Taps: if implants with an endosteal diameter of 4.8 mm are inserted, <strong>the</strong> thread<br />
should be pretapped. in <strong>the</strong> case of implants with a smaller endosteal diameter<br />
(Ø 3.3 mm or Ø 4.1 mm), tapping can be omitted, depending <strong>on</strong> <strong>the</strong> situati<strong>on</strong><br />
(see also page 31, B<strong>on</strong>e classes).<br />
Important<br />
Please note that <strong>the</strong> taps for S/SP differ from <strong>the</strong> TE taps (see below). Depending<br />
<strong>on</strong> b<strong>on</strong>e class, Tapered Effect implants can be implanted without or<br />
with prior tapping.<br />
<strong>the</strong> thread can be tapped using a handpiece (max. 15 rpm) or ratchet. <strong>the</strong>re are<br />
<strong>the</strong>refore two tap variants for each implant diameter and <strong>the</strong> tap for manual use is<br />
available in a short and a l<strong>on</strong>g versi<strong>on</strong>.<br />
Taps for S and SP implants<br />
p<br />
S/SP for ratchet<br />
(short)<br />
Machine tapping<br />
for ratchet<br />
(l<strong>on</strong>g)<br />
for handpiece<br />
Ø 3,3 mm 044.590 044.591 044.575<br />
Ø 4,1 mm 044.592 044.593 044.577<br />
Ø 4,8 mm 044.594 044.595 044.579<br />
Recommended speed: max. 15 rpm<br />
<strong>the</strong>re is an adapter tap for each implant diameter, which is attached to <strong>the</strong><br />
handpiece adapter.<br />
Handpiece adapter<br />
machine<br />
tapping<br />
Taps<br />
Taps for TE implants<br />
manual<br />
tapping<br />
TE for handpiece<br />
Ø 3,3 mm 80171<br />
Ø 4,1 mm 80169<br />
Ø 4,8 mm 80170