Y-<strong>CLICK</strong><br />
Y Connector for PTCA Application<br />
Introduction<br />
<strong>Elcam</strong>’s new Y-Click connector provides an easy-to-use hemostatic<br />
device for both diagnostic and interventional procedures.<br />
Mechanism<br />
Open position<br />
2 clicks in<br />
The Y-Click consists of two channels; a straight<br />
primary channel which contains the silicone<br />
rubber sealing gasket that opens to accept<br />
the catheter or guide wire, and an angled<br />
secondary channel that attaches to a manifold<br />
for pressure monitoring, contrast injection<br />
and flushing or to automatic contrast media<br />
injectors. The Y-Click prevents back-bleeding<br />
while the balloon catheter / guide wire is inserted<br />
into the guiding catheter, and provides easy manipulation<br />
during placement while securing catheter’s position.<br />
Revolutionary New Design<br />
The ergonomically designed Y-Click significantly reduces the blood loss during<br />
interventional procedures using a three stage valve: open, semi-open and closed positions.<br />
With the easy-to-use lever you can use Y-Click in the following positions:<br />
• Open: For insertion of guide wire / catheter<br />
• Semi-open: Allows for maneuverability with minimal friction while preventing back bleeding<br />
• Closed: Seals around the catheter or wire prevents blood loss still allowing maneuverability of guide<br />
wires and small diameter catheters<br />
Lever holds the product at the open position<br />
Semi-Open position<br />
Closed position<br />
Single click<br />
Provides Convenience<br />
at device insertion and<br />
removal stage<br />
Used when you want to manipulate the catheter with<br />
no blood loss<br />
• Easier Catheter maneuverability<br />
• No blood loss<br />
Single click to close<br />
Press Lever - the product returns<br />
to closed position<br />
Unique Advantages<br />
Self closing mechanism<br />
During injection of contrast media, if the device is in semi-open<br />
position, its mechanism will shut itself before leakage occurs<br />
through the proximal side. The device, then, will move to<br />
‘closed’ position.<br />
Insertion of catheters<br />
1<br />
Push cup for 2 clicks and<br />
leave in open position<br />
2<br />
Insert Catheter<br />
One-Handed Operation<br />
Compared to current devices on the market which<br />
use either a Tuohy-Borst valve or simple open/close<br />
mechanism, <strong>Elcam</strong>’s Y-Click allows the operator total<br />
control over the manipulation of catheters and wires<br />
without concern of bleedback.<br />
The Y-Click utilizes a one hand operation both for opening<br />
the device by pushing the valve or closing the device by<br />
pressing on the lever with one’s thumb.<br />
3<br />
Single click lever to lock