13.07.2015 Views

A Digital-to-Channel Interface for In-Line Sample ... - Diyhpl.us

A Digital-to-Channel Interface for In-Line Sample ... - Diyhpl.us

A Digital-to-Channel Interface for In-Line Sample ... - Diyhpl.us

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

experiments, two solutions were <strong>us</strong>ed, containing the analytesand the label, respectively. For on-chip labeling of amino acids,the <strong>for</strong>mer solution comprised glycine, alanine and valine standards(20 mM each with 4 mM KCN in run buffer 2), while thelatter comprised NDA (2 mM in run buffer 2). For on-chiplabeling of cell constituents, the analyte solution was <strong>for</strong>med bydiluting a thawed aliquot of lysate 1 : 10 in run buffer 1 containingKCN (20 mM final concentration), while the reagentsolution was NDA (10 mM) in neat ace<strong>to</strong>nitrile. <strong>In</strong> each case,droplets of analyte and reagent solution were moved, merged,incubated (2 min), and then delivered <strong>to</strong> the digital-channelinterface by DMF. <strong>Sample</strong>s were then loaded and injected electrokinetically,followed by a separation <strong>us</strong>ing MEKC in runbuffer 1 (lysate) or 2 (standards).Lysate peaks were tentatively identified by standard additionsof NDA-labeled amino acids. Briefly, amino acid standards (50mM) were reacted off-line with KCN (2 mM) and NDA (1 mM)in borate buffer (50 mM, pH 9, 30% ACN). Lysate was labeledoff-line <strong>us</strong>ing the same concentrations described above anddiluted 1 : 10 in run buffer 1 prior <strong>to</strong> analysis. Aliquots of lysate(48 mL) were combined with aliquots of amino acidstandards (2 mL) and separated by MEKC in run buffer 1 <strong>to</strong>identify co-eluting lysate analytes.Tryptic digestionTwo solutions were <strong>us</strong>ed <strong>for</strong> evaluation of on-chip digestion:FITC-<strong>In</strong>s (50 mg/mL) in borate buffer (50 mM, pH 9), andtrypsin (100 mg/mL) in Tris-HCl buffer (10 mM, 1 mM CaCl 2 ,0.08% pluronic F-127, pH 8.5). Droplets of each solution weremoved, merged, and mixed by DMF, and allowed <strong>to</strong> react <strong>for</strong>a designated period of time (1, 5, 15, and 30 min). Duringincubation, <strong>to</strong> limit the effects of evaporation, the reactingdroplet was enclosed in a PDMS cover. After reaction, the processedinsulin was delivered <strong>to</strong> the digital-channel interface byDMF, and loaded, injected, and separated by MEKC in runbuffer 1.electrokinetic flow (as described in the methods section, anelectrode was positioned such that it penetrated in<strong>to</strong> the dropletat the interface); however, we also demonstrated that: (a)a droplet containing run buffer delivered <strong>to</strong> empty channels willfill the network by capillary action; and (b) a droplet delivered <strong>to</strong>pre-filled channels will spontaneo<strong>us</strong>ly load by LaPlace pressure(similar <strong>to</strong> what has been reported previo<strong>us</strong>ly 10 ). We <strong>us</strong>ed electrokineticloading beca<strong>us</strong>e this was the most reliable method <strong>to</strong>inject samples on<strong>to</strong> the separation column. We note that in thismethod, only a small fraction of the fluid manipulated by DMF(5 mL merged droplets) is sampled in<strong>to</strong> the channels, and aneven smaller fraction (1 min),evaporation was minimized by enclosing the droplet undera PDMS cover. <strong>In</strong> designing and building the devices reportedhere, we chose <strong>to</strong> <strong>us</strong>e the single-plate DMF <strong>for</strong>mat, which isconvenient <strong>for</strong> droplet delivery <strong>to</strong> the interface with microchannels.A drawback of the single-plate <strong>for</strong>mat is incompatibilitywith droplet dispensing from reservoirs; in future designs,that capacity might be added by including a two-plate <strong>to</strong> singleplateDMF interface, which has been reported previo<strong>us</strong>ly. 35After delivering a droplet <strong>to</strong> the channel inlet, the droplet’scontents were loaded in<strong>to</strong> the microchannel <strong>for</strong> separations. <strong>In</strong>most cases, reagents were loaded in<strong>to</strong> pre-filled channels byFig. 3 Analysis of hybrid digital-channel microfluidic device reproducibility.<strong>In</strong> replicate experiments, droplets containing rhodamine 123 andfluorescein were combined on-chip and then evaluated by MEKC. Arepresentative electropherogram is shown in (a), and reproducibility dataare listed in (b). For comparison, two off-chip controls comprisingreplicate analyses of different mixtures (‘‘individual samples’’) and ofa single mixture (‘‘common sample’’) were carried out, and the data arelisted in (b). The observation that, in comparison with data generatedfrom samples prepared on-chip, that the individual samples control hassimilar peak area reproducibility suggests that the main source of thisvariance is sample dispensing (i.e., pipetting), rather than from chipoperation.This journal is ª The Royal Society of Chemistry 2009 Lab Chip, 2009, 9, 1046–1051 | 1049


the electropherogram. When evaluated in replicate trials, themethod was characterized by excellent retention time reproducibility(20 000 plates overthe 4.5 cm channel), but the relative peak area variation waslarger than expected (7% RSD). To probe the source of thiserror, we evaluated two controls, prepared off-chip. The firstcontrol (called ‘‘individual samples’’), implemented <strong>to</strong> determinethe variance ca<strong>us</strong>ed by sample dispensing (i.e., pipetting the twodye solutions), involved replicate measurements made from fiveindividually prepared samples. Each sample contained run buffer(45 mL), and rhodamine 123 and fluorescein (2.5 mL each). Thesecond control (called ‘‘common sample’’), implemented <strong>to</strong>determine the variance ca<strong>us</strong>ed by mixing, injection, and separation,involved replicate measurements from a single mixture ofthe two dyes. Car<strong>to</strong>ons depicting the controls and their respectivedata are listed in Fig. 3(b).<strong>In</strong> comparing the per<strong>for</strong>mance of the on-chip method <strong>to</strong> thetwo controls, it appears that the primary source of peak areavariance is sample dispensing (i.e., pipetting). As listed inFig. 3(b), the peak area reproducibility in the ‘‘individualsamples’’ control is similar <strong>to</strong> that observed <strong>for</strong> the on-chipmethod, suggesting that the on-chip method and this controlshare the primary source of variance. <strong>In</strong> contrast, the peak areareproducibility in the ‘‘common sample’’ control is significantlyimproved, suggesting that the contribution <strong>to</strong> variance frommixing, injection, and separation, is much lower. Th<strong>us</strong>, wespeculate that in future experiments with on-chip dispensingfrom reservoirs (instead of pipetting <strong>to</strong> the surface), the peak areareproducibility of the on-chip method will be substantiallyimproved.On-chip sample processingTo demonstrate the utility of the new device <strong>for</strong>mat <strong>for</strong> integratedsample processing, we <strong>us</strong>ed it <strong>to</strong> fluorescently label aminoacid standards and cell lysate on-chip, followed by separations.For the <strong>for</strong>mer, a droplet containing the fluorogenic reagent,NDA, was merged with a droplet containing a mixture of threeamino acid standards, glycine (Gly), alanine (Ala) and valine(Val), on the digital plat<strong>for</strong>m. The merged droplet was actuatedbetween adjacent electrodes <strong>to</strong> mix its contents <strong>for</strong> 2 min(comparable <strong>to</strong> reaction times reported <strong>for</strong> NDA labeling inmicrochannels 3 ) and was then delivered <strong>to</strong> the interface where itscontents were sampled in<strong>to</strong> the channel by EOF. Fig. 4(a) showsan electropherogram generated <strong>us</strong>ing this method. Under theseconditions, the three species separate in less than one minute.A similar on-chip pro<strong>to</strong>col was <strong>us</strong>ed <strong>to</strong> label the amines ina solution of cell lysate. As shown in Fig. 4(b), the constituentpeaks are partially resolved in less than two minutes, and severalof the peaks were tentatively assigned by spiking lysate mixtureswith NDA-labeled amino acid standards. As expected, the morehydrophobic amino acids (e.g., leucine) migrated slowly beca<strong>us</strong>eof interaction with the micelles. Basic species such as argininewere likewise slowed as a function of electrophoretic migration inthe opposite direction of the cathodic EOF.To demonstrate a second sample processing application <strong>for</strong>the new device, we <strong>us</strong>ed it <strong>to</strong> digest a proteomic analyte prior <strong>to</strong>separation. Singly tagged FITC-<strong>In</strong>sulin (FITC-<strong>In</strong>s) was a <strong>us</strong>efulmodel <strong>for</strong> this work, as the single label simplifies the number ofFig. 4 Electropherograms generated after on-chip NDA labeling ofamino acids and cell lysate. (a) Glycine (Gly), alanine (Ala) and valine(Val) (10 mM ea.) were labeled with NDA <strong>for</strong> two minutes and theninjected and separated by MEKC. (b) HeLa cell lysate labeled with NDA<strong>for</strong> one minute and then separated by MECK. The inset was generated<strong>us</strong>ing an identical pro<strong>to</strong>col, but with no lysate, and the Y-axis was scaledidentically <strong>to</strong> that of the main panel. Peaks were assigned by spiking withNDA labeled standards.Fig. 5 Electropherograms generated by on-chip digestions of singlylabeled FITC-<strong>In</strong>sulin (offset vertically <strong>for</strong> clarity). As shown, as thedigestion progresses <strong>for</strong> longer periods, peak(s) representing tryptic digestfragment(s) appear at 94 s, and the parent FITC-<strong>In</strong>sulin peak at 110 sdisappears. After fifteen minutes the digestion has neared completion.detectable species. Droplets containing FITC-<strong>In</strong>s were mergedwith droplets containing trypsin and incubated <strong>for</strong> differentperiods prior <strong>to</strong> being driven <strong>to</strong> the interface <strong>to</strong> be sampled in<strong>to</strong>1050 | Lab Chip, 2009, 9, 1046–1051 This journal is ª The Royal Society of Chemistry 2009


the microchannels. Upon injection in<strong>to</strong> the channels, the reactionwas quenched, as the surfactant denatures the enzyme; th<strong>us</strong>, thisserves as a metric <strong>for</strong> moni<strong>to</strong>ring reaction progress as a functionof time. Fig. 5 shows four electropherograms generated afterprogressively longer digestion times. As shown, as digestion timeincreases, the primary FITC-<strong>In</strong>s peak (retention time 110 s)disappears, while a new peak belonging <strong>to</strong> a digest fragmentappears at retention time 94 s.Tryptic digestion of FITC-<strong>In</strong>s can create two labeled fragmentsthrough cleavage at the B-Chain Lys or B-chain Argresidues <strong>for</strong>ming peptides that are 1- or 8-residues shorter thanthe parent molecule. The digest fragment peak in the electropherogramsappears <strong>to</strong> have a shoulder which may correspond <strong>to</strong>detection of both fragments. Overall, the time required <strong>for</strong>complete digestion (15 min) is short relative <strong>to</strong> conventionalsolution-phase digestion pro<strong>to</strong>cols which require longer times(12 hours) and elevated temperature (37 C). 7 These resultsshow great promise <strong>for</strong> our plans <strong>to</strong> build devices with muchlarger electrode arrays <strong>for</strong> integration of multistep proteomicprocessing regimens. 39Concl<strong>us</strong>ionWe have proposed, fabricated, and tested a hybrid device thatintegrates digital microfluidics with microchannels on onesubstrate. The hybrid device <strong>us</strong>es digital microfluidics <strong>to</strong> per<strong>for</strong>mchemical processing on samples prior <strong>to</strong> transporting them <strong>to</strong>microchannels <strong>for</strong> analytical separations. The method wasdemonstrated <strong>to</strong> be capable of on-chip labeling of amino acidsand amines in cell lysate, as well as enzymatic digestion ofpeptide standards, followed by MEKC separation in microchannels.We believe this marriage of digital microfluidics andmicrochannels is an important step <strong>to</strong>ward fully integrated labson-a-chip<strong>for</strong> in-line sample processing and separation.AcknowledgementsWe thank Irena Barbulovic-Nad <strong>for</strong> assistance with cell culture.We acknowledge the Natural Sciences and Engineering ResearchCouncil (NSERC) and the Canadian Cancer Society (CCS) <strong>for</strong>financial support. M.A. and M.W.L.W. on thank the OntarioGraduate Scholarship program and A.R.W. thanks the CanadaResearch Chair program.References1 S. C. Jacobson, R. Hergenroder, L. B. Koutny and J. M. Ramsey,Anal. Chem., 1994, 66, 1114–1118.2 D. Wu, J. Qin and B. Lin, J. Chroma<strong>to</strong>gr. A, 2008, 1184, 542–559.3 N. Gottschlich, C. T. Culbertson, T. E. McKnight, S. C. Jacobsonand J. M. Ramsey, J. Chroma<strong>to</strong>gr. B, 2000, 745, 243–249.4 M. Brivio, R. H. Fokkens, W. Verboom, D. N. Reinhoudt, N. R. Tas,M. Goedbloed and A. Van den Berg, Anal. Chem., 2002, 74, 3972–3976.5 S. C. Jacobson, R. Hergenroder, A. W. Moore, Jr. and J. M. Ramsey,Anal. Chem., 1994, 66, 4127–4132.6 K. W. Ro, K. Lim, H. Kim and J. H. Hahn, Electrophoresis, 2002, 23,1129–1137.7 M. P. Washburn, D. Wolters and J. R. Yates III, Nat. Biotechnol.,2001, 19, 242–247.8 H. J. Crabtree, E. C. S. Cheong, D. A. Tilroe and C. J. Backho<strong>us</strong>e,Anal. Chem., 2001, 73, 4079–4086.9 D. Sin<strong>to</strong>n and D. Li, Colloids Surf., A, 2003, 222, 273–283.10 G. M. Walker and D. J. Beebe, Lab Chip, 2002, 2, 131–134.11 M. A. Unger, H. P. Chou, T. Thorsen, A. Scherer and S. R. Quake,Science, 2000, 288, 113–116.12 G. T. Roman and R. T. Kennedy, J. Chroma<strong>to</strong>gr. A, 2007, 1168, 170–188.13 S.-Y. Teh, R. Lin, L.-H. Hung and A. P. Lee, Lab Chip, 2008, 8, 198–220.14 J. S. Edgar, C. P. Pabbati, R. M. Lorenz, M. He, G. S. Fiorini andD. T. Chiu, Anal. Chem., 2006, 78, 6948–6954.15 G. T. Roman, M. Wang, K. N. Shultz, C. Jennings andR. T. Kennedy, Anal. Chem., 2008, 80, 8231–8238.16 M. G. Pollack, R. B. Fair and A. D. Shenderov, Appl. Phys. Lett.,2000, 77, 1725–1726.17 J. Lee, H. Moon, J. Fowler, T. Schoellhammer and C.-J. Kim, Sens.Actua<strong>to</strong>rs, A, 2002, 95, 259–268.18 I. Barbulovic-Nad, H. Yang, P. S. Park and A. R. Wheeler, Lab Chip,2008, 8, 519–526.19 V. N. Luk, G. C. Mo and A. R. Wheeler, Langmuir, 2008, 24, 6382–6389.20 E. M. Miller and A. R. Wheeler, Anal. Chem., 2008, 80, 1614–1619.21 D. Chatterjee, B. Hetayothin, A. R. Wheeler, D. J. King andR. L. Garrell, Lab Chip, 2006, 6, 199–206.22 V. Srinivasan, V. K. Pamula and R. B. Fair, Lab Chip, 2004, 4, 310–315.23 D. Brassard, L. Malic, F. Normandin, M. Tabrizian and T. Veres,Lab Chip, 2008, 8, 1342–1349.24 S. K. Cho, H. Moon and C. J. Kim, J. MEMS, 2003, 12, 70–80.25 M. Abdelgawad and A. R. Wheeler, Adv. Mater., DOI: 10.1002/adma.200802244.26 A. R. Wheeler, Science, 2008, 322, 539–540.27 F. Malloggi, H. Gu, A. G. Banpurkar, S. A. Vanapalli and F. Mugele,Eur. Phys. J. E, 2008, 26, 91–96.28 W. Sa<strong>to</strong>h, H. Hosono and H. Suzuki, Anal. Chem., 2005, 77, 6857–6863.29 W. Sa<strong>to</strong>h, H. Yokomaku, H. Hosono, N. Ohnishi and H. Suzuki,J. Appl. Phys., 2008, 103, 034903.30 D. Huh, A. H. Tkaczyk, J. H. Bahng, Y. Chang, H. H. Wei,J. B. Grotberg, C. J. Kim, K. Kurabayashi and S. Takayama,J. Am. Chem. Soc., 2003, 125, 14678–14679.31 U.-C. YiW. LiuP.-P. d. Guzman and C.-J. Kimin, Proceedings of 2006Hil<strong>to</strong>n Head Solid-State Sensor, Actua<strong>to</strong>r and MicrosystemsWorkshop,IEEE, Piscataway, NJ, 2006, pp. 128–131.32 D. C. Duffy, J. C. McDonald, O. J. A. Schueller andG. M. Whitesides, Anal. Chem., 1998, 70, 4974–4984.33 M. Abdelgawad and A. R. Wheeler, Adv. Mater., 2007, 19, 133–137.34 C. Cooney, C.-Y. Chen, M. Emerling, A. Nadim and J. Sterling,Microfluid. Nanofluid., 2006, 2, 435–446.35 J. Berthier, Microdrops and digital microfluidics, William AndrewPub.Norwich, NY, 2008.36 K. Sueyoshi, F. Kitagawa and K. Otsuka, J. Sep. Sci., 2008, 31, 2650–2666.37 Z. Long, Z. Shen, D. Wu, J. Qin and B. Lin, Lab Chip, 2007, 7, 1819–1824.38 W. Yang, X. Sun, T. Pan and A. T. Woolley, Electrophoresis, 2008,29, 3429–3435.39 M. Jebrail and A. R. Wheeler, Anal. Chem., 2009, 81, 330–335.This journal is ª The Royal Society of Chemistry 2009 Lab Chip, 2009, 9, 1046–1051 | 1051

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!