<strong>Pronto</strong>-7 <strong>Accuracy</strong><br />
25%<br />
20%<br />
Frequency of Occurence (%)<br />
15%<br />
10%<br />
5%<br />
0%<br />
-9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9<br />
Difference between <strong>Pronto</strong>-7 SpHb and Laboratory Reference tHb (g/dL)<br />
Figure 3. Normal Sensitivity Mode: Histogram of the <strong>Pronto</strong>-7 SpHb measurements vs. venous tHb measurements on the<br />
Beckman Coulter Model LH500 (N=11,335)<br />
25%<br />
20%<br />
Frequency of Occurence (%)<br />
15%<br />
10%<br />
5%<br />
0%<br />
-9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9<br />
Difference between <strong>Pronto</strong>-7 SpHb and Laboratory Reference tHb (g/dL)<br />
Figure 4. Maximum Sensitivity Mode: Histogram of the <strong>Pronto</strong>-7 SpHb measurements vs. venous tHb measurements on the<br />
Beckman Coulter Model LH500 (N=12,622)<br />
STUDY #2: accuracy of the PRONTO-7 pulse co-oximeter IN OUTPATIENT SETTING<br />
Methods<br />
All data was collected under institutional review board approval and all patients provided written, informed<br />
consent. Study subjects were adult and pediatric patients presenting to two outpatient clinics and one health screening<br />
event. Following written informed consent, patient demographics (age, gender, body mass index, ethnicity, smoking<br />
status, and skin pigmentation), medical condition, and medications were recorded for each patient. The Massey and<br />
Martin NIS Skin Color Scale 7 was used to categorize skin pigmentation with light pigmentation considered to be 1-3,<br />
medium pigmentation 4-7, and dark pigmentation 8-10. Following measurement and recording of finger diameter, a<br />
noninvasive hemoglobin test device (<strong>Pronto</strong>-7, <strong>Masimo</strong> Corporation, Irvine CA) with a reusable finger clip sensor (rainbow<br />
4D) was placed on the patient’s non-dominant ring finger. The sensor was covered with an opaque shield to prevent