<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> <strong>Software</strong><br />
<strong>Instruction</strong> <strong>Manual</strong><br />
Version No. 1.24<br />
Page 0<br />
TOPCON Medical Systems, Inc.<br />
TOPCON Corporation
Introduction<br />
Thank you very much for purchasing the TOPCON <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>.<br />
Please read this <strong>Instruction</strong> <strong>Manual</strong> with care before using this <strong>Software</strong>.<br />
A good understanding and working knowledge of this software will guarantee<br />
your satisfaction for many years to come.<br />
Keep this <strong>Manual</strong> at hand for further reference.<br />
Precautions<br />
• TOPCON has the copyright of the <strong>Software</strong>, "<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>".<br />
• Reproduction and copying of the software and the <strong>Instruction</strong> manual, in<br />
whole or in part, without our permission is strictly prohibited.<br />
• As a rule, one set of the software must be purchased for each PC installed.<br />
• For the handling procedures of KR-7000P/ KR-7100P/ KR-8000PA/ KR-<br />
8100PA/ KR-8100P, refer to its instruction manual.<br />
• This <strong>Instruction</strong> <strong>Manual</strong> does not describe details about the handling<br />
procedures of PC and Microsoft Windows. This <strong>Instruction</strong> <strong>Manual</strong> is<br />
made on an assumption that the user already has knowledge about the PC<br />
and Microsoft Windows.<br />
• For the handling procedures of PC and Microsoft Windows, refer to the<br />
pertaining manuals.<br />
• Incorrect handling or operation of the PC may damage it or its software.<br />
• We shall not take any responsibility for the effects of or as a result of<br />
using the <strong>Instruction</strong> <strong>Manual</strong> and <strong>Software</strong>.<br />
• Specifications of the <strong>Software</strong> and descriptions of the <strong>Instruction</strong> <strong>Manual</strong><br />
are subject to change without prior notice.<br />
KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, and<br />
CV Power Supply Unit are trademarks of the TOPCON Corporation.<br />
Microsoft, MS, and MS-DOS are registered trademarks, and Windows is a trademark of<br />
Microsoft Corporation.<br />
PC-DOS is a registered trademark of International Business Machines Corporation<br />
Page 1<br />
TOPCON Corporation
NOTATIONS<br />
The following notations are used in the <strong>Instruction</strong> <strong>Manual</strong>.<br />
MENU, COMMAND AND DIALOG BOX<br />
Expression Description<br />
[File]menu Menu name is expressed in [ ].<br />
[Open] of [File] menu Command name is expressed in [ ] with menu name in which it<br />
is included.<br />
[Open File]dialog box Dialog box name displayed on title bar is expressed in [ ].<br />
[OK] Command button name is expressed in [ ].<br />
MOUSE OPERATION<br />
Expression Description<br />
Point Move mouse and bring mouse pointer onto object of operation.<br />
Click Press and release mouse button.<br />
Double-click Click mouse button twice.<br />
Drag Press and move mouse button and release it at final position.<br />
KEY OPERATION<br />
Expression Description<br />
Enter Key name is expressed with bold type.<br />
Shift+F1 Plus symbol (+) means to press multiple keys; for example,<br />
Shift+F1 means to press Shift key and F1 key.<br />
Direction key Generic name of →, ←, ↑ and ↓ keys.<br />
OTHER EXPRESSIONS<br />
Expression Description<br />
Memo Gives hints for operation.<br />
Page 2<br />
TOPCON Corporation
SAFETY PRECATIONS<br />
Items Prevention items Page<br />
Note that the necessary memory capacity and hard disk capacity<br />
may differ for each PC.<br />
Contact Lens Simulation<br />
7<br />
Physician can not use report data for lens prescription without<br />
43<br />
their personal validation.<br />
Contents of the package<br />
The following are standard accessories. Make sure that all these items are attached (quantity).<br />
CD-ROM (1)<br />
<strong>Instruction</strong> <strong>Manual</strong> (1)<br />
CAUTION<br />
Page 3<br />
RS2<strong>32</strong>C Cable (1)<br />
(D-sub9P-DIN8P)<br />
TOPCON Corporation
Table of Contents<br />
Chapter 1 Overview...........................................................................6<br />
Chapter 2 Setup <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> System .....................................7<br />
2.1 <strong>Software</strong> Requirements ....................................................7<br />
2.2 Installing <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> ..............................................8<br />
2.3 Communication Setting...................................................14<br />
2.4 Preparation for KR-7000P...............................................16<br />
2.5 Preparation for KR-8000P...............................................19<br />
Chapter 3 Getting Started................................................................23<br />
3.1 Starting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> ..............................................23<br />
3.2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Components ......................................23<br />
3.3 Operating <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>...........................................26<br />
3.4 Exiting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>................................................26<br />
Chapter 4 <strong>Color</strong> Map Report Data...................................................28<br />
4.1 Receiving <strong>Color</strong> Map Report Data ..................................28<br />
4.2 Display <strong>Color</strong> Map Report Data ......................................29<br />
4.3 Patient Registration ........................................................29<br />
4.4 Saving Report Data ........................................................31<br />
4.5 <strong>Color</strong> Map Report Window Control .................................<strong>32</strong><br />
4.6 Opening Saved Report Data...........................................33<br />
4.7 Viewing Report Data from the Database ........................34<br />
Chapter 5 Option Settings ...............................................................37<br />
5.1 Map Method....................................................................37<br />
5.2 Scale Unit .......................................................................37<br />
5.3 Map Type........................................................................38<br />
5.4 Overlays .........................................................................38<br />
5.5 Peripheral .......................................................................39<br />
5.6 Others.............................................................................39<br />
5.7 Image View.....................................................................40<br />
Chapter 6 Contact Lens Simulation Report Data.............................42<br />
6.1 Display Contact Lens Simulation Report Data ................42<br />
6.2 Parameters of the Contact Lens Simulation Report........45<br />
6.3 Handling Each Lens Type...............................................45<br />
6.4 Parameter Setting...........................................................46<br />
6.5 Contact Lens Utility.........................................................47<br />
Page 4<br />
TOPCON Corporation
Chapter 7 Printing Reports ..............................................................50<br />
7.1 Print Preview Screen ......................................................50<br />
7.2 Print Setup Dialog Box....................................................51<br />
Chapter 8 Other Features................................................................52<br />
8.1 Hiding Tool Bar and Status Bar ......................................52<br />
8.2 Cascade and Tile............................................................52<br />
8.3 Select COM Port.............................................................52<br />
8.4 Reset COM Port .............................................................52<br />
<strong>Software</strong> License Terms .......................................................................54<br />
Page 5<br />
TOPCON Corporation
Chapter 1<br />
1.1 Overview<br />
Topcon’s <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is a <strong>32</strong>bit multiple-document report Windows application<br />
running under Microsoft Windows 95/98/NT and Microsoft Windows XP.<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> works with the tri-function Topcon KR-7000P/ KR-7100P/ KR-<br />
8000PA/ KR-8100PA/ KR-8100P, auto-refractometer, expanded auto-keratometer and<br />
placido disk-based corneal mapping system. The data captured by the KR-7000P/ KR-<br />
7100P/ KR-8000PA/ KR-8100PA/ KR-8100P is transferred to <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> via an<br />
RS-2<strong>32</strong>C interface, and is then reconstructed into corneal topography color map data.<br />
The multiple documents include refractometer data, keratometer data, and a corneal<br />
topography color map. To send refractometer and/or keratometer data to the TOPCON<br />
CV Power Supply Unit, the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> creates the necessary data for the<br />
TOPCON CV Power Supply Unit and sends it using the second RS-2<strong>32</strong>C port of the PC.<br />
After reconstructing the corneal topography color map data, the multiple document report<br />
may be viewed on the Windows screen or printed on a color printer.<br />
The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> provides various options for evaluating cornea, such as tangential<br />
/ sagittal map, adjustable / normalized / absolute map, diopter / mm display, overlay<br />
information display, auto-print, contact lens fitting simulation, etc. Default option<br />
parameters may also be set.<br />
An important feature of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is that it creates and supports a patient<br />
information database. The database services integrated in the software speeds-up patient<br />
data storage/ retrieval and simplifies data analysis.<br />
If the auto-print option is enabled, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> prints a multiple-document report<br />
to a color printer automatically using default option parameters when new data is<br />
received from the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P.<br />
Page 6<br />
TOPCON Corporation
Chapter 2<br />
Setup <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> System<br />
2.1 System Requirements<br />
2.1.1 Hardware Requirements<br />
The following hardware is necessary to run <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>:<br />
· 100% IBM PC compatible computer with i-486DX or better<br />
· Minimum 8MB of system memory<br />
· Minimum 1MB of video memory<br />
· 2MB of free disk space at installation<br />
· A color printer supported by Windows<br />
· An available COM port*<br />
· A KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to PC interface<br />
cable**<br />
* To send the refractometer and/or keratometer data to the TOPCON CV Power Supply<br />
Unit, an additional COM port is needed.<br />
** To send the refractometer and/or keratometer data to the TOPCON CV Power Supply<br />
Unit, the PC to CV Power Supply Unit interface cable is needed.<br />
In the above specifications, the video memory and the system memory may be the most<br />
important hardware for <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>. For a computer with 1MB of video memory,<br />
the video mode can be set to 1024x768 256 colors and 640x480 16-million colors.<br />
The faster the computer CPU, the better the performance of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>. In<br />
general, performance of Windows is influenced most by the amount of system memory<br />
and the speed of the hard disk. If the system appears to be running slow, try adding<br />
system memory first. If it still seems slow, upgrade the computer with a faster CPU and<br />
much larger hard disk.<br />
Note that the necessary memory capacity and hard disk capacity<br />
may differ for each PC.<br />
2.1.2 <strong>Software</strong> Requirement<br />
To run <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, the following software is necessary:<br />
· Microsoft Windows 95/98/NT or Microsoft Windows XP.<br />
· The video mode of Windows should be set to at least 640x480 x 256 colors.<br />
Page 7<br />
TOPCON Corporation
2.2 Installing <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
2.2.1 Running Setup Program<br />
If the previous version of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> was installed, uninstalling is always<br />
recommended. See section 2.2.2 for Uninstalling.<br />
Follow the instructions bellow to install <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> for PC with WINDOWS 95:<br />
Quit any applications that are currently running.<br />
If you use the Microsoft Plus! System Agent, right-click the System Agent icon on the<br />
toolbar, and then click Suspend System Agent.<br />
• Insert the floppy disk into your computer’s drive;<br />
1. Click the [Start] button of task bar and click [Run] menu.<br />
2. The [Run] dialog box is displayed. Enter [A:\Setup.exe], insert the floppy disk #1/1<br />
and click the [OK] button.<br />
Page 8<br />
TOPCON Corporation
3. Once the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> installation program is started, it displays the following<br />
‘Welcome’ dialog box:<br />
Click the ‘Next’ button to continue the installation<br />
4. In this dialog box, specify the destination where <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is to be installed.<br />
It is not recommended to change the default destination directory. Click the ‘Next’<br />
button to continue.<br />
Page 9<br />
TOPCON Corporation
5. Specify Disk Name, you will use for the Patient Information Database.<br />
6. Select directory for storage of the color map image files that Patient Database<br />
Records refer to.<br />
Page 10<br />
TOPCON Corporation
7. In this dialog box, click the ‘Finish’ button to finish the installation.<br />
When the Installation is done click on the shortcut icon to start the <strong>Color</strong> <strong>Mapping</strong><br />
<strong>32</strong>.The shortcut icon must be created by the Installation program on your desktop screen<br />
when the installation is completed(see section 3-1)<br />
2.2.2 Removing The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
1. Click the [Start] button, point to [Settings], and then click [Control Panel].<br />
Page 11<br />
TOPCON Corporation
2. Double-click [Add/Remove Programs].<br />
Page 12<br />
TOPCON Corporation
3. Click the [Install/Uninstall] tab, click [KRP<strong>32</strong>], and then click [Add/Remove].<br />
4. The following dialog box will appear in order to make sure you are uninstalling the<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. Click ‘Yes’ button to start the uninstalling.<br />
Page 13<br />
TOPCON Corporation
5. In this dialog box, click the ‘OK’ button to finish the uninstalling.<br />
2.3. Communication Setting<br />
2.3.1. Port Setting<br />
If it is necessary to select the communication port for the KR-7000P/ KR-7100P/ KR-<br />
8000PA/ KR-8100PA/ KR-8100P, choose ‘Select Com Port for KR’ in the [Port] menu to<br />
get the Select COM Port dialog box. (Figure 2-1) To change the COM port number,<br />
select COM# (# is number of communication port) and click the ‘OK’ switch. The<br />
selected communication port will be initialized automatically. If None was selected,<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> displays only saved data.<br />
Figure 2-1 KR port selection Dialog Box<br />
Page 14<br />
TOPCON Corporation
If it is necessary to select the communication port for the CV Power Supply Unit, choose<br />
‘Select Com Port for CV’ in the [Port] menu to get the Select COM Port dialog box.<br />
(Figure 2-2) To change the COM port number, select COM# and click the “OK” switch.<br />
Figure 2-2 CV port selection Dialog Box<br />
If None was selected, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> does not send data to the TOPCON CV Power<br />
Supply Unit. Information about the old and new formats can be found in the TOPCON<br />
CV Power Supply Unit documentation.<br />
2.3.2 Communication Format Instrument Setting<br />
<strong>Software</strong> supports data communication with two types of instruments:<br />
• KR7000P(Include KR7100P)<br />
• KR8000P/19K(Include KR-8100PA/ KR-8100P with 19200Bps)<br />
• KR8000P/38K(Include KR-8100PA/ KR-8100P with 38400Bps)<br />
To receive measurements the type of connected instrument must be selected:<br />
1. Select pull-down Port Menu by clicking on the Port button on the Menu Bar.<br />
2. Set the instrument type by selecting appropriate option KR7000P/KR8000P from the<br />
[Select Instrument] pull-down menu.<br />
Note: The previous setting is indicated with the check mark next to the instrument type<br />
in the [Select Instrument] pull-down menu.<br />
2.3.3 Cable Connection<br />
For a KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to PC interface<br />
cable: (A round connector 8-PIN DIN on one end and a 9-PIN D-SUB connector on the<br />
other end.)<br />
To connect the interface cable between the KR-7000P/ KR-7100P/ KR-8000P/ KR-<br />
8100PA/ KR-8100P and the PC:<br />
1. Turn off the power switch of the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />
KR-8100P and the PC.<br />
2. Align the round cable connector (8-PIN DIN) with the KR-7000P/ KR-7100P/ KR-<br />
8000PA/ KR-8100PA/ KR-8100P’s OUT port of “DATA TERMINAL” and press the<br />
round cable connector into the port.<br />
3. Connect the other connector (9-PIN DSUB) to the PC COM port.<br />
Page 15<br />
TOPCON Corporation
2.4 Preparation for KR-7000P<br />
2.4.1 Setting Up KR-7000P<br />
Make sure the following settings of the KR-7000P/ KR-7100P are correctly configured:<br />
RS2<strong>32</strong>C FORM EXT<br />
BAUD-RATE 9600<br />
To change settings of the KR-7000P/ KR-7100P:<br />
1. Turn on the power switch of the KR-7000P/ KR-7100P while pressing the<br />
“CONTACT / GLASSES” switch. The initial setting screen appears on the screen.<br />
2. Release the “CONTACT / GLASSES” switch<br />
3. “INITIAL SETTING” screen appears on the screen (Figure 2-3).<br />
INITIAL SET<br />
NUMBER SET<br />
PRINT OUT<br />
ONLINE<br />
PLACIDO SET<br />
Figure 2-3 Initial Setting Screen<br />
4. Bring the cursor to “ONLINE” by pressing the measuring switch.<br />
5. Press the PRINT switch. The on-line screen appears on the screen.<br />
CL INPUT NO<br />
RS-2<strong>32</strong>C FORM EXT<br />
BAUD-RATE 9600<br />
RS2<strong>32</strong>C OUT TYPE REF<br />
RS2<strong>32</strong>C KRT CENT<br />
Figure 2-4 Online Setting Screen<br />
6. Bring the cursor to “RS-2<strong>32</strong>C FORM” by pressing the measuring switch.<br />
7. Bring the cursor to “RS-2<strong>32</strong>C FORM” to “EXT” by pressing the PRINT switch. The<br />
“BAUD-RATE will be set to 9600 automatically. Also change the “RS2<strong>32</strong>C OUT<br />
TYPE” to “ALL” and the “RS2<strong>32</strong>C KRT” to “BOTH”.<br />
8. Bring the cursor to “NEXT by pressing the measuring switch.<br />
9. Press the PRINT switch.<br />
10. Bring the cursor to “END” by pressing the measuring switch.<br />
11. Press the PRINT switch. “END OK?” appears on the screen.<br />
12. Press the PRINT switch. “SETTING MENU DATA” will be printed out, and the<br />
measurement screen will appear.<br />
Page 16<br />
TOPCON Corporation
SETTING MENU DATA<br />
VD 12.00<br />
STEP 0.25<br />
…………………………………<br />
………… ……<br />
RS2<strong>32</strong>C FORM EXT<br />
BAUD-RATE 9600<br />
…… ……<br />
…… ……<br />
LIGHTING RING<br />
Figure 2-5 Printout Setting Data<br />
13. Please make sure settings of the KR-7000P/ KR-7100P are correctly configured:<br />
2.4.2 <strong>Cal</strong>ibration Data Registration<br />
Before transferring the calibration data of the KR-7000P/ KR-7100P to the PC, proceed<br />
with following steps:<br />
3.2 Start the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. When starting the software, the message,<br />
“<strong>Cal</strong>ibration Data does not exist. Please setup <strong>Cal</strong>ibration data” may appear until the<br />
calibration data has been transferred from the KR-7000P/ KR-7100P (Figure 2-6).<br />
Figure 2-6 No calibration data message box<br />
Please click the ‘OK’ button to prepare to receive data from the KR-7000P/ KR-<br />
7100P.<br />
3.2 Set the instrument's type by clicking on following controls from the main menu<br />
bar:<br />
• Select [Port]-[Select Instrument] from Menu Bar.<br />
• Select KR7000P from [Select Instrument].<br />
If the KR-7000P/ KR-7100P unit is replaced, it is necessary to transfer the new unit’s<br />
calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the registered calibration data.<br />
In this case, the message in Figure 2-6 may not appear, because the old calibration data is<br />
already registered.<br />
Page 17<br />
TOPCON Corporation
2.4.3 <strong>Cal</strong>ibration Data Transfer Procedure<br />
The following steps are required for the KR-7000P/ KR-7100P to transfer the calibration<br />
data:<br />
1. Dip switch 7 & 8 at position ON and 1 through 6 at position OFF. Rotary switch to<br />
position 7.<br />
Figure 2-7 Switches Condition (<strong>Cal</strong>ibration Data Registration)<br />
2. Turn on the KR-7000P/ KR-7100P, (a “PLACIDO CHECK” message will appear).<br />
3. Press the joystick once to enter the submenu.<br />
4. Press the PRINT switch and move the cursor to “5:CALIB DATA OUT”.<br />
5. Press the joystick. The KR-7000P/ KR-7100P will send the calibration data to the<br />
PC.<br />
6. Turn off the KR-7000P/ KR-7100P.<br />
2.4.4 KR-7000P Measurement mode Setting<br />
In order to transfer measurement data to the PC, please make sure Dip switches are in the<br />
following positions:<br />
Dip switch 7 position OFF<br />
Dip switch 8 position ON<br />
Rotary switch position ‘0’<br />
Page 18<br />
TOPCON Corporation
Turn on the KR-7000P/ KR-7100P.<br />
Figure 2-8 Switches Condition (Measurement Mode)<br />
After measuring the patient’s eyes, press the GRAPHIC PRINT SWITCH on the KR-<br />
7000P/ KR-7100P to transfer the necessary data to the PC.<br />
Memo:<br />
1. If “MAPPING PRT” on the KR-7000P/ KR-7100P is set for PRINT, you may also<br />
use the PRINT switch to send data.<br />
2. If a printout from the KR-7000P/ KR-7100P is not necessary, lift the paper retainer<br />
lever.<br />
3. Usually, registering calibration data needs to be done only once. However, when the<br />
KR-7000P/ KR-7100P unit is replaced or returned from repair, it is necessary to<br />
transfer the new calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the<br />
registered calibration data.<br />
4. When a KR-7000P/ KR-7100P keeps displaying the “WAITING DIPSW8” message<br />
for a long time, please make sure the Dip switch 8 is ON.<br />
5. When a KR-7000P/ KR-7100P keeps displaying the “CALIBRATING” message for<br />
a long time, please check the interface cable connection.<br />
6. When a <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> displays error messages such as “CEbreak”, “CEerror”,<br />
etc, please check the KR-7000P/ KR-7100P settings are “RS2<strong>32</strong>C FORM EXT”,<br />
“BAUD RATE 9600”.<br />
2.5 Preparation for KR-8000P<br />
2.5.1 Setting up KR-8000P<br />
The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />
8100P types of the instruments.<br />
While pressing MENU SWITCH on the control panel of the KR-8000PA/ KR-8100PA/<br />
KR-8100P, turn on the KR-8000PA, "INITIAL MENU” screen appears.<br />
Page 19<br />
TOPCON Corporation
1. Select “ON-LINE" option from the “INITIAL MENU” screen to bring up the “ON-<br />
LINE” screen.<br />
Figure 2-9 “ON-LINE” Screen<br />
2. Press the joystick and bring the cursor to “RS2<strong>32</strong>C FORM”.<br />
3. Press the PRINT switch to highlight EXT (available for later version of <strong>Color</strong><br />
<strong>Mapping</strong> <strong>32</strong> Version 1.24).<br />
4. Press the joystick and bring the cursor to “BAUD-RATE”.<br />
5. Press the PRINT switch and highlight 38400 or 19200. (For saving transfer time,<br />
select 38400Bps).<br />
6. Press the PRINT switch and bring the cursor to “EXIT”.<br />
7. Press the PRINT switch to exit "ON-LINE" screen.<br />
2.5.2 <strong>Cal</strong>ibration Data Registration<br />
Before transferring the calibration data of the KR-8000PA/ KR-8100PA/ KR-8100P to<br />
the PC, proceed with following steps:<br />
1. Start the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. When starting the software, the message,<br />
“<strong>Cal</strong>ibration data does not exist. Please setup <strong>Cal</strong>ibration data” may appear until the<br />
calibration data has been transferred from the KR-8000PA/ KR-8100PA/ KR-8100P<br />
(Figure 2-6). Please click the “OK” button to prepare to receive data from the KR-<br />
8000PA/ KR-8100PA/ KR-8100P.<br />
2. Set the instrument's type by clicking on following controls from the main menu bar:<br />
• Select [Port]-[Select Instrument] from Menu Bar.<br />
• Select KR8000P/38K or /19K from [Select Instrument].<br />
Selection must correspond to the instrument’s “BAUD-RATE” setting, please refer to<br />
the Section 2.5.1 (Step 4)<br />
If the KR-8000PA/ KR-8100PA/ KR-8100P unit is replaced, it is necessary to transfer<br />
the new unit’s calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the registered<br />
calibration data.<br />
2.5.3 KR-8000P Setting and <strong>Cal</strong>ibration Data Transfer<br />
The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />
8100P types of the instruments.<br />
1. Turn on KR-8000PA/ 8100PA/ 8100P "PLACIDO SET" screen from "INITIAL<br />
MENU".<br />
Page 20<br />
TOPCON Corporation
Figure 2-10 PLACIDO SET Screen<br />
2. Press the joystick and bring the cursor to “PC S/W SETUP”.<br />
3. Press the PRINT switch. <strong>Cal</strong>ibration data is transferred automatically and setup of<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is done.<br />
4. Press the joystick and bring the cursor to “EXIT”.<br />
5. Press the PRINT switch. “EXIT OK?” appears on the screen. Press the PRINT<br />
switch. “SETTING MENU DATA” will be printed out, and the measurement<br />
screen will appear<br />
SETTING MENU DATA<br />
VD 12.00<br />
STEP 0.25<br />
…………………………………<br />
………… ……<br />
RS2<strong>32</strong>C FORM EXT2<br />
BAUD-RATE 38400<br />
…… ……<br />
…… ……<br />
LIGHTING RING<br />
Figure 2-11 SETTING MENU DATA Screen<br />
6. Please make sure settings of the KR-8000PA are correctly configured.<br />
2.5.4 KR-8000P Measurement Mode Setting<br />
The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />
8100P types of the instruments.<br />
In order to transfer measurement data to the PC, configure the instrument RS2<strong>32</strong>C output<br />
port by performing the following steps:<br />
1. Press “MENU SWITCH” of the control panel on measurement mode.<br />
"SETTING MENU" screen appears.<br />
2. Press the joystick and bring the cursor to “RS2<strong>32</strong>C”.<br />
3. Press the PRINT switch to highlight “YES”.<br />
Page 21<br />
TOPCON<br />
TOPCON Corporation
Figure 2-12 SETTING MENU Screen<br />
Press the joystick and bring the cursor to “EXIT”. Press the PRINT switch to exit<br />
"SETTING MENU" screen<br />
After measuring the patient’s eyes, press the GRAPHIC PRINT SWITCH on the control<br />
panel of KR-8000PA/ KR-8100PA/ KR-8100P to transfer the necessary data to the PC.<br />
Memo:<br />
1. If “MAPPING PRT” on the KR-8000PA/ KR-8100PA/ KR-8100P is set to<br />
“PRINT”, you may also use the PRINT switch to send data.<br />
2. If a printout from the KR-8000PA/ KR-8100PA/ KR-8100P is not necessary, lift the<br />
paper retainer lever.<br />
3.2 Usually, registering calibration data needs to be done only once. However, when<br />
the KR-8000PA/ KR-8100PA/ KR-8100P unit is replaced or returned from repair, it<br />
is necessary to transfer the new calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to<br />
update the registered calibration data.<br />
3.2 If no data is transferred from a KR-8000PA/ KR-8100PA/ KR-8100P, Please<br />
check the setting of port. “RS2<strong>32</strong>C” is “YES” and “RS2<strong>32</strong>C FORM EXT”, “BAUD<br />
RATE 19200 or 38400”.<br />
3.2 When a KR-8000PA/ KR-8100PA/ KR-8100P keeps displaying the “CABLE<br />
UNCONNECTED” message, please check the interface cable connection and port<br />
setting.<br />
Page 22<br />
TOPCON Corporation
Chapter 3<br />
Getting Started<br />
3.1 Starting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
To start <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> click on the shortcut icon labeled KRP<strong>32</strong><br />
(Figure 3-1) in your Desktop Window.<br />
KRP<strong>32</strong><br />
Figure 3-1 Shortcut Icon<br />
3.2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Components<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Window<br />
The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window consists of several elements such as the Title Bar, the<br />
Menu Bar, the Toolbar and the Status Bar. The Title Bar is located at the top of the<br />
window. The Menu Bar is just below the Title Bar, and the Toolbar below that. The<br />
Status Bar is located at the bottom of the window.<br />
Figure 3-2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Window<br />
Page 23<br />
TOPCON Corporation
<strong>Color</strong> maps are displayed in windows referred to here as “Report Windows”. A report<br />
window is created when a <strong>Color</strong> Map file is opened and is removed when the file is<br />
closed. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window is maximally sized on start up to make room for<br />
report windows to be as large as possible. It is possible to make the window smaller,<br />
however, the report windows can never be larger than the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window.<br />
Title Bar The title bar is used to show the report currently displayed.<br />
Menu Bar The menu bar consists of several items. By clicking these items,<br />
the pull-down menu related to the item will be opened. By<br />
choosing an item in a pull-down menu, you can call the function<br />
assigned to the item.<br />
[File] Open...<br />
Close*<br />
Close All*<br />
Save*<br />
Save As*<br />
Open from Database<br />
Save to Database*<br />
Print...*<br />
Print Preview*<br />
Print Setup...<br />
Exit<br />
[Patient] Select Patient<br />
Register Patient<br />
Deleted Patient<br />
* Item in the [Patient] menu is enabled only if the<br />
Patient Registration Database is not empty<br />
[View] Toolbar<br />
Status Bar<br />
[Port] Select Com Port for KR<br />
Select Com Port for CV<br />
Reset Port<br />
Select Instrument***<br />
KR7000P<br />
KR8000P/19K<br />
KR8000P/38K<br />
[Utility] Map Options*<br />
<strong>Cal</strong>l Map*<br />
<strong>Color</strong> Map*<br />
Contact Lens Fitting [OD(Right)]*<br />
Contact Lens Fitting [OS(Left)]*<br />
Contact Lens Utility<br />
Delete Data from Database<br />
Date Format Settings<br />
Change Parameters**<br />
Page 24<br />
TOPCON Corporation
Note:<br />
Pan & Tilt**<br />
Profile Graph**<br />
Isometric Graph**<br />
[Window]* Cascade<br />
Tile<br />
[Help] About <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>...<br />
* Items in the Menu Bar or in the each pull-down menu are available only,<br />
when the <strong>Color</strong> Map Report is displayed.<br />
** Items in the Menu Bar or in the each pull-down menu are available only,<br />
when the Contact Lens Fitting Report is displayed.<br />
*** Items in the Menu Bar or in the pull-down menu are available only, when<br />
there is no document (report) in the main frame window.<br />
Tool Bar The tool bar provides quick access to items in pull-down menus of<br />
the menu bar. By clicking an icon on the tool bar, the function<br />
assigned to the icon is directly called.<br />
Note:<br />
Open Icon Open [File]<br />
Save Icon Save [File]<br />
Save to DB Save to the Database [Utility]<br />
Open from DB Retrieve from the Database [Utility]<br />
Map Options Map Options Settings * [Utility]<br />
<strong>Color</strong> Map/Contact Lens <strong>Color</strong> /Simulation Maps* [Utility]<br />
Pan & Tilt Icon Contact Lens Fitting** [Utility]<br />
Printer Icon Print [File]<br />
Info Icon About <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> [Help]<br />
Page 25<br />
TOPCON Corporation
* Items in the Menu Bar or in the each pull-down menu are available only,<br />
when the <strong>Color</strong> Map Report is displayed.<br />
** Items in the Menu Bar or in the each pull-down menu are available only,<br />
when the Contact Lens Fitting Report is displayed.<br />
Status Bar The status bar shows the status of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>.<br />
Report Window When a report is displayed on the screen, a new window for the<br />
report is created in the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window. Although two<br />
or more report windows may be open simultaneously, only one<br />
may be active. Most functions work on the report in the active<br />
report window. Each report window has the title on the top to<br />
show identification information of the report.<br />
3.3 Operating <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
Once <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is started, the communication port will be initialized<br />
automatically and ready to receive transfer data from the KR-7000P/ KR-7100P/KR-<br />
8000PA/ KR-8100PA/ KR-8100P. When the data is received, the corneal topography<br />
mapping data is reconstructed, and a new report window is created automatically. Use<br />
the menu icons on the tool bar as a short cut for some menu items. A menu item or a<br />
menu icon displayed in gray means that the item or the icon is not available at that time.<br />
If the port for the TOPCON CV Power Supply Unit has been set, the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
creates the necessary data using the specified data format and sends it to the TOPCON<br />
CV Power Supply Unit.<br />
To display reports, choose [Open] from the [File] menu. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display<br />
a dialog box to get necessary information. Type or select items in the dialog box, then<br />
click “OK” to continue. To apply any options to the displayed report, choose [Option<br />
Setting] from the [Utility] menu. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display the option setting<br />
dialog box to change option parameters. Change parameters in the dialog box, then click<br />
“OK” to continue. The changed parameters will be immediately applied to the report.<br />
To get contact lens simulation report window, choose “Contact Lens Fitting” from the<br />
[Utility] menu. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display the Contact Lens Parameter setting<br />
dialog box to set for perform the contact lens fitting simulation.<br />
To print a report, choose [Print] from the [File] menu. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will send the<br />
report information to the printer. If the “Auto Print” check box in the option setting<br />
dialog box has been checked, the color map report information will be sent automatically<br />
as soon as the program reconstructs the new report using the data transferred from the<br />
KR-7000P/ KR-7100P/KR-8000PA/ KR-8100PA/ KR-8100P.<br />
3.4 Exiting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
To exit <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, choose [Exit] from the [File] menu. Upon exit, the program<br />
closes all report windows and releases resources. Before exiting, save any important new<br />
Page 26<br />
TOPCON Corporation
eports that were created during the session, or they will be lost. The program does not<br />
prompt to save new reports when exiting.<br />
Page 27<br />
TOPCON Corporation
Chapter 4<br />
<strong>Color</strong> Map Report Data<br />
4.1 Receiving <strong>Color</strong> Map Report Data<br />
To receive measured data from the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />
KR-8100P, press the GRAPHIC PRINT SWITCH (open the cover/door of the control panel<br />
to use this switch) on the instrument. After receiving data, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
reconstructs the corneal structure data.<br />
4.1.1 KR- 8000P Data Transfer Setting<br />
To transfer measurement data from the KR-8000PA/KR-81000PA/KR-8100P proceed<br />
with flowing steps.<br />
1. Press the GRAPHIC PRINT SWITCH, “IMAGE DATAOUT MENU” screen will<br />
appear (Figure 4-1).<br />
2. Press the joystick to highlight one of the menu items (Figure 4-1).<br />
Each item indicates type of transferring image that will be sent to the <strong>Color</strong> <strong>Mapping</strong><br />
<strong>32</strong> with numeric data.<br />
Figure 4-1 Image Data Out Screen<br />
Each menu item corresponds to the type of image that will be attached to the numeric<br />
data package transferring to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, as specify below:<br />
• NO IMAGE data transfer without image<br />
• BOTH data transfer with the images for Left and Right eyes<br />
• RIGHT data transfer with the Right eye’s image only<br />
• LEFT data transfer with the Left eye’s image only<br />
3. Press GRAPHIC PRINT SWITCH to conform that the menu items’ selection is<br />
completed. Instrument must display “RS2<strong>32</strong> DATA OUT” message.<br />
Each item indicates type of transferring image that will be sent to the <strong>Color</strong> <strong>Mapping</strong><br />
<strong>32</strong> with numeric data.<br />
Page 28<br />
TOPCON Corporation
4.2 Display <strong>Color</strong> Map Report Data<br />
Using the transferred report data, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> creates a new color map report<br />
window. This report window consists of < Ref Data >, < Kerato Data >, < Peripheral<br />
Data >, and the color map reports.<br />
4.3 Patient Registration<br />
Figure 4-2 <strong>Color</strong> map report display<br />
The patient registration service allows the user to create a database of records with<br />
patient information on an available storage media.<br />
� There are two alternatives to access the Patient Registration dialog box:<br />
• By selecting the [Patient] item from the Menu Bar;<br />
• By selecting [Save to Database] from the [File] pull-down menu (see Section 4.4)<br />
� To register a patient using the Menu Bar:<br />
Page 29<br />
TOPCON Corporation
• Select the [Patient] item from the Menu Bar;<br />
• Select [Register Patient] item from [Patient] menu (Figure 4-3).<br />
• Fill in the dialog box form, click on “Register” button to create patient record in<br />
the Database;<br />
Figure 4-3 Register Patient Dialog Box<br />
� To retrieve and update patient information from the Database:<br />
• Select [Patient] item in the Menu Bar;<br />
• Click on the [Select Patient] item from [Patient] menu. (Figure 4-4);<br />
Figure 4-4 Select Patient Dialog Box<br />
• Enter a search key in ID. Code and/or Last Name lines of the dialog box:<br />
• Click the “Information” button to retrieve a patient record in the Patient<br />
Information dialog box. To store modified record, click the “Update” button.<br />
• Click the “OK” button to retrieve a color map report image associated with<br />
the entered search key;<br />
� To delete record with patient information from the Database:<br />
Page 30<br />
TOPCON Corporation
• Select [Patient] item in the Menu Bar.<br />
• Select [Delete Patient] item from [Patient] menu.<br />
• Select a record from the list and press the “Delete” button on the dialog box;<br />
4.4 Saving Report Data<br />
Figure 4-5 Delete Patient Dialog Box<br />
� <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> support two types of report data storage:<br />
• Saving report data as a regular file with extension *.krp/*.kp8 for data received<br />
from KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P instrument on<br />
any available storage media.<br />
• Saving report data as a database file in database disk area allocated on available<br />
archive media. Patient information registration records and associated <strong>Color</strong><br />
<strong>Mapping</strong> <strong>32</strong> reports are stored in the created archive.<br />
� To save a document as a regular file from the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Report Window:<br />
• Select the [File] item in the Menu Bar.<br />
• Select the [Save As] item. (Do not change extension ‘.krp or .kp8’)<br />
Figure 4-6 Save Report Dialog Box<br />
When saving the report data as a file, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> attaches the calibration data to<br />
the file. This means that even if the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />
Page 31<br />
TOPCON Corporation
KR-8100P unit is replaced, the saved report data file may still be used in the new<br />
configuration.<br />
� To save a document to the database file from the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Report Window:<br />
• Select the [File] item in the Menu Bar.<br />
• Select [Save to Database] (see Section 3.2).<br />
• Fill patient registration form in the dialog box.<br />
• Click “OK” (Figure 4-7).<br />
• Fill dialog box form and click “OK” to save the report.<br />
Figure 4-7 Save Data Dialog Box<br />
4.5 <strong>Color</strong> Map Report Window Control<br />
Closing <strong>Color</strong> Map Report Window<br />
To close a report window, make the window active, then choose [Close] from the [File]<br />
menu or use the system menu of the report window. Alternatively, access the system<br />
menu of the window by clicking the square at the upper left corner, the left end of the<br />
title of the window.<br />
Putting <strong>Color</strong> Map Report Window on Top<br />
Many report windows can be opened simultaneously. The number of report windows<br />
one can open is only limited by the amount of system memory. The report window<br />
opened last is displayed on top. To place a different report window on top, click<br />
anywhere on that report window using the left mouse button.<br />
� To enable this service:<br />
Page <strong>32</strong><br />
TOPCON Corporation
• Click on [Open] from DB icon in the Tool Bar to display the Select Data dialog<br />
box.<br />
• Select data (see Section 4.6) to bring up a color map report data from the database.<br />
• Select a measurement point by moving the mouse on the image map and clicking<br />
the left button. Measurements associated with position of cross-hair marker on the<br />
color image will be displayed at the left top corner of the <strong>Color</strong> Map Report<br />
Window.<br />
4.6 Opening Saved Report Data<br />
To open a saved report data file, choose [Open] from the [File] menu, or click the open<br />
icon on the tool bar. The following dialog box will be displayed to select the saved<br />
report file:<br />
Figure 4-8 Open Report Dialog Box<br />
After choosing a report file and clicking “OK”, the dialog box will be hidden, and the<br />
selected report will be displayed in a new report window.<br />
The above dialog box can be displayed anytime by choosing [Open] in the [File] menu or<br />
by clicking the open file icon on the tool bar.<br />
Page 33<br />
TOPCON Corporation
4.7 Viewing Report Data from the Database<br />
� Report data saved on the database disk as archive can be retrieved in the following<br />
formats:<br />
• “Single map” is a view of one image with associated measurements and patient<br />
data (Figure 4-9). Selection of more then one file from the Record List Box<br />
(Figure 4-8) list disables the “OK” button of the Select Data dialog box.<br />
• “Multi map” image is a side by side view of several images with associated<br />
measurements and patient data (Figure 4-9). Patient records may be selected for<br />
the same eye as well as for different eyes from the database Record List Box.<br />
Selection of more records then the map can display disables the “OK” button of<br />
the Select Data dialog box (Figure 4-9);<br />
• “Difference Map” view is a color presentation of the difference between two<br />
images retrieved from database with associated measurement difference display<br />
box (Figure 4-10). The compared images are also presented side by side on the<br />
map. Selecting records of different eyes from the Select Data dialog box disables<br />
the “OK” button.<br />
� To retrieve <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> report image from the Database:<br />
• Select the [Open] from DB icon from the Tool Bar or click [Open] from Database<br />
on the [File] pull-down menu (see Section 3.2). The [Select Data] dialog box<br />
(Figure 4-9) will be displayed on the screen<br />
• Use the scroll list or press the [Select Patient] button to enter a search key;<br />
• Click the “OK” button to retrieve the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> report from the database<br />
Figure 4-9 Select Data Dialog Box<br />
Page 34<br />
TOPCON Corporation
Multi Map View<br />
Map view displays running measurements associated with the position of the cursor on<br />
the selected color map image. Measurement display in the bottom right corner of each<br />
image box (Figure 4-10) contains of following information:<br />
PWR Power in diopters at the cursor position;<br />
RAD Radius of curvature at the cursor position<br />
DIST Distance of cursor from the center of the grid in mm<br />
AXIS Angle of the axis connecting the cursor with the center of the grid<br />
in degrees (0°-360°);<br />
DISTP* Distance of cursor from the center of the pupil in mm<br />
AXISP* Angle of the axis connecting the cursor with the center of the pupil<br />
in degree (0°-360°)<br />
� To enable Multi Map View:<br />
• Click the [Open] from DB icon in the Tool Bar to bring up “Select Data” dialog<br />
box;<br />
• Select one of the “Multi Map” options from the “Select Map” box of the<br />
“Select Data” dialog box to bring up a “Multi Map” report data from the<br />
database;<br />
• Select measurement point by moving the mouse on the image map and clicking<br />
the left button. Measurements associated with the position of cross-hair marker on<br />
the color image will be displayed<br />
Figure 4-10 Multi Images Map View<br />
Page 35<br />
TOPCON Corporation
Difference Map View<br />
Map view displays running measurements associated with position of the cursor on the<br />
selected color map image. Measurement display is in the bottom, right corner of each<br />
comparing image box (Figure 4-10). It contains the same information type that the<br />
images on the Multi Map View along with the difference between the two. The following<br />
measurement information is provided for each map:<br />
PWR Power measurements of each map and the difference of the two.<br />
RAD Radius of curvature for each map and the difference of the two (mm).<br />
DIST Distance from the center of the grid for each map in mm.<br />
The difference map reports the distance of the first image in mm.<br />
AXIS The angle between the line connecting the cursor with the center of the<br />
grid and X-axis of the greed in degrees (0°-360°). The difference map<br />
reports the angle value of the first image.<br />
DISTP* Distance from the center of the pupil for each map in mm.<br />
AXISP* The angle between the line connecting the cursor with the center of the<br />
pupil and X-axis of the greed in degrees (0°-360°). The difference map<br />
reports the angle value of the first image.<br />
* Measurement is displayed if pupil detection data is available<br />
in the image file.<br />
To enable “Difference Map” View follow the steps from the paragraph “Multiple Map<br />
View” for “Difference Map” selection in the “Select Map” box of “Select Data” dialog<br />
box. (Figure 4-9).<br />
Figure 4-11 Difference Image View<br />
Page 36<br />
TOPCON Corporation
Chapter 5<br />
Option Settings<br />
When a color map report is open, the options may be changed for that report. To get the<br />
option dialog box menu, choose [Map Options] from the [Utility] menu, or click the<br />
option icon on the tool bar. The following dialog box will be displayed to change<br />
options:<br />
5.1 Map Method<br />
Figure 5-1 Option Setting Dialog Box<br />
Sagittal: This power is based on the normal distance from the cornea to the<br />
refractive axis (also known as Axial Power).<br />
Tangential: This power is based on the surface radius for each data point (also known<br />
as Instantaneous Power).<br />
5.2 Scale Unit<br />
Diopter: Displayed color map based on Diopter power scale.<br />
Page 37<br />
TOPCON Corporation
mm: Displayed color map based on mm power scale.<br />
5.3 Map Type<br />
Adjustable: Use the color map center and step values to adjust the color levels. By<br />
using appropriate values, more detail can be seen in the map. (The center<br />
value will appear as the bottom of the eleventh color.)<br />
Normalized: Minimum and Maximum levels of the data are used for limits of the color<br />
map. Thus, data will stretch over entire map. In cases of extreme data<br />
range, the top 1% and bottom 1% are excluded.<br />
Absolute: Standard scale ranging from 35.5D to 52D with a step size of 1/2D, or<br />
6.0mm to 9.3mm with a step size of 0.1mm. 33 different levels are used<br />
for this map by 22 colors.<br />
5.4 Overlays<br />
Cross Scale: X and Y axes overlay<br />
Angle Scale: Displays angle measurement around the color maps.<br />
mm Grid Scale: Overlay of a 1 mm grid<br />
Kerato Axis:<br />
None: No Kerato Axis displayed.<br />
Kerato Axis: Kerato Axis displayed based on the Keratometer axis.<br />
Sim-K Axis: Shows the keratometric circle (radius 1.5mm) along with the<br />
strongest axis ( as a black line ) and weakest axis ( as a gray line ).<br />
Pupil: Displays pupil boundary curve. Pupil check box is enabled, if pupil edge<br />
detection points are available in the opened map document file (Figure 5-2)<br />
This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />
Page 38<br />
TOPCON Corporation
5.5 Peripheral<br />
Figure 5-2 <strong>Color</strong> map with pupil overlays<br />
When measuring a patient eye with 5 point peripheral measurement, this switch becomes<br />
available along with peripheral display options:<br />
Figure 5-3 Peripheral Point 5 Option Dialog Box<br />
Display Point 5 Map: If this box is checked, the peripheral data will be displayed as a 15<br />
degree step map. If this box is unchecked, the peripheral data will be displayed as<br />
numeric data.<br />
Eccentricity Options<br />
No E Value: No eccentricity data displayed.<br />
Meridian E Value: Meridian eccentricity data displayed. (2 directions)<br />
Semi meridian E Value: Semi-meridian eccentricity data displayed. (4 directions)<br />
5.6 Others<br />
Page 39<br />
TOPCON Corporation
Initial Display Top: If this box is checked, the top of the new report window will be<br />
displayed. If this box is not checked, the bottom ( <strong>Color</strong> Map ) of<br />
the new report window will be displayed.<br />
Save Settings: If this box is checked, and the ‘OK’ switch is clicked, <strong>Color</strong><br />
<strong>Mapping</strong> <strong>32</strong> saves and uses these settings as the defaults for new<br />
reports.<br />
Auto Print: If this box is checked, data will automatically be sent to the printer<br />
(using default option settings) as soon as the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
receives and reconstructs the color map structure.<br />
5.7 Image View<br />
This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />
Map: If this button is active color map with pupil center marker will be<br />
displayed, every time user opens map document file. Pupil center marker<br />
will appear only for document files containing pupil detection points.<br />
Image: If this button is active eyes’ color map reports will be replaced with its<br />
images in the color map report window (Figure 5-4). If image is available<br />
only for one eye in the opened document, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display<br />
image only for this eye and color map for the eye without image.<br />
Overlay: This button repeats the same logic as the “Image” button. The difference is<br />
that image is overlaid with color map for each eye which image data is<br />
available.<br />
Side by Side: The button is enabled only for the document, which was opened from the<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Database and its display format to “Single Map”<br />
(Section 4.7). Also, eye image data must be available in the database<br />
document.<br />
Page 40<br />
TOPCON Corporation
Figure 5-4 Report window (Image view – Image)<br />
Page 41<br />
TOPCON Corporation
Chapter 6<br />
Contact Lens Simulation Report Data<br />
6.1 Display Contact Lens Simulation Report Data<br />
To get a Contact Lens Simulation Report data when a <strong>Color</strong> map Report window is<br />
activated, perform following steps:<br />
• Choose “Contact Lens Fitting” from the [Utility] “<strong>Cal</strong>l Map” menu, or click the<br />
[Contact Lens] Icon on the Tool bar. When the Contact Lens Icon was clicked and a<br />
<strong>Color</strong> map Report has data for both eyes, the following “Select Processing Eye”<br />
dialog box will be displayed.<br />
Figure 6-1 Select processing Eye Dialog Box<br />
The contact lens simulation Parameter Setting dialog box will appear (Figure 6-4).<br />
• Select simulation parameters and simulation report view from the Parameter Setting<br />
dialog box, as specify in Section 6.4 later in this chapter.<br />
• Click “OK” to display the contact lens simulation report. The image view type in the<br />
Fitting <strong>Color</strong> Map box of the report is defined by the settings of the Image View<br />
control buttons in the “Parameter Setting” dialog box (Section 6.4).<br />
• The contact lens simulation report window will display Fitting <strong>Color</strong> Map view<br />
(Figure 6-2).<br />
Page 42<br />
TOPCON Corporation
Figure 6-2 Contact Lens Fitting Map display with pupil overlay<br />
6.1.1 Contact Lens Simulation Precaution<br />
Contact Lens Simulation<br />
Physician can not use report data for lens prescription without<br />
their personal validation.<br />
Page 43<br />
TOPCON Corporation
Figure 6-3 Contact Lens Fitting Map with pupil and eye image overlay display<br />
Page 44<br />
TOPCON Corporation
6.2 Parameters of the Contact Lens Simulation Report<br />
Abbreviations for the Contact Lens simulation report:<br />
BC Base curve (mm)<br />
BC2 Base curve 2 for toric lens (mm)BC must be greater then BC2<br />
E1 Eccentricity of aspheric lens<br />
TD Total diameter of a lens (mm)<br />
OZ Optic zone (mm) (which affects the optical function of a lens)<br />
SC Secondary curve (mm)<br />
SCW Secondary curve width (mm)<br />
E2 Eccentricity for secondary curve<br />
PC Peripheral curve (mm)<br />
PCW Peripheral curve width (mm)<br />
PC2 Peripheral curve 2 for toric lens (mm) PC must be greater than PC2<br />
AEL Axial edge lift (mm)<br />
VD Vertex distance of spectacles (mm)<br />
PWR Power of the contact lens to correct refraction spherical data (mm)<br />
PanH amount of horizontal lens lag from the corneal (mm)<br />
PanV amount of vertical lens lag from the corneal center (mm)<br />
TilH Horizontal lens tilt angle (mm)<br />
TilV Vertical tilt angle (mm)<br />
For a three curve contact lens, BC, TD, OZ, SC, SCW, PC, and PCW are necessary. If<br />
for a two-curve lens, SC and SCW, or PC and PCW may be omitted. Either SC or PC can<br />
be ignored if their width field (SCW or PCW) is zero.<br />
6.3 Handling Each Lens Type<br />
Lens types can be set in the parameter setting dialog box.<br />
RGP For a RGP lenses, a total of three curves including a base curve ( BC ), a<br />
secondary curve ( SC ), and a peripheral curve ( PC ), are available.<br />
Alternatives ( i.e. BC+SC only, or BC+PC only ) are also available.<br />
TORIC When a TORIC lens type has been selected, the BC2 and PC2 data fields<br />
will be activated. There are basic curve 2 and peripheral curve2,<br />
respectively. Listed below are two available lens structures in the TORIC<br />
lens type.<br />
BC (toric) + SC (spheric) + PC (spheric)<br />
BC (toric) + PC (toric)<br />
If the PC value is same as the PC2 value, the contact lens simulation<br />
function assumes the lens has spherical peripheral curves. In this case, the<br />
SC, and SCW fields will be activated. However, if PC differs from PC2,<br />
since this is a toric peripheral, the SC and SCW fields become disabled.<br />
ASPHERIC When an ASPHERIC lens type has been selected, the E1 and E2 fields<br />
will be activated. The meanings of these fields are eccentricity 1 and 2<br />
Page 45<br />
TOPCON Corporation
espectively. These values are calculated from the corneal topography,<br />
and initially, the two values are the same. One or both of them may be<br />
changed according to the manufacture’s lens specification. For a spherical<br />
secondary curve, set E2 equal to 0 to make it represent a spherical<br />
secondary curve. Use the two lens structure as follows:<br />
BC (aspheric) + SC (aspheric) + PC (spheric)<br />
BC (aspheric) + SC (spheric) + PC (spheric)<br />
In both cases, PC can be omitted when PCW is zero.<br />
SOFT Most SOFT lenses have one BC and One PC, so secondary curve fields<br />
are disabled. The fluorescein simulation is not available for soft lenses.<br />
PMMA For a PMMA lens type, the software treats the lens structure in the same<br />
way as a RGP lens.<br />
6.4 Parameter Setting<br />
When a contact lens simulation report is open or the ‘Contact Lens Fitting’ function from<br />
a color map report has been called for the first time, the parameters may be changed for<br />
that report. To get the parameter setting dialog box menu, choose ‘Change Parameters’<br />
from the [Utility] menu, or click the parameter icon on the tool bar. The following dialog<br />
box will be displayed to change parameters. (‘Select Data Set’ item will be appeared only<br />
when some contact lens data were registered – See Section 6.5 ):<br />
Figure 6-4 Parameter Setting dialog box<br />
Lens Type<br />
Please refer to section “6.3 Handling each lens type”.<br />
Page 46<br />
TOPCON Corporation
Map Type<br />
Absolute map type connects tear layer thickness to an actual fluorescent color.<br />
Normalized map type assigns colors to express the entire thickness range.<br />
Overlays<br />
The functions of the Overlay group of controls are the same as functions of the Overlay<br />
control group in the Map Option Settings dialog box for the <strong>Color</strong> Map Report document<br />
type, refer to the Section 5.4 of the manual.<br />
Reference BC<br />
Parameter helps to find a base curve that meets a certain apical clearance demand. The<br />
software finds the radius of curvature of a sphere, which is defined by three points.<br />
These points are the center of the lens with a certain apical clearance, and two optic zone<br />
end points, which touch the cornea.<br />
Sagittal Vault<br />
If the sagittal vault value is supposed to be a certain value, input that value. An<br />
appropriate TD value will be calculated for this sagittal vault.<br />
Axial Edge Lift<br />
The required AEL (Axial Edge Lift) value is specified, the software finds the appropriate<br />
PC value.<br />
Graph Setting: Profile Graph<br />
The tear layer thickness profile graph is displayed at the specified angle value.<br />
Graph Setting: Isometric Graph<br />
The circular tear layer variation graph is displayed at the specified radius value.<br />
Select Data Set<br />
The pull down scroll list appears when more then one contact lens data sets have been<br />
registered. The software will set the parameters based on the selected contact lens data<br />
set.<br />
Image View<br />
This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />
The functions of the Image View control group are the same as the functions of its<br />
counterpart in the Map Option Settings dialog box of the <strong>Color</strong> Map Report document<br />
type, refer to the Section 5.7 of the manual. There are three alternative views including<br />
the default Fitting <strong>Color</strong> Map view available and may be selected by setting the Image<br />
View control group in on of the following options:<br />
• MAP option to display Fitting <strong>Color</strong> Map(Figure 6-2)<br />
• IMAGE option to display eye image the same as in the <strong>Color</strong> Map Report, refer to<br />
Section 5.7(Figure 5-4)<br />
• OVERLAY option to display Fitting <strong>Color</strong> Map overlaid with eye image (Figure 6-3)<br />
6.5 Contact Lens Utility<br />
Page 47<br />
TOPCON Corporation
To get the contact lens utility dialog box menu, choose ‘Contact Lens Utility’ from the<br />
[Utility] menu. The following dialog box will be displayed to set the default and register<br />
the contact lens parameters sets. The number of the registered contact lens sets is up to 10.<br />
Figure 6-5 Contact Lens Utility dialog box<br />
Available setting parameters for the default and register contact lens sets are shown<br />
below.<br />
Contact Lens Parameters Default Setting<br />
When you have a commonly used contact lens specification, you may define these<br />
parameters. Definable parameters are “Lens Type”, “Map Type”, “Total Diameter of the<br />
contact lens”, “Optical Zone of the contact lens”, “Vertex Distance”, and “Overlays for<br />
the simulation report”. If you have registered some manufacture’s contact lens data sets,<br />
a “Select Data Set” pull-down list will appear.<br />
Figure 6-6 Contact Lens Parameters Default Setting dialog box<br />
CL Advice Enable<br />
This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />
A check box is located in the ‘Contact Lens Parameters Default Setting’ dialog box<br />
(Figure 6-6).<br />
This check box enables automatic setting of the simulation parameters for selected lens<br />
type (Section6.2) to the values calculated by the instrument. These values are displayed<br />
in the ‘’ section of the <strong>Color</strong> Map Report, please refer to the Section<br />
4.2(Figure 4-2). CL Advice data is available only for RGP and ASPHERIC type of lens.<br />
The other lens type selections are disabled as long as the check box is checked.<br />
Contact Lens Data Register<br />
It is possible to register some contact lens model parameters sets using the ‘Contact Lens<br />
Data Register function. You may register up to 10 contact lens model parameter sets.<br />
Page 48<br />
TOPCON Corporation
Figure 6-7 Contact Lens Data Register dialog box<br />
Page 49<br />
TOPCON Corporation
Chapter 7<br />
Printing Reports<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> allows printing of reports displayed on the screen using a Windows<br />
printer. There are 3 menu items related to the print operation in the [File] menu.<br />
• Print... Make a print of the active report window.<br />
• Print preview Preview print on the screen.<br />
• Print setup... Set up the Windows printer.<br />
7.1 Print Preview Screen<br />
To check how the report will be printed before making a print, choose [Print Preview]<br />
from the [File] menu.<br />
Figure 7-1 Print Preview Screen<br />
Page 50<br />
TOPCON Corporation
7.2 Print Setup Dialog Box<br />
The ‘Print Setup’ dialog box is used for changing the printer or paper to be used.<br />
Figure 7-2 Print Setup Dialog Box<br />
Page 51<br />
TOPCON Corporation
Chapter 8<br />
Other Features<br />
8.1 Hiding Tool Bar and Status Bar<br />
It is possible to hide the tool bar and status bar to make more space in the <strong>Color</strong> <strong>Mapping</strong><br />
<strong>32</strong> window. To hide the tool bar, choose “Toolbar” in the [View] menu. The check mark<br />
in front of “Toolbar” in the pull-down menu will be erased, and the tool bar will<br />
disappear. To hide the status bar, choose “Status Bar” in the [View] menu. The check<br />
mark at the head of “Status Bar” in the pull-down menu will be erased, and the status bar<br />
will disappear. To display these bars again, choose the same menu item, “Toolbar” or<br />
“Status Bar” in the [View] menu again. These menu items work as a toggle to hide and<br />
show the bars.<br />
8.2 Cascade and Tile<br />
Arrange report windows by choosing “Cascade” or ‘Tile” in the [Window] menu.<br />
“Cascade” arranges all report windows in an overlapping pattern. “Tile” arranges all<br />
report windows in a tile pattern.<br />
8.3 Select COM Port<br />
See Section 2.2.3 Port Setting.<br />
8.4 Reset COM Port<br />
If the communication is disrupted during the report data transfer from the KR-7000P/<br />
KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, choosing “Reset<br />
Port” in the [Port] menu can reset the communication port. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> reinitializes<br />
the communication port.<br />
Page 52<br />
TOPCON Corporation
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />
<strong>Software</strong> License Terms<br />
PLEASE READ THE FOLLOWING "LICENSE TERMS" CAREFULLY BEFORE USING (OR<br />
INSTALLING) THIS SOFTWARE!<br />
ANY AND ALL TERMS AND CONDITIONS OF THE "LICENCE TERMS" SHALL BE DEEMED TO<br />
BE ACCEPTED AND AGREED BY YOU IF YOU USE (OR INSTALL) THIS SOFTWARE.<br />
This “<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" <strong>Software</strong>, including its instruction manual and other documents, contained in this<br />
package comprises copyrightable works of TOPCON CORPORATION protected by copyright laws, and is<br />
licensed, not sold.<br />
TOPCON grants to you the right to use this “<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" <strong>Software</strong> under the terms and conditions<br />
set out below.<br />
LICENSE TERMS<br />
1. LICENSE<br />
Subject to the terms and conditions set out in this LICENSE TERMS, TOPCON CORPORATION<br />
("TOPCON") hereby grants to you (the "USER") a non-exclusive and non-transferable license (the<br />
"LICENSE") to use the "<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" including its instruction manual and other documents<br />
contained in this package (separatively or collectively the "SOFTWARE") only on a single computer<br />
which is a stand-alone or a terminal of a computer network (the “COMPUTER”). USER may not use the<br />
SOFTWARE on more than one computer or computer terminal at the same time.<br />
2. RESTRICTION<br />
2.1 USER shall NOT:<br />
(a) use the SOFTWARE on or with more than one computer or terminal at a time; or<br />
(b) make any copy of the SOFTWARE, except for making one copy of the SOFTWARE<br />
solely for back up purpose, or installing of the SOFTWARE into the COMPUTER,<br />
or its internal or external hard disc drive; or<br />
(c) sell, assign, exchange, pass, rent, lease or otherwise dispose of the SOFTWARE, or assign or sublicense<br />
any right to use of the SOFTWARE to anyone else; or<br />
(d) adapt, modify or alter the SOFTWARE, or merge any program with the SOFTWARE, or make any<br />
derivative from the SOFTWARE; or<br />
(e) remove, erase or make unclear any copyright notice of the SOFTWARE; or<br />
(f) decompile, disassemble, analyze, or make reverse-engineering to, the SOFTWARE; or<br />
(g) copy its instruction manual and other documents contained in this package.<br />
2.2 In any event that USER exchanges, or sells, assigns, passes, rents, leases or otherwise dispose of to<br />
any third party, or scraps the COMPUTER or any its internal or external hard disc drive or other<br />
memory means in which the SOFTWARE has been installed, USER shall destroy or erase any and all<br />
of the said installed SOFTWARE prior to the said exchange, etc.<br />
Page 53<br />
TOPCON Corporation
3. COPYRIGHT<br />
The SOFTWARE (including its recording Media) is the exclusive property of TOPCON, and full title,<br />
ownership and interests and copyrights and other intellectual property thereof shall remain vested with<br />
TOPCON at any and all times.<br />
4. TERM<br />
The LICENSE shall commence on the date when the SOFTWARE is first used (or installed) by USER, and<br />
shall remain in full force until the termination hereof pursuant to the provisions set out in Section 5.1<br />
below.<br />
5. TERMINATION<br />
5.1 The LICENSE shall be forthwith terminated if USER fails to comply with any term or condition of<br />
this LICENSE TERMS.<br />
5.2 Upon any termination of the LICENSE, USER shall forthwith cease using the SOFTWARE, and<br />
immediately destroy the original and any and all copies of the SOFTWARE (if any).<br />
6. LIMITED WARRANTY<br />
THE SOFTWARE IS PROVIDED ON AN "AS IS" BASIS WITHOUT WARRANTY OF ANY KIND.<br />
TOPCON FURTHER DISCLAIMS ALL WARRANTIES, INCLUDING, WITHOUT LIMITATION,<br />
ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR<br />
PURPOSE AND NONINFRINGEMENT.<br />
THE ENTIRE RISK ARISING OUT OF, RESULTING FROM OR IN CONNECTION WITH THE USE<br />
OR PERFORMANCE OF THE SOFTWARE REMAINS WITH THE USER. IN NO EVENT SHALL<br />
TOPCON OR ITS SUBSIDIARIES, AFFILIATES, DISTRIBUTORS OR AGENTS BE LIABLE FOR<br />
ANY CONSEQUENTIAL, INCIDENTAL, DIRECT, INDIRECT, SPECIAL, PUNITIVE OR OTHER<br />
DAMAGES WHATSOEVER (INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF<br />
BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS OF MEDICAL, CLINICAL OR BUSINESS<br />
INFORMATION, OR OTHER PECUNIARY LOSS, OR PERSONAL INJURY OR PROPERTY<br />
DAMAGE) ARISING OUT OF, RESULTING FROM OR INCONNECTION WITH THIS LICENSE<br />
TERMS OR THE USE OF OR INABILITY TO USE THE SOFTWARE.<br />
7. COMPLIANCE WITH EXPORT CONTROL LAWS<br />
Any and all exports or reexports of the SOFTWARE are subject to Japanese Exchange and Foreign Trade<br />
Law and other Japanese related laws and regulations, and/or the United States Export Administration Act,<br />
its implementing regulations as well as other U.S. export laws and regulations. Export or reexport of the<br />
SOFTWARE in a manner contrary to either Japanese or U.S. law or regulation is strictly prohibited.<br />
USER shall not export nor reexport to any country the SOFTWARE or any part or copy thereof without<br />
prior obtaining all required authorization from the Japanese or the United States government.<br />
TOPCON OR ITS SUBSIDIARIES, AFFILIATES, DISTRIBUTORS AND AGENTS SHALL NOT BE<br />
RESPONSIBLE NOR LIABLE TO USER FOR ANY LIABILITY, LOSS, DAMAGE, PUNISHMENT,<br />
OR PENALTY ARISING OUT OF OR RESULTING FROM USER ´ S VIOLATION OF OR<br />
NONCOMPLIANCE WITH ANY JAPANESE OR U.S. EXPORT CONTROL LAWS OR<br />
REGULATIONS.<br />
Page 54<br />
TOPCON Corporation
8. GOVERNING LAW<br />
The validity, performance, construction, interpretation, effect and enforcement of this LICENSE TERMS<br />
shall be governed by and under the laws of Japan.<br />
9. ENTIRE AGREEMENT<br />
This LICENSE TERMS constitutes the entire agreement between USER and TOPCON with respect to the<br />
subject matter hereof, and shall supersede and cancel any and all prior written or oral agreements,<br />
undertakings, negotiations, communications, commitments, representations, publications and advertisement,<br />
etc. not specifically and expressly incorporated herein.<br />
Page 55<br />
TOPCON Corporation
TOPCON Medical Systems, Inc.<br />
<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> <strong>Software</strong><br />
<strong>Instruction</strong> <strong>Manual</strong><br />
Version No. 1.24<br />
TOPCON MEDICAL SYSTEMS, INC.<br />
CORPORATE OFFICE: 37 West Century Road, Paramus, New Jersey 07652, U.S.A. Phone: 201-261-9450 Fax:201-387-2710 www.topcon.com<br />
TOPCON OMNI SYSTEMS, INC.<br />
Valley Forge Business Center, 2430 Blvd. of the Generals, Norristown, PA 19403, U.S.A. Phone:610-630-9200 Fax:610-630-6428 www.topconomni.com<br />
TOPCON CANADA INC.<br />
110 Provencher Avenue, Boisbriand, QC J7G 1N1 CANADA Phone:450-430-7771 Fax:450-430-6457 www.topcon.ca<br />
TOPCON EUROPE B.V.<br />
ESSE Baan 11, 2908 LJ Capelle a/d IJssel, THE NETHERLANDS Phone:010-4585077 Fax: 010-4585045 www.topconeurope.com<br />
ITALY OFFICE<br />
Via Montalcone 39, 20092 Cinisello B, mo(MI), ITALY Phone: 02-61-25-583 Fax: 02-61-25-927 www.topcon.it<br />
DENMARK OFFICE<br />
Ringstodvej 125, DK-4000 Roskilde, DENMARK Phone: 046-<strong>32</strong>75000 Fax: 046-<strong>32</strong>7555<br />
IRELAND OFFICE<br />
Unit 69 Western Parkway Business Centre, Lower Ballymount Road, Dublin 12, IRELAND Phone: 01460-0021 Fax: 01460-0129<br />
TOPCON S.A.R.L.<br />
89, Rue de Paris 92585 Clichy, Cedex, FRANCE Phone: 01-4106-9494 Fax: 01-4739-0251<br />
TOPCON DEUTSCHLAND G.m.b.H.<br />
Gieberallee 31-33, 47877 Willich, GERMANY Phone: 49-2154-8850 Fax: 49-2154-885111 E-mail: info@topcon.de www.topcon.de<br />
TOPCON SCANDINAVIA A.B.<br />
Neongatan 2 S-43151 Molndal, SWEDEN Phone: 31-7109200 Fax: 31-268607 www.topcon.se<br />
TOPCON ESPAÑA S.A.<br />
Frederic Mompou 5, ED. Euro 3, 08960, Sant Just Desvern Barcelona, SPAIN Phone: 34-93-473-4057 Fax: 34-93-473-39<strong>32</strong> www.topcon.es<br />
TOPCON ( GREAT BRITAIN ) LTD.<br />
Topcon House, Kennet Side, Bone Lane, Newbury, Berkshire RG14 5PX UNITED KINGDOM Phone: 01635-551120 Fax: 01635-551170 www.topcon.co.uk<br />
TOPCON SOUTH ASIA PTE. LTD.<br />
Blk 192 Pandan Loop, #07-01, Pantech Industrial Complex, SINGAPORE 128381 Phone: 62780222 Fax: 62733540 www.topcon.com.sg<br />
TOPCON AUSTRALIA PTY. LTD.<br />
Unit 18, 4 Avenue of Americas, Newington, NSW 2127 AUSTRALIA Phone: 61-2-8748-8777 Fax: 61-2-9647-2926 www.topcon.com.au<br />
TOPCON INSTRUMENTS ( MALAYSIA ) SDN. BHD.<br />
Excella Business Park Block C, Ground & 1 st Floor, Jalan Ampang Putra, Taman Ampang Hilir, 55100 Kuala Lumpur, MALAYSIA Phone:03-42701068 Fax:03-4274508<br />
TOPCON INSTRUMENTS ( THAILAND ) CO., LTD.<br />
77/162 Sinn Sathorn Tower, 37th Fl., Krungdhonburi Rd., Klongtonsai, Klongsarn, Bangkok 10600, THAILAND Phone: 02-440-1152~7 Fax: 02-440-1158<br />
TOPCON KOREA CORPORATION<br />
2F Yooseoung Bldg., 1595-3, Seocho-Dong, Seocho-Gu, Seoul, 137-876, KOREA Phone: 02-2055-0<strong>32</strong>1-7 Fax: 02-2055-0319 www.topcon.co.kr<br />
TOPCON OPTICAL ( H.K. ) LTD.<br />
2/F, Meeco Industrial Bldg., No.53-55 Au Pui Wan Street, Fo Tan Road, Shatin, N.T., HONG KONG Phone: 852-2690-1<strong>32</strong>8 Fax: 852-2690-2221 www.topcon.com.hk<br />
TOPCON CORPORATION BEIJING OFFICE<br />
Room 8A Poly Plaza Building, 14 Dongzhimen Nandajie, Dongcheng District, Beijing, 100027, CHINA Phone:10-6501-4191~2 Fax:10-6501-4190<br />
TOPCON CORPORATION SHANGHAI OFFICE<br />
Room 901~903 Lippo Plaza No. 222, Huai Hai Zhong Lu, Shanghai, 20021 CHINA Phone: 21-62701877 Fax: 21-53966485<br />
TOPCON CORPORATION BEIRUT OFFICE<br />
P.O.Box 70-1002 Antelias, BEIRUT-LEBANON Phone: 961-4-523525/523526 Fax: 961-4-521119<br />
TOPCON CORPORATION DUBAI OFFICE<br />
c/o Atlas Medical FZCO., P.O. Box 54304, C-25, Dubai Airport Free Zone, UNITED ARAB EMIRATES Phone: 971-4-2995900 Fax: 971-4-2995901<br />
TOPCON CORPORATION<br />
75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580 JAPAN.<br />
Phone:3-3558-2520 Fax:3-3960-4214 www.topcon.co.jp<br />
Page 56<br />
© 2004 TOPCON CORPORATION All Rights Reserved<br />