Y-CLICK - Elcam Medical
Y-CLICK - Elcam Medical
Y-CLICK - Elcam Medical
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Y-<strong>CLICK</strong><br />
Y Connector for PTCA Application<br />
Product Specifications<br />
Can be used with automatic contrast media injectors employing a flow rate of up to 15cc per<br />
second and setting pressure of up to 600 psi.<br />
Compatible with any angiography device that goes through 9fr or<br />
smaller guide catheters.<br />
Compatible with 0.014” – 0.038”<br />
diameter guide wires.<br />
Male luer is designed according<br />
to ISO 594-1 & 2.<br />
Regulatory information<br />
Quality system requirements:<br />
<strong>Elcam</strong> <strong>Medical</strong> manufactures in a clean<br />
room environment (Class 100.000, according to<br />
federal standard 209E and Class 8, according to ISO<br />
14644).<br />
U.S. regulations:<br />
Production, warehousing, documentation and traceability meet Quality Systems<br />
Regulation, Title 21 Code of Federal Regulations Part 820.<br />
International standards:<br />
<strong>Elcam</strong>’s Quality Management System is ISO<br />
13485 (2003) and ISO 9001 (2000) certified.<br />
Biocompatibility requirements:<br />
All “Y-Click” components meet biological<br />
requirements according to ISO 10993-1,<br />
under ‘blood contact, indirect, limited contact<br />
duration’.<br />
Regulatory Status:<br />
510(k) Cleared (#K060759)<br />
CE Mark approved<br />
<strong>Elcam</strong> is a Leading OEM Provider of High<br />
Quality <strong>Medical</strong> Devices<br />
<strong>Elcam</strong>, the leading OEM producer of stopcocks and manifolds, is also an<br />
innovator with a number of technologically advanced devices for interventional<br />
cardiology, radiology, anesthesia, and critical care. <strong>Elcam</strong> provides solutions<br />
for over 300 companies throughout the world.<br />
Y-<strong>CLICK</strong><br />
Y C o n n e c t o r f o r P T C A A p p l i c a t i o n<br />
Head Office / Manufacturer<br />
<strong>Elcam</strong> <strong>Medical</strong><br />
BarAm 13860<br />
Israel<br />
Tel: 972-4-698-8120/1/2<br />
Fax: 972-4-698-0777<br />
sales@elcam.co.il<br />
www.elcam-medical.com<br />
USA<br />
<strong>Elcam</strong> <strong>Medical</strong> INC.<br />
2 University Plaza, Suite 620<br />
07601 Hackensack, NJ<br />
Tel: +1 201 4571120<br />
Fax: +1 201 4571125<br />
info@elcam-medical.com<br />
www.elcam-medical.com<br />
Europe<br />
MedNet GmbH<br />
Borkstrasse 10<br />
48163 Muenster Germany<br />
Tel: +49 (0) 251 32266 0<br />
Fax: +49 (0) 251 32266 22<br />
info@medneteurope.com<br />
www.medneteurope.com<br />
China<br />
<strong>Elcam</strong> China Office<br />
Room 25B, Building B, No. 1118 of Changshou Rd.,<br />
Shanghai 200042, China<br />
<br />
Tel: +86 2152375561<br />
Fax: +86 2152375562<br />
johnliu02@126.com<br />
0483<br />
ISO 13485<br />
ISO 9001<br />
REV-2, 01/2009<br />
www.elcam-medical.com
Y-<strong>CLICK</strong><br />
Y Connector for PTCA Application<br />
Introduction<br />
<strong>Elcam</strong>’s new Y-Click connector provides an easy-to-use hemostatic<br />
device for both diagnostic and interventional procedures.<br />
Mechanism<br />
Open position<br />
2 clicks in<br />
The Y-Click consists of two channels; a straight<br />
primary channel which contains the silicone<br />
rubber sealing gasket that opens to accept<br />
the catheter or guide wire, and an angled<br />
secondary channel that attaches to a manifold<br />
for pressure monitoring, contrast injection<br />
and flushing or to automatic contrast media<br />
injectors. The Y-Click prevents back-bleeding<br />
while the balloon catheter / guide wire is inserted<br />
into the guiding catheter, and provides easy manipulation<br />
during placement while securing catheter’s position.<br />
Revolutionary New Design<br />
The ergonomically designed Y-Click significantly reduces the blood loss during<br />
interventional procedures using a three stage valve: open, semi-open and closed positions.<br />
With the easy-to-use lever you can use Y-Click in the following positions:<br />
• Open: For insertion of guide wire / catheter<br />
• Semi-open: Allows for maneuverability with minimal friction while preventing back bleeding<br />
• Closed: Seals around the catheter or wire prevents blood loss still allowing maneuverability of guide<br />
wires and small diameter catheters<br />
Lever holds the product at the open position<br />
Semi-Open position<br />
Closed position<br />
Single click<br />
Provides Convenience<br />
at device insertion and<br />
removal stage<br />
Used when you want to manipulate the catheter with<br />
no blood loss<br />
• Easier Catheter maneuverability<br />
• No blood loss<br />
Single click to close<br />
Press Lever - the product returns<br />
to closed position<br />
Unique Advantages<br />
Self closing mechanism<br />
During injection of contrast media, if the device is in semi-open<br />
position, its mechanism will shut itself before leakage occurs<br />
through the proximal side. The device, then, will move to<br />
‘closed’ position.<br />
Insertion of catheters<br />
1<br />
Push cup for 2 clicks and<br />
leave in open position<br />
2<br />
Insert Catheter<br />
One-Handed Operation<br />
Compared to current devices on the market which<br />
use either a Tuohy-Borst valve or simple open/close<br />
mechanism, <strong>Elcam</strong>’s Y-Click allows the operator total<br />
control over the manipulation of catheters and wires<br />
without concern of bleedback.<br />
The Y-Click utilizes a one hand operation both for opening<br />
the device by pushing the valve or closing the device by<br />
pressing on the lever with one’s thumb.<br />
3<br />
Single click lever to lock