Multifunctional Microplate Reader with<br />
I<strong>Infinite</strong> fi it <strong>200</strong>® PRO<br />
injector option
Measurement techniques supported by the<br />
<strong>Infinite</strong> M<strong>200</strong> <strong>Pro</strong>:<br />
Absorbance:<br />
‐ Single point and scanning<br />
‐ Single point and scanning<br />
with injectors<br />
Fluorescence:<br />
‐ Fluorescence Intensity top and bottom<br />
‐ Time‐Resolved Time Resolved Fluorescence<br />
‐ Fluorescence Resonance Energy Transfer<br />
‐ Flash Fluorescence (with injectors)<br />
TECAN<br />
I<strong>Infinite</strong> fi it<br />
M<strong>200</strong> <strong>Pro</strong><br />
Luminescence:<br />
‐ Gl Glow LLuminescence i<br />
‐ Flash Luminescence<br />
‐ Bioluminescence Resonance Energy Transfer<br />
Absorbance and<br />
Fluorescence:<br />
NNanoquant: t<br />
Small volume measurements,<br />
2 ul sample
Technical specifications p <strong>Infinite</strong> M<strong>200</strong> <strong>Pro</strong><br />
General specifications:<br />
� Microplates to be measured: from 66‐well well to 384‐well 384 well plates<br />
� Temperature Control: from 5 °C above ambient up to 42 °C<br />
� Plate Shaking: Linear and orbital shaking, amplitude from 1 –6 mm in 0.5 mm steps<br />
� Light Source: High energy xenon flash lamp<br />
� Detectors: Fluorescence ‐> PMT; Luminescence ‐> Low dark count PMT photon‐<br />
counting electronics; Absorbance ‐> Silicon photodiode<br />
Ab Absorbance b specifications: ifi ti<br />
� Wavelength range: 230 – 1000 nm<br />
� Accuracy: < ±(1 % + 10 mOD); ≤ ±1.5 nm λ > 315 nm; ≤ ±0.8 nm λ ≤ 315 nm;<br />
Fluorescence specifications:<br />
� Wavelength Range: Excitation: 230 – 850 nm; Emission: 280 – 850 nm<br />
� Detection limit (fluorescein, 96 well plate, <strong>200</strong> μl): < 20 pM (top); 100 pM (bottom)<br />
� Europium detection limit (96 well plate, <strong>200</strong> μl ): < 5 pM (typical value)<br />
� Reproducability (at 25nM fluorescein) < 2%
Technical specifications p <strong>Infinite</strong> M<strong>200</strong> <strong>Pro</strong><br />
Luminescence specifications:<br />
�� Wavelength Range: 380 – 600 nm<br />
� Linear dynamic range: at least 6 orders of magnitude<br />
� Integration time per well: 100 – 20,000 ms<br />
�� Cross Talk (black plate): 00.01% 01%<br />
� Attenuation of light: 100 (OD2 attenuation filter), 1 (no attenuation)<br />
� ATP detection limit (96 well plate, <strong>200</strong> μl volume, 1000 ms integration time):<br />
3 fmol/well (glow); 80 amol/well (flash)<br />
Injector specifications:<br />
� Accuracy: < 10% for injection volume of 10 μl; < 2% for injection volume of 100 μl;<br />
< 0.7% for injection volume of 450 μl<br />
� Precision: < 10% for injection volume of 10 μl; < 2% for injection volume of 100 μl;<br />
< 0.7% for injection volume of 450 μl
Comparison measurements on <strong>Tecan</strong> multiplate<br />
reader vs Thermo “Nanodrop”<br />
Measurements of DNA at different known concentrations, 2 μl volumes:<br />
TECAN multiplate reader<br />
Thermo Nanodrop<br />
Known<br />
CConcentration i<br />
[ng/ml]<br />
Concentration<br />
measured d by b<br />
NanoDrop [ng/ml]<br />
Concentration<br />
measured d by b<br />
<strong>Tecan</strong> [ng/ml]<br />
0.05 0.84 0<br />
0.5 3.85 0.1<br />
5 8.61 5.4<br />
50 48.71 52.2<br />
500 443.83 492.7<br />
Note: with the TECAN Nanoquant, 16 samples can be measured in parallel vs 1 on Nanodrop.<br />
Note: fluorescence and absorbance can be measured on TECAN Nanoquant (e.g. for labelling efficiency)<br />
vs absorbance only on Nanodrop.
Typical Readout for ELISA (absorbance)<br />
Absorbance measurement at 405nm with a reference wavelength of 650nm<br />
Calculated difference between measurement and reference measurement<br />
Mycoplasma species (Sample) Neg. control Pos. control<br />
M. arginini 0.1747 0.1334 2.7127<br />
M. hyorhinis 0.22325 0.152 1.2394<br />
A. laidlawii 0.1817 0.1063 1.3544<br />
M. orale 0.16455 0.145 2.6902<br />
Sample
Typical Readout for luminescence<br />
Luminescence readings for a Mycoplasma detection test based on a bioluminescent<br />
reaction: Reading A before substrate addition, Reading B after substrate addition<br />
� Transparent plate reading<br />
Reading A Reading B Ratio (B/A)<br />
Pos. control 537 22624 42.13035382<br />
Neg. control 618 132 0.213592233<br />
bEND3 (Sample) 311 144 0.463022508<br />
The bEND3 brain endothelial cells are negative for Mycoplasmal enzymes.<br />
� Black plate reading<br />
Reading A Reading B Ration (B/A)<br />
Pos. control 42 2767 65.88095238<br />
Neg. control 29 ‐1 ‐0.034482759<br />
Both transparent and black plates can be adequately used for measurements. Black plates have<br />
lower overall signal, but better signal to background ratio.
Typical Readout for fluorescence<br />
Standard curve for fluorescence reading of different known concentrations of<br />
fluorescein sodium salt<br />
� Excitation: 485 nm<br />
� Emission: 535 nm<br />
Fluorescennce<br />
intensity<br />
45000<br />
40000<br />
35000<br />
30000<br />
25000<br />
<strong>200</strong>00<br />
15000<br />
10000<br />
5000<br />
0<br />
1 5 50 100 250 500 1000<br />
Fluorescein concentration (µM)
Summary<br />
The TECAN <strong>Infinite</strong> <strong>200</strong> <strong>Pro</strong> at the BNL lab offers:<br />
� High accuracy multiplate readings for fluorescence,<br />
absorbance and/or luminescence<br />
� Temperature control, dual injectors, shaking<br />
� Software with a user friendly interface<br />
� Nanoquant option for 2 μl sample volumes<br />
(absorbance and/or fluorescence), equal or<br />
superior accuracy to “ “Nanodrop” d ” instrument.<br />
Link to recharge information:<br />
http://www.cnsi.ucsb.edu/facilities/nanostructures/biological/recharge‐rates.pdf