<strong>Automatic</strong> <strong>Optimization</strong> <strong>Tutorial</strong><br />
Table 1-7 Product Ion Auto Selection Criteria Dialog Parameters (Continued)<br />
Parameter Value Description<br />
Exclude Product<br />
Ions within ± (Da of<br />
Precursor Ion m/z)<br />
Min. Mass for<br />
Product Ion (amu)<br />
Threshold for<br />
Product Ion (cps)<br />
20.000 The Da value that defines the exclusion window<br />
around the precursor ion so that fragment ions<br />
that fall within this window are not selected for<br />
MRM optimization. For example, if you type ± 5<br />
Da for a precursor ion of 500 m/z, any fragment<br />
ions within the 495 to 505 m/z region are<br />
excluded. This prevents the precursor ion from<br />
being optimized.<br />
60.000 The lowest fragment mass to be considered for<br />
optimization. Use this option to narrow or widen<br />
the window of fragment ions to be considered<br />
from the precursor mass.<br />
100.000 Minimum number of counts for a product ion to<br />
be considered.<br />
7. Click OK to save the changes to the selection criteria.<br />
8. Click Next.<br />
9. In the Target Components dialog, type the parameters from the following table.<br />
Note: The compound name must be unique for each compound or<br />
transition.<br />
Table 1-8 Target Compounds Dialog Parameters<br />
Target Compound Field Value*<br />
Reserpine Compound Name Reserpine<br />
MW (Da) 609.4<br />
No. Charges 1<br />
Minoxidil Compound Name Minoxidil<br />
MW (Da) 210.2<br />
No. Charges 1<br />
Tolbutamide Compound Name Tolbutamide<br />
MW (Da) 271.3<br />
No. Charges 1<br />
Rescinnamine (IS) Compound Name Rescinnamine<br />
MW (Da) 635.4<br />
No. Charges 1<br />
* You must type the exact ion mass.<br />
10. Click Finish to begin the optimization process.<br />
Analyst ® <strong>1.6.1</strong> <strong>Software</strong><br />
10 of 16<br />
<strong>Automatic</strong> <strong>Optimization</strong> <strong>Tutorial</strong><br />
Release Date: March 2012