7. Symptom: Unexpected sharp peaks.<strong>HT</strong> <strong>DNA</strong> <strong>High</strong> <strong>Sensitivity</strong> <strong>LabChip</strong> ® <strong>Kit</strong><strong>LabChip</strong> <strong>GX</strong>/<strong>GX</strong>II <strong>User</strong> <strong>Guide</strong>The electropherogram of a <strong>DNA</strong> <strong>High</strong> <strong>Sensitivity</strong> ladder is shown below with an unexpected sharp peakcaused by dust or other particulates.LowermarkerPeak caused byparticulatesUppermarkerPossible causes:A) Dust or other particulates introduced through sample or reagents.What to do:A) Replace molecular biology grade water used for chip preparation. Replace buffer used for sampleand reagent preparation. Use a 0.22-micron filter for all water and buffers used for chip, sample, andreagent preparation.Caliper Life Sciences Page 16 of 22 Revised Nov 201068 Elm StreetHopkinton, MA 01748-16681-877-LABCHIP (522-2447)