444 Instrumentation and Quality Assurance59) A dose calibrator is received in the <strong>Nuclear</strong> <strong>Medicine</strong>department following repair. A linearity test isperformed on the dose calibrator, and the results areas follows.33.9 mCi 17-Jan 08:0030.3 mCi 17-Jan 09:009.2 mCi 17-Jan 19:002.3 mCi 18-Jan 07:000.5 mCi 18-Jan 20:000.2 mCi 19-Jan 03:00What is the next step the technologist should take?a. send the instrument for further repairb. accept the instrument for usec. reject the instrument for used. change the battery60) What is the sensitivity of a camera that registers28,137 cpm from a 120 µCi source if the background is705 cpm?a. 40 cpm/µCib. 203 cpm/µCic. 228 cpm/µCid. 234 cpm/µCie. 240 cpm/µCi61) A check source is calibrated for June 13, 1995 at noonand contains 145 µCi of 137 Cs (Tp = 30 years). If a dosecalibrator is used to measure that dose on June 13, 2001,what is the range of acceptable activities to prove thatthe accuracy is within 10%?a. 115–130 µCib. 119–125 µCic. 120–133 µCid. 125–140 µCi