20.04.2022 Views

Brochure-testo-saveris-1-2022-EN

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>testo</strong> Saveris 1 environmental monitoring system<br />

Technical data<br />

Technical data for<br />

NTC probes / Pt100 probes<br />

NTC probes<br />

Penetration probe NTC with ribbon<br />

cable, cable length 2 m, IP 54<br />

External temperature probe 12 mm,<br />

plug-in, without cable<br />

Stub probe, IP 54<br />

Accurate immersion/penetration<br />

probe, cable length 6 m, IP 67,<br />

connection: fixed cable; cable<br />

length: 6 m<br />

Stationary probe with aluminium<br />

sleeve, IP 65, connection: fixed<br />

cable; cable length: 2.4 m<br />

Pipe wrap probe with Velcro tape<br />

for pipe diameters up to max. 75<br />

mm, Tmax +75 °C, NTC, connection:<br />

fixed cable; cable length: 1.5 m<br />

Probe for surface measurement,<br />

fixed cable, 2 m<br />

Probe shaft/probe shaft tip dimensions<br />

105 mm<br />

Ø 20 mm<br />

90 mm<br />

Ø 5 mm<br />

Ø 3 mm<br />

Ø 12 mm<br />

Measuring<br />

range<br />

Accuracy t 90<br />

Order no.<br />

30 mm<br />

Ø 3.6 mm<br />

-40 to +125 °C ±0.5% of m.v.<br />

(+100 to +125 °C)<br />

±0.2 °C (-25 to +80 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

-30 to +50 °C ±0.2 °C (-30 to +50 °C) 240<br />

sec<br />

35 mm -20 to +70 °C ±0.2 °C (-20 to +40 °C)<br />

±0.4 °C (+40.1 to +70 °C)<br />

40 mm<br />

Ø 3 mm<br />

Ø 3 mm<br />

-35 to +80 °C ±0.2 °C (-25 to +74.9 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

40 mm -30 to +90 °C ±0.2 °C (0 to +70 °C)<br />

±0.5 °C (remaining meas.<br />

Ø 6 mm<br />

range)<br />

300 mm<br />

40 mm<br />

30 mm<br />

8 x 8 mm<br />

-50 to +70 °C ±0.2 °C (-25 to +70 °C)<br />

±0.4 °C (-50 to -25.1 °C)<br />

8 sec 0572 1001<br />

0572 2153<br />

15 sec 0628 7510<br />

5 sec 0610 1725<br />

190<br />

sec<br />

-50 to +80 °C ±0.2 °C (0 to +70 °C) 150<br />

sec<br />

0628 7503<br />

60 sec 0613 4611<br />

0628 7516<br />

Wall surface temperature probe, e.<br />

g. for proof of structural damage in<br />

buildings, connection: fixed cable;<br />

cable length: 3 m<br />

Stainless steel NTC food probe<br />

(IP65) with PUR cable, connection:<br />

fixed cable; cable length: 1.6 m<br />

Waterproof NTC immersion/<br />

penetration probe, fixed cable 1.2 m<br />

Accurate immersion/penetration<br />

probe, cable length 1.5 m, IP 67,<br />

connection: fixed cable; cable<br />

length: 1.5 m<br />

Waterproof NTC immersion/<br />

penetration probe, fixed cable 1.2 m<br />

Robust air probe NTC, fixed cable<br />

1.2 m<br />

125 mm<br />

Ø 4 mm<br />

115 mm 50 mm<br />

Ø 5 mm<br />

40 mm<br />

Ø 3 mm<br />

Ø 4 mm<br />

115 mm 50 mm<br />

Ø 5 mm<br />

Ø 4 mm<br />

115 mm 50 mm<br />

Ø 5 mm<br />

Ø 4 mm<br />

-50 to +80 °C ±0.2 °C (-25 to +80 °C)<br />

±0.5 °C (-40 to -25.1 °C)<br />

15 mm<br />

Ø 3 mm<br />

-50 to +150 °C ±0.5% of m.v. (+100 to<br />

+150 °C)<br />

±0.2 °C (-25 to +74.9 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

Ø 3 mm<br />

-50 to +150 °C ±0.5% of m.v. (+100 to<br />

+150 °C)<br />

±0.2 °C (-25 to +74.9 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

-35 to +80 °C ±0.2 °C (-25 to +74.9 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

-50 to +150 °C ±0.5% of m.v. (+100 to<br />

+150 °C)<br />

±0.2 °C (-25 to +74.9 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

-50 to +125 °C ±0.2 °C (-25 to +80 °C)<br />

±0.4 °C (remaining meas.<br />

range)<br />

20 sec 0628 7507<br />

8 sec 0613 2211<br />

10 sec 0613 1212<br />

5 sec 0628 0006<br />

10 sec 0615 1212<br />

60 sec 0615 1712<br />

Temperature probe with Velcro<br />

(NTC), fixed cable 1.4 m<br />

300 mm<br />

30 mm<br />

The standard temperature probes from the Testo range can be individually tailored to your application.<br />

For more information please contact your Testo partner.<br />

Pt100 probes<br />

-50 to +70 °C ±0.2 °C (-25 to +70 °C)<br />

±0.4 °C (-50 to -25.1 °C)<br />

60 sec 0615 4611<br />

Penetration probe Pt100 with ribbon<br />

cable, cable length 2 m, IP54<br />

Robust, waterproof Pt100<br />

immersion/penetration probe, fixed<br />

cable<br />

Robust Pt100 stainless steel food<br />

probe (IP65), connection: fixed cable<br />

60 mm<br />

Ø 5 mm<br />

30 mm -85 to +150 °C Class A* 35 sec 0572 7001<br />

Ø 3.6 mm<br />

114 mm 50 mm<br />

Ø 5 mm<br />

125 mm<br />

Ø 4 mm<br />

Ø 3.7 mm<br />

-50 to +400 °C Class A* (-50 to +300 °C)<br />

Class B* (remaining meas.<br />

range)<br />

15 mm -50 to +400 °C Class A* (-50 to +300 °C)<br />

Class B* (remaining meas.<br />

Ø 3 mm<br />

range)<br />

12 sec 0609 1273<br />

10 sec 0609 2272<br />

* According to standard <strong>EN</strong> 60751, the accuracies of Classes A and B refer to -200 to +600 °C (Pt100).<br />

30<br />

31

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!