Introduction<br />
TD_HeliosC_08.doc<br />
Rev. Janyary 08<br />
<strong>Angelantoni</strong> <strong>Industrie</strong> S.p.A. - Steril Division<br />
Export Office<br />
Via Achille Grandi 16, 20017 Mazzo di Rho (MI) – ITALY<br />
+39 02 939 701 43 / +39 02 939 701 21<br />
Head Quarter<br />
Località Cimacolle 464 – 06056 Massa Martana (PG) - ITALY<br />
+39 075 489551 +39 075 8955200<br />
HELIOS C<br />
HORIZONTAL LAMINAR AIRFLOW CABINET<br />
ARTICLE DESCRIPTION CODE<br />
HELIOS 48 C 00275400000<br />
HELIOS 72 C 00277100000<br />
Helios are <strong>horizontal</strong> <strong>laminar</strong> <strong>airflow</strong> <strong>cabinet</strong>s produced<br />
by Steril and designed for the situation where is<br />
requested the product protection from dangerous effects<br />
due to uncontrolled diffusion of air-transported<br />
contaminants from the environmental to the product<br />
during its handling.<br />
Applications<br />
Normally used in laboratory of microbiology and cell<br />
culture.<br />
Furthermore, used in innovative fields of scientific<br />
research and above all in the manipulation of:<br />
• Quality Control in food industries<br />
• Micro mechanical assembly<br />
• Preparation of nutritional parentheral bags<br />
• IVF<br />
Performances<br />
Product Protection<br />
Sterility in the working area higher than ISO 5 class according to ISO EN 14644-1 @ 0,3 and 0,5 µm particles size (Class<br />
100/M3.5 according to Federal Standard 209E) with a <strong>laminar</strong> air flow velocity of 0,40 m/s +/-20%.<br />
Since HELIOS is not a biosafety <strong>cabinet</strong>, it is not suitable for handling pathogens components.<br />
Norms and directives<br />
As far as electrical safety is concerned, the <strong>cabinet</strong> is manufactured according to EN 61010 norm and in particular:<br />
EN 292-1 EN 292-2 EN 61010-1 EN 50081-1 EN 50082-1 EN 50081-2<br />
EN 50082-2 IEC 801-2 IEC 801-4 ENV 50140 ENV 50141<br />
Furthermore it has been designed to fulfil the European Directives like:<br />
Machinery Directive 89/392/EEC, 91/368/EEC, 93/44/EEC, 93/68 EEC.<br />
Electromagnetic Compatibility Directive 89/336/EEC, 92/31/EEC, 93/68/EEC.<br />
The Masini Institute of Milan tested it and copy of the certification is available on request.<br />
Low voltage Directive 72/23/EEC, 93/68/EEC.<br />
IMPORTANT !!<br />
The <strong>cabinet</strong>s we produce, according to the above statements, are manufactured in compliance with the laws on safety<br />
into force.
Particularly, they are manufactured in compliance with Art. 6 Dlgs 626/94 and its modification (Italian law in compliance<br />
with European Directive EEC/89/391) on the DUTY OF DESIGNERS, OF MANUFACTURERS, OF SUPPLIERS AND<br />
INSTALLATORS.<br />
The <strong>cabinet</strong>s are NOT submitted to the directive EEC 93/42 on Medical Devices, as indicated in the definition of “Medical<br />
Devices” reported in art. 1 point 2a of the Directive mentioned.<br />
STERIL IS NOT responsible for damages to people and objects due to a non proper use of the <strong>cabinet</strong>, and due to a fail<br />
to comply with the user instructions and maintenance that are always supplied with the <strong>cabinet</strong>s.<br />
Technical Specification<br />
Working surface in stainless steel AISI 304L “2B” finishing with 15 mm lift (see<br />
picture).<br />
Side walls in toughened safety glass with 3 holes 19 mm diameter for the easy<br />
passage of external connection of fluids and/or cables without necessity of factory<br />
assembly.<br />
Lighting on working surface by means of 2 fluorescent lamps of 36W fitted in noncontaminated<br />
area for HELIOS 48 and 58W for HELIOS 72.<br />
Construction structure in steel epoxy-polyester powder coated, resistant to the most<br />
common industrial disinfectants.<br />
Ventilation system by means of n°1 motor blower for HELIOS 48 dedicated to the<br />
one-way <strong>airflow</strong> in the working chamber.<br />
HELIOS 72 is supplied with n° 2 motor blowers in parallel way system<br />
The motor blower is centrifugal type, direct driven motor with double aspiration and<br />
protection factor IP55.<br />
Air filtration by means of disposable high retention capacity pre-filters and H14<br />
HEPA filters with efficiency 99,995% MPPS according to EN 1822 (99,999% tested<br />
with DOP @ 0,3 µm).<br />
Protection grid, for HEPA filter made by anodized aluminium.<br />
Command and Controls<br />
Air ventilation control by means of automatic regulation of the revolution velocity of the motor blower. High-resolution<br />
flow rate volumetric device, directly interfaced to microprocessor, achieve the automatic regulation. The microprocessor<br />
guarantees the activity of the motor blower and controls the optimal function even in presence of effects caused by<br />
progressive clogging of the HEPA filters.<br />
Alarm device optical (red light signal visible from working position and display message) and acoustic (buzzer) type,<br />
activated, at real time, by the microprocessor, with indication of type of alarm activate showing on a wide alpha numeric<br />
display.<br />
Monitoring with alarms of <strong>laminar</strong> air flow velocity out of the pre-set range;<br />
Alpha numeric display showing:<br />
• Laminar air flow speed in m/s;<br />
• Date and time<br />
• Hours counters for <strong>cabinet</strong> life<br />
• UV Lamp run hours<br />
• HEPA filters run hours<br />
• Date of last HEPA filters change<br />
• Date of last service<br />
• Change language<br />
• Change password<br />
• Setting UV timer program<br />
• Date of last power failure<br />
Tests control<br />
In our Quality Control Dept., according to our procedure WI-MFG-006 included in the Quality manual of Steril<br />
Equipment Division. The Quality System of Steril Equipment Division is certified in compliance to EN ISO 9001 –2000<br />
with a Certificate n° CERT-00669-95-AQ-MIL-SINCERT/B released by “DNV – Det Norske Veritas”.<br />
On request, validation tests at site with IQ/OQ protocols are available.<br />
TD_HeliosC_08.doc<br />
Rev. January 08
Standard features<br />
• Automatic air velocity regulation<br />
• Fluorescent lights<br />
• Power socket 4 A UNEL IP 55<br />
• DOP/DOS 100% test inlet hose-barb port<br />
Optional features<br />
• Steel support stand epoxy polyester coated<br />
• Chest of 3 drawers on pivoting wheels in steel epoxy coated<br />
• 30W germicidal lamp with front screen, wavelength 253.7 nm (UV-C)<br />
• Gas manual tap manufactured in compliance with UNI/CIG regulations. Tap is provided with the automatic safety valve<br />
for interruption in case of electric blackout and/or air velocity out of pre-set range. Terminal connection for 6 mm<br />
diameter gas line.<br />
• Bunsen burner with automatic switch ignition<br />
• Bunsen burner with automatic foot switch ignition<br />
• Compressed air line with manual tap<br />
• Nitrogen line with manual tap<br />
• Vacuum line with manual tap<br />
Maintenance<br />
The replacement of filters can be done from the front of the <strong>cabinet</strong>. Any other maintenance activity (control of the<br />
electric board, replacement of UV lamp) can be done from the top.<br />
TD_HeliosC_08.doc<br />
Rev. January 08
Technical characteristics<br />
Model Unit Helios 48 C Helios 72 C<br />
Overall dimension (WxDxH) mm 1310x875x1284 1920x875x1284<br />
Working area dimension (WxDxH) mm 1270x537x814 1880x537x814<br />
Package dimension (WxDxH) mm 1430x980x1700 2010x980x1700<br />
Front opening height mm 814 814<br />
Net weight Kg 160 240<br />
System ventilation data<br />
Flow rate m 3 /h 1316 1975<br />
Heat emission at 25 °C (1)<br />
Kcal/h 345 525<br />
Noise level dB (A)
M2754-02.DGN<br />
TD_HeliosC_08.doc<br />
Rev. January 08<br />
A<br />
HELIOS 48<br />
HELIOS 72<br />
A LA R M<br />
A B<br />
1310<br />
1920<br />
Esc<br />
NETZSPANNUNG<br />
BRENNG AS<br />
6<br />
M O DUS O<br />
LI NK<br />
0<br />
7<br />
M O DUS I<br />
RECHTS<br />
1<br />
M O DUS I I<br />
AUF<br />
2<br />
8<br />
BELEUCHTUNG<br />
AB<br />
3<br />
9<br />
UV LAM PE<br />
ENTER<br />
4<br />
0 1 2 3 4 5 6 7 8 9<br />
NEBENSTECKER E s c ESC- M UTE<br />
5<br />
1430<br />
2010<br />
829 (useful)<br />
25<br />
1284<br />
785<br />
1700<br />
B<br />
870<br />
reggia<br />
537 (useful)<br />
FRAGILE<br />
980