select 3D ® and select 3D ® ii<br />
ArtEriAl CAnnulAE<br />
lAsEr DOPPlEr AnEmOmEtry<br />
(lDA) PErFOrmAnCE †<br />
laser doppler anemometry testing demonstrates that<br />
the select <strong>3D®</strong> arterial cannula features significantly<br />
reduced velocity compared to a standard open<br />
tip arterial cannula while also maintaining a more<br />
centralized flow pattern compared to competitive<br />
diffuse flow arterial cannulae.<br />
Comparison of flow velocity for 24 Fr arterial cannulae<br />
at 6 lPm. velocity measurements at 12mm from<br />
cannula tip for the select 3D (diffuse flow), a standard<br />
open tip angled (non-diffuse flow), terumo®-sarns<br />
angled soft-Flow® (diffuse flow).<br />
<strong>Medtronic</strong>-DLP 78624 – 24 Fr<br />
<strong>Medtronic</strong>-DLP 72524 – 24 Fr<br />
Terumo ® -Sarns Soft-Flow - 5768 – 24 Fr - 8 mm<br />
† Data on file at medtronic, inc.<br />
FlOw PErFOrmAnCE<br />
COmPArisOn<br />
Select 3D and Select 3D II Arterial Cannulae<br />
Pressure Loss (mm Hg)<br />
100<br />
80<br />
60<br />
40<br />
20<br />
18 FR<br />
20 FR<br />
22 FR<br />
24 FR<br />
0<br />
0 1 2 3 4<br />
Flow Rate (L/min. of water)<br />
5 6<br />
the unique design of the 3D arterial cannulae<br />
tip results in lower arterial line pressures than<br />
a conventional open tip arterial cannula or<br />
competitive diffuse flow arterial cannula.<br />
Line Pressure (mm of Hg)<br />
Average Line Pressure – 24 Fr Arterial Cannulae<br />
Average Line Pressure – 24 Fr Arterial Cannula<br />
(Closed Circuit Using Blood)<br />
250<br />
200<br />
150<br />
100<br />
50<br />
0<br />
112.20<br />
Select 3D<br />
24 Fr Angled<br />
(<strong>Medtronic</strong>-<br />
DLP 78624)<br />
(Closed Circuit Using Blood)<br />
127.73<br />
Open Tip<br />
24 Fr Curved<br />
(<strong>Medtronic</strong>-<br />
DLP 83024)<br />
156.47<br />
Soft-�ow®<br />
24 Fr Angled<br />
Terumo®-Sarns<br />
5768<br />
laboratory comparison of arterial line pressures at 6 liters<br />
per minute*<br />
*Data on file at medtronic, inc.