Technical Bulletin, Pronto-7 Accuracy - Masimo

Technical Bulletin, Pronto-7 Accuracy - Masimo

Technical Bulletin, Pronto-7 Accuracy - Masimo

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Pronto</strong>-7 <strong>Accuracy</strong><br />

25%<br />

20%<br />

Frequency of Occurence (%)<br />

15%<br />

10%<br />

5%<br />

0%<br />

-9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9<br />

Difference between <strong>Pronto</strong>-7 SpHb and Laboratory Reference tHb (g/dL)<br />

Figure 3. Normal Sensitivity Mode: Histogram of the <strong>Pronto</strong>-7 SpHb measurements vs. venous tHb measurements on the<br />

Beckman Coulter Model LH500 (N=11,335)<br />

25%<br />

20%<br />

Frequency of Occurence (%)<br />

15%<br />

10%<br />

5%<br />

0%<br />

-9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9<br />

Difference between <strong>Pronto</strong>-7 SpHb and Laboratory Reference tHb (g/dL)<br />

Figure 4. Maximum Sensitivity Mode: Histogram of the <strong>Pronto</strong>-7 SpHb measurements vs. venous tHb measurements on the<br />

Beckman Coulter Model LH500 (N=12,622)<br />

STUDY #2: accuracy of the PRONTO-7 pulse co-oximeter IN OUTPATIENT SETTING<br />

Methods<br />

All data was collected under institutional review board approval and all patients provided written, informed<br />

consent. Study subjects were adult and pediatric patients presenting to two outpatient clinics and one health screening<br />

event. Following written informed consent, patient demographics (age, gender, body mass index, ethnicity, smoking<br />

status, and skin pigmentation), medical condition, and medications were recorded for each patient. The Massey and<br />

Martin NIS Skin Color Scale 7 was used to categorize skin pigmentation with light pigmentation considered to be 1-3,<br />

medium pigmentation 4-7, and dark pigmentation 8-10. Following measurement and recording of finger diameter, a<br />

noninvasive hemoglobin test device (<strong>Pronto</strong>-7, <strong>Masimo</strong> Corporation, Irvine CA) with a reusable finger clip sensor (rainbow<br />

4D) was placed on the patient’s non-dominant ring finger. The sensor was covered with an opaque shield to prevent