong><br />

Control Panel (fig. 5). The first in <strong>the</strong> left is for setting <strong>the</strong><br />

“brightness” of <strong>the</strong> image, or voltage of <strong>the</strong> detector. The<br />

second is for setting <strong>the</strong> offset value, to acquire, for example,<br />

nicely black background for areas <strong>with</strong>out any sample in<br />

scanned area.<br />

Now, scan <strong>the</strong> sample <strong>with</strong> Quick LUT-button activated.<br />

Rotate <strong>the</strong> first knob in left to get brighter or dimmer image.<br />

The intended value is such, that <strong>the</strong>re is little blue to be seen<br />

somewhere in <strong>the</strong> image (fig. 25). The blue pixels are<br />

saturated, and represent <strong>the</strong> maximal value 255 in a digitalized<br />

8-bit image. Note! While scanning, move also in z-direction<br />

<strong>with</strong> <strong>the</strong> first knob in right to reveal any brighter areas in differing depths of <strong>the</strong> sample.<br />

Continue scanning, and rotate <strong>the</strong> second, offset knob. In case you have a totally orange-brown image, rotate <strong>the</strong> knob counter<br />

clockwise until <strong>the</strong> image turns to green. Green colour is black pixels, representing zero values in digitalized 8-bit image. The<br />

intended value is such, that <strong>the</strong>re are some green pixels somewhere in <strong>the</strong> image (fig. 25). Quite often, however, <strong>the</strong> green pixels<br />

cover large area of <strong>the</strong> image, and what is more, <strong>the</strong> limiting value between no green and a green cover can be hard to detect.<br />

Eventually you have both blue and green pixels in scanned image (fig. 25). Click once <strong>the</strong> Quick LUT-button, and <strong>the</strong> image<br />

turns to black-and-white. Click once more, and <strong>the</strong> chosen scan colours appear on <strong>the</strong> screen.<br />

Always use as low laser power as possible. When setting <strong>the</strong> dynamic range, try also reducing laser excitation power when<br />

setting <strong>the</strong> voltage of <strong>the</strong> detector. Usually, <strong>the</strong> best values for PMT voltage are between 800 and 900, which means quite<br />

reduced laser power <strong>with</strong> bright samples. Use <strong>the</strong> slider of <strong>the</strong> specific laser excitation line in top right corner of <strong>the</strong> left screen to<br />

control laser power. The optimal brightness of <strong>the</strong> image is set <strong>with</strong> both laser excitation power and PMT voltage. Nei<strong>the</strong>r should<br />

ever be in, or even near maximal setting!<br />

Averaging an optical plane<br />

Try also averaging function to get better image/noise-ratio in blacker background and<br />

brighter specific staining. In practice averaging means how many times a single<br />

optical plane is scanned before it will be <strong>the</strong> final image. Scanning <strong>the</strong> same optical<br />

plane twice, for example, reduces some 25% of unspecific brightness and noise. This<br />

is worth using especially when <strong>the</strong>re is clear “stardust” in background. In best case<br />

background becomes black after averaging enough. Frame averaging is <strong>the</strong> mode for<br />

fixed samples, line averaging for live cell imaging. Choose a value for “Frame<br />

Average” (figs. 19 & 26), and click “Capture Image” scan option on lower right<br />

corner. Chosen number of scans will be performed on <strong>the</strong> one optical plane visible on<br />

<strong>the</strong> screen. Choose a different value to try ano<strong>the</strong>r amount of averaging, and compare<br />

results of <strong>the</strong> scanned images in “Experiments” to find out <strong>the</strong> best option.<br />

Page 9 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Bidirectional scan<br />

30.<br />

You can also use “Bidirectional X” option to double scan speed. Usually it does not<br />

deteriorate image quality <strong>with</strong>out fur<strong>the</strong>r modifications, but always check quality when<br />

changing to bidirectional scan mode. When bidirectional option is activated, also a “Phase<br />

correction X” slider is visible (fig. 26). It is a bit too rough tool for phase setting, though.<br />

So, first click words Control Panel (fig. 27), and a USB Control Panel window will open<br />

(fig. 28). With this tool you can change <strong>the</strong> status of each rotating knob in <strong>the</strong> Control Panel<br />

(fig. 5). You need to change function of one of <strong>the</strong> knobs to “Phase”, for setting <strong>the</strong> optimal<br />

phase for bidirectional scan. The Pinhole knob is converted to Phase knob in figures 29 &<br />

Start scan by clicking “Live”, and move in z direction in your sample to choose a bright image to be scanned. Click a few times<br />

<strong>the</strong> magnifying glass (+) on <strong>the</strong> left side of <strong>the</strong> image on <strong>the</strong> screen, and check shape of a contrastingly bright area. Turn <strong>the</strong><br />

phase knob to see modification on <strong>the</strong> screen. The bright edge will turn from even to more saw edged <strong>the</strong> fur<strong>the</strong>r you go from <strong>the</strong><br />

optimal phase, and finally even forms double image. Try phase setting also to <strong>the</strong> o<strong>the</strong>r direction to find <strong>the</strong> best position, which<br />

produces <strong>the</strong> most even image. Click 1:1, and continue scan procedure.<br />

Page 10 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

Zooming<br />

Scanning can be concentrated on an interesting, smaller area of <strong>the</strong> image even on <strong>the</strong> fly. Just tick <strong>the</strong> “Zoom in” box (fig. 26),<br />

and draw a square on <strong>the</strong> image screen. The scan area and scan information will immediately be updated on <strong>the</strong> screens.<br />

Although a <strong>confocal</strong> <strong>microscope</strong> is an excellent optical instrument, <strong>the</strong> limits of resolution must be taken into account. In<br />

practice this means using no smaller pixel size than some 100 nm, even <strong>with</strong> a high-resolution, N.A. 1.40 objective. Zooming<br />

above this limit will only get “larger view”, but <strong>the</strong> resolution will remain <strong>the</strong> same, and image becomes <strong>the</strong> fuzzier <strong>the</strong> more you<br />

zoom.<br />

Saving images<br />

LAS AF scanning software creates a folder named “Experiment” where <strong>the</strong> scanned images are saved. This is only a buffer<br />

memory, though, and you have to rename and save <strong>the</strong> images to defined storage yourself. In case you use “save” command, <strong>the</strong><br />

image format will be dot-lif. In case you export <strong>the</strong> images, you can use o<strong>the</strong>r formats, such as TIFF, JPEG or AVI.<br />

It is advisable to rename scanned images right away, as well as delete needless images. In that way you can save or export <strong>the</strong><br />

whole Experiment folder whenever.<br />

Experiment folders can be found under Experiments-tab, but <strong>the</strong> Acquisition-tab is in use while scanning.<br />

Known bugs<br />

The beam splitter 488/543/633 is chosen in factory defined Beam Path Setting for FITC. However, it is better to use <strong>the</strong> R<strong>SP5</strong>00<br />

beam splitter for single green emission dye.<br />

LEICA <strong>TCS</strong> <strong>SP5</strong> - SHUTDOWN PROCEDURE<br />

After your scanning session always log your reservation as used in <strong>the</strong> Reservation Book, for example through <strong>the</strong> link that was<br />

e-mailed to you in context your reservation. It is a four-step process where <strong>the</strong> submit command is showed on <strong>the</strong> last page. Do<br />

not forget to inform on any technical problems or malfunction of <strong>the</strong> system.<br />

Make sure you have transferred all your important images to external memory device from <strong>the</strong> buffer memory.<br />

Check <strong>the</strong> Reservation Book for any reservations after your session.<br />

In case <strong>the</strong>re is a reservation during <strong>the</strong> next three hours:<br />

I) Configuration Hardware Configuration Laser (figs 13 & 14)<br />

Decrease Ar-laser power to 0% <strong>with</strong> <strong>the</strong> slider. Remove ticks in <strong>the</strong> boxes of lasers in use:<br />

II)<br />

III)<br />

Exit <strong>the</strong> scanning software LAS AF: click <strong>the</strong> cross on top right corner of <strong>the</strong> left screen. A warning message for<br />

unsaved images appears on <strong>the</strong> screen, in case <strong>the</strong>re are any.<br />

Log off Windows WP.<br />

Leave all o<strong>the</strong>r instrumentation running. This way <strong>the</strong> computer is in XP-user accounts view, and hardware ready to be used.<br />

In case you are <strong>the</strong> last user of <strong>the</strong> day:<br />

1) Switch off <strong>the</strong> external light source (fig. 7.).<br />

2) Configuration Hardware Configuration Laser (figs 13 & 14)<br />

Decrease Ar-laser power to 0% <strong>with</strong> <strong>the</strong> slider. Remove ticks in <strong>the</strong> boxes of lasers in use:<br />

Page 11 of 12


<strong>Getting</strong> <strong>started</strong> v. 1.7<br />

September 12th, 2011<br />

3) Turn <strong>the</strong> Laser Emission key to “Off-0”-position (fig. 2.). Yellow indicator light goes out. NOTE! Leave <strong>the</strong> right<br />

green button “Laser Power” on until <strong>the</strong> air cooling of <strong>the</strong> Ar-laser automatically goes out after a few minutes. You will<br />

hear <strong>the</strong> difference, once it is silent. In any case, five minutes is long enough.<br />

4) Exit <strong>the</strong> scanning software LAS AF: click <strong>the</strong> cross on top right corner of <strong>the</strong> left screen.<br />

A warning message for unsaved images appears on <strong>the</strong> screen, in case <strong>the</strong>re are any.<br />

5) First log off Windows XP!<br />

6) Thereafter turn off <strong>the</strong> computer.<br />

7) When <strong>the</strong> screens become black, press off <strong>the</strong> middle green button “Scanner Power” (fig. 2). The scanner becomes<br />

quiet.<br />

8) Press off <strong>the</strong> left green button “PC, Microscope” (fig. 2).<br />

9) Remove <strong>the</strong> sample. In case you used oil immersion objective(s), clean <strong>the</strong> lense(s). First wipe away <strong>the</strong> remaining oil<br />

from <strong>the</strong> lens <strong>with</strong> a dry lens tissue. Take a new lens tissue, dampen it <strong>with</strong> isopropanol (bottle on <strong>the</strong> table), and wipe<br />

<strong>the</strong> lens again. Use lens tissue only for cleaning objective lenses.<br />

What is more, clean all o<strong>the</strong>r stains of oil or cell medium etc. on <strong>microscope</strong>, tables and everywhere. Use <strong>the</strong> roll of<br />

Katrin tissue for <strong>the</strong>se cleaning purposes. Always keep <strong>the</strong> <strong>microscope</strong> facility clean!<br />

Fig. 31. A cleaned oil-immersion objective lens after use.<br />

Fig. 32. Partially dried oil on immersion objective lens<br />

some 48 hours after use.<br />

10) Press off <strong>the</strong> right green button “Laser Power” before you leave <strong>the</strong> room, in case some five minutes have passed since<br />

you turned <strong>the</strong> Laser Emission key to OFF-0.<br />

11) Place <strong>the</strong> Zeiss eyepiece cover on.<br />

12) Place <strong>the</strong> “Please take care on <strong>microscope</strong>s” note on <strong>the</strong> specimen stage. Make sure <strong>the</strong> room is as tidy as it was when<br />

you entered. Don’t leave any of your belongings lying around on <strong>the</strong> tables. Sharp waste into special waste dishes, vinyl<br />

gloves to <strong>the</strong> bucket, and o<strong>the</strong>r waste to trash bin.<br />

Fig. 33 Not a pretty sight.<br />

Specimen stage smeared <strong>with</strong> cell medium and oil immersion objective lens<br />

not cleaned.<br />

Note that continuous carelessness will lead to sanctions in usage.<br />

Page 12 of 12

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!