Instruction Manual - Days Healthcare
Instruction Manual - Days Healthcare
Instruction Manual - Days Healthcare
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Spare parts list and additional documentation is available from the supplier<br />
or specialty shop at your request.<br />
All repairs must be carried out by an authorized specialty shop. Please<br />
contact the supplier for a list of authorized repair centers nearest you.<br />
6. Troubleshooting<br />
In the event that your<br />
is not functioning properly, please refer to<br />
the following table. If you can not remedy the fault, please contact an authorized<br />
service dealer directly.<br />
Fault Possible Cause Remedy<br />
The red light on the hand Electronic battery charge Fully charge the battery<br />
control is on and the lift<br />
does not move in a<br />
downward position<br />
control is activated. Batter<br />
charge inadequate.<br />
Battery is fully charged Battery can be put into<br />
The charger is plugged<br />
into the mains but the red<br />
charging light is not lit<br />
The bath lift does not<br />
move. No sound from the<br />
motor<br />
Control Light is broken<br />
The plug from the battery<br />
charger is not inserted<br />
completely in the battery.<br />
Battery contacts are dirty<br />
or battery is not fully<br />
charged<br />
No battery in the battery<br />
slot in the backrest<br />
Battery or main control<br />
unit not inserted completely<br />
in the backrest<br />
The battery is defect<br />
The actuator is defect<br />
Main control unit is defect<br />
normal use<br />
Insert battery and perform<br />
a trial run. If problems<br />
persist, replace the<br />
battery charger.<br />
Insert the plug as far as<br />
it will go<br />
Remove the battery,<br />
clean the contacts and<br />
recharge the battery<br />
Insert a fully charged<br />
battery in the backrest<br />
Completely insert the<br />
battery and main control<br />
unit<br />
Replace battery<br />
Send to authorized service<br />
center<br />
Remove main control<br />
unit and send to authorized<br />
repair center<br />
Seite 15