20.04.2022 Views

Brochure-testo-saveris-1-2022-EN

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>testo</strong> Saveris 1 environmental monitoring system<br />

Technical data<br />

<strong>testo</strong> Saveris 1 environmental monitoring system<br />

Technical data<br />

Technical data for<br />

Base station and Gateway<br />

Technical data for<br />

digital analog coupler<br />

Physical specifications<br />

Housing material<br />

Dimensions (L x W x H)<br />

Base station <strong>testo</strong> Saveris Base V3.0<br />

ABS/PC plastic<br />

193 x 112 x 46 mm<br />

<strong>testo</strong> UltraRange Gateway<br />

Weight Approx. 370 g Approx. 314 g<br />

IP protection class<br />

Operating and storage conditions<br />

Storage temperature -20 to +60 °C -20 to +80 °C<br />

Operating temperature +5 to +35 °C 0 to +50 °C<br />

Power<br />

Power supply PoE class 0; optionally via mains unit & micro USB cable (order no. 0572 5004)<br />

Rechargeable battery type<br />

Interfaces<br />

Connections<br />

Li-Ion rechargeable battery, 3.7 V / 2.6 Ah,<br />

Order no. 0515 0107<br />

(for data backup and emergency alarm in the event of<br />

power failure)<br />

2x USB<br />

LAN/PoE:<br />

Transfer rate 10/100 Mbit<br />

PoE class 0<br />

micro USB<br />

alarm relay connection<br />

IP20<br />

–<br />

1x USB<br />

LAN/PoE:<br />

Transfer rate 10/100 Mbit<br />

PoE class 0<br />

micro USB<br />

Channels per Base 3,000 –<br />

Loggers per Gateway – 40<br />

Measurement data storage<br />

Memory Circular buffer memory –<br />

Max. number of<br />

measurement values<br />

Internal memory (per<br />

channel)<br />

Other<br />

120,000,000 –<br />

40,000 –<br />

Alarm relay Connection for external alarm relay available –<br />

GSM module via LTE stick –<br />

Physical specifications<br />

Housing material<br />

Size (W x H x L)<br />

Weight<br />

IP protection class<br />

Operating and storage conditions<br />

Plastic<br />

85 x 100 x 38 mm<br />

240 g<br />

Storage temperature -25 to +60 °C<br />

Operating temperature +5 to +45 °C<br />

Power<br />

Power supply<br />

Interfaces<br />

Connections<br />

Connectible loggers<br />

Measurement data storage<br />

Measuring range<br />

Measuring interval<br />

Internal memory (per<br />

channel)<br />

IP54<br />

Power supply via <strong>testo</strong> 150 TUC4 logger<br />

2- or 4-wire current/voltage input<br />

<strong>testo</strong> 150 TUC4<br />

4 to 20 mA; 0 to 10 V<br />

1 min to 24 h<br />

6,000 readings<br />

Communication cycle depends on method of communication of <strong>testo</strong> 150<br />

Accuracy<br />

Power<br />

Maximum error: ±0.03 mA<br />

Resolution (min. error): 0.75 µA (16 bit) typical error: 5 µA<br />

Voltage<br />

0 to 1 V maximum error: ±1.5 mV resolution (min. error): 39 µV (16 bit)<br />

Typical error: 250 µV<br />

0 to 5 V maximum error: ±7.5 mV resolution (min. error): 0.17 mV typical error: 1.25 mV<br />

0 to 10 V maximum error: ±15 mV<br />

Resolution (min. error): 0.34 mV<br />

Typical error: 2.50 mV<br />

26<br />

27

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!