Vaporizers are agent specific<br />

Only the anesthetic agent they are built for should be used in them<br />

Use a different agent will change the output of the vaporizer<br />

Isoflurane in a sevoflurane vaporizer higher output<br />

Sevoflurane in a isoflurane vaporizer lower output<br />

The dial of a vaporizer is graduated in percent concentrations<br />

of inhalant in the inspired gas<br />

This is called volume percent<br />

Volume percent = partial pressure/ambient pressure x 100%<br />

VAPORIZERS<br />