Analysis Tech Inc. is a designer, manufacturer and global supplier of electronic instruments associated with reliability testing of electronic packaging. Founded in 1983 and headquartered ten miles north of Boston, Massachusetts, Analysis Tech has developed instruments that have traced the evolution of electronic packaging technology.

Analysis Tech Inc. is a designer, manufacturer and global supplier of electronic instruments associated with reliability testing of electronic packaging. Founded in 1983 and headquartered ten miles north of Boston, Massachusetts, Analysis Tech has developed instruments that have traced the evolution of electronic packaging technology.

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

epeatability, the Sample Positioning Fixture can be quite useful, particularly with type III samples<br />

(incompressible samples), The positioning fixture eliminates sample-position variation by<br />

providing a firm "corner" into which the test sample disks can be located with ease, consistently,<br />

each and every test.<br />

Positioning without alignment fixture (type I and II samples) :<br />

The positioning fixture also eliminates the problem of type III samples migrating with the initial<br />

application of the pressure when oil or grease is used to reduce the surface-contact-resistance.<br />

The TIM Positioning Fixture is included with the TIM Accessory Pack.<br />

TIM Tester Calibration <strong>Package</strong>: Calibration Instruments & Reference Sample<br />

Set<br />

TIM Tester instrument calibrations can be conveniently performed "in-lab" without ever moving<br />

the unit. The TIM Calibrator and TC Calibrator accessories are provided for this purpose. The entire<br />

bi-annual calibration of the TIM Tester is a user-friendly, 10 minute procedure handled by WinTIM<br />

Software. During calibration, the TIM Calibrator is<br />

connected to the rear panel of the TIM Tester and<br />

provides calibration signals which can be measured with<br />

a calibrated lab-voltmeter (not included). The TC<br />

Calibrator provides automatic calibration of the TIM<br />

thermocouple ports in conjunction with WinTIM calroutines<br />

and a USB communication link. WinTIM also<br />

provides complete help files and on-screen instructions<br />

for all aspects of instrument calibration.The Tc Calibrator<br />

and TIM Calibrator should be sent for yearly calibration and re-certification to Analysis Tech.