<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> <strong>Software</strong><br />

<strong>Instruction</strong> <strong>Manual</strong><br />

Version No. 1.24<br />

Page 0<br />

TOPCON Medical Systems, Inc.<br />

TOPCON Corporation


Introduction<br />

Thank you very much for purchasing the TOPCON <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>.<br />

Please read this <strong>Instruction</strong> <strong>Manual</strong> with care before using this <strong>Software</strong>.<br />

A good understanding and working knowledge of this software will guarantee<br />

your satisfaction for many years to come.<br />

Keep this <strong>Manual</strong> at hand for further reference.<br />

Precautions<br />

• TOPCON has the copyright of the <strong>Software</strong>, "<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>".<br />

• Reproduction and copying of the software and the <strong>Instruction</strong> manual, in<br />

whole or in part, without our permission is strictly prohibited.<br />

• As a rule, one set of the software must be purchased for each PC installed.<br />

• For the handling procedures of KR-7000P/ KR-7100P/ KR-8000PA/ KR-<br />

8100PA/ KR-8100P, refer to its instruction manual.<br />

• This <strong>Instruction</strong> <strong>Manual</strong> does not describe details about the handling<br />

procedures of PC and Microsoft Windows. This <strong>Instruction</strong> <strong>Manual</strong> is<br />

made on an assumption that the user already has knowledge about the PC<br />

and Microsoft Windows.<br />

• For the handling procedures of PC and Microsoft Windows, refer to the<br />

pertaining manuals.<br />

• Incorrect handling or operation of the PC may damage it or its software.<br />

• We shall not take any responsibility for the effects of or as a result of<br />

using the <strong>Instruction</strong> <strong>Manual</strong> and <strong>Software</strong>.<br />

• Specifications of the <strong>Software</strong> and descriptions of the <strong>Instruction</strong> <strong>Manual</strong><br />

are subject to change without prior notice.<br />

KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, and<br />

CV Power Supply Unit are trademarks of the TOPCON Corporation.<br />

Microsoft, MS, and MS-DOS are registered trademarks, and Windows is a trademark of<br />

Microsoft Corporation.<br />

PC-DOS is a registered trademark of International Business Machines Corporation<br />

Page 1<br />

TOPCON Corporation


NOTATIONS<br />

The following notations are used in the <strong>Instruction</strong> <strong>Manual</strong>.<br />

MENU, COMMAND AND DIALOG BOX<br />

Expression Description<br />

[File]menu Menu name is expressed in [ ].<br />

[Open] of [File] menu Command name is expressed in [ ] with menu name in which it<br />

is included.<br />

[Open File]dialog box Dialog box name displayed on title bar is expressed in [ ].<br />

[OK] Command button name is expressed in [ ].<br />

MOUSE OPERATION<br />

Expression Description<br />

Point Move mouse and bring mouse pointer onto object of operation.<br />

Click Press and release mouse button.<br />

Double-click Click mouse button twice.<br />

Drag Press and move mouse button and release it at final position.<br />

KEY OPERATION<br />

Expression Description<br />

Enter Key name is expressed with bold type.<br />

Shift+F1 Plus symbol (+) means to press multiple keys; for example,<br />

Shift+F1 means to press Shift key and F1 key.<br />

Direction key Generic name of →, ←, ↑ and ↓ keys.<br />

OTHER EXPRESSIONS<br />

Expression Description<br />

Memo Gives hints for operation.<br />

Page 2<br />

TOPCON Corporation


SAFETY PRECATIONS<br />

Items Prevention items Page<br />

Note that the necessary memory capacity and hard disk capacity<br />

may differ for each PC.<br />

Contact Lens Simulation<br />

7<br />

Physician can not use report data for lens prescription without<br />

43<br />

their personal validation.<br />

Contents of the package<br />

The following are standard accessories. Make sure that all these items are attached (quantity).<br />

CD-ROM (1)<br />

<strong>Instruction</strong> <strong>Manual</strong> (1)<br />

CAUTION<br />

Page 3<br />

RS2<strong>32</strong>C Cable (1)<br />

(D-sub9P-DIN8P)<br />

TOPCON Corporation


Table of Contents<br />

Chapter 1 Overview...........................................................................6<br />

Chapter 2 Setup <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> System .....................................7<br />

2.1 <strong>Software</strong> Requirements ....................................................7<br />

2.2 Installing <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> ..............................................8<br />

2.3 Communication Setting...................................................14<br />

2.4 Preparation for KR-7000P...............................................16<br />

2.5 Preparation for KR-8000P...............................................19<br />

Chapter 3 Getting Started................................................................23<br />

3.1 Starting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> ..............................................23<br />

3.2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Components ......................................23<br />

3.3 Operating <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>...........................................26<br />

3.4 Exiting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>................................................26<br />

Chapter 4 <strong>Color</strong> Map Report Data...................................................28<br />

4.1 Receiving <strong>Color</strong> Map Report Data ..................................28<br />

4.2 Display <strong>Color</strong> Map Report Data ......................................29<br />

4.3 Patient Registration ........................................................29<br />

4.4 Saving Report Data ........................................................31<br />

4.5 <strong>Color</strong> Map Report Window Control .................................<strong>32</strong><br />

4.6 Opening Saved Report Data...........................................33<br />

4.7 Viewing Report Data from the Database ........................34<br />

Chapter 5 Option Settings ...............................................................37<br />

5.1 Map Method....................................................................37<br />

5.2 Scale Unit .......................................................................37<br />

5.3 Map Type........................................................................38<br />

5.4 Overlays .........................................................................38<br />

5.5 Peripheral .......................................................................39<br />

5.6 Others.............................................................................39<br />

5.7 Image View.....................................................................40<br />

Chapter 6 Contact Lens Simulation Report Data.............................42<br />

6.1 Display Contact Lens Simulation Report Data ................42<br />

6.2 Parameters of the Contact Lens Simulation Report........45<br />

6.3 Handling Each Lens Type...............................................45<br />

6.4 Parameter Setting...........................................................46<br />

6.5 Contact Lens Utility.........................................................47<br />

Page 4<br />

TOPCON Corporation


Chapter 7 Printing Reports ..............................................................50<br />

7.1 Print Preview Screen ......................................................50<br />

7.2 Print Setup Dialog Box....................................................51<br />

Chapter 8 Other Features................................................................52<br />

8.1 Hiding Tool Bar and Status Bar ......................................52<br />

8.2 Cascade and Tile............................................................52<br />

8.3 Select COM Port.............................................................52<br />

8.4 Reset COM Port .............................................................52<br />

<strong>Software</strong> License Terms .......................................................................54<br />

Page 5<br />

TOPCON Corporation


Chapter 1<br />

1.1 Overview<br />

Topcon’s <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is a <strong>32</strong>bit multiple-document report Windows application<br />

running under Microsoft Windows 95/98/NT and Microsoft Windows XP.<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> works with the tri-function Topcon KR-7000P/ KR-7100P/ KR-<br />

8000PA/ KR-8100PA/ KR-8100P, auto-refractometer, expanded auto-keratometer and<br />

placido disk-based corneal mapping system. The data captured by the KR-7000P/ KR-<br />

7100P/ KR-8000PA/ KR-8100PA/ KR-8100P is transferred to <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> via an<br />

RS-2<strong>32</strong>C interface, and is then reconstructed into corneal topography color map data.<br />

The multiple documents include refractometer data, keratometer data, and a corneal<br />

topography color map. To send refractometer and/or keratometer data to the TOPCON<br />

CV Power Supply Unit, the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> creates the necessary data for the<br />

TOPCON CV Power Supply Unit and sends it using the second RS-2<strong>32</strong>C port of the PC.<br />

After reconstructing the corneal topography color map data, the multiple document report<br />

may be viewed on the Windows screen or printed on a color printer.<br />

The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> provides various options for evaluating cornea, such as tangential<br />

/ sagittal map, adjustable / normalized / absolute map, diopter / mm display, overlay<br />

information display, auto-print, contact lens fitting simulation, etc. Default option<br />

parameters may also be set.<br />

An important feature of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is that it creates and supports a patient<br />

information database. The database services integrated in the software speeds-up patient<br />

data storage/ retrieval and simplifies data analysis.<br />

If the auto-print option is enabled, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> prints a multiple-document report<br />

to a color printer automatically using default option parameters when new data is<br />

received from the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P.<br />

Page 6<br />

TOPCON Corporation


Chapter 2<br />

Setup <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> System<br />

2.1 System Requirements<br />

2.1.1 Hardware Requirements<br />

The following hardware is necessary to run <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>:<br />

· 100% IBM PC compatible computer with i-486DX or better<br />

· Minimum 8MB of system memory<br />

· Minimum 1MB of video memory<br />

· 2MB of free disk space at installation<br />

· A color printer supported by Windows<br />

· An available COM port*<br />

· A KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to PC interface<br />

cable**<br />

* To send the refractometer and/or keratometer data to the TOPCON CV Power Supply<br />

Unit, an additional COM port is needed.<br />

** To send the refractometer and/or keratometer data to the TOPCON CV Power Supply<br />

Unit, the PC to CV Power Supply Unit interface cable is needed.<br />

In the above specifications, the video memory and the system memory may be the most<br />

important hardware for <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>. For a computer with 1MB of video memory,<br />

the video mode can be set to 1024x768 256 colors and 640x480 16-million colors.<br />

The faster the computer CPU, the better the performance of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>. In<br />

general, performance of Windows is influenced most by the amount of system memory<br />

and the speed of the hard disk. If the system appears to be running slow, try adding<br />

system memory first. If it still seems slow, upgrade the computer with a faster CPU and<br />

much larger hard disk.<br />

Note that the necessary memory capacity and hard disk capacity<br />

may differ for each PC.<br />

2.1.2 <strong>Software</strong> Requirement<br />

To run <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, the following software is necessary:<br />

· Microsoft Windows 95/98/NT or Microsoft Windows XP.<br />

· The video mode of Windows should be set to at least 640x480 x 256 colors.<br />

Page 7<br />

TOPCON Corporation


2.2 Installing <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

2.2.1 Running Setup Program<br />

If the previous version of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> was installed, uninstalling is always<br />

recommended. See section 2.2.2 for Uninstalling.<br />

Follow the instructions bellow to install <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> for PC with WINDOWS 95:<br />

Quit any applications that are currently running.<br />

If you use the Microsoft Plus! System Agent, right-click the System Agent icon on the<br />

toolbar, and then click Suspend System Agent.<br />

• Insert the floppy disk into your computer’s drive;<br />

1. Click the [Start] button of task bar and click [Run] menu.<br />

2. The [Run] dialog box is displayed. Enter [A:\Setup.exe], insert the floppy disk #1/1<br />

and click the [OK] button.<br />

Page 8<br />

TOPCON Corporation


3. Once the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> installation program is started, it displays the following<br />

‘Welcome’ dialog box:<br />

Click the ‘Next’ button to continue the installation<br />

4. In this dialog box, specify the destination where <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is to be installed.<br />

It is not recommended to change the default destination directory. Click the ‘Next’<br />

button to continue.<br />

Page 9<br />

TOPCON Corporation


5. Specify Disk Name, you will use for the Patient Information Database.<br />

6. Select directory for storage of the color map image files that Patient Database<br />

Records refer to.<br />

Page 10<br />

TOPCON Corporation


7. In this dialog box, click the ‘Finish’ button to finish the installation.<br />

When the Installation is done click on the shortcut icon to start the <strong>Color</strong> <strong>Mapping</strong><br />

<strong>32</strong>.The shortcut icon must be created by the Installation program on your desktop screen<br />

when the installation is completed(see section 3-1)<br />

2.2.2 Removing The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

1. Click the [Start] button, point to [Settings], and then click [Control Panel].<br />

Page 11<br />

TOPCON Corporation


2. Double-click [Add/Remove Programs].<br />

Page 12<br />

TOPCON Corporation


3. Click the [Install/Uninstall] tab, click [KRP<strong>32</strong>], and then click [Add/Remove].<br />

4. The following dialog box will appear in order to make sure you are uninstalling the<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. Click ‘Yes’ button to start the uninstalling.<br />

Page 13<br />

TOPCON Corporation


5. In this dialog box, click the ‘OK’ button to finish the uninstalling.<br />

2.3. Communication Setting<br />

2.3.1. Port Setting<br />

If it is necessary to select the communication port for the KR-7000P/ KR-7100P/ KR-<br />

8000PA/ KR-8100PA/ KR-8100P, choose ‘Select Com Port for KR’ in the [Port] menu to<br />

get the Select COM Port dialog box. (Figure 2-1) To change the COM port number,<br />

select COM# (# is number of communication port) and click the ‘OK’ switch. The<br />

selected communication port will be initialized automatically. If None was selected,<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> displays only saved data.<br />

Figure 2-1 KR port selection Dialog Box<br />

Page 14<br />

TOPCON Corporation


If it is necessary to select the communication port for the CV Power Supply Unit, choose<br />

‘Select Com Port for CV’ in the [Port] menu to get the Select COM Port dialog box.<br />

(Figure 2-2) To change the COM port number, select COM# and click the “OK” switch.<br />

Figure 2-2 CV port selection Dialog Box<br />

If None was selected, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> does not send data to the TOPCON CV Power<br />

Supply Unit. Information about the old and new formats can be found in the TOPCON<br />

CV Power Supply Unit documentation.<br />

2.3.2 Communication Format Instrument Setting<br />

<strong>Software</strong> supports data communication with two types of instruments:<br />

• KR7000P(Include KR7100P)<br />

• KR8000P/19K(Include KR-8100PA/ KR-8100P with 19200Bps)<br />

• KR8000P/38K(Include KR-8100PA/ KR-8100P with 38400Bps)<br />

To receive measurements the type of connected instrument must be selected:<br />

1. Select pull-down Port Menu by clicking on the Port button on the Menu Bar.<br />

2. Set the instrument type by selecting appropriate option KR7000P/KR8000P from the<br />

[Select Instrument] pull-down menu.<br />

Note: The previous setting is indicated with the check mark next to the instrument type<br />

in the [Select Instrument] pull-down menu.<br />

2.3.3 Cable Connection<br />

For a KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to PC interface<br />

cable: (A round connector 8-PIN DIN on one end and a 9-PIN D-SUB connector on the<br />

other end.)<br />

To connect the interface cable between the KR-7000P/ KR-7100P/ KR-8000P/ KR-<br />

8100PA/ KR-8100P and the PC:<br />

1. Turn off the power switch of the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />

KR-8100P and the PC.<br />

2. Align the round cable connector (8-PIN DIN) with the KR-7000P/ KR-7100P/ KR-<br />

8000PA/ KR-8100PA/ KR-8100P’s OUT port of “DATA TERMINAL” and press the<br />

round cable connector into the port.<br />

3. Connect the other connector (9-PIN DSUB) to the PC COM port.<br />

Page 15<br />

TOPCON Corporation


2.4 Preparation for KR-7000P<br />

2.4.1 Setting Up KR-7000P<br />

Make sure the following settings of the KR-7000P/ KR-7100P are correctly configured:<br />

RS2<strong>32</strong>C FORM EXT<br />

BAUD-RATE 9600<br />

To change settings of the KR-7000P/ KR-7100P:<br />

1. Turn on the power switch of the KR-7000P/ KR-7100P while pressing the<br />

“CONTACT / GLASSES” switch. The initial setting screen appears on the screen.<br />

2. Release the “CONTACT / GLASSES” switch<br />

3. “INITIAL SETTING” screen appears on the screen (Figure 2-3).<br />

INITIAL SET<br />

NUMBER SET<br />

PRINT OUT<br />

ONLINE<br />

PLACIDO SET<br />

Figure 2-3 Initial Setting Screen<br />

4. Bring the cursor to “ONLINE” by pressing the measuring switch.<br />

5. Press the PRINT switch. The on-line screen appears on the screen.<br />

CL INPUT NO<br />

RS-2<strong>32</strong>C FORM EXT<br />

BAUD-RATE 9600<br />

RS2<strong>32</strong>C OUT TYPE REF<br />

RS2<strong>32</strong>C KRT CENT<br />

Figure 2-4 Online Setting Screen<br />

6. Bring the cursor to “RS-2<strong>32</strong>C FORM” by pressing the measuring switch.<br />

7. Bring the cursor to “RS-2<strong>32</strong>C FORM” to “EXT” by pressing the PRINT switch. The<br />

“BAUD-RATE will be set to 9600 automatically. Also change the “RS2<strong>32</strong>C OUT<br />

TYPE” to “ALL” and the “RS2<strong>32</strong>C KRT” to “BOTH”.<br />

8. Bring the cursor to “NEXT by pressing the measuring switch.<br />

9. Press the PRINT switch.<br />

10. Bring the cursor to “END” by pressing the measuring switch.<br />

11. Press the PRINT switch. “END OK?” appears on the screen.<br />

12. Press the PRINT switch. “SETTING MENU DATA” will be printed out, and the<br />

measurement screen will appear.<br />

Page 16<br />

TOPCON Corporation


SETTING MENU DATA<br />

VD 12.00<br />

STEP 0.25<br />

…………………………………<br />

………… ……<br />

RS2<strong>32</strong>C FORM EXT<br />

BAUD-RATE 9600<br />

…… ……<br />

…… ……<br />

LIGHTING RING<br />

Figure 2-5 Printout Setting Data<br />

13. Please make sure settings of the KR-7000P/ KR-7100P are correctly configured:<br />

2.4.2 <strong>Cal</strong>ibration Data Registration<br />

Before transferring the calibration data of the KR-7000P/ KR-7100P to the PC, proceed<br />

with following steps:<br />

3.2 Start the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. When starting the software, the message,<br />

“<strong>Cal</strong>ibration Data does not exist. Please setup <strong>Cal</strong>ibration data” may appear until the<br />

calibration data has been transferred from the KR-7000P/ KR-7100P (Figure 2-6).<br />

Figure 2-6 No calibration data message box<br />

Please click the ‘OK’ button to prepare to receive data from the KR-7000P/ KR-<br />

7100P.<br />

3.2 Set the instrument's type by clicking on following controls from the main menu<br />

bar:<br />

• Select [Port]-[Select Instrument] from Menu Bar.<br />

• Select KR7000P from [Select Instrument].<br />

If the KR-7000P/ KR-7100P unit is replaced, it is necessary to transfer the new unit’s<br />

calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the registered calibration data.<br />

In this case, the message in Figure 2-6 may not appear, because the old calibration data is<br />

already registered.<br />

Page 17<br />

TOPCON Corporation


2.4.3 <strong>Cal</strong>ibration Data Transfer Procedure<br />

The following steps are required for the KR-7000P/ KR-7100P to transfer the calibration<br />

data:<br />

1. Dip switch 7 & 8 at position ON and 1 through 6 at position OFF. Rotary switch to<br />

position 7.<br />

Figure 2-7 Switches Condition (<strong>Cal</strong>ibration Data Registration)<br />

2. Turn on the KR-7000P/ KR-7100P, (a “PLACIDO CHECK” message will appear).<br />

3. Press the joystick once to enter the submenu.<br />

4. Press the PRINT switch and move the cursor to “5:CALIB DATA OUT”.<br />

5. Press the joystick. The KR-7000P/ KR-7100P will send the calibration data to the<br />

PC.<br />

6. Turn off the KR-7000P/ KR-7100P.<br />

2.4.4 KR-7000P Measurement mode Setting<br />

In order to transfer measurement data to the PC, please make sure Dip switches are in the<br />

following positions:<br />

Dip switch 7 position OFF<br />

Dip switch 8 position ON<br />

Rotary switch position ‘0’<br />

Page 18<br />

TOPCON Corporation


Turn on the KR-7000P/ KR-7100P.<br />

Figure 2-8 Switches Condition (Measurement Mode)<br />

After measuring the patient’s eyes, press the GRAPHIC PRINT SWITCH on the KR-<br />

7000P/ KR-7100P to transfer the necessary data to the PC.<br />

Memo:<br />

1. If “MAPPING PRT” on the KR-7000P/ KR-7100P is set for PRINT, you may also<br />

use the PRINT switch to send data.<br />

2. If a printout from the KR-7000P/ KR-7100P is not necessary, lift the paper retainer<br />

lever.<br />

3. Usually, registering calibration data needs to be done only once. However, when the<br />

KR-7000P/ KR-7100P unit is replaced or returned from repair, it is necessary to<br />

transfer the new calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the<br />

registered calibration data.<br />

4. When a KR-7000P/ KR-7100P keeps displaying the “WAITING DIPSW8” message<br />

for a long time, please make sure the Dip switch 8 is ON.<br />

5. When a KR-7000P/ KR-7100P keeps displaying the “CALIBRATING” message for<br />

a long time, please check the interface cable connection.<br />

6. When a <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> displays error messages such as “CEbreak”, “CEerror”,<br />

etc, please check the KR-7000P/ KR-7100P settings are “RS2<strong>32</strong>C FORM EXT”,<br />

“BAUD RATE 9600”.<br />

2.5 Preparation for KR-8000P<br />

2.5.1 Setting up KR-8000P<br />

The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />

8100P types of the instruments.<br />

While pressing MENU SWITCH on the control panel of the KR-8000PA/ KR-8100PA/<br />

KR-8100P, turn on the KR-8000PA, "INITIAL MENU” screen appears.<br />

Page 19<br />

TOPCON Corporation


1. Select “ON-LINE" option from the “INITIAL MENU” screen to bring up the “ON-<br />

LINE” screen.<br />

Figure 2-9 “ON-LINE” Screen<br />

2. Press the joystick and bring the cursor to “RS2<strong>32</strong>C FORM”.<br />

3. Press the PRINT switch to highlight EXT (available for later version of <strong>Color</strong><br />

<strong>Mapping</strong> <strong>32</strong> Version 1.24).<br />

4. Press the joystick and bring the cursor to “BAUD-RATE”.<br />

5. Press the PRINT switch and highlight 38400 or 19200. (For saving transfer time,<br />

select 38400Bps).<br />

6. Press the PRINT switch and bring the cursor to “EXIT”.<br />

7. Press the PRINT switch to exit "ON-LINE" screen.<br />

2.5.2 <strong>Cal</strong>ibration Data Registration<br />

Before transferring the calibration data of the KR-8000PA/ KR-8100PA/ KR-8100P to<br />

the PC, proceed with following steps:<br />

1. Start the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> program. When starting the software, the message,<br />

“<strong>Cal</strong>ibration data does not exist. Please setup <strong>Cal</strong>ibration data” may appear until the<br />

calibration data has been transferred from the KR-8000PA/ KR-8100PA/ KR-8100P<br />

(Figure 2-6). Please click the “OK” button to prepare to receive data from the KR-<br />

8000PA/ KR-8100PA/ KR-8100P.<br />

2. Set the instrument's type by clicking on following controls from the main menu bar:<br />

• Select [Port]-[Select Instrument] from Menu Bar.<br />

• Select KR8000P/38K or /19K from [Select Instrument].<br />

Selection must correspond to the instrument’s “BAUD-RATE” setting, please refer to<br />

the Section 2.5.1 (Step 4)<br />

If the KR-8000PA/ KR-8100PA/ KR-8100P unit is replaced, it is necessary to transfer<br />

the new unit’s calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to update the registered<br />

calibration data.<br />

2.5.3 KR-8000P Setting and <strong>Cal</strong>ibration Data Transfer<br />

The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />

8100P types of the instruments.<br />

1. Turn on KR-8000PA/ 8100PA/ 8100P "PLACIDO SET" screen from "INITIAL<br />

MENU".<br />

Page 20<br />

TOPCON Corporation


Figure 2-10 PLACIDO SET Screen<br />

2. Press the joystick and bring the cursor to “PC S/W SETUP”.<br />

3. Press the PRINT switch. <strong>Cal</strong>ibration data is transferred automatically and setup of<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is done.<br />

4. Press the joystick and bring the cursor to “EXIT”.<br />

5. Press the PRINT switch. “EXIT OK?” appears on the screen. Press the PRINT<br />

switch. “SETTING MENU DATA” will be printed out, and the measurement<br />

screen will appear<br />

SETTING MENU DATA<br />

VD 12.00<br />

STEP 0.25<br />

…………………………………<br />

………… ……<br />

RS2<strong>32</strong>C FORM EXT2<br />

BAUD-RATE 38400<br />

…… ……<br />

…… ……<br />

LIGHTING RING<br />

Figure 2-11 SETTING MENU DATA Screen<br />

6. Please make sure settings of the KR-8000PA are correctly configured.<br />

2.5.4 KR-8000P Measurement Mode Setting<br />

The following setting procedure is applicable for the KR-8000PA/ KR-8100PA/ KR-<br />

8100P types of the instruments.<br />

In order to transfer measurement data to the PC, configure the instrument RS2<strong>32</strong>C output<br />

port by performing the following steps:<br />

1. Press “MENU SWITCH” of the control panel on measurement mode.<br />

"SETTING MENU" screen appears.<br />

2. Press the joystick and bring the cursor to “RS2<strong>32</strong>C”.<br />

3. Press the PRINT switch to highlight “YES”.<br />

Page 21<br />

TOPCON<br />

TOPCON Corporation


Figure 2-12 SETTING MENU Screen<br />

Press the joystick and bring the cursor to “EXIT”. Press the PRINT switch to exit<br />

"SETTING MENU" screen<br />

After measuring the patient’s eyes, press the GRAPHIC PRINT SWITCH on the control<br />

panel of KR-8000PA/ KR-8100PA/ KR-8100P to transfer the necessary data to the PC.<br />

Memo:<br />

1. If “MAPPING PRT” on the KR-8000PA/ KR-8100PA/ KR-8100P is set to<br />

“PRINT”, you may also use the PRINT switch to send data.<br />

2. If a printout from the KR-8000PA/ KR-8100PA/ KR-8100P is not necessary, lift the<br />

paper retainer lever.<br />

3.2 Usually, registering calibration data needs to be done only once. However, when<br />

the KR-8000PA/ KR-8100PA/ KR-8100P unit is replaced or returned from repair, it<br />

is necessary to transfer the new calibration data to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> in order to<br />

update the registered calibration data.<br />

3.2 If no data is transferred from a KR-8000PA/ KR-8100PA/ KR-8100P, Please<br />

check the setting of port. “RS2<strong>32</strong>C” is “YES” and “RS2<strong>32</strong>C FORM EXT”, “BAUD<br />

RATE 19200 or 38400”.<br />

3.2 When a KR-8000PA/ KR-8100PA/ KR-8100P keeps displaying the “CABLE<br />

UNCONNECTED” message, please check the interface cable connection and port<br />

setting.<br />

Page 22<br />

TOPCON Corporation


Chapter 3<br />

Getting Started<br />

3.1 Starting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

To start <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> click on the shortcut icon labeled KRP<strong>32</strong><br />

(Figure 3-1) in your Desktop Window.<br />

KRP<strong>32</strong><br />

Figure 3-1 Shortcut Icon<br />

3.2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Components<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Window<br />

The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window consists of several elements such as the Title Bar, the<br />

Menu Bar, the Toolbar and the Status Bar. The Title Bar is located at the top of the<br />

window. The Menu Bar is just below the Title Bar, and the Toolbar below that. The<br />

Status Bar is located at the bottom of the window.<br />

Figure 3-2 <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Window<br />

Page 23<br />

TOPCON Corporation


<strong>Color</strong> maps are displayed in windows referred to here as “Report Windows”. A report<br />

window is created when a <strong>Color</strong> Map file is opened and is removed when the file is<br />

closed. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window is maximally sized on start up to make room for<br />

report windows to be as large as possible. It is possible to make the window smaller,<br />

however, the report windows can never be larger than the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window.<br />

Title Bar The title bar is used to show the report currently displayed.<br />

Menu Bar The menu bar consists of several items. By clicking these items,<br />

the pull-down menu related to the item will be opened. By<br />

choosing an item in a pull-down menu, you can call the function<br />

assigned to the item.<br />

[File] Open...<br />

Close*<br />

Close All*<br />

Save*<br />

Save As*<br />

Open from Database<br />

Save to Database*<br />

Print...*<br />

Print Preview*<br />

Print Setup...<br />

Exit<br />

[Patient] Select Patient<br />

Register Patient<br />

Deleted Patient<br />

* Item in the [Patient] menu is enabled only if the<br />

Patient Registration Database is not empty<br />

[View] Toolbar<br />

Status Bar<br />

[Port] Select Com Port for KR<br />

Select Com Port for CV<br />

Reset Port<br />

Select Instrument***<br />

KR7000P<br />

KR8000P/19K<br />

KR8000P/38K<br />

[Utility] Map Options*<br />

<strong>Cal</strong>l Map*<br />

<strong>Color</strong> Map*<br />

Contact Lens Fitting [OD(Right)]*<br />

Contact Lens Fitting [OS(Left)]*<br />

Contact Lens Utility<br />

Delete Data from Database<br />

Date Format Settings<br />

Change Parameters**<br />

Page 24<br />

TOPCON Corporation


Note:<br />

Pan & Tilt**<br />

Profile Graph**<br />

Isometric Graph**<br />

[Window]* Cascade<br />

Tile<br />

[Help] About <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>...<br />

* Items in the Menu Bar or in the each pull-down menu are available only,<br />

when the <strong>Color</strong> Map Report is displayed.<br />

** Items in the Menu Bar or in the each pull-down menu are available only,<br />

when the Contact Lens Fitting Report is displayed.<br />

*** Items in the Menu Bar or in the pull-down menu are available only, when<br />

there is no document (report) in the main frame window.<br />

Tool Bar The tool bar provides quick access to items in pull-down menus of<br />

the menu bar. By clicking an icon on the tool bar, the function<br />

assigned to the icon is directly called.<br />

Note:<br />

Open Icon Open [File]<br />

Save Icon Save [File]<br />

Save to DB Save to the Database [Utility]<br />

Open from DB Retrieve from the Database [Utility]<br />

Map Options Map Options Settings * [Utility]<br />

<strong>Color</strong> Map/Contact Lens <strong>Color</strong> /Simulation Maps* [Utility]<br />

Pan & Tilt Icon Contact Lens Fitting** [Utility]<br />

Printer Icon Print [File]<br />

Info Icon About <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> [Help]<br />

Page 25<br />

TOPCON Corporation


* Items in the Menu Bar or in the each pull-down menu are available only,<br />

when the <strong>Color</strong> Map Report is displayed.<br />

** Items in the Menu Bar or in the each pull-down menu are available only,<br />

when the Contact Lens Fitting Report is displayed.<br />

Status Bar The status bar shows the status of <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>.<br />

Report Window When a report is displayed on the screen, a new window for the<br />

report is created in the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> window. Although two<br />

or more report windows may be open simultaneously, only one<br />

may be active. Most functions work on the report in the active<br />

report window. Each report window has the title on the top to<br />

show identification information of the report.<br />

3.3 Operating <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

Once <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> is started, the communication port will be initialized<br />

automatically and ready to receive transfer data from the KR-7000P/ KR-7100P/KR-<br />

8000PA/ KR-8100PA/ KR-8100P. When the data is received, the corneal topography<br />

mapping data is reconstructed, and a new report window is created automatically. Use<br />

the menu icons on the tool bar as a short cut for some menu items. A menu item or a<br />

menu icon displayed in gray means that the item or the icon is not available at that time.<br />

If the port for the TOPCON CV Power Supply Unit has been set, the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

creates the necessary data using the specified data format and sends it to the TOPCON<br />

CV Power Supply Unit.<br />

To display reports, choose [Open] from the [File] menu. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display<br />

a dialog box to get necessary information. Type or select items in the dialog box, then<br />

click “OK” to continue. To apply any options to the displayed report, choose [Option<br />

Setting] from the [Utility] menu. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display the option setting<br />

dialog box to change option parameters. Change parameters in the dialog box, then click<br />

“OK” to continue. The changed parameters will be immediately applied to the report.<br />

To get contact lens simulation report window, choose “Contact Lens Fitting” from the<br />

[Utility] menu. The <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display the Contact Lens Parameter setting<br />

dialog box to set for perform the contact lens fitting simulation.<br />

To print a report, choose [Print] from the [File] menu. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will send the<br />

report information to the printer. If the “Auto Print” check box in the option setting<br />

dialog box has been checked, the color map report information will be sent automatically<br />

as soon as the program reconstructs the new report using the data transferred from the<br />

KR-7000P/ KR-7100P/KR-8000PA/ KR-8100PA/ KR-8100P.<br />

3.4 Exiting <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

To exit <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, choose [Exit] from the [File] menu. Upon exit, the program<br />

closes all report windows and releases resources. Before exiting, save any important new<br />

Page 26<br />

TOPCON Corporation


eports that were created during the session, or they will be lost. The program does not<br />

prompt to save new reports when exiting.<br />

Page 27<br />

TOPCON Corporation


Chapter 4<br />

<strong>Color</strong> Map Report Data<br />

4.1 Receiving <strong>Color</strong> Map Report Data<br />

To receive measured data from the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />

KR-8100P, press the GRAPHIC PRINT SWITCH (open the cover/door of the control panel<br />

to use this switch) on the instrument. After receiving data, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

reconstructs the corneal structure data.<br />

4.1.1 KR- 8000P Data Transfer Setting<br />

To transfer measurement data from the KR-8000PA/KR-81000PA/KR-8100P proceed<br />

with flowing steps.<br />

1. Press the GRAPHIC PRINT SWITCH, “IMAGE DATAOUT MENU” screen will<br />

appear (Figure 4-1).<br />

2. Press the joystick to highlight one of the menu items (Figure 4-1).<br />

Each item indicates type of transferring image that will be sent to the <strong>Color</strong> <strong>Mapping</strong><br />

<strong>32</strong> with numeric data.<br />

Figure 4-1 Image Data Out Screen<br />

Each menu item corresponds to the type of image that will be attached to the numeric<br />

data package transferring to the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, as specify below:<br />

• NO IMAGE data transfer without image<br />

• BOTH data transfer with the images for Left and Right eyes<br />

• RIGHT data transfer with the Right eye’s image only<br />

• LEFT data transfer with the Left eye’s image only<br />

3. Press GRAPHIC PRINT SWITCH to conform that the menu items’ selection is<br />

completed. Instrument must display “RS2<strong>32</strong> DATA OUT” message.<br />

Each item indicates type of transferring image that will be sent to the <strong>Color</strong> <strong>Mapping</strong><br />

<strong>32</strong> with numeric data.<br />

Page 28<br />

TOPCON Corporation


4.2 Display <strong>Color</strong> Map Report Data<br />

Using the transferred report data, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> creates a new color map report<br />

window. This report window consists of < Ref Data >, < Kerato Data >, < Peripheral<br />

Data >, and the color map reports.<br />

4.3 Patient Registration<br />

Figure 4-2 <strong>Color</strong> map report display<br />

The patient registration service allows the user to create a database of records with<br />

patient information on an available storage media.<br />

� There are two alternatives to access the Patient Registration dialog box:<br />

• By selecting the [Patient] item from the Menu Bar;<br />

• By selecting [Save to Database] from the [File] pull-down menu (see Section 4.4)<br />

� To register a patient using the Menu Bar:<br />

Page 29<br />

TOPCON Corporation


• Select the [Patient] item from the Menu Bar;<br />

• Select [Register Patient] item from [Patient] menu (Figure 4-3).<br />

• Fill in the dialog box form, click on “Register” button to create patient record in<br />

the Database;<br />

Figure 4-3 Register Patient Dialog Box<br />

� To retrieve and update patient information from the Database:<br />

• Select [Patient] item in the Menu Bar;<br />

• Click on the [Select Patient] item from [Patient] menu. (Figure 4-4);<br />

Figure 4-4 Select Patient Dialog Box<br />

• Enter a search key in ID. Code and/or Last Name lines of the dialog box:<br />

• Click the “Information” button to retrieve a patient record in the Patient<br />

Information dialog box. To store modified record, click the “Update” button.<br />

• Click the “OK” button to retrieve a color map report image associated with<br />

the entered search key;<br />

� To delete record with patient information from the Database:<br />

Page 30<br />

TOPCON Corporation


• Select [Patient] item in the Menu Bar.<br />

• Select [Delete Patient] item from [Patient] menu.<br />

• Select a record from the list and press the “Delete” button on the dialog box;<br />

4.4 Saving Report Data<br />

Figure 4-5 Delete Patient Dialog Box<br />

� <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> support two types of report data storage:<br />

• Saving report data as a regular file with extension *.krp/*.kp8 for data received<br />

from KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P instrument on<br />

any available storage media.<br />

• Saving report data as a database file in database disk area allocated on available<br />

archive media. Patient information registration records and associated <strong>Color</strong><br />

<strong>Mapping</strong> <strong>32</strong> reports are stored in the created archive.<br />

� To save a document as a regular file from the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Report Window:<br />

• Select the [File] item in the Menu Bar.<br />

• Select the [Save As] item. (Do not change extension ‘.krp or .kp8’)<br />

Figure 4-6 Save Report Dialog Box<br />

When saving the report data as a file, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> attaches the calibration data to<br />

the file. This means that even if the KR-7000P/ KR-7100P/ KR-8000PA/ KR-8100PA/<br />

Page 31<br />

TOPCON Corporation


KR-8100P unit is replaced, the saved report data file may still be used in the new<br />

configuration.<br />

� To save a document to the database file from the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Report Window:<br />

• Select the [File] item in the Menu Bar.<br />

• Select [Save to Database] (see Section 3.2).<br />

• Fill patient registration form in the dialog box.<br />

• Click “OK” (Figure 4-7).<br />

• Fill dialog box form and click “OK” to save the report.<br />

Figure 4-7 Save Data Dialog Box<br />

4.5 <strong>Color</strong> Map Report Window Control<br />

Closing <strong>Color</strong> Map Report Window<br />

To close a report window, make the window active, then choose [Close] from the [File]<br />

menu or use the system menu of the report window. Alternatively, access the system<br />

menu of the window by clicking the square at the upper left corner, the left end of the<br />

title of the window.<br />

Putting <strong>Color</strong> Map Report Window on Top<br />

Many report windows can be opened simultaneously. The number of report windows<br />

one can open is only limited by the amount of system memory. The report window<br />

opened last is displayed on top. To place a different report window on top, click<br />

anywhere on that report window using the left mouse button.<br />

� To enable this service:<br />

Page <strong>32</strong><br />

TOPCON Corporation


• Click on [Open] from DB icon in the Tool Bar to display the Select Data dialog<br />

box.<br />

• Select data (see Section 4.6) to bring up a color map report data from the database.<br />

• Select a measurement point by moving the mouse on the image map and clicking<br />

the left button. Measurements associated with position of cross-hair marker on the<br />

color image will be displayed at the left top corner of the <strong>Color</strong> Map Report<br />

Window.<br />

4.6 Opening Saved Report Data<br />

To open a saved report data file, choose [Open] from the [File] menu, or click the open<br />

icon on the tool bar. The following dialog box will be displayed to select the saved<br />

report file:<br />

Figure 4-8 Open Report Dialog Box<br />

After choosing a report file and clicking “OK”, the dialog box will be hidden, and the<br />

selected report will be displayed in a new report window.<br />

The above dialog box can be displayed anytime by choosing [Open] in the [File] menu or<br />

by clicking the open file icon on the tool bar.<br />

Page 33<br />

TOPCON Corporation


4.7 Viewing Report Data from the Database<br />

� Report data saved on the database disk as archive can be retrieved in the following<br />

formats:<br />

• “Single map” is a view of one image with associated measurements and patient<br />

data (Figure 4-9). Selection of more then one file from the Record List Box<br />

(Figure 4-8) list disables the “OK” button of the Select Data dialog box.<br />

• “Multi map” image is a side by side view of several images with associated<br />

measurements and patient data (Figure 4-9). Patient records may be selected for<br />

the same eye as well as for different eyes from the database Record List Box.<br />

Selection of more records then the map can display disables the “OK” button of<br />

the Select Data dialog box (Figure 4-9);<br />

• “Difference Map” view is a color presentation of the difference between two<br />

images retrieved from database with associated measurement difference display<br />

box (Figure 4-10). The compared images are also presented side by side on the<br />

map. Selecting records of different eyes from the Select Data dialog box disables<br />

the “OK” button.<br />

� To retrieve <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> report image from the Database:<br />

• Select the [Open] from DB icon from the Tool Bar or click [Open] from Database<br />

on the [File] pull-down menu (see Section 3.2). The [Select Data] dialog box<br />

(Figure 4-9) will be displayed on the screen<br />

• Use the scroll list or press the [Select Patient] button to enter a search key;<br />

• Click the “OK” button to retrieve the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> report from the database<br />

Figure 4-9 Select Data Dialog Box<br />

Page 34<br />

TOPCON Corporation


Multi Map View<br />

Map view displays running measurements associated with the position of the cursor on<br />

the selected color map image. Measurement display in the bottom right corner of each<br />

image box (Figure 4-10) contains of following information:<br />

PWR Power in diopters at the cursor position;<br />

RAD Radius of curvature at the cursor position<br />

DIST Distance of cursor from the center of the grid in mm<br />

AXIS Angle of the axis connecting the cursor with the center of the grid<br />

in degrees (0°-360°);<br />

DISTP* Distance of cursor from the center of the pupil in mm<br />

AXISP* Angle of the axis connecting the cursor with the center of the pupil<br />

in degree (0°-360°)<br />

� To enable Multi Map View:<br />

• Click the [Open] from DB icon in the Tool Bar to bring up “Select Data” dialog<br />

box;<br />

• Select one of the “Multi Map” options from the “Select Map” box of the<br />

“Select Data” dialog box to bring up a “Multi Map” report data from the<br />

database;<br />

• Select measurement point by moving the mouse on the image map and clicking<br />

the left button. Measurements associated with the position of cross-hair marker on<br />

the color image will be displayed<br />

Figure 4-10 Multi Images Map View<br />

Page 35<br />

TOPCON Corporation


Difference Map View<br />

Map view displays running measurements associated with position of the cursor on the<br />

selected color map image. Measurement display is in the bottom, right corner of each<br />

comparing image box (Figure 4-10). It contains the same information type that the<br />

images on the Multi Map View along with the difference between the two. The following<br />

measurement information is provided for each map:<br />

PWR Power measurements of each map and the difference of the two.<br />

RAD Radius of curvature for each map and the difference of the two (mm).<br />

DIST Distance from the center of the grid for each map in mm.<br />

The difference map reports the distance of the first image in mm.<br />

AXIS The angle between the line connecting the cursor with the center of the<br />

grid and X-axis of the greed in degrees (0°-360°). The difference map<br />

reports the angle value of the first image.<br />

DISTP* Distance from the center of the pupil for each map in mm.<br />

AXISP* The angle between the line connecting the cursor with the center of the<br />

pupil and X-axis of the greed in degrees (0°-360°). The difference map<br />

reports the angle value of the first image.<br />

* Measurement is displayed if pupil detection data is available<br />

in the image file.<br />

To enable “Difference Map” View follow the steps from the paragraph “Multiple Map<br />

View” for “Difference Map” selection in the “Select Map” box of “Select Data” dialog<br />

box. (Figure 4-9).<br />

Figure 4-11 Difference Image View<br />

Page 36<br />

TOPCON Corporation


Chapter 5<br />

Option Settings<br />

When a color map report is open, the options may be changed for that report. To get the<br />

option dialog box menu, choose [Map Options] from the [Utility] menu, or click the<br />

option icon on the tool bar. The following dialog box will be displayed to change<br />

options:<br />

5.1 Map Method<br />

Figure 5-1 Option Setting Dialog Box<br />

Sagittal: This power is based on the normal distance from the cornea to the<br />

refractive axis (also known as Axial Power).<br />

Tangential: This power is based on the surface radius for each data point (also known<br />

as Instantaneous Power).<br />

5.2 Scale Unit<br />

Diopter: Displayed color map based on Diopter power scale.<br />

Page 37<br />

TOPCON Corporation


mm: Displayed color map based on mm power scale.<br />

5.3 Map Type<br />

Adjustable: Use the color map center and step values to adjust the color levels. By<br />

using appropriate values, more detail can be seen in the map. (The center<br />

value will appear as the bottom of the eleventh color.)<br />

Normalized: Minimum and Maximum levels of the data are used for limits of the color<br />

map. Thus, data will stretch over entire map. In cases of extreme data<br />

range, the top 1% and bottom 1% are excluded.<br />

Absolute: Standard scale ranging from 35.5D to 52D with a step size of 1/2D, or<br />

6.0mm to 9.3mm with a step size of 0.1mm. 33 different levels are used<br />

for this map by 22 colors.<br />

5.4 Overlays<br />

Cross Scale: X and Y axes overlay<br />

Angle Scale: Displays angle measurement around the color maps.<br />

mm Grid Scale: Overlay of a 1 mm grid<br />

Kerato Axis:<br />

None: No Kerato Axis displayed.<br />

Kerato Axis: Kerato Axis displayed based on the Keratometer axis.<br />

Sim-K Axis: Shows the keratometric circle (radius 1.5mm) along with the<br />

strongest axis ( as a black line ) and weakest axis ( as a gray line ).<br />

Pupil: Displays pupil boundary curve. Pupil check box is enabled, if pupil edge<br />

detection points are available in the opened map document file (Figure 5-2)<br />

This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />

Page 38<br />

TOPCON Corporation


5.5 Peripheral<br />

Figure 5-2 <strong>Color</strong> map with pupil overlays<br />

When measuring a patient eye with 5 point peripheral measurement, this switch becomes<br />

available along with peripheral display options:<br />

Figure 5-3 Peripheral Point 5 Option Dialog Box<br />

Display Point 5 Map: If this box is checked, the peripheral data will be displayed as a 15<br />

degree step map. If this box is unchecked, the peripheral data will be displayed as<br />

numeric data.<br />

Eccentricity Options<br />

No E Value: No eccentricity data displayed.<br />

Meridian E Value: Meridian eccentricity data displayed. (2 directions)<br />

Semi meridian E Value: Semi-meridian eccentricity data displayed. (4 directions)<br />

5.6 Others<br />

Page 39<br />

TOPCON Corporation


Initial Display Top: If this box is checked, the top of the new report window will be<br />

displayed. If this box is not checked, the bottom ( <strong>Color</strong> Map ) of<br />

the new report window will be displayed.<br />

Save Settings: If this box is checked, and the ‘OK’ switch is clicked, <strong>Color</strong><br />

<strong>Mapping</strong> <strong>32</strong> saves and uses these settings as the defaults for new<br />

reports.<br />

Auto Print: If this box is checked, data will automatically be sent to the printer<br />

(using default option settings) as soon as the <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

receives and reconstructs the color map structure.<br />

5.7 Image View<br />

This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />

Map: If this button is active color map with pupil center marker will be<br />

displayed, every time user opens map document file. Pupil center marker<br />

will appear only for document files containing pupil detection points.<br />

Image: If this button is active eyes’ color map reports will be replaced with its<br />

images in the color map report window (Figure 5-4). If image is available<br />

only for one eye in the opened document, <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> will display<br />

image only for this eye and color map for the eye without image.<br />

Overlay: This button repeats the same logic as the “Image” button. The difference is<br />

that image is overlaid with color map for each eye which image data is<br />

available.<br />

Side by Side: The button is enabled only for the document, which was opened from the<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> Database and its display format to “Single Map”<br />

(Section 4.7). Also, eye image data must be available in the database<br />

document.<br />

Page 40<br />

TOPCON Corporation


Figure 5-4 Report window (Image view – Image)<br />

Page 41<br />

TOPCON Corporation


Chapter 6<br />

Contact Lens Simulation Report Data<br />

6.1 Display Contact Lens Simulation Report Data<br />

To get a Contact Lens Simulation Report data when a <strong>Color</strong> map Report window is<br />

activated, perform following steps:<br />

• Choose “Contact Lens Fitting” from the [Utility] “<strong>Cal</strong>l Map” menu, or click the<br />

[Contact Lens] Icon on the Tool bar. When the Contact Lens Icon was clicked and a<br />

<strong>Color</strong> map Report has data for both eyes, the following “Select Processing Eye”<br />

dialog box will be displayed.<br />

Figure 6-1 Select processing Eye Dialog Box<br />

The contact lens simulation Parameter Setting dialog box will appear (Figure 6-4).<br />

• Select simulation parameters and simulation report view from the Parameter Setting<br />

dialog box, as specify in Section 6.4 later in this chapter.<br />

• Click “OK” to display the contact lens simulation report. The image view type in the<br />

Fitting <strong>Color</strong> Map box of the report is defined by the settings of the Image View<br />

control buttons in the “Parameter Setting” dialog box (Section 6.4).<br />

• The contact lens simulation report window will display Fitting <strong>Color</strong> Map view<br />

(Figure 6-2).<br />

Page 42<br />

TOPCON Corporation


Figure 6-2 Contact Lens Fitting Map display with pupil overlay<br />

6.1.1 Contact Lens Simulation Precaution<br />

Contact Lens Simulation<br />

Physician can not use report data for lens prescription without<br />

their personal validation.<br />

Page 43<br />

TOPCON Corporation


Figure 6-3 Contact Lens Fitting Map with pupil and eye image overlay display<br />

Page 44<br />

TOPCON Corporation


6.2 Parameters of the Contact Lens Simulation Report<br />

Abbreviations for the Contact Lens simulation report:<br />

BC Base curve (mm)<br />

BC2 Base curve 2 for toric lens (mm)BC must be greater then BC2<br />

E1 Eccentricity of aspheric lens<br />

TD Total diameter of a lens (mm)<br />

OZ Optic zone (mm) (which affects the optical function of a lens)<br />

SC Secondary curve (mm)<br />

SCW Secondary curve width (mm)<br />

E2 Eccentricity for secondary curve<br />

PC Peripheral curve (mm)<br />

PCW Peripheral curve width (mm)<br />

PC2 Peripheral curve 2 for toric lens (mm) PC must be greater than PC2<br />

AEL Axial edge lift (mm)<br />

VD Vertex distance of spectacles (mm)<br />

PWR Power of the contact lens to correct refraction spherical data (mm)<br />

PanH amount of horizontal lens lag from the corneal (mm)<br />

PanV amount of vertical lens lag from the corneal center (mm)<br />

TilH Horizontal lens tilt angle (mm)<br />

TilV Vertical tilt angle (mm)<br />

For a three curve contact lens, BC, TD, OZ, SC, SCW, PC, and PCW are necessary. If<br />

for a two-curve lens, SC and SCW, or PC and PCW may be omitted. Either SC or PC can<br />

be ignored if their width field (SCW or PCW) is zero.<br />

6.3 Handling Each Lens Type<br />

Lens types can be set in the parameter setting dialog box.<br />

RGP For a RGP lenses, a total of three curves including a base curve ( BC ), a<br />

secondary curve ( SC ), and a peripheral curve ( PC ), are available.<br />

Alternatives ( i.e. BC+SC only, or BC+PC only ) are also available.<br />

TORIC When a TORIC lens type has been selected, the BC2 and PC2 data fields<br />

will be activated. There are basic curve 2 and peripheral curve2,<br />

respectively. Listed below are two available lens structures in the TORIC<br />

lens type.<br />

BC (toric) + SC (spheric) + PC (spheric)<br />

BC (toric) + PC (toric)<br />

If the PC value is same as the PC2 value, the contact lens simulation<br />

function assumes the lens has spherical peripheral curves. In this case, the<br />

SC, and SCW fields will be activated. However, if PC differs from PC2,<br />

since this is a toric peripheral, the SC and SCW fields become disabled.<br />

ASPHERIC When an ASPHERIC lens type has been selected, the E1 and E2 fields<br />

will be activated. The meanings of these fields are eccentricity 1 and 2<br />

Page 45<br />

TOPCON Corporation


espectively. These values are calculated from the corneal topography,<br />

and initially, the two values are the same. One or both of them may be<br />

changed according to the manufacture’s lens specification. For a spherical<br />

secondary curve, set E2 equal to 0 to make it represent a spherical<br />

secondary curve. Use the two lens structure as follows:<br />

BC (aspheric) + SC (aspheric) + PC (spheric)<br />

BC (aspheric) + SC (spheric) + PC (spheric)<br />

In both cases, PC can be omitted when PCW is zero.<br />

SOFT Most SOFT lenses have one BC and One PC, so secondary curve fields<br />

are disabled. The fluorescein simulation is not available for soft lenses.<br />

PMMA For a PMMA lens type, the software treats the lens structure in the same<br />

way as a RGP lens.<br />

6.4 Parameter Setting<br />

When a contact lens simulation report is open or the ‘Contact Lens Fitting’ function from<br />

a color map report has been called for the first time, the parameters may be changed for<br />

that report. To get the parameter setting dialog box menu, choose ‘Change Parameters’<br />

from the [Utility] menu, or click the parameter icon on the tool bar. The following dialog<br />

box will be displayed to change parameters. (‘Select Data Set’ item will be appeared only<br />

when some contact lens data were registered – See Section 6.5 ):<br />

Figure 6-4 Parameter Setting dialog box<br />

Lens Type<br />

Please refer to section “6.3 Handling each lens type”.<br />

Page 46<br />

TOPCON Corporation


Map Type<br />

Absolute map type connects tear layer thickness to an actual fluorescent color.<br />

Normalized map type assigns colors to express the entire thickness range.<br />

Overlays<br />

The functions of the Overlay group of controls are the same as functions of the Overlay<br />

control group in the Map Option Settings dialog box for the <strong>Color</strong> Map Report document<br />

type, refer to the Section 5.4 of the manual.<br />

Reference BC<br />

Parameter helps to find a base curve that meets a certain apical clearance demand. The<br />

software finds the radius of curvature of a sphere, which is defined by three points.<br />

These points are the center of the lens with a certain apical clearance, and two optic zone<br />

end points, which touch the cornea.<br />

Sagittal Vault<br />

If the sagittal vault value is supposed to be a certain value, input that value. An<br />

appropriate TD value will be calculated for this sagittal vault.<br />

Axial Edge Lift<br />

The required AEL (Axial Edge Lift) value is specified, the software finds the appropriate<br />

PC value.<br />

Graph Setting: Profile Graph<br />

The tear layer thickness profile graph is displayed at the specified angle value.<br />

Graph Setting: Isometric Graph<br />

The circular tear layer variation graph is displayed at the specified radius value.<br />

Select Data Set<br />

The pull down scroll list appears when more then one contact lens data sets have been<br />

registered. The software will set the parameters based on the selected contact lens data<br />

set.<br />

Image View<br />

This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />

The functions of the Image View control group are the same as the functions of its<br />

counterpart in the Map Option Settings dialog box of the <strong>Color</strong> Map Report document<br />

type, refer to the Section 5.7 of the manual. There are three alternative views including<br />

the default Fitting <strong>Color</strong> Map view available and may be selected by setting the Image<br />

View control group in on of the following options:<br />

• MAP option to display Fitting <strong>Color</strong> Map(Figure 6-2)<br />

• IMAGE option to display eye image the same as in the <strong>Color</strong> Map Report, refer to<br />

Section 5.7(Figure 5-4)<br />

• OVERLAY option to display Fitting <strong>Color</strong> Map overlaid with eye image (Figure 6-3)<br />

6.5 Contact Lens Utility<br />

Page 47<br />

TOPCON Corporation


To get the contact lens utility dialog box menu, choose ‘Contact Lens Utility’ from the<br />

[Utility] menu. The following dialog box will be displayed to set the default and register<br />

the contact lens parameters sets. The number of the registered contact lens sets is up to 10.<br />

Figure 6-5 Contact Lens Utility dialog box<br />

Available setting parameters for the default and register contact lens sets are shown<br />

below.<br />

Contact Lens Parameters Default Setting<br />

When you have a commonly used contact lens specification, you may define these<br />

parameters. Definable parameters are “Lens Type”, “Map Type”, “Total Diameter of the<br />

contact lens”, “Optical Zone of the contact lens”, “Vertex Distance”, and “Overlays for<br />

the simulation report”. If you have registered some manufacture’s contact lens data sets,<br />

a “Select Data Set” pull-down list will appear.<br />

Figure 6-6 Contact Lens Parameters Default Setting dialog box<br />

CL Advice Enable<br />

This option Is available for KR-8000PA/ KR-8100PA/ KR-8100P<br />

A check box is located in the ‘Contact Lens Parameters Default Setting’ dialog box<br />

(Figure 6-6).<br />

This check box enables automatic setting of the simulation parameters for selected lens<br />

type (Section6.2) to the values calculated by the instrument. These values are displayed<br />

in the ‘’ section of the <strong>Color</strong> Map Report, please refer to the Section<br />

4.2(Figure 4-2). CL Advice data is available only for RGP and ASPHERIC type of lens.<br />

The other lens type selections are disabled as long as the check box is checked.<br />

Contact Lens Data Register<br />

It is possible to register some contact lens model parameters sets using the ‘Contact Lens<br />

Data Register function. You may register up to 10 contact lens model parameter sets.<br />

Page 48<br />

TOPCON Corporation


Figure 6-7 Contact Lens Data Register dialog box<br />

Page 49<br />

TOPCON Corporation


Chapter 7<br />

Printing Reports<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> allows printing of reports displayed on the screen using a Windows<br />

printer. There are 3 menu items related to the print operation in the [File] menu.<br />

• Print... Make a print of the active report window.<br />

• Print preview Preview print on the screen.<br />

• Print setup... Set up the Windows printer.<br />

7.1 Print Preview Screen<br />

To check how the report will be printed before making a print, choose [Print Preview]<br />

from the [File] menu.<br />

Figure 7-1 Print Preview Screen<br />

Page 50<br />

TOPCON Corporation


7.2 Print Setup Dialog Box<br />

The ‘Print Setup’ dialog box is used for changing the printer or paper to be used.<br />

Figure 7-2 Print Setup Dialog Box<br />

Page 51<br />

TOPCON Corporation


Chapter 8<br />

Other Features<br />

8.1 Hiding Tool Bar and Status Bar<br />

It is possible to hide the tool bar and status bar to make more space in the <strong>Color</strong> <strong>Mapping</strong><br />

<strong>32</strong> window. To hide the tool bar, choose “Toolbar” in the [View] menu. The check mark<br />

in front of “Toolbar” in the pull-down menu will be erased, and the tool bar will<br />

disappear. To hide the status bar, choose “Status Bar” in the [View] menu. The check<br />

mark at the head of “Status Bar” in the pull-down menu will be erased, and the status bar<br />

will disappear. To display these bars again, choose the same menu item, “Toolbar” or<br />

“Status Bar” in the [View] menu again. These menu items work as a toggle to hide and<br />

show the bars.<br />

8.2 Cascade and Tile<br />

Arrange report windows by choosing “Cascade” or ‘Tile” in the [Window] menu.<br />

“Cascade” arranges all report windows in an overlapping pattern. “Tile” arranges all<br />

report windows in a tile pattern.<br />

8.3 Select COM Port<br />

See Section 2.2.3 Port Setting.<br />

8.4 Reset COM Port<br />

If the communication is disrupted during the report data transfer from the KR-7000P/<br />

KR-7100P/ KR-8000PA/ KR-8100PA/ KR-8100P to <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>, choosing “Reset<br />

Port” in the [Port] menu can reset the communication port. <strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> reinitializes<br />

the communication port.<br />

Page 52<br />

TOPCON Corporation


<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong><br />

<strong>Software</strong> License Terms<br />

PLEASE READ THE FOLLOWING "LICENSE TERMS" CAREFULLY BEFORE USING (OR<br />

INSTALLING) THIS SOFTWARE!<br />

ANY AND ALL TERMS AND CONDITIONS OF THE "LICENCE TERMS" SHALL BE DEEMED TO<br />

BE ACCEPTED AND AGREED BY YOU IF YOU USE (OR INSTALL) THIS SOFTWARE.<br />

This “<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" <strong>Software</strong>, including its instruction manual and other documents, contained in this<br />

package comprises copyrightable works of TOPCON CORPORATION protected by copyright laws, and is<br />

licensed, not sold.<br />

TOPCON grants to you the right to use this “<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" <strong>Software</strong> under the terms and conditions<br />

set out below.<br />

LICENSE TERMS<br />

1. LICENSE<br />

Subject to the terms and conditions set out in this LICENSE TERMS, TOPCON CORPORATION<br />

("TOPCON") hereby grants to you (the "USER") a non-exclusive and non-transferable license (the<br />

"LICENSE") to use the "<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong>" including its instruction manual and other documents<br />

contained in this package (separatively or collectively the "SOFTWARE") only on a single computer<br />

which is a stand-alone or a terminal of a computer network (the “COMPUTER”). USER may not use the<br />

SOFTWARE on more than one computer or computer terminal at the same time.<br />

2. RESTRICTION<br />

2.1 USER shall NOT:<br />

(a) use the SOFTWARE on or with more than one computer or terminal at a time; or<br />

(b) make any copy of the SOFTWARE, except for making one copy of the SOFTWARE<br />

solely for back up purpose, or installing of the SOFTWARE into the COMPUTER,<br />

or its internal or external hard disc drive; or<br />

(c) sell, assign, exchange, pass, rent, lease or otherwise dispose of the SOFTWARE, or assign or sublicense<br />

any right to use of the SOFTWARE to anyone else; or<br />

(d) adapt, modify or alter the SOFTWARE, or merge any program with the SOFTWARE, or make any<br />

derivative from the SOFTWARE; or<br />

(e) remove, erase or make unclear any copyright notice of the SOFTWARE; or<br />

(f) decompile, disassemble, analyze, or make reverse-engineering to, the SOFTWARE; or<br />

(g) copy its instruction manual and other documents contained in this package.<br />

2.2 In any event that USER exchanges, or sells, assigns, passes, rents, leases or otherwise dispose of to<br />

any third party, or scraps the COMPUTER or any its internal or external hard disc drive or other<br />

memory means in which the SOFTWARE has been installed, USER shall destroy or erase any and all<br />

of the said installed SOFTWARE prior to the said exchange, etc.<br />

Page 53<br />

TOPCON Corporation


3. COPYRIGHT<br />

The SOFTWARE (including its recording Media) is the exclusive property of TOPCON, and full title,<br />

ownership and interests and copyrights and other intellectual property thereof shall remain vested with<br />

TOPCON at any and all times.<br />

4. TERM<br />

The LICENSE shall commence on the date when the SOFTWARE is first used (or installed) by USER, and<br />

shall remain in full force until the termination hereof pursuant to the provisions set out in Section 5.1<br />

below.<br />

5. TERMINATION<br />

5.1 The LICENSE shall be forthwith terminated if USER fails to comply with any term or condition of<br />

this LICENSE TERMS.<br />

5.2 Upon any termination of the LICENSE, USER shall forthwith cease using the SOFTWARE, and<br />

immediately destroy the original and any and all copies of the SOFTWARE (if any).<br />

6. LIMITED WARRANTY<br />

THE SOFTWARE IS PROVIDED ON AN "AS IS" BASIS WITHOUT WARRANTY OF ANY KIND.<br />

TOPCON FURTHER DISCLAIMS ALL WARRANTIES, INCLUDING, WITHOUT LIMITATION,<br />

ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR<br />

PURPOSE AND NONINFRINGEMENT.<br />

THE ENTIRE RISK ARISING OUT OF, RESULTING FROM OR IN CONNECTION WITH THE USE<br />

OR PERFORMANCE OF THE SOFTWARE REMAINS WITH THE USER. IN NO EVENT SHALL<br />

TOPCON OR ITS SUBSIDIARIES, AFFILIATES, DISTRIBUTORS OR AGENTS BE LIABLE FOR<br />

ANY CONSEQUENTIAL, INCIDENTAL, DIRECT, INDIRECT, SPECIAL, PUNITIVE OR OTHER<br />

DAMAGES WHATSOEVER (INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF<br />

BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS OF MEDICAL, CLINICAL OR BUSINESS<br />

INFORMATION, OR OTHER PECUNIARY LOSS, OR PERSONAL INJURY OR PROPERTY<br />

DAMAGE) ARISING OUT OF, RESULTING FROM OR INCONNECTION WITH THIS LICENSE<br />

TERMS OR THE USE OF OR INABILITY TO USE THE SOFTWARE.<br />

7. COMPLIANCE WITH EXPORT CONTROL LAWS<br />

Any and all exports or reexports of the SOFTWARE are subject to Japanese Exchange and Foreign Trade<br />

Law and other Japanese related laws and regulations, and/or the United States Export Administration Act,<br />

its implementing regulations as well as other U.S. export laws and regulations. Export or reexport of the<br />

SOFTWARE in a manner contrary to either Japanese or U.S. law or regulation is strictly prohibited.<br />

USER shall not export nor reexport to any country the SOFTWARE or any part or copy thereof without<br />

prior obtaining all required authorization from the Japanese or the United States government.<br />

TOPCON OR ITS SUBSIDIARIES, AFFILIATES, DISTRIBUTORS AND AGENTS SHALL NOT BE<br />

RESPONSIBLE NOR LIABLE TO USER FOR ANY LIABILITY, LOSS, DAMAGE, PUNISHMENT,<br />

OR PENALTY ARISING OUT OF OR RESULTING FROM USER ´ S VIOLATION OF OR<br />

NONCOMPLIANCE WITH ANY JAPANESE OR U.S. EXPORT CONTROL LAWS OR<br />

REGULATIONS.<br />

Page 54<br />

TOPCON Corporation


8. GOVERNING LAW<br />

The validity, performance, construction, interpretation, effect and enforcement of this LICENSE TERMS<br />

shall be governed by and under the laws of Japan.<br />

9. ENTIRE AGREEMENT<br />

This LICENSE TERMS constitutes the entire agreement between USER and TOPCON with respect to the<br />

subject matter hereof, and shall supersede and cancel any and all prior written or oral agreements,<br />

undertakings, negotiations, communications, commitments, representations, publications and advertisement,<br />

etc. not specifically and expressly incorporated herein.<br />

Page 55<br />

TOPCON Corporation


TOPCON Medical Systems, Inc.<br />

<strong>Color</strong> <strong>Mapping</strong> <strong>32</strong> <strong>Software</strong><br />

<strong>Instruction</strong> <strong>Manual</strong><br />

Version No. 1.24<br />

TOPCON MEDICAL SYSTEMS, INC.<br />

CORPORATE OFFICE: 37 West Century Road, Paramus, New Jersey 07652, U.S.A. Phone: 201-261-9450 Fax:201-387-2710 www.topcon.com<br />

TOPCON OMNI SYSTEMS, INC.<br />

Valley Forge Business Center, 2430 Blvd. of the Generals, Norristown, PA 19403, U.S.A. Phone:610-630-9200 Fax:610-630-6428 www.topconomni.com<br />

TOPCON CANADA INC.<br />

110 Provencher Avenue, Boisbriand, QC J7G 1N1 CANADA Phone:450-430-7771 Fax:450-430-6457 www.topcon.ca<br />

TOPCON EUROPE B.V.<br />

ESSE Baan 11, 2908 LJ Capelle a/d IJssel, THE NETHERLANDS Phone:010-4585077 Fax: 010-4585045 www.topconeurope.com<br />

ITALY OFFICE<br />

Via Montalcone 39, 20092 Cinisello B, mo(MI), ITALY Phone: 02-61-25-583 Fax: 02-61-25-927 www.topcon.it<br />

DENMARK OFFICE<br />

Ringstodvej 125, DK-4000 Roskilde, DENMARK Phone: 046-<strong>32</strong>75000 Fax: 046-<strong>32</strong>7555<br />

IRELAND OFFICE<br />

Unit 69 Western Parkway Business Centre, Lower Ballymount Road, Dublin 12, IRELAND Phone: 01460-0021 Fax: 01460-0129<br />

TOPCON S.A.R.L.<br />

89, Rue de Paris 92585 Clichy, Cedex, FRANCE Phone: 01-4106-9494 Fax: 01-4739-0251<br />

TOPCON DEUTSCHLAND G.m.b.H.<br />

Gieberallee 31-33, 47877 Willich, GERMANY Phone: 49-2154-8850 Fax: 49-2154-885111 E-mail: info@topcon.de www.topcon.de<br />

TOPCON SCANDINAVIA A.B.<br />

Neongatan 2 S-43151 Molndal, SWEDEN Phone: 31-7109200 Fax: 31-268607 www.topcon.se<br />

TOPCON ESPAÑA S.A.<br />

Frederic Mompou 5, ED. Euro 3, 08960, Sant Just Desvern Barcelona, SPAIN Phone: 34-93-473-4057 Fax: 34-93-473-39<strong>32</strong> www.topcon.es<br />

TOPCON ( GREAT BRITAIN ) LTD.<br />

Topcon House, Kennet Side, Bone Lane, Newbury, Berkshire RG14 5PX UNITED KINGDOM Phone: 01635-551120 Fax: 01635-551170 www.topcon.co.uk<br />

TOPCON SOUTH ASIA PTE. LTD.<br />

Blk 192 Pandan Loop, #07-01, Pantech Industrial Complex, SINGAPORE 128381 Phone: 62780222 Fax: 62733540 www.topcon.com.sg<br />

TOPCON AUSTRALIA PTY. LTD.<br />

Unit 18, 4 Avenue of Americas, Newington, NSW 2127 AUSTRALIA Phone: 61-2-8748-8777 Fax: 61-2-9647-2926 www.topcon.com.au<br />

TOPCON INSTRUMENTS ( MALAYSIA ) SDN. BHD.<br />

Excella Business Park Block C, Ground & 1 st Floor, Jalan Ampang Putra, Taman Ampang Hilir, 55100 Kuala Lumpur, MALAYSIA Phone:03-42701068 Fax:03-4274508<br />

TOPCON INSTRUMENTS ( THAILAND ) CO., LTD.<br />

77/162 Sinn Sathorn Tower, 37th Fl., Krungdhonburi Rd., Klongtonsai, Klongsarn, Bangkok 10600, THAILAND Phone: 02-440-1152~7 Fax: 02-440-1158<br />

TOPCON KOREA CORPORATION<br />

2F Yooseoung Bldg., 1595-3, Seocho-Dong, Seocho-Gu, Seoul, 137-876, KOREA Phone: 02-2055-0<strong>32</strong>1-7 Fax: 02-2055-0319 www.topcon.co.kr<br />

TOPCON OPTICAL ( H.K. ) LTD.<br />

2/F, Meeco Industrial Bldg., No.53-55 Au Pui Wan Street, Fo Tan Road, Shatin, N.T., HONG KONG Phone: 852-2690-1<strong>32</strong>8 Fax: 852-2690-2221 www.topcon.com.hk<br />

TOPCON CORPORATION BEIJING OFFICE<br />

Room 8A Poly Plaza Building, 14 Dongzhimen Nandajie, Dongcheng District, Beijing, 100027, CHINA Phone:10-6501-4191~2 Fax:10-6501-4190<br />

TOPCON CORPORATION SHANGHAI OFFICE<br />

Room 901~903 Lippo Plaza No. 222, Huai Hai Zhong Lu, Shanghai, 20021 CHINA Phone: 21-62701877 Fax: 21-53966485<br />

TOPCON CORPORATION BEIRUT OFFICE<br />

P.O.Box 70-1002 Antelias, BEIRUT-LEBANON Phone: 961-4-523525/523526 Fax: 961-4-521119<br />

TOPCON CORPORATION DUBAI OFFICE<br />

c/o Atlas Medical FZCO., P.O. Box 54304, C-25, Dubai Airport Free Zone, UNITED ARAB EMIRATES Phone: 971-4-2995900 Fax: 971-4-2995901<br />

TOPCON CORPORATION<br />

75-1 Hasunuma-cho, Itabashi-ku, Tokyo, 174-8580 JAPAN.<br />

Phone:3-3558-2520 Fax:3-3960-4214 www.topcon.co.jp<br />

Page 56<br />

© 2004 TOPCON CORPORATION All Rights Reserved<br />