Measuring using Placebo wavelength.<br />
�� External Data Input of HPLC or Photometric data. (copy/paste from MS - Excel)<br />
�� From not validated Analysis, after a Run the Registration Data can me modified. (Must be<br />
activated by the administrator)<br />
�� After 3 wrong Logins the user will be locked from the system.<br />
�� Analysis Recalculation possible for dosage / purity.<br />
�� Through a copy of the original analysis data, changes from Analysis data are possible.<br />
�� Calibration Module Data can be printed without being saved to the database.<br />
�� The information that a recalibration of the analysis is being performed can be found on<br />
each report page.<br />
�� Listing of serial numbers of devices ( Bath / Photometer) in the analysis report is possible.<br />
�� CE 70 Systems: Implementation: Closed System double loop.<br />
�� Perkin Elmer Photometer driver: Improved Error handing/recognition / Improved<br />
Initialisation of device.<br />
�� AT 70 Systems: Implementation: Medium Replacement using Bottom valve possible.<br />
�� AT 7X Systems: Implementation: Fill in closed Vials. ( Piercing System )<br />
Bug fixes:<br />
General Corrections<br />
�� Report Generator properties could cause ( without Method band ) errors in shown<br />
corrected dosage volumes. ( Only AT 70 )<br />
�� After the Login for the validation the validator stayed in the system as the new actual user.<br />
�� Sorting of component & Medium doesn't work ( even if so registered ) not using name.<br />
�� Using Cell Grouping ( stirring ) protocoling could fail for 2 nd. Analysis Group.<br />
�� Deactivation of Blanc cell for Analysis view could cause failures in the statistical evaluation<br />
of analysis data.<br />
�� Agilent Photometer driver problems are solved.<br />
�� Baseline absorbance has not been calculated correctly using analysis with placebo<br />
substraction and single beam photometer.<br />
<strong>Release</strong> <strong>Note</strong> 2G (WinSOTAX EXE Version: 2.3.1)<br />
New Functionality<br />
General:<br />
�� Implementation CE 7smart.<br />
�� Analysis executed on AT 7x a manual media change is possible.<br />
�� Analysis executed on AT 7x a media change may be executed independently from a<br />
sampling.<br />
Bug fixes:<br />
General Corrections<br />
�� Negative absorbance values on Standard readings cause wrong calculation of %-<br />
Dissolved.<br />
�� Multi component analysis: It could happen that on the sampling reading the Absorbance<br />
value from cuvette 1 was not recorded.<br />
<strong>Release</strong> <strong>Note</strong> 2F (WinSOTAX EXE Version: 2.2.5)<br />
New Functionality