22.09.2021 Views

Arro_User_Manual

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

230 V Arro Bed Manual - 1-150-041-REL

Instructions and Warranty

Please Read Carefully Before Using Your New Bed

Uncontrolled Document - For Reference Only


Arro Bed Instructions and Warranty Content

FIVE (5) YEAR WARRANTY...................................................................................................................................................................... 2

STANDARD CONVENTIONS USED IN THIS MANUAL ....................................................................................................................... 3

SYMBOLS USED ON THE ARRO BED ............................................................................................................................................................................................ 3

TECHNICAL SPECIFICATIONS - CARROLL HEALTHCARE ARRO BED...................................................................................... 4

IMPORTANT ELECTRICAL SAFEGUARDS .......................................................................................................................................... 5

GROUNDING INSTRUCTIONS.................................................................................................................................................................. 5

IMPORTANT MECHANICAL SAFEGUARDS......................................................................................................................................... 7

GENERAL PRECAUTIONS......................................................................................................................................................................... 7

MATTRESS SPECIFICATIONS.................................................................................................................................................................. 8

UNPACKING INSTRUCTIONS .................................................................................................................................................................. 8

UNPACKING NON TUCKABLE SIDE RAILS (PARTIAL AND FULL RAIL OPTION) ......................................................................................................................... 9

ASSEMBLY INSTRUCTIONS ................................................................................................................................................................... 10

INSTALL THE MATTRESS KEEPERS TO THE HEAD AND FOOT DECKS....................................................................................................................................... 10

ATTACH HEAD BOARD AND FOOT BOARD ............................................................................................................................................................................... 11

INSTALL REMOVABLE HEAD BOARD BRACKET WITH I.V. - OPTIONAL................................................................................................................................... 12

CARROLL ASSIST RAIL – INSTALLATION REQUIRED ................................................................................................................................................................ 13

MOUNTING THE ASSIST RAIL .................................................................................................................................................................................................... 13

MINIMUM SAFE GAP BETWEEN HEAD BOARD AND ASSIST RAIL............................................................................................................................................. 14

BED OPERATION - MOBILIZATION SYSTEM / HILO FUNCTION ................................................................................................ 15

ARRO BED MOBILIZATION ........................................................................................................................................................................................................ 15

ARRO BED STABILIZATION ....................................................................................................................................................................................................... 15

PENDANT STOWAGE .................................................................................................................................................................................................................. 16

PENDANT LOCATION.................................................................................................................................................................................................................. 16

BED ELEVATION - HILO FUNCTION ......................................................................................................................................................................................... 17

HEAD ELEVATION - HEAD MOTOR FUNCTION.......................................................................................................................................................................... 17

KNEE-FOOT ELEVATION - FOOT MOTOR FUNCTION ................................................................................................................................................................ 17

TRENDELENBURG AND REVERSE TRENDELENBURG................................................................................................................................................................. 18

CARDIOPULMONARY RESUSCITATION (CPR) MODE................................................................................................................................................................ 18

ACTIVATE CPR MODE .............................................................................................................................................................................................................. 18

PENDANT LOCK OUT VIA ATTENDANT CONTROL PANEL ......................................................................................................... 19

LOCK-OUT HEAD FUNCTION ON HAND PENDANT .................................................................................................................................................................... 19

LOCK-OUT HILO FUNCTION ON HAND PENDANT .................................................................................................................................................................... 19

LOCK-OUT FOOT FUNCTION ON HAND PENDANT..................................................................................................................................................................... 19

ACTIVATE TOTAL LOCK-OUT.................................................................................................................................................................................................... 19

SIDE RAIL OPERATION (OPTIONAL NON-TUCKABLE RAILS).................................................................................................... 20

MOVE SIDE RAIL FROM DOWN TO UP POSITION (NON-TUCKABLE RAILS ONLY)............................................................................................................... 20

MOVE SIDE RAIL FROM UP TO DOWN POSITION (NON-TUCKABLE RAILS ONLY)............................................................................................................... 20

ASSIST RAIL OPERATION (STANDARD)............................................................................................................................................. 20

OPERATION OF THE STANDARD ASSIST RAIL ........................................................................................................................................................................... 20

CHANGE ASSIST RAIL FROM “GUARD” TO “ASSIST” ............................................................................................................................................................... 21

CARE AND MAINTENANCE.................................................................................................................................................................... 21

STERILISATION METHOD........................................................................................................................................................................................................... 21

ANNUAL MAINTENANCE CHECK............................................................................................................................................................................................... 21

SEMI-ANNUAL MAINTENANCE CHECK ..................................................................................................................................................................................... 22

STORAGE ENVIRONMENT CONDITIONS..................................................................................................................................................................................... 22

TRANSPORTATION ENVIRONMENTAL CONDITIONS .................................................................................................................................................................. 22

DECK REMOVAL - FOOT DECK.................................................................................................................................................................................................. 22

DECK REMOVAL - KNEE DECK ................................................................................................................................................................................................. 22

DECK REMOVAL - HEAD DECK................................................................................................................................................................................................. 23

DECK REMOVAL - SEAT DECK.................................................................................................................................................................................................. 23

URINE GUARD REMOVAL.......................................................................................................................................................................................................... 23

BEFORE CALLING THE FACTORY ...................................................................................................................................................... 24

OPTIONS INFORMATION........................................................................................................................................................................ 25

HEAD & FOOT BOARD OPTIONS.......................................................................................................................................................... 26

ARRO 171 HEAD BOARD OPTIONS............................................................................................................................................................................................ 26

ARRO 171 FOOT BOARD OPTIONS ............................................................................................................................................................................................ 26

ARRO 171 WIRING SCHEMATIC ........................................................................................................................................................... 27

ARRO BED - FULL ASSEMBLY............................................................................................................................................................... 28

ARRO BED - FRAME ASSEMBLY........................................................................................................................................................... 29

ARRO BED - FRAME ASSEMBLY BILL OF MATERIALS.............................................................................................................................................................. 30

ASSIST RAIL ASSEMBLY......................................................................................................................................................................... 31

ARRO BED - NON TUCKABLE FULL & PARTIAL RAIL ASSEMBLY............................................................................................ 32

SERVICE PART SUB-ASSEMBLIES ....................................................................................................................................................... 33

HEAD GATCH SOCKET STYLE ASSEMBLY ................................................................................................................................................................................ 33

FOOT GATCH SOCKET STYLE ASSEMBLY ................................................................................................................................................................................. 34

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 1 of 34


FIVE (5) YEAR WARRANTY

Carroll’s pro-rated warranty covers manufacturing defect, or failure due to workmanship or materials.

Electrical components are guaranteed to be free from manufacturing defect for one (1) year from the

date of purchase, covered by the motor manufacturer’s own warranty. Carroll Healthcare Inc. warrants

to the original purchaser of its products that the product or any part thereof, in the judgement of Carroll

Healthcare Inc. will not be defective in material or workmanship for one (1) year from date of original

purchase. Carroll Healthcare Inc. specifically warrants to the original purchaser of its products that the

reinforced fibreglass bed deck will not be defective in material or workmanship for twenty (20) years

from date of original purchase. Such defects will be repaired or replaced (at the company’s option) free

of charge for parts only, provided the defective product is shipped to the factory or an authorized depot

at the customer’s own expense. Shipping charges for return of the product to the customer will also be

at the customer’s own expense.

This warranty does not include any labour or shipping charges incurred in replacement part installation

or repair of any such part. Carroll Healthcare’s sole obligation and your exclusive remedy under this

warranty shall be limited to such repair and/or replacement.

After one year, for the next four years, frame parts only will be covered by the warranty according to the

following fee schedule: 2 nd year – frame parts (list price less 60%); 3 rd year – frame parts only (list price

less 40%); 4 th year – frame parts (list price less 20%); 5 th year frame parts (list price less 10%). Parts in

the judgement of Carroll Healthcare Inc. found to be damaged through normal wear and tear, impact,

abuse, overloading, fire, lightning, water damage or power surges, brown-outs, improper supply voltage

or use of other manufacturer’s head and foot boards or accessories will not be covered and parts will be

charged out at the full list price. Note also that any alteration of the bed or one of its components voids

this warranty. A Return Goods Authorization number (RGA) must be obtained before returning any

products or parts to the factory. We regret that no products will be accepted at the factory without an

RGA number. No other warranty, expressed or implied by Carroll Healthcare Inc. or its agents shall be

considered valid.

Carroll Healthcare Inc.

Mailing Address: 994 Hargrieve Road, London, Ontario, CANADA, N6E 1P5.

(519) 659 1395

1-800-668-2337

www.carrollhealthcare.com

email: service@carrollhealthcare.com

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 2 of 34


STANDARD CONVENTIONS USED IN THIS MANUAL

This manual includes information essential to the safety of personnel and equipment. As you read

through this manual be alert to the four signal words.

NOTE:

DANGER

Information appearing under the DANGER caption concerns the protection of

personnel from the immediate and imminent hazards that, if not avoided, will result in

immediate, serious personal injury or loss of life in addition to equipment damage.

WARNING

Information appearing under the WARNING caption concerns the protection of

personnel from potential hazards that can result in personal injury or loss of life in

addition to equipment damage.

CAUTION

Information appearing under the CAUTION caption concerns the protection of

personnel from potential hazards that can result in minor personal injury or equipment

damage.

Information appearing in a NOTE caption provides additional information, which is

helpful in understanding the item being explained.

Symbols Used on the Arro Bed

PROTECTIVE (EARTH GROUND)

POTENTIAL EQUALIZATION

WARNING - SAFE WORKING LOAD

THE MAXIMUM ALLOWABLE LOAD ON EQUIPMENT

IF THE INSTRUCTIONS FOR INSTALLATION

AND USE ARE FOLLOWED

INTERNATIONAL PROTECTION CODE (IP CODE)

TYPE B EQUIPMENT

EQUIPMENT CLASSIFICATION IN TERMS OF PROTECTION

AGAINST ELECTRICAL SHOCK

CARDIOPULMONARY RESUSCITATION (CPR)

ONE-TOUCH TO LOWER ALL PANELS TO

THEIR LOWEST POSITION

TRENDELENBURG CONTROL

MOVE THE BED INTO TRENDELENBURG

POSITION

REVERSE TRENDELENBURG CONTROL

MOVE THE BED INTO REVERSE

TRENDELENBURG POSITION

INITIATE ACTION TO HEAD SECTION OF

MATTRESS SUPPORT PLATFORM

INITIATE ACTION TO THE ENTIRE MATTRESS

SUPPORT PLATFORM (HILO)

INITIATE ACTION TO THE KNEE-FOOT SECTION

OF THE MATTRESS SUPPORT PLATFORM

CONTOUR

TOGGLE SWITCH FOR ONE-TOUCH HEAD

AND KNEE-FOOT RAISE OPERATION

FUNCTION LOCK-OUT

ACTIVATE TOGGLE SWITCH TO DENY FUNCTION

ACCESS ON THE HAND PENDANT

MOBILITY

ACTIVATE BED MOBILITY POSITION

UP CONTROL

RAISE THE CORRESPONDING SECTION OF THE

MATTRESS SUPPORT PLATFORM

Uncontrolled Document - For Reference Only

DOWN CONTROL

LOWER THE CORRESPONDING SECTION OF

THE MATTRESS SUPPORT PLATFORM

DIA 01

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 3 of 34


TECHNICAL SPECIFICATIONS - CARROLL HEALTHCARE ARRO BED

Arro 171

Elevation Range (floor to deck)

Stability Mode (off casters) 6-3/4" to 28"

Transition from Stability to Mobility 28" to 33"

Mobility Mode (on Casters) 33"

Head Deck Angle 0° to 67°

Knee - Foot Angle 0° to 25°

Trendelenburg 0° to 14°

Reverse Trendelenburg 0° to 14°

Under-bed patient lift clearance

13" @ highest stable pos.

Electrical Specifications

Control Box Mains Voltage

230V

Control Box Input Current

1.5 A

Control Box Current Frequency

50 Hz

Control Box Output Voltage

24 Volts DC (70W max)

Control Box Duty Cycle

10% (6 min/hr max)

Motorised Linear Head Actuator

24 Volts DC

Motorised Linear HILO Actuator (head end 24 Volts DC

mount)

Motorised Linear HILO Actuator (foot end 24 Volts DC

mount)

Motorised Linear Foot Actuator

24 Volts DC

Six function hand pendant

standard

Staff Control Keypad in foot board

standard

Intermittent Operation or Maximum Continuous 6 min/hr

Operation

Safety Features

Protective Earth Ground

Class I

Electric shock protection

Type B

Enclosure protection

IPX4

Mobility Access Safety Feature

standard

Built-in Battery Back-up

standard

CPR Function

standard

Security - Pendant Lock-Out

Bed Elevation (HILO)

standard

Head Section

standard

Knee-Foot Section

standard

Total Function lock-Out

standard

Dimensions and Weight

Length (overall) 87"

Length - Between head & foot board 83"

Length - Mattress Platform 80"

Width (with Assist Rails) 41"

Width - Mattress Platform 35"

Weight (pounds)

120 kg (265 lb.)

Load Capacity - Safe Working Load

230 kg (500 lb.)

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 4 of 34


IMPORTANT ELECTRICAL SAFEGUARDS

• Plug the three prong electrical cord ONLY into a properly grounded 230 VAC, 50 cycle

outlet that is rated to at least five amps.

• DO NOT use an extension cord.

• DO NOT permit the cord, wall outlet, electric motors, control box or pendant to become wet

or submerged.

• DO NOT operate the bed if the cord or outlet is damaged. DO NOT operate the bed if any

motor has malfunctioned or is damaged in any manner.

• DO NOT use this bed in Active Treatment "Ether Environments".

GROUNDING INSTRUCTIONS

This electric bed must be grounded.

In the event of a malfunction or breakdown, grounding provides a path of least resistance for

electric current, thereby reducing the risk of electric shock.

This product is equipped with a cord having an equipment-grounding conductor and a

grounded plug. The plug must be inserted into an appropriate electric outlet that is properly

installed and grounded in accordance with all local codes and ordinances.

DANGER

Improper connection of the equipment-grounding conductor can result in

electrocution. Check with a qualified electrician or service person if you are

doubtful that the electrical outlet is properly grounded.

DO NOT modify the three prong plug provided. If it will not fit into the electrical

outlet, have a proper outlet installed by a qualified electrician.

DANGER

Electric components, such as motors, exposed to water, may cause severe

injury including electrocution.

During routine bed operation or when cleaning, DO NOT expose electrical

components to water. Unplug and de-energize electrical components when

water products are near these components.

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 5 of 34


DANGER

EXPLOSION Hazard

DO NOT operate this bed in the presence of a flammable anaesthetic mixture

with air or with nitrous oxide.

DANGER

Possible FIRE hazard exists if the hand control pendant is not disabled when

using oxygen administering equipment. Disconnect the pendant from the foot

end of the bed and hang it on the pendant hook (in holder) when not in use.

Keep pendant cord clear of moving equipment.

DANGER

Possible FIRE hazard exists when connecting the electric cord to a 230V

electrical outlet if oxygen administering equipment is in use.

Use either the nasal mask or 1/2 length tent, ensuring the tent never extends

below the mattress support level, when administering oxygen.

WARNING

DO NOT roll the bed over the power cord. DO NOT entangle the cord on

other objects. A pinched electric cord is dangerous and can become

damaged. Be aware of the power cord location especially when moving the

bed.

This equipment may cause electromagnetic disturbances to it's environment.

The bed has NOT been evaluated for electromagnetic compatibility (emc).

The performance of this equipment may be degradation when exposed to

electromagnetic disturbances and / or electrostatic discharge.

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 6 of 34


IMPORTANT MECHANICAL SAFEGUARDS

• DO NOT stand on the bed or concentrate weight on the fibreglass deck panels.

• DO NOT use the bed without a mattress at least 6" (152 mm) thick and of recommended

dimensions. Refer to Mattress Specifications on page 8.

• DO NOT use without a special mattress specifically designed to bend and conform to the

shape of the bed.

• DO NOT use a water filled or gel filled mattress on this bed.

• Ensure that the head board and foot board are properly attached to the frame. These parts

assist in stabilising the bed.

GENERAL PRECAUTIONS

• Do not use head/foot panels from other manufacturers.

• Make sure lobes on deck hinge have not shifted in transit. Refer to Deck Removal on page

22 for instructions.

WARNING

This bed is rated to 500 lbs. (230 kg). DO NOT overload the bed. The

combined weight of patients, visitors, mattress and bedding must NOT exceed

500 lbs. (230 kg). The maximum load on either side rail is 170 lbs. (77 kg).

WARNING

DO NOT drop the bed.

DO NOT allow patients to fall on the bed.

DO NOT jump on the bed.

Each of the preceding impacts can permanently damage either HILO motor

resulting in an inoperable bed.

In the case of an inoperable bed due to a damaged HILO motor, the HILO

motor must be replaced immediately. If left unattended, a damaged HILO

motor can cause the bed to fall from the highest extended position, resulting in

patient injury.

DO NOT exceed the maximum continuous mode of operation. Refer to

Technical Specifications on page 4.

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 7 of 34


MATTRESS SPECIFICATIONS

A mattress is not included with this bed. It is highly recommended that you use a 35.5" (900 mm) wide

mattress that is designed to conform to the bed length of 80" (2032 mm), such as a Carroll Healthcare

Thera Max Mattress. The mattress must be at least 6" (150 mm) in height.

WARNING

Possible patient ENTRAPMENT Hazard if using non-specified mattress.

Patient entrapment may result in injury or death.

Use only a mattress of recommended specifications with this bed.

UNPACKING INSTRUCTIONS

The Arro bed is shipped in a vertical position with the head end on the ground. Follow these

instructions to unpack the bed.

CAUTION

Equipment DAMAGE may result from improper wire tie removal.

Do not cut any trimmed plastic wire ties (see step 2 below). (Trimmed wire

ties are permanent features on the Arro bed).

1. Cut all untrimmed wire ties. 2. Do NOT cut trimmed wire

ties.

3. With the bed upright, cut the

wire tie holding the mattress

keeper bag and remove.

Uncontrolled Document - For Reference Only

4. Lower the bed to the floor. 5. Cut the plastic strapping

around the head and foot

sections of the deck.

6. Lift the foot deck and cut the

wire tie around the foam foot

pad.

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 8 of 34


7. Lift the head deck and cut the

wire tie used to stabilise the

caster.

10. Cut the wire tie around the

pendant wire bundle.

NOTE:

8. Cut the wire tie around the

foam foot pad - head end.

11. Plug the bed into a properly

grounded 230V outlet. Press

the HILO UP button on the

pendant and raise bed to a

comfortable working height.

9. Cut the wire tie around the

main power cord bundle.

All bolts, nuts and washers used to secure the mattress keepers to the deck and head &

foot boards to the frame are included inside the mattress keeper bag at the foot end of

the bed. See Step 3 above.

Unpacking Non Tuckable Side Rails (Partial and Full Rail Option)

Carroll side-rails are pre-installed at the factory. They are held in place, for shipping purposes, with tape

and plastic strapping.

Cut and remove the

wire ties around the

foam (see DIA 13) and

cut the filament tape

(see DIA 14). The rails

can now be raised from

the stored position. (see

DIA 15).

Uncontrolled Document - For Reference Only

DIA 13 DIA 14 DIA 15

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 9 of 34


ASSEMBLY INSTRUCTIONS

Install the Mattress Keepers to the Head and Foot Decks

The mattress keepers hold the mattress in place. Install them on either side of the head and foot deck

panels at each end of the bed. Refer to DIA 16.

NOTE:

1. Align mattress keeper square holes with the

moulded holes in the foot deck.

2. Press a 1/4-20 X 5/8 long carriage bolt

through the top of the mattress keeper hole.

3. Tighten, from below, a 1/4-20 hex nut using

a 7/16 wrench or socket wrench.

4. Repeat for the second hole and the

remaining three mattress keepers.

CAUTION

DAMAGE to deck may occur.

Over-tight fasteners could cause the deck to crack.

Do not over-tighten the mattress keeper hardware.

There are two left and two right side mattress keepers. When assembling, make sure

the high back faces either the front or back of the bed. The tapered side should face

the sides of the bed.

1

DIA 16

Uncontrolled Document - For Reference Only

No. Description Part Number QTY

1 Mattress Keeper B MATTKP-RPL-CN RF-R 2

2 Mattress Keeper A MATTKP-RPL-CN RF-L 2

3 Carriage Bolt 1/4-20 X 5/8 long BLT145-8C 8

4 Hex Nut 1/4-20 X 5/8 NUT14 8

5 Foot Deck Section DECK00-05 1

6 Head Deck Section - for 80" bed (Shown) DECK00-07 1

2

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 10 of 34


Attach Head Board and Foot Board

NOTE:

The head and foot boards are packaged separately from the Arro bed either

individually blanket-wrapped on a skid or in separate cardboard boxes.

NOTE:

Always attach the boards so that the posts are on the head end side of the boards.

DIA 17

No. PART NAME PART NUMBER QTY

1 Head & Foot Post Plug - IV Rod Holder PLG343-421-4 4

2 Torx® Drive Machine Screw for 3/4 Board With Square Post

Torx® Drive Machine Screw for 1 1/8 Board With Square Post

PHMS14-201-14

PHMS14-201-12 8

3 Foot Post PST002-9FC-BP-NT 2

4 Attendant Control Pad for Arro 171 (Trendelenburg) JNCTND-C20-A 1

5 Membrane Foot Board Various - See DIA 91 1

6 Bronze Coated Insert Nut (dark wood) OR Zinc Coated Insert Nut (light wood) NUT14I-NSE-RT-BR or NUT14I-NSE-RT-ZN 8

7 Washer 1 1/4 OD X 11/32 ID X 1/8 Thick WSR-114-113-21-8 4

8 Nylon Locking Hex Nut 5/16-18 NUT516-NL 4

9 Head Board Various - See DIA 90 1

10 Head Post PST002-9HC-BP-NT 2

1. Insert the treaded rod of the posts through the holes in the bed frame. See DIA

17.

Uncontrolled Document - For Reference Only

2. Slip the 1-1/4" washer (item 7) on the treaded rod then fasten the 5/16-18 nylon

locking hex nut (item 8) to the rod. Tighten the hex nut with a 1/2" wrench.

3. Connect the phone wire at the foot end of the main rail to the phone junction

connection box on the back of the foot board. See DIA 18.

DIA 18

10

7

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 11 of 34


Install Removable Head Board Bracket with I.V. - Optional

CAUTION

Incorrect assembly of the Removable Board Bracket to the Arro Bed may

create a PINCH-POINT.

There is a left and a right side bracket. When assembled correctly, the smaller

diameter I.V. hole in the Black Insert will be on the outside of the bed frame.

DIA 19

REF Description: Part No. Qty

1 Removable Board Bracket - Right Side RD0055-55P-NT-AR 1

2 Removable Board Bracket - Left Side RD0055-55P-NT-AL 1

3 Hex Head Bolt 5/16-18 X 1-3/4 Long - Partial Threaded BLT516 2

4 Flat Washer for 5/16 Fastener WSR516-SAE 2

5 Black Insert for Removable Board Bracket MISC11-4 2

6 Removable Head Post (shown on ThinkSoft Head Board) PST002-9HP 2

1. Slide the Removable Board Bracket on the end of the Main Rail as shown in DIA 19. Do not

remove the Main Rail End Plug.

2. Fasten from below a 5/16-18 X 1-3/4 Hex Bolt. The bolt will fasten to an internal weld nut in the

bracket.

3. Place the Head Board Bracket with Head Board into the large round holes in the two brackets.

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 12 of 34


Carroll Assist Rail – Installation Required

Assist rails are packaged separately from the bed in the Assist Rail box. Remove the rails and fit them

to the main rails on either side of the bed. The left and Right Side Assist Rails are not interchangeable.

Mounting the Assist Rail

WARNING

Incorrect Assist Rail mounting may result in PATIENT ENTRAPMENT.

To comply with FDA Rail Guidelines, which specify a minimum distance

between the rail and the head board, mount the Assist Rail bracket ONLY in

the pre-punched mounting holes on the main rail. When mounted correctly,

the distance between the Assist Rail and the head board will exceed the

minimum requirement of 318 mm.

Never Install two Assist Rails on the same side of the bed.

1. Mounting holes on the main rail are

below the deck. See DIA 20. Rail

shown in Guard position.

2. Place one 3/8 Flat washer (item 4)

on the 3/8-16 X 2" long hex bolt

(item 2) and push this from outside

the bracket, through the mounting

hole in the main rail. Place another

3/8 washer over the bolt on the

inside main rail mounting hole then

secure with a 3/8-16 jam nut. Use

two 9/16" wrenches to tighten the

bolt while holding the nut. See DIA

20.

3. Place one 3/8 Flat washer (item 4)

on the 5/16-18 X 2" long hex bolt

(item 1) and push this from outside

the bracket, through the mounting

hole in the main rail. Place another

3/8 washer over the bolt on the

inside main rail mounting hole then

secure with a 5/16-18 nylon lock

nut. Tighten with two 1/2"

wrenches.

Uncontrolled Document - For Reference Only

4. Repeat the installation for the

opposite side Assist Rail.

DIA 20

No Description Part Number Qty

1 Hex Bolt 5/16-18 X 2 Long BLT516-2P 2

2 Hex Bolt 3/8-16 X 2 Long BLT382-P 2

3 Assist Rail Side Mount - Left Side (Not Shown) /

Assist Rail Side Mount - Right Side (Shown) /

RL3-010-26-1

RR3-010-26-1

1

1

4 Flat Washer for 3/8 Bolt WSR38U-SS 8

5 Locking Jam Nut 3/8-16 NUT38N-L 2

6 Nylon Lock Nut 5/16-18 NUT516-NL 2

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 13 of 34


Minimum Safe Gap between Head Board and Assist Rail

When installed properly, the space between the Assist Rail and the Head Board should be as shown:

ThinkSoft Boards (Non-Removable)

In Assist Position (UP): 13-1/8”

In Guard Position (DOWN) 27-9/16”

Arro Designer Deluxe Series Head Board

In Assist Position (UP): 13-9/16”

In Guard Position (DOWN) 28”

DIA 21

Uncontrolled Document - For Reference Only

27

16 9 "

28"

13 1 8 "

13

16 9 "

DIA 22

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 14 of 34


BED OPERATION - MOBILIZATION SYSTEM / HILO FUNCTION

WARNING

Unintended bed movement may occur if bed is left in the mobilized position.

Unintended bed movement may cause property DAMAGE or resident

INJURY.

Never leave the bed unattended while the bed is in the Mobilized position.

Arro Bed Mobilization

The mobilized position enables bed movement once the wheels are on the floor. The Arro bed is mobile

in its highest position, 33".

Press the mobilization button on the Attendant Control Pad and hold for 15 seconds. See DIA 23.

The hand pendant HILO control will not operate when the bed is in the Transitional phase between

Stability and Mobility (above 28" on 171). See DIA 24, 25 and 26.

Attendant Control Pad Hand Pendant Operation Foot / Wheel Position

DIA 23 - Hold for 15 seconds DIA 24 DIA 25 DIA 26

Arro Bed Stabilization

CAUTION

Moving the bed while in the Stabilized position may DAMAGE the bed.

Do not move the bed unless the Mobilization system has been activated.

The Arro bed is stable from the lowest position, 6-3/4" to the highest stable position, 28".

While in the mobilized or transitional position, press the HILO down button on the Attendant

Control Pad and hold until the Mobility LED display stops blinking and turns off. The hand pendant

HILO control will operate once the bed is in Stability mode. See DIA 28, 29 and 30.

Attendant Control Pad Hand Pendant Operation Foot / Wheel Position

Uncontrolled Document - For Reference Only

DIA 27 - Hold until Mobility LED is off DIA 28 DIA 29 DIA 30

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 15 of 34


Pendant Stowage

Pendant can stow in three ways:

1. Inside the Pendant holster. Clip the holster between the Assist Rail bars on the outside of the bed.

See DIA 32. The Pendant slips into the holster for easy access. See DIA 33.

2. Secured with the spring loaded linen clip on the bedding. See DIA 34.

3. Placed on the adhesive Velcro® strip at the back of the head board. See DIA 35.

Pendant location

DIA 32 DIA 33

DIA 31 DIA 34 DIA 35

DANGER

Electrical SHOCK Hazard.

Before working on any electrical part, such as the 9-pin pendant coupling,

ensure the main power to the bed is disconnected and the capacitor inside the

control box is fully discharged.

Full capacitor discharge can be achieved by sequentially depressing the 'Up'

and 'Down' buttons on the pendant at least 20 times, after main electrical

power to the control box is cut.

The Hand Control Pendant can be located on either

side of the bed, just beneath the seat deck panel.

To move the pendant to the opposite side of the

bed, first, disconnect the 9-pin pendant wire

coupling, then reconnect the pendant cable to the

female 9-pin coupling on the other side of the bed.

See DIA 36 and 37.

Uncontrolled Document - For Reference Only

DIA 36 DIA 37

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 16 of 34


Bed Elevation - HILO Function

To raise the bed press the middle left button

on the hand pendant or the top second

column button on the attendant control pad.

See DIA 38 and 39. Release the button

when the desired level is reached. See DIA

40.

To lower the bed press the middle right

button on the hand pendant or the middle

second column button on the attendant

control pad. See DIA 41 and 42. Release

the button when the desired level is reached.

See DIA 43.

Head Elevation - Head Motor Function

NOTE:

DIA 38 DIA 39 DIA 40

DIA 41 DIA 42 DIA 43

Contour Control enables both the head and knee-foot section to raise simultaneously.

The Contour button can be toggled-on or off.

To raise the head section press the top left

button on the hand pendant or the top

leftmost button on the attendant control pad.

See DIA 44 and 45. Release the button

when the desired head panel angle between

0 and 67° is reached. See DIA 46.

DIA 44 DIA 45 DIA 46

To raise the head and knee-foot section

simultaneously, first toggle-on the Contour Button on

the attendant control pad. See DIA 47. Next press the

top left button on the attendant control pad. See DIA

48. Release the button when the desired head and

knee-foot panel combined level is reached. See DIA

49.

DIA 47 DIA 48 DIA 49

To lower the bed press the top right button

on the hand pendant or the middle left

button on the attendant control pad. See

DIA 50 and 51. Release the button when

the desired head panel angle between 0 and

67° is reached. See DIA 52.

DIA 50 DIA 51 DIA 52

Knee-Foot Elevation - Foot Motor Function

NOTE:

It may be beneficial to elevate the foot and knee sections when the head is elevated, in

order to prevent the resident from sliding down the mattress.

To raise the knee-foot section press the

bottom left button on the hand pendant or

the bottom leftmost button on the attendant

control pad. See DIA 53 and 54. Release

the button when the desired head panel

angle between 0 and 25° is reached. See

DIA 55.

DIA 53 DIA 54 DIA 55

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 17 of 34


To lower the bed press the bottom right

button on the hand pendant or the bottom

second column button on the attendant

control pad. See DIA 56 and 57. Release

the button when the desired head panel

angle between 0 and 25° is reached. See

DIA 58.

DIA 56 DIA 57 DIA 58

Trendelenburg and Reverse Trendelenburg

CAUTION

Use of Trendelenburg or Reverse Trendelenburg feature may not be

appropriate for certain patients.

Trendelenburg and Reverse Trendelenburg should only be used on the advice

of a medical practitioner.

To activate Trendelenburg, press and hold the

top right-most button on the attendant control

pad. See DIA 59. Release the button when

the desired bed angle between 0 and 14° is

reached. See DIA 60. Press and hold either

HILO function buttons to deactivate

Trendelenburg.

To activate Reverse Trendelenburg, press and

hold the bottom right-most button on the

attendant control pad. See DIA 61. Release

the button when the desired bed angle between

0 and 14° is reached. See DIA 62. Press and

hold either HILO function buttons to

deactivate Reverse Trendelenburg.

Cardiopulmonary Resuscitation (CPR) Mode

DIA 59 - Hold for 3 sec to activate DIA 60

DIA 61 - Hold for 3 sec to activate DIA 62

CAUTION

Activation of CPR mode may cause patient injury.

Once activated, CPR function can not be aborted. Caregiver should NEVER leave the

patient’s bedside during operation of the CRR function.

Activation of the CPR control allows one-touch flattening of all bed surfaces.

Activate CPR Mode

Uncontrolled Document - For Reference Only

Press the CPR button on the attendant control pad one

time. The head and / or knee-foot panels will lower to

their lowest heights. See DIA 63 and 64.

The CPR mode does not need deactivation. Simply

resume head and knee-foot operation once the CPR

action is complete.

DIA 63 DIA 64

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 18 of 34


PENDANT LOCK OUT VIA ATTENDANT CONTROL PANEL

NOTE:

By default, all pendant control is initially locked-out, after connection to a wall outlet.

To deactivate the lock out feature, all three lock out LED's must be toggled on

Pendant functions may be locked out individually or in tandem by toggling the appropriate Lock Out

buttons on the attendant control pad. When the corresponding LED is illuminated, the lock out is

deactivated. When the corresponding LED is not illuminated, the pendant feature is locked-out.

Lock-out Head Function on

Hand Pendant

Press the bottom left button on the

attendant control pad. Lock-out is

working when the LED is ON.

Lock-out HILO Function on

Hand Pendant

Press the bottom middle button on the

attendant control pad. Lock-out is

working when the LED is ON.

Lock-out Foot Function on

Hand Pendant

Press the bottom button in the third

column on the attendant control pad.

Lock-out is working when the LED is

DIA 65 DIA 66 DIA 67

DIA 68 DIA 69 DIA 70

ON. DIA 71 DIA 72 DIA 73

Activate Total Lock-out

Press and hold all three lock-out buttons for 3 seconds. All functions on the

attendant control pad and the hand pendant are unavailable. Lock-out is

working when all three LED’s are flashing.

To deactivate total lock-out, press and hold all three lock-out buttons for three

seconds.

Uncontrolled Document - For Reference Only

DIA 74

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 19 of 34


SIDE RAIL OPERATION (OPTIONAL NON-TUCKABLE RAILS)

All Carroll side rails comply to FDA guidelines concerning rail entrapment. With proper monitoring

and the right health care bed, bed-rail entrapment can be prevented. For further information, refer to

documentation from the Hospital Bed Safety Workgroup.

Move Side Rail From DOWN to UP position (Non-Tuckable Rails ONLY)

Gently lift the rail to the UP position. The rail will first arc toward the foot end of the bed then the head

end of the bed as it moves to the UP position. See DIA 75. (An audible "click" can be heard when the

UP position is reached).

DIA 75 DIA 76 DIA 77

Move Side Rail From UP to DOWN position (Non-Tuckable Rails ONLY)

Pull the release knob while pushing down on the side rail. See DIA 76. The rail will first arc toward the

foot end then the head end as it moves to the DOWN position. See DIA 77.

ASSIST RAIL OPERATION (STANDARD)

It is important that caregivers know how to operate the Assist Rails safely. Assist Rails lock in two

positions and allow patient transfer in a third unlocked position:

Operation of the Standard Assist Rail

NOTE:

The Assist Rail can travel beyond the "Assist" position to allow patient transfer to and

from the bed with the rail out of the way.

WARNING

Patient ENTRAMENT may occur with assist rail in the transfer position.

The transfer position should be used only while attending to the resident.

When unattended, the assist rail must be returned to either of the locking

positions: Guard or Assist.

Uncontrolled Document - For Reference Only

1. In the “Guard” position, the patient is contained in the bed with a subtle reminder of the bed edge

location.

2. In the “Assist” position, the patient is aided in entering and leaving the bed.

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 20 of 34


Change Assist Rail From “Guard” to “Assist”

DIA 76 DIA 77 DIA 78

1. Pull the release knob at the bottom of the rail and rotate the rail around the pivot point. See DIA 76.

2. As soon as the rotation begins, you may let go of the pull knob. See DIA 77.

3. Continue to rotate the assist rail until it 'clicks' into the ASSIST position. See DIA 78. To change

position back to GUARD, pull the release knob as shown in DIA 77, and reverse the process.

CARE AND MAINTENANCE

Very little maintenance is required to keep the Arro bed and rails in good working order.

CAUTION

Equipment or property DAMAGE or resident INJURY may occur if the

following points are not followed

DO NOT use solvents, alcohol or petroleum products on the surface.

DO NOT submerge the bed frame or electrical parts.

DO NOT allow liquids to enter electrical components.

DO NOT Steam Clean or Pressure Wash.

Sanitizing Method

All surfaces may be cleaned with:

• soapy water or

• any common consumer disinfecting product, or

• most non-abrasive commercially available disinfecting products.

Annual Maintenance Check

• Inspect for damaged or loose wiring. Have a qualified electrician replace any frayed or damaged

cords.

• Inspect for secure grounding.

• Ensure that the rails engage and lock as specified.

• Tighten, adjust or replace any parts or bolts etc. that are loose or show signs of wear.

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 21 of 34


• If epoxy coated rails become scratched or chipped, use a matching enamel touch up paint. (Refer to

Options Information on page 25 for paint specifications).

Semi-Annual Maintenance Check

• Unplug bed from wall outlet and verify battery function (171 Only).

Storage Environment Conditions

• The Arro Bed should be stored in a clean and dry environment with an ambient temperature range

between 15 and 30 Celsius (59° to 86° F).

Transportation Environmental Conditions

• The environment for transportation of the Arro bed should be clean and dry.

Deck Removal - Foot Deck

1. The foot deck can be removed by lifting the foot and knee decks together until the angle between

them is 90º. See DIA 81.

2. Lift the foot deck away from the knee deck at the hinge, to separate the decks. See DIA 82.

3. Bend the foot deck over the mattress keeper - foot bale assembly to expose the underside of the

deck. Compress the foot bale ends inward to disengage the 3/8" foot bale from the mounting holes.

See DIA 83.

4. If necessary, re-install in the reverse order.

DIA 81 DIA 82 DIA 83

Deck Removal - Knee Deck

Once the foot deck is removed,

the knee deck can be separated

from the stationary seat deck.

Lift the knee until it is at 90º to

the seat deck. See DIA 84.

Using a horizontal force, strike

the knee deck near the hinge to

separate the decks. See DIA 85.

Uncontrolled Document - For Reference Only

DIA 84 DIA 85

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 22 of 34


Deck Removal - Head Deck

The head deck is removable for thorough cleaning or for service access to the lower frame mechanical

components.

Lift the head deck until it is at a

90-degree angle to the stationary

(bolted) seat section deck. See

DIA 86. The head deck can now

be lifted straight up, detaching it

from bed. See DIA 87.

DIA 86 DIA 87

To re-install the panel, hold it at 90 degrees to the seat panel and re-align the lobes of the hinge with the

continuous hinge of the seat section.

NOTE:

• When reinstalling the deck, Make sure all the lobes are completely engaged

(especially in the centre).

• Check for engagement by gently moving the deck up and down.

• The deck should swing freely without binding.

Deck Removal - Seat Deck

The seat panel is attached to the bed frame with two 5/16 carriage bolts and hex nut fasteners. Although

it is not designed for routine removal, it can be removed, if necessary, using a 1/2" wrench.

Unscrew the hex nut-from the carriage bolt beneath the main rail.

See DIA 88. Repeat for the opposite side, being careful to save the

fasteners. The seat deck can now be lifted off the frame.

Urine Guard Removal

The urine guard is held in place beneath the seat deck with adhesive

Velcro® strips. It is accessible once all the deck panels are

removed. Lift the urine guard up from the bed frame. See DIA 89.

Uncontrolled Document - For Reference Only

When reapplying the urine guard, align the hole cut-out to the

elevator bolt and press into place.

DIA 88

DIA 89

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 23 of 34


BEFORE CALLING THE FACTORY

To ensure prompt and accurate attention to your concerns please have the following information

available before you call:

Your facility’s phone number:

(Where you can be reached):

Your facility's fax number (If Available):

Bed(s) Model and Type:

Bed(s) Serial Number

Date of Manufacture:

List of defective part(s) (Identify by part number,

or describe relative to nearest numbered part):

List of deficiencies or Type of problem:

Arro _____

NOTE: Left/right determination is made standing at the foot of the bed, facing the resident.

If you have carefully followed all of these instructions, you should have many years of trouble free

service. If you’re not 100% satisfied with your bed, please call your sales representative or the factory

immediately. It's always better to ask if you have any questions such as hole drilling locations in head

and foot boards, electrical service or modifications, or any minor damage that could lead to more serious

consequences.

If you require further information regarding circuit diagrams, component part lists, descriptions etc.

please call the number below.

Call Carroll Healthcare Inc. at (519) 659-1395 or Toll Free 1-800-668-2337.

Mailing Address: 994 Hargrieve Road, London, Ontario, CANADA, N6E 1P5.

www.carrollhealthcare.com

email: service@carrollhealthcare.com

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 24 of 34


OPTIONS INFORMATION

1. Patient Restraint Bracket. Part # 5301 (order as KIT 18).

(Attaches to the head deck panel). See DIA 90 and 91.

Ø 1 4

3

32

5

1

1

Ø 1 4

3 5 32

1

1

A

HEAD DECK PANEL

PART NO. DECK00-07 (80" BED)

PART NO. DECK00-02 (76" BED)

B

SEAT DECK PANEL

PART NO. DECK00-16

C

D

1. Drill four Ø1/4 holes in the head deck section, as

shown in DIA 90 (Holes A, B, C and D). The hole

centres should be 1" in from the outside edge,

spaced 3-5/32" apart.

2. Assemble the Restraint Bracket (Item 2 in DIA 91)

to the deck panel. Use the 1/4-20 X 3/4 long hex

bolt, flat washer and locking hex nut to fasten the

Restraint Bracket to the head deck. Do Not Overtighten.

DIA 90 DIA 91

2. Touch up paint Sherwin Williams Auto #35-30041 – Cream.

No Description Part Number Qty

1 1/4-20 x 3/4 long Hex Bolt BLT143-4 4

2 Restraint Bracket MFOOTM-ANL-OO-P 2

3 Washer 1 1/4 OD X 11/32 ID X 1/8 Thick WSR-114-113-21-8 4

4 1/4-20 Nylon Lock Nut NUT14L 4

1,2,3,4 Patient Restraint Bracket Kit KIT18 -

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 25 of 34


HEAD & FOOT BOARD OPTIONS

Head and Foot Boards are ordered in pairs. Your Carroll Healthcare Customer Service Representative

can elaborate on the various options. Call Carroll Healthcare Inc. at (519) 659-1395 or Toll Free

1-800-668-2337.

Arro 171 Head Board Options

Style

ARRO - REMOVABLE

THINKSOFT - REGULAR

1-1/8" THICK 1-1/8" THICK

DIA 92

Arro 171 Foot Board Options

Style

ARRO - MEMBRANE

1-1/8" THICK

DIA 93

Available Colours

• Solar Oak

• Regal Mahogany

• Honey Maple

• Carmel Cherry

Available Colours

• Solar Oak

• Regal Mahogany

• Honey Maple

• Carmel Cherry

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 26 of 34


ARRO 171 WIRING SCHEMATIC

NOTE: DECK PANELS AND URINE GUARD NOT SHOWN FOR CLARITY.

HEAD END

HILO ACTUATOR

ITEM 2

FOOT-KNEE

ACTUATOR

ITEM 4

18

10

2

12

7

20

13

15

17

3

6

10

8

19

10

16

14

11

7

1

4

5

7

FOOT END

9

MODULAR

IN-LINE PHONE

JUNCTION BOX

ITEM 19

H1 ITEM 13

H2 ITEM 14

L1 ITEM 15

L2 ITEM 16

ATTENDANT

CONTROL PAD

ITEM 18

LIMIT SWITCHES

I/O CONTROLLER

ITEM 6

BATTERY BACK-UP

(ITEM 17 - OPTIONAL

ON ARRO 171 ONLY)

DIA 94

HILO

ACTUATOR

ITEM 1

PORTABLE

BATTERY BACK-UP

(ITEM 9 - OPTIONAL)

HEAD

ACTUATOR

ITEM 3

HAND CONTROL

PENDANT

ITEM 8

CB09

ACTUATOR

CONTROL

ITEM 5

Y CABLE

CONNECTION

ITEM 7

TO WALL

OUTLET

COLOUR DOTS FOR CBO9 COUPLINGS

1 YELLOW - HILO MOTOR (FOOT END MOUNT)

2 WHITE - HILO MOTOR (HEAD END MOUNT)

3 RED - HEAD MOTOR

4 BLUE - KNEE-FOOT MOTOR

Ref Description Part No. Qty

1 Bed HILO Actuator (motor - mounted at the foot end of the bed) MTRM25-471 1

2 Bed HILO Actuator (motor - mounted at the head end of the bed) MTRM25-472 1

3 Head Section Actuator MTRM25-03 1

4 Knee-Foot Section Actuator MTRM25-02 1

5 CB09 Actuator Control Box (For 230V Arro Model 171 - Trendelenburg) JNCTND-C17-A 1

6 I/O Control Box - Arro 151 ONLY

JNCTND-C18 or

1

I/O Control Box - Arro 171 ONLY

JNCTND-C18-A

7 Five Foot Y cable with Two Male Ends PNDNTT-CAB-LE 1

8 6 Function Hand Control Pendant PNDNT6-FNC-B 1

9 Portable Battery Back-up (Optional) KITARR-O 1

10 Ground Wire 4.75 long c/w

JNCTNG-RND-04-34 c/w

3

(2X) #10 X 1/2 long TEK Screw and (2X) #10 Star Washer

(2X) TEKSCR-HX#-10-12 and (2X) WSRSTA-R10

11 Ground Wire 7.75 long c/w

JNCTNG-RND-07-34 c/w

1

(2X) #10 X 1/2 long TEK Screw and (2X) #10 Star Washer

(2X) TEKSCR-HX#-10-12 and (2X) WSRSTA-R10

12 Ground Wire 14.0 long c/w

JNCTNG-RND-14 c/w

1

#10 X 1/2 long TEK Screw and #10 Star Washer

TEKSCR-HX#-10-12 and WSRSTA-R10

13 Bed Height Limit Switch - H1 JNCTND-C6 1

14 Bed Height Limit Switch - H2 JNCTND-C6 1

15 Bed Height Limit Switch - L1 JNCTND-C6 1

16 Bed Height Limit Switch - L2 JNCTND-C6 1

17 Battery Back-up (Optional on Arro 171 Only) BA1800 1

18 Attendant Control Pad for Arro 171 (Trendelenburg) JNCTND-C20-A 1

19 Modular In-line Phone Coupler JNCTND-C12 1

20 Plug for Panel Receptacle c/w Coloured Washer for Panel Recept. and Serrated

Lock Washer for Panel Recept. and Nut for Panel Recept.

JNCTND-CTR-PL-UG c/w JNCTND-CTR-CW-SR and

JNCTND-CTR-LW-SR and JNCTND-CTR-NU-T

1

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 27 of 34


ARRO BED - FULL ASSEMBLY

19

23

29

31

DIA 95

Ref Description Part No. Qty

1 Arro Bed Frame Assembly See DIA 96 1

2 Arro Foot Bail - Plated BAIL00-02P-L 1

3 Head & Foot Post Plug IV Rod Holder PLG343-421-4 4

4 Standard Foot Post (Shown - Other available styles are not shown) C/W

PST002-9FC-BP-NT (Shown) c/w

2

Flat Washer 1-1/4 OD X 11/32 ID X 1/8 Thick AND

Nylon Locking Hex Nut 5/16-18

WSR114-113-21-8P and

NUT516-NL

5 Pan Head Torx® Machine Screw T-20 X 1-1/2 LG PHMS14-201-12 8

6 Decorative Cap Insert Nut - Bronze Coated for Dark Wood OR

NUT14I-NSE-RT-BR or

8

Decorative Cap Insert Nut - Zinc Coated for Light Wood

NUT14I-NSE-RT-ZN

7 Arro Foot Board See DIA 93 for Options 1

8 Modular In-line Phone Coupler (Arro 171 ONLY) JNCTND-C12 1

9 Attendant Control Pad - Arro 171 ONLY- Trendelenburg or JNCTND-C20-A (171 ONLY) 1

10 Bullnose Banding 1-1/8 X 66 long - various colours n/a - order with bed board (item 7 or 20) 2

11 Mattress Keeper A (foot left) MATTKP-RCN-RF-L 2

12 Mattress Keeper B (foot right) MATTKP-RCN-RF-R 2

13 Mattress Platform - Foot Section 35 X 22 Composite DECK00-05 1

14 Mattress Platform - Knee Section 35 X 18 Composite DECK00-11 1

15 Mattress Platform - Seat Section 35 X 8 Composite DECK00-16 1

16 Mattress Platform - Head Section 35 X 32 Composite DECK00-07 1

17 Fully Assembled Assist Rail - Left Side - (Order as L & R Set) (See DIA 97 for Assembly) RL3010-261 (Order with item 22 as RAILSA-EPR) 1

18 Carriage Bolt 1/4-20 X 5/8 Long C/W 1/4-20 Hex Bolt BLT145-8C c/w NUT14 4

19 Carriage Bolt 5/16-18 X 2-1/4 long C/W

BLT516-214-C c/w

2

Flat Washer 1-1/4 OD X 11/32 ID X 1/8 Thick and Hex Nut 5/16-18

WSR114-113-21-8P and NUT516

20 Arro Head Board See DIA 92 for Options 1

21 Standard Head Post (Shown - Other available styles are not shown) C/W

PST002-9HC-BP-NT (shown) c/w

2

Flat Washer 1-1/4 OD X 11/32 ID X 1/8 Thick AND

Nylon Locking Hex Nut 5/16-18

WSR114-113-21-8P and

NUT516-NL

22 Fully Assembled Assist Rail - Right Side - (Order as L & R Set) (See DIA 97 for Assembly) RR3010-261 (Order with item 17 as RAILSA-EPR) 1

23 Six Function Hand Pendant - Velcro® (for Assist Rail) or

PNDNT6-FNC-IV or

1

Six Function Hand Pendant - Strap (for Partial and Full Rails Only)

PNDNT6-FNC-IS

24 Pendant Holster with Warning Label PNDNTH-OLS-TE-RS 1

25 Hex Bolt 1/4-20 X 1 long C/W P Clip AND Hex Nut 1/4-20 BLT141 c/w BAILCL-IP AND NUT14 2

26 Hex Bolt 3/8-16 X 2 LG C/W Flat Washer for 3/8 bolt (X2) AND Locking Jam Nut 3/8-16 (AE Only) BLT382-P c/w WSR38U-SS & NUT38N-L 2

27 Hex Bolt 5/16-18 X 2 LG C/W Flat Washer for 3/8 bolt (X2) AND Nylon Lock Nut 5/16-18 (AE Only) BLT516-2P c/w WSR38U-SS & NUT516-NL 2

28 4 IN Extension Mattress Support - Painted (Optional) MNTY84-EXT-P 1

29 Hex Bolt 5/16-18 X 1-1/2 LG C/W Flat Washer for 5/16 bolt & 5/16-18 Locking Flange Nut (Optional) BLT516-112 c/w WSR516-SAE & NUT516-FL 2

30 11 GA Extension Plug for Arro (Optional) PLG114-PB 2

31 4 In Bed Extension Tube - Painted (Optional) TBS114-114-P 2

32 Square Tube End Cap PLG114-114-11 4

33 Non-Tuckable Side Rail Arm Holder (PE & FE Only) RAILBR-ACK-NT 4

34 Hex Bolt 5/16-18 X 2-1/4 LG C/W 1-1/4 OD Flt Wsr & 5/16-18 Lock’g Fl Nut (PE & FE Only) BLT516-112 c/w WSR114-113-21-8P & NUT516-FL 4

Uncontrolled Document - For Reference Only

17

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 28 of 34


ARRO BED - FRAME ASSEMBLY

NOTE:

54

53

DIA 96

Bill of Material for DIA 96 is on next page.

Uncontrolled Document - For Reference Only

25

35

63

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 29 of 34


Arro Bed - Frame Assembly Bill of Materials

NOTE:

Assembly Drawing for Bill of Materials is on previous page.

Ref Description Part No. Qty

1 Head Gatch Painted (See DIA 99 for Service Assembly) GATHEA-DAP 1

2 Clevis Pin Ø3/8 X 2.0 Long comes with Ring Cotter Pin PIN380-2 c/w PINRIN-GCO-TT-ER 2

3 Ground Wire 4.75 long c/w

JNCTNG-RND-04-34 c/w

4

(2X) #10 X 1/2 long TEK Screw and (2X) #10 Star Washer

(2X) TEKSCR-HX#-10-12 and (2X) WSRSTA-R10

4 Arro Bed Stabilizer Leg LEG002-3PN-T 2

5 11 GA Square Tube End Cap PLG114-114-11 4

6 Plug Rectangular 1 X 2 - 14 Gauge Black PLG121-4 4

7 Hex Bolt 5/16-18 X 2-1/4 LG (Grade 5) c/w Flat Washer 5/16 ID BLT516-2T c/w WSR516-SAE 4

8 Arro Bed Leg LEG002-2PN-T 2

9 Arro Caster Foot - Head End Right Side GLDARR-OFT-RH-M 1

10 Arro Wall Bumper GLDBUM-PER 2

11 Hex Bolt 1/4-20 X 3/4 long c/w Washer SAE 1/4 BLT143-4 c/w WSR14S-AE 2

12 Non locking 4 Inch Swivel Caster WHLRHO-M4 2

13 Hex Head Cap Screw 1/4-20 X 1/2 long c/w Lock Washer 1/4 BLT141-2HS-CS c/w WSR14L 16

14 Socket Head Cap Screw 1/2-20 X 1 long BLT121-BSH-CS-B 4

15 Arro Rubber Bed Foot GLDARR-OBE-DF-OO 4

16 Hex Bolt 1/4-20 X 3/4 long c/w Washer SAE 1/4 and

BLT143-4 c/w WSR14S-AE and

4

Flat Washer 1-1/4 OD X 11/32 ID X 1/8 Thick

WSR114-113-21-8P

17 HILO Motor - Mounted at the Head End of the Bed (short cord) MTRM25-471 1

18 Hex Head Shoulder Bolt 10mm X 30mm c/w 8mm Lock Nut Yellow Zinc BLT103-0HS-SB c/w NUT08M-ML 4

19 Arro Assist Spring SPRARR-O 2

20 Arro Bed Linear Guide Rod ROD119-47P 2

21 Arro Bed Linear Guide Bearing BRGGUI-DEA 2

22 Hex Head Shoulder Bolt 3/8-16 X 3-3/4 Lg c/w Nylon Lock Nut 5/16-18 BLT383-34H-SS-B c/w NUT516-NL 2

23 Hex Bolt M8 X 50mm X 1.25 pitch c/w Lock Washer 1/4 and

BLTM81-255-0 c/w WSR516-L and

2

Flat Washer 1-1/4 OD X 11/32 ID X 1/8 Thick

WSR114-113-21-8P

24 Hex Bolt 5/16-18 X 2 long Grade 5 c/w Flat Washer 1-1/4 OD X 11/32 ID BLT516-2T c/w WSR114-113-21-8P 4

25 Arro Bed Spring Bushing SPRARR-OBU-SH-IN 2

26 Plastic Cap RH 20 CAP14R-H20 8

27 Oillite Bushing 1/2 ID X 5/8 OD X 1/2 long BSH125-812 8

28 Socket Shoulder Bolt 1/2 X 1-1/4 X 3/8-16 c/w Washer RHW-207 and BLT121-14H-SS-B c/w WSRRHW-207 and

4

Washer RHW-205 and Hex Nut 3/8-16 Centre Lock

WSRRHW-205 and NUT38L

29 Washer 0.5 ID X 1.125 OD X 0.063 Thick Nylon WSR121-181-16-N 4

30 Arro Caster Foot - Foot End Right Side GLDARR-OFT-RF-M 1

31 Arro Caster Foot - Foot End Left Side GLDARR-OFT-LF-M 1

32 RHOM 4" Total Lock PB Plate WHLRHO-M4T-L 2

33 DIA 1-1/2 - 11 Gauge Round Socket PLG112-11R-DA 4

34 Pivot Pin PIN122-516-NC-F 8

35 Hex Head Elevator Bolt 5/16-18 X 3 C/W (2X) Hex Nut 5/16-18 AND BLT516-3EL-EV c/w (2X) NUT516 and WSR516-L 1

Lock Washer 5/16

36 Foot Gatch Painted (See DIA 100 for Service Assembly) GATFOO-TP 1

37 Gatch Wheel - Ø2.0 X 3/4 WHL623-4 4

38 Hex Bolt 1/4-20 X 1-1/2 long c/w Locking Hex Nut 1/4-20 BLT141-12 c/w NUT14L 4

39 Arro 230V Trendelenburg Junction Box for 171 ONLY JNCTND-C17-A 1

40 Set Screw M3 X 10mm X 0.5 Pitch SETSCR-M30-51-0 4

41 ABS Urine Guard Almond 25-1/2 X 28-1/8 ABS251-222-18-18 1

42 Arro Head Motor MTRM25-031 1

43 Rolled Spacer 10mm X 10mm WSRSPA-101-0 2

44 Ground Wire 14.0 long c/w

JNCTNG-RND-14 c/w

1

#10 X 1/2 long TEK Screw and #10 Star Washer

TEKSCR-HX#-10-12 and WSRSTA-R10

45 Arro Foot Motor MTRM25-021 1

46 Plug for Panel Receptacle c/w Coloured Washer for Panel Recept. and JNCTND-CTR-PL-UG c/w JNCTND-CTR-CW-SR and

1

Serrated Lock Washer for Panel Recept. and Nut for Panel Recept. JNCTND-CTR-LW-SR and JNCTND-CTR-NU-T

47 Arro 80 inch Main Frame UPRFRP-NT1 1

48 Hex Bolt 5/16-18 X 1 long c/w Lock Washer 5/16-18 BLT516-1 c/w WSR516-L 4

50 3" Velcro Strip Beige, Mated PNDNTV-ELC-RO 7

51 Arro Caster Foot - Head End Left Side GLDARR-OFT-LH-M 1

52 Hex Bolt 1/4-20 X 3/4 long c/w Nylon lock Nut 1/4-20 BLT143-4 c/w NUT14L 2

53 Button Head Socket Cap Screw #8-32 X 1/2 long BLT832-12 4

54 Arro Limit Switch Mounting Bracket JNCTND-CLS-BR-P 2

55 Arro HILO limit Switch JNCTND-C6 4

56 Hex Bolt 1/4-20 X 3/4 long c/w Nylon Lock Nut 1/4-20 BLT143-4 c/w NUT14L 2

57 Arro Trendelenburg Battery Bracket JNCTND-CTR-BR-AC 2

58 HILO Motor - Mounted at the Foot End of the Bed MTRM25-472 1

59 Pan-H M-Screws #8-32 X 3/4 lg c/w Lock Washer #10 and Nut #8-32 PHMS34-832 c/w WSR#10-L and NUT832 4

60 Arro Trendelenburg Battery Backup JNCTND-CTR-BAT 1

61 I/O DC Controller Trendelenburg JNCTND-C18 1

62 Hex Head Tek Screw #10 X 3/4 long TEKSCR-HX#-10-34 4

63 Pendant T Cable PNDNTY-CAB-LE 1

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 30 of 34


ASSIST RAIL ASSEMBLY

LEFT SIDE ASSIST RAIL

14

7

DIA 97

RIGHT SIDE ASSIST RAIL

Full Assemblies:

Ref Description Part No. Qty

- Fully Assembled Assist Rail Pair - Left & Right Side RAILSA-EPR 1

- Fully Assembled Assist Rail - Right Side RR3010-261 1

- Fully Assembled Assist Rail - Left Side RL3010-261 1

Components:

Ref Description Part No. Qty

1 Socket Shoulder Bolt BLT121-2SO-CK-SH 2

2 Bellville Washer WSRBEL-VIL-1 2

3 18 Gauge Plug PLG1RD-18 4

4 Washer - Nylon 1/2 ID, 2 1/2 OD WSR122-123-64-N 2

5 3/8 Nylon Lock Nut NUT38N-L 2

6 Right Side Assist Rail Bracket - Plated RRBRAC-KAS-SI-ST 1

7 Left Side Assist Rail Bracket - Plated RLBRAC-KAS-SI-ST 1

8 Slotted Shoulder Bolt BLT121-2SL-OT-SH 2

9 Compression Spring SPRASS-IST 2

10 Washer 3/8 ID X 3/4 OD X 1/16 Thick Nylon WSR383-411-6N 2

11 Black Ball Knob LATCH0-002 2

12 Pendant Holster PNDNTH-OLS-TE-RS 1

13 Side Mount Arro Assist Rail Member - Right Side RR1010-26 1

14 Side Mount Arro Assist Rail Member - Left Side RL1010-26 1

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 31 of 34


ARRO BED - NON TUCKABLE FULL & PARTIAL RAIL ASSEMBLY

2

1

21

3

4

11

7

6 5

19

8

DIA 98

20

10

2

9

NOTE:

NON-TUCKABLE PARTIAL

RAIL RIGHT SIDE SHOWN

Full Assemblies:

Ref Description Part No. Qty

n/a Fully Assembled Non-Tuckable Full Rail Pair - Left & Right Side RAILSE-T01-06-7A -

Components:

Ref Description Part No. Qty

1 Black Ball Pull Knob LATCH0-002 1

2 Nylon Spacer 3/8 X 3/4 X 1/16 Thick WSR383-411-6N 5

3 Compression Springs SPRASS-IST 1

4 Slotted Shoulder Bolt 1/2-13 X 1/2 Long BLT121-2SL-OT-SH 1

5 Steel Washer 3/8 ID WSRRHW-204 2

6 Plastic Cap RH-20 CAP14R-H20 2

7 Locking Flange Nut 5/16-18 NUT516-FLL 2

8 Nylon Washer 1.46 OD X 0.515 ID X 0.050 Thick WSRNYL-ONF-LA-T 2

9 Rail Pin 1/2 X 7/8 Plated RPIN12-78P 2

10 Non-Tuckable Short Rail Arm Welded - Right Side (Shown) or

SRA001-RA (Shown) or

1

Non-Tuckable Short Rail Arm Welded - Left Side (Not Shown)

SRA001-LA (Not Shown)

11 Non-Tuckable Long Rail Arm Welded - Right Side (Shown) or

LRA001-RA (Shown) or

1

Non-Tuckable Long Rail Arm Welded - Left Side (Not Shown)

LRA001-LA (Not Shown)

12 Set Screw 5/16-18 NC SETSCR-516-51-6N 2

13 Locking Set Collar 3/4 ID with 1/4-20 NC Set Screw COL34R 2

14 Non-Tuckable Side Rail Bracket RAILBR-ACK-NT 2

15 Nylon Bushing - 3/4 X 1-1/2 Long WSR341-141-2N 4

16 Nylon Spacer - 3/4 X 1-11/16 Thick WSR341-116-4N 2

17 5/16-18 Centre Lock Nut NUT516-L 2

18 Nylon Bushing - 3/8 X 3/4 X 5/16 long WSR383-451-6N 4

19 Socket Head Shoulder Bolt 3/8-16 X 1 long BLT381-HSS-B 2

20 Non-Tuckable Partial Rail link Bar (Shown) or

RLNK00-01P-AP-NT (Shown) or

1

Non-Tuckable Full Rail link Bar (Not Shown)

RLNK00-01F-AP-NT (Not Shown)

21 Partial Rail Welded and Epoxy Painted - Right Side (Shown) or

Partial Rail Welded and Epoxy Painted - Left Side (Not Shown) or

Full Rail Welded and Epoxy Painted - Right Side (Shown) or

Full Rail Welded and Epoxy Painted - Left Side (Not Shown

RR1010-52 (Shown); or

RL1010-52 (Not Shown); or

RR1010-67 (Not Shown); or

RL1010-67 (Not Shown)

1

Uncontrolled Document - For Reference Only

18

15

17

16

14

12

13

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 32 of 34


SERVICE PART SUB-ASSEMBLIES

Head Gatch Socket Style Assembly

DIA 99

Full Assembly:

Ref Description Part No. Qty

n/a Head Gatch Socket Style Assembly with Wheel and Inserts GATHEA-DAA-SS-Y -

Components:

Ref Description Part No. Qty

1 Arro Head Gatch - Painted (11 GA) GATHEA-DAP 1

2 Round Tube Gatch Socket Ø1-1/2 - 11 GA PLG112-11R-DA 2

3 Pivot Pin PIN122-516-NC-F 2

4 Hex Head Bolt 1/4-20 X 1-1/2 BLT141-12 2

5 Gatch Wheel 2 X 3/4 WHL623-4 2

6 1/4-20 Nylon Lock Nut NUT14L 2

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 33 of 34


Foot Gatch Socket Style Assembly

DIA 100

Full Assembly:

Ref Description Part No. Qty

n/a Foot Gatch Socket Style Assembly With Inserts and Fasteners GATFOO-TAS-SY -

Components:

Ref Description Part No. Qty

1 Foot Gatch - Painted (11 GA) GATFOO-TP 1

2 Round Tube Gatch Socket Ø1-1/2 - 11 GA PLG112-11R-DA 2

3 Pivot Pin PIN122-516-NC-F 2

4 Hex Head Bolt 1/4-20 X 1-1/2 BLT141-12 2

5 Gatch Wheel 2 X 3/4 WHL623-4 2

6 1/4-20 Nylon Lock Nut NUT14L 2

Uncontrolled Document - For Reference Only

1-150-041-REL 230 V Arro Bed Manual.doc © Carroll Healthcare Inc. Page 34 of 34