<strong>266</strong> Plus Electrical <strong>Medical</strong> Safety AnalyserTechnical <strong>Data</strong> <strong>Sheet</strong> 2009Tried. Tested. Trusted.


Tried. Tested. Trusted.A compact, easy-to-use electricalmedical safety analyser designed to testin accordance with IEC/EN 60601-1,MDA DB9801, and AS/NZ 3200.The <strong>Rigel</strong> <strong>266</strong> plus electrical medical safety analyser, whenused in the Semi Automatic mode, will automatically runthrough the applicable Single Fault Conditions for eachindividual Leakage Test.Key features■ Portable & LightweightSmall and lightweight design for easytransportation and fast operationIn addition to the semi-automatic test, the<strong>Rigel</strong> <strong>266</strong> plus provides fully manual testroutines. Test routines are easily selectablevia the rotary selector switch and dedicatedpush buttons are provided for settingindividual Single Fault Conditions.A large LCD display provides clearindication of test results, Single FaultCondition, AC or DC measurements andpass/fail limits.IEC test leads are easily tested with thededicated IEC lead test socket. A serialoutput allows for data transfer to printeror PC.With its compact design and full IEC60601-1 features, the <strong>Rigel</strong> <strong>266</strong> plusprovides the user with a highly effective andportable test solutions.Semi Automatic or Fully Manual?Conventional safety testing on medicalequipment can often result in a number ofpower breaks caused by the various testconditions set by local or internationalstandards.Power breaks are not always withoutdanger to the equipment under test, forexample PC based equipment / systems orimaging equipment such as ultrasound.By using the <strong>Rigel</strong> <strong>266</strong> plus in full manualmode, the user is not only in control of whenthe power breaks occurs, the user can alsolimit the number of power breaks to aminimum of just two when testing in fullcompliance with IEC 60601-1.Dedicated buttons provide an easy meansof selecting the required fault condition ormains reversal.The semi-automatic mode automaticallyactivates each single fault condition in a logicsequence for each individual leakage setting,thus speeding up your test routine whentesting most common medical andnon medial devices.■ Full IEC 60601 Leakage TestsEarth, Enclosure, Patient, Patient Auxiliary,F-type and SIP/SOP leakage tests■ Up to 25A AC Test CurrentUnique choice of 1A, 10A and 25A testcurrent■ Large LCD DisplayFast and Easy interpretation of testresults, PASS/FAIL limits, test conditionand more■ Separate AC & DC ValuesMeasures both the AC and DC leakage asper IEC 60601-1 requirements■ Full Manual ControlFast and Accurate control and Safety Tests■ Insulation TestOn device under test or applied partsto earth■ 5 Patient ConnectionsProvide an easy and fast way of testingmedical electronic equipment withpatient connections■ Dedicated IEC Lead TestFor easy testing of IEC power cables


Tried. Tested. Trusted.Preset Test Limits in the <strong>Rigel</strong> <strong>266</strong> plusTYPE B TYPE BF TYPE CFLeakage Current NC SFC NC SFC NC SFCEarth 500μA 1000μA 500μA 1000μA 500μA 1000μAEnclosure 100μA 500μA 100μA 500μA 100μA 500μAPatient (dc) 10μA 50μA 10μA 50μA 10μA 50μAPatient (ac) 100μA 500μA 100μA 500μA 10μA 50μAPatient Auxiliary (dc) 10μA 50μA 10μA 50μA 10μA 50μAPatient Auxiliary (ac) 100μA 500μA 100μA 500μA 10μA 50μASIP / SOP 5000μA NA NAF-Type NA 5000μA 50μATable T1; IEC 60601-1 limits for equipment designed to IEC 60601ApplicationsIn-service Safety Testing of:patient monitors, infusion pumps, hospitalbeds, X-ray and ultrasound equipment, videoequipment, PC based medical systems,defibrillators and many other medical andnon-medical electrical equipment andsystems.Fast & Efficient Testing of IEC leadsEarth bond testing on (medical)installations and non-medical equipmentPortable Appliance TestingWhat comes in the box?■ Calibration Certificate<strong>Rigel</strong> <strong>266</strong> plus Design PhilosophyThe <strong>Rigel</strong> <strong>266</strong> plus was designed to addressthe increasing demand for smaller testequipment in the healthcare industry. Thechallenge was not to compromise the fullIEC 60601-1 leakage measurements andflexibility in test currents up to 25A.Most available menu driven or switchoperated safety testers are cumbersome inuse and for this reason, great attention wasgiven to providing an easy and fast means ofoperating the <strong>Rigel</strong> <strong>266</strong>. A large rotaryswitch and LCD display give the<strong>Rigel</strong> <strong>266</strong> plus its unique identity.After more than 9 years, the <strong>Rigel</strong> <strong>266</strong> plushas become an industry standard and iswidely considered as the easiest and fastestIEC60601 tester currently on the market.■ Earthbond Test Lead + Clip■ Applied Part Harness■ 5 x Applied Part Adaptor■ Instruction Manual■ Rugged Carrying Case


Tried. Tested. Trusted.SPECIFICATIONSAC Earth Bond TestOpen Test Voltage 6V rms nominalTest Current1A, 10A, 25A ACRange40mohms – 19.99ohmsResolution0.01ohmsAccuracy+/- 5% of reading, +/- 2 countsMeasurement4-wire measurement, floating earthInsulation TestTest Voltage500V d.c. nominal (0.5Mohms load)Range100kohms - 100 MohmsResolution0.01 Mohms (1MohmFrequency Response as per IEC 60601-1 requirementsAccuracy+/- 5% of reading, +/- 10 countsRangeEarth Leakage0.04mA – 9.999mAEnclosure Leakage 0.04mA – 9.999mAPatient Leakage0.004mA – 9.999mAPatient Auxiliary Leakage 0.004mA – 9.999mAPatient SIP / SOP Leakage 0.04mA – 9.999mAPatient F-Type Leakage 0.025mA – 9.999mAF-Type Open Circuit Voltage 110% +/-20% of Mains Input VoltageLoad TestMeasured Load0.06 - 4kVAMeasured Voltage Mains Supply +/-10%IEC Lead TestTest40V a.c., 1mA nominalDetectsGood, Open, Short, ReverseAccessories:■ Mini Thermal Printer■ Rubber Holster■ PASS / FAIL Labels■ ASSET ID Labels■ <strong>Rigel</strong> 601 CheckboxFrom <strong>Rigel</strong> <strong>Medical</strong>Also available■ <strong>Rigel</strong> 277 Plus Automatic Safety Analyser■ <strong>Rigel</strong> 288 Plus Hand-held Safety Analyser■ <strong>Rigel</strong> 601 Checkbox■ <strong>Rigel</strong> BP-Sim NIBP Simulator■ <strong>Rigel</strong> SP-Sim SpO2 Simulator■ <strong>Rigel</strong> 333 Patient Simulator■ <strong>Rigel</strong> 344 Defibrillator Tester■ <strong>Rigel</strong> 355 Ventilator Tester■ <strong>Rigel</strong> 377 Electrosurgical Analyser■ Med-eBase – Software ApplicationFrom the Seaward Group■ Portable Appliance Testers■ IEC Lead Tester■ Insulation Resistance Testers■ RCD Testers■ Earth Loop Impedance Testers■ Installation Testers■ Multimeters■ Current Clamps■ Hipot Testers■ Earthbond Testers■ Micro OhmmetersMechanicalSizeWeight310mm x 210mm x 80mm4kgEnvironmentalOperating0ºC to 40ºC (non condensing)Storage-10ºC to 50ºC (non condensing)Maximum R.H 90%Part number: 350A910350A913350A915350A917350A918UK (13A)Schuko (16A)Danish (10A)French (16A)Czech / Polish (16A)For orders orenquiries call+44 (0) 191 587 8730<strong>Rigel</strong> <strong>Medical</strong>, Bracken Hill,South West Industrial Estate,Peterlee, County DurhamSR8 2SW United KingdomFax: +44 (0) 191 586 0227Email: sales@rigelmedical.comWeb: www.rigelmedical.comPart of