ELECTROCHEMICAL GAS SENSORS4 <strong>to</strong> <strong>20</strong> <strong>mA</strong> <strong>Transmitter</strong> <strong>Board</strong> <strong>for</strong> <strong>Electrochemical</strong> <strong>Gas</strong> SensorSuitable <strong>for</strong> Compact-Sensors (Pin layout: 7-Series)3 5 4 88 45317235523177565756517.0 mmPCD17551743431971.7All dimensions in mm10<strong>Transmitter</strong> SpecificationsInput voltage requiredOutput at zero gas concentrationOutput at full-scaleSockets PlatingCompliance+12 <strong>to</strong> +36 Volts DC4 <strong>mA</strong> (40 mV)<strong>20</strong> <strong>mA</strong> (<strong>20</strong>0 mV)GoldRoHS CompliantOperation of the <strong>Transmitter</strong>The transmitter needs a power supply of 12 – 36 V DC <strong>to</strong> operate. Connect the power source <strong>to</strong>the connec<strong>to</strong>r (V/C). Pay attention <strong>to</strong> the correct polarity, otherwise the transmitter can bedestroyed. If the transmitter is not powered the inserted sensor is short-circuited (exception:NO-Sensors).The voltage output signal can be collected between pins 1 und 2. To obtain the current outputsignal, refer <strong>to</strong> the last section. The output signal has a linear range from 40 – <strong>20</strong>0 mV and 4 –<strong>20</strong> <strong>mA</strong> respectively.REV.: 09/<strong>20</strong>09 Page 1 of 2Phone: +41 43 311 72 00MEMBRAPOR AGFax: +41 43 311 72 01 Birkenweg 2Email: info@membrapor.chCH-8304 Wallisellen /Zurichwww.membrapor.chSwitzerland


<strong>Transmitter</strong>s <strong>for</strong> NO-Sensors are equipped with a potentiometer <strong>to</strong> adjust the bias-voltage. Thisvoltage can be measured between position 4 and 5. <strong>Transmitter</strong>s <strong>for</strong> NO-Sensors are deliveredwith an adjusted bias-voltage of 300 mV.7543VC2 14Output 40 -<strong>20</strong>0 mV+ 24 VoltPower Supply0 VoltSpan- +Zero5To adjust Bias-Voltage(with NO and ETO-Sensors)To control adjusted Bias-Voltage(with NO and ETO-Sensors)4: COM5: 0 - +/- 350 mVCalibrationThe spring on the sensor must beremoved be<strong>for</strong>e insertion in theelectronic circuit.There are two potentiometers on thecircuit board <strong>to</strong> calibrate thetransmitter <strong>to</strong>gether with a sensor. Toadjust the zero point, purge thesensor with synthetic air at a flow ofaround 30 l/h <strong>for</strong> at least 3 minutes.Adjust the potentiometer “zero” <strong>to</strong> getan output signal of 40 mV.Use the maximum gas concentration allowed by the specification of the particular sensor <strong>to</strong>adjust the rang of the output signal. Purge the sensor with this gas mixture. When the signal isstable, adjust with the potentiometer “span” the gain <strong>to</strong> reach an output signal of <strong>20</strong>0 mV.If you use a gas mixture with an analyte gas concentration lower than the maximumconcentration, calculate the appropriate output signal using a linear relation ship.Operating the <strong>Transmitter</strong> in current modeThe transmitter can be operated in a 2-wire mode. The current being drawn (4-<strong>20</strong> <strong>mA</strong>) is linear<strong>to</strong> the sensor signal and there<strong>for</strong>e the sensor signal can be obtained from the supply current.This setup is widely used in (Multi-Channel) <strong>Gas</strong> Detec<strong>to</strong>rs:Point of measurement Distance Your device- +ZeroGain12Signal Output1 - 5 VoltPower Supply250 Ohm 0 Volt+ 24 VoltIo : 4 - <strong>20</strong> <strong>mA</strong>Current is proportional<strong>to</strong> the Sensor SignalREV.: 09/<strong>20</strong>09 Page 2 of 2Phone: +41 43 311 72 00MEMBRAPOR AGFax: +41 43 311 72 01 Birkenweg 2Email: info@membrapor.chCH-8304 Wallisellen /Zurichwww.membrapor.chSwitzerland

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!