Hematocrit effects For a rapid test, WB is preferable if not necessary Most systems now report “plasma equivalent” Systems are calibrated at normal hematocrit. WB sample hematocrits may vary significantly(~15 to >70) Glucose content <strong>of</strong> whole blood as compared toplasma is inversely related with hematocrit.03162010 Ken Ervin Consulting Services