Powerlink ® <strong>System</strong> <strong>for</strong> <strong>AAA</strong><br />

<strong>Visiflex</strong> <strong>SurePass</strong> <strong>Delivery</strong> <strong>System</strong><br />

Deployment Technique


Powerlink ®<br />

<strong>Visiflex</strong> <strong>SurePass</strong><br />

<strong>Delivery</strong> <strong>System</strong> Components<br />

<strong>SurePass</strong> Wire - connects to the Contralateral<br />

Limb Cover. Pull it to deploy the contralateral limb.<br />

Ability to pass a guide<br />

wire through a hollow<br />

lumen to maintain<br />

contralateral access.<br />

Pusher Rod – connects to the<br />

Front Sheath. Advance it to<br />

deploy the main body.<br />

Front Sheath –<br />

covers the main<br />

body of the stent<br />

graft.<br />

White Hub – connects to the Outer<br />

Sheath. Pull it to release the limbs.<br />

1<br />

Red Hub – connects to the Radiopaque Front<br />

Stop on the Middle Core, and facilitates<br />

retraction of the Front Sheath.


Deployment Procedure<br />

First, per<strong>for</strong>m a cut-down on the access vessel using standard<br />

surgical technique.<br />

Insert a 12.5F tear-away sheath into the ipsi vessel, and a 9F<br />

sheath into the contra side.<br />

If necessary, insert a marker pigtail catheter and per<strong>for</strong>m an<br />

angiogram to measure vessel lengths and scout areas of potential<br />

troubleshooting.<br />

2


Insert a snare into the aorta through the ipsi tear-away sheath, and<br />

advance a floppy wire up the contra side.<br />

Snare and pull the wire through the ipsi side.<br />

3


Advance the Dual Lumen Catheter (DLC) up the contra and out the<br />

ipsi side.<br />

Advance a stiff wire through the DLC skives into the thoracic<br />

aorta.<br />

Position the DLC skive marker at the ipsi common iliac artery just<br />

over the bifurcation.<br />

4


Load the delivery system onto the stiff wire, and keep the side<br />

port medial.<br />

Remove the DLC from the contra side, and snap and peel away<br />

the ipsi sheath.<br />

Remove the floppy wire from the DLC, and advance the <strong>SurePass</strong><br />

wire up the ipsi DLC central lumen.<br />

5


Advance the delivery system into the aorta under fluoro, and<br />

position the stent graft distal end above the aortic bifurcation.<br />

Release the limbs:<br />

Loosen the White Hub, then stabilize the Red Hub while pulling<br />

White completely to Red.<br />

6


Re-advance the White Hub to recapture the ipsi limb, and tighten<br />

the White Hub.<br />

Gently pull out slack in the <strong>SurePass</strong> wire.<br />

While buttressing the White Hub with your thumb, pull the entire<br />

delivery system to bring the Front Stop below the lowest renal<br />

artery.<br />

Guide the contra limb into the contra iliac.<br />

7


Deploy the main body:<br />

Loosen the Red Hub, then stabilize it and advance the Pusher Rod<br />

completely. Tighten the Red Hub.<br />

Advance a .014” wire through the <strong>SurePass</strong> wire into the thoracic<br />

aorta to maintain contralateral wire access.<br />

Retract the Front Sheath:<br />

Stabilize the White Hub, and pull the Red Hub until the Front Stop<br />

is slightly above the aortic bifurcation.<br />

8


Deploy the ipsi limb in steps:<br />

Loosen the White Hub, then stabilize the Red Hub and pull White<br />

to deploy half of the ipsi limb.<br />

Repeat this maneuver to fully deploy the ipsi limb, then pull the<br />

Red Hub until both radiopaque markers are 1cm apart over the<br />

Front Stop.<br />

Pull the Red Hub to bring the Front Stop inside the limb.<br />

9


Pull the <strong>SurePass</strong> wire until the contra limb is fully deployed, and<br />

maintain .014” wire position while removing it.<br />

Exchange a pigtail catheter over the .014” wire, and per<strong>for</strong>m a<br />

completion angiogram.<br />

Remove the delivery system by pulling only on the Red Hub.<br />

10


The fully deployed Powerlink stent graft delivered with the <strong>Visiflex</strong><br />

<strong>SurePass</strong> <strong>Delivery</strong> <strong>System</strong>.<br />

11


Telephone: 949.595.7200<br />

Customer Service: 800.983.2284<br />

Fax: 949.830.4463<br />

customerservice@endologix.com<br />

www.endologix.com<br />

See product “Instructions <strong>for</strong> Use” <strong>for</strong> specific and<br />

complete prescribing in<strong>for</strong>mation.<br />