Factory Communication - Thermo Fisher Scientific

Factory Communication - Thermo Fisher Scientific

Factory Communication - Thermo Fisher Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Factory</strong> <strong>Communication</strong><br />

<strong>Factory</strong> :Madison Date:October, 26, 2009 #: IR 2309<br />

Model: FTIR Product Line: FTIR<br />

Instrument: Nicolet 6700 Revision Level: Release<br />

Author: Andy Bradley<br />

Priority<br />

□ Urgent<br />

□ Next visit<br />

□ Not required<br />

Category<br />

□ Safety<br />

□ Upgrade<br />

□ Fix<br />

□ Info<br />

□ New Feature<br />

Issue<br />

□ Hardware<br />

□ Software<br />

□ Application<br />

□ Documentation<br />

Implementation<br />

□ Field<br />

□ Depot<br />

Sensitivity<br />

□ General<br />

Distribution<br />

□ Preferred<br />

Customer<br />

□ Internal<br />

Title: Nicolet 6700 on-axis interferometer optics<br />

Issue:<br />

The laser beampath through the beamsplitter and interferometer will<br />

change from Off-Axis to On-Axis starting with 6700 serial number prefix<br />

APW. This change affects the location of the interferometer laser<br />

pick off mirror, the beamsplitter style, laser detector preamp and the<br />

laser detector preamp cable.<br />

Also, there is a small cosmetic change to the beamsplitter manufacture<br />

process. This is described in detail with a picture on Page 2.<br />

Background:<br />

Magna 550, 750, 560, 760 and Nexus 470, 670, 870 models all used<br />

On-Axis interferometer optics. The Nicolet 6700 changed the laser<br />

beampath to Off-Axis for the Continuum XL and Micro Probe microscope.<br />

Off-Axis optics placed the laser signal and laser preamp off the<br />

center of the beamsplitter. This provided consistent illumination to the<br />

pixels of the array detector. These microscope models are either discontinued<br />

or in retirement. Off axis interferometer optics do not add any<br />

other value to the 6700 model. Therefore, the Nicolet 6700 will migrate<br />

back to On-axis beamsplitter optics.<br />

Upgraded systems:<br />

Magna, Protégé and Nexus systems that have been upgraded to the<br />

6700 are all On-Axis interferometer systems. The Magna / Protégé upgraded<br />

6700 systems have the AKL serial number prefix. This change<br />

will not affect the upgraded systems since they are already On-Axis.<br />

Action Required<br />

Logistics: ___________________________ _________<br />

Institute: _____________________________________<br />

Order Entry: __________________________________<br />

Technical Support: _____________________________<br />

Confidentiality Notice:<br />

This document contains confidential or legally privileged information that is intended for the exclusive use of<br />

<strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employees. If you are not a <strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employee, then you are hereby notified that any<br />

disclosure, copying, distribution, or reliance upon the contents of this document is strictly prohibited. If you have received this<br />

document in error, then please destroy it immediately.<br />

1


Solution:<br />

<strong>Factory</strong> <strong>Communication</strong><br />

Nicolet 6700 and 8700 systems with Serial number APW and forward will have On-Axis interferometers,<br />

laser preamps and beamsplitters. This change will be transparent to most customers.<br />

The greatest impact will affect what part to order when placing service orders for interferometers<br />

and beamplitters. There should be no need to change systems in the field from Off-<br />

Axis to On-Axis.<br />

Visual change to beamsplitter mask<br />

The On-Axis beamsplitter mask has been updated. The new on-axis beam splitters will show<br />

one slight visual difference from the older “Magna” and “Nexus” on-axis beam splitters. A thin<br />

uncoated slot will run from the central laser hole to the right side (pictured below). This is due<br />

to the use of a new metal mask for production, which greatly improves the yield in beam splitter<br />

manufacturing. This will not affect performance since the slot is in the shadow of the laser detector<br />

arm and the laser pick off mirror. The Solid Substrate beam splitter will not change. The<br />

Si beamsplitter has a drilled hole for the laser window with no slot, regardless of part number.<br />

Service LRU part numbers<br />

Interferometer service orders for new Nicolet 6700 systems with On-Axis optics, AKL and<br />

APW serial number prefix, should order the On-Axis Nexus interferometer and<br />

beamsplitters.<br />

Beamsplitters<br />

Beam Splitter substrate<br />

On-Axis (Magna, Nexus,<br />

6700 AKL, APW prefix)<br />

Off-Axis (Original 6700)<br />

Ge on KBr 840-128900 840-150700<br />

Quartz 840-129300 840-129300<br />

Solid Substrate Far-IR 840-129400 840-151200<br />

Silicon on CaF2 840-129500 840-151000<br />

Ge on CsI 840-133800 840-150900<br />

XT-KBr 840-140900 840-150800<br />

Confidentiality Notice:<br />

This document contains confidential or legally privileged information that is intended for the exclusive use of<br />

<strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employees. If you are not a <strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employee, then you are hereby notified that any<br />

disclosure, copying, distribution, or reliance upon the contents of this document is strictly prohibited. If you have received this<br />

document in error, then please destroy it immediately.<br />

2


Interferometers<br />

On-Axis interferometer - Magna, Nexus, 6700 - AKL, APW serial number prefix<br />

Gold optics<br />

470-247000 X700 INFR GOLD ON-AXIS TALON BSID PIEZO<br />

470-246800 X700 INFR GOLD ON-AXIS TALON BSID<br />

470-246900 X700 INFR GOLD ON-AXIS<br />

Aluminum optics<br />

470-102600 X700 INFR ALUM ON-AXIS TALON BSID PIEZO<br />

470-102200 X700 INFR ALUM ON-AXIS TALON BSID<br />

470-075900 X700 INFR ALUM ON-AXIS<br />

On-Axis interferometer includes:<br />

410-024100 Nexus / Antaris laser preamp<br />

Off-Axis interferometer - Original 6700<br />

<strong>Factory</strong> <strong>Communication</strong><br />

Interferometer service orders for Off-Axis optics continue to order the documented part numbers.<br />

The descriptions have been updated.<br />

Gold optics<br />

470-229100 X700 INFR GOLD OFF-AXIS TALON BSID PIEZO<br />

470-228600 X700 INFR GOLD OFF-AXIS TALON BSID<br />

470-228700 X700 INFR GOLD OFF-AXIS<br />

Aluminum optics<br />

470-229600 X700 INFR ALUM OFF-AXIS TALON BSID PIEZO<br />

470-229500 X700 INFR ALUM OFF-AXIS TALON BSID<br />

470-229400 X700 INFR ALUM OFF-AXIS<br />

Off-Axis interferometers include:<br />

410-045600 N2 laser preamp<br />

Part number Description of terms:<br />

Piezo –Dithering fixed mirror. Used for 8700 only<br />

Talon – Beamsplitter interlock T-handle<br />

BSID – Beamsplitter Identification Reader<br />

Confidentiality Notice:<br />

This document contains confidential or legally privileged information that is intended for the exclusive use of<br />

<strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employees. If you are not a <strong>Thermo</strong> <strong>Fisher</strong> <strong>Scientific</strong> employee, then you are hereby notified that any<br />

disclosure, copying, distribution, or reliance upon the contents of this document is strictly prohibited. If you have received this<br />

document in error, then please destroy it immediately.<br />