Introduction<br />

TD_HeliosC_08.doc<br />

Rev. Janyary 08<br />

<strong>Angelantoni</strong> <strong>Industrie</strong> S.p.A. - Steril Division<br />

Export Office<br />

Via Achille Grandi 16, 20017 Mazzo di Rho (MI) – ITALY<br />

+39 02 939 701 43 / +39 02 939 701 21<br />

Head Quarter<br />

Località Cimacolle 464 – 06056 Massa Martana (PG) - ITALY<br />

+39 075 489551 +39 075 8955200<br />

HELIOS C<br />

HORIZONTAL LAMINAR AIRFLOW CABINET<br />

ARTICLE DESCRIPTION CODE<br />

HELIOS 48 C 00275400000<br />

HELIOS 72 C 00277100000<br />

Helios are <strong>horizontal</strong> <strong>laminar</strong> <strong>airflow</strong> <strong>cabinet</strong>s produced<br />

by Steril and designed for the situation where is<br />

requested the product protection from dangerous effects<br />

due to uncontrolled diffusion of air-transported<br />

contaminants from the environmental to the product<br />

during its handling.<br />

Applications<br />

Normally used in laboratory of microbiology and cell<br />

culture.<br />

Furthermore, used in innovative fields of scientific<br />

research and above all in the manipulation of:<br />

• Quality Control in food industries<br />

• Micro mechanical assembly<br />

• Preparation of nutritional parentheral bags<br />

• IVF<br />

Performances<br />

Product Protection<br />

Sterility in the working area higher than ISO 5 class according to ISO EN 14644-1 @ 0,3 and 0,5 µm particles size (Class<br />

100/M3.5 according to Federal Standard 209E) with a <strong>laminar</strong> air flow velocity of 0,40 m/s +/-20%.<br />

Since HELIOS is not a biosafety <strong>cabinet</strong>, it is not suitable for handling pathogens components.<br />

Norms and directives<br />

As far as electrical safety is concerned, the <strong>cabinet</strong> is manufactured according to EN 61010 norm and in particular:<br />

EN 292-1 EN 292-2 EN 61010-1 EN 50081-1 EN 50082-1 EN 50081-2<br />

EN 50082-2 IEC 801-2 IEC 801-4 ENV 50140 ENV 50141<br />

Furthermore it has been designed to fulfil the European Directives like:<br />

Machinery Directive 89/392/EEC, 91/368/EEC, 93/44/EEC, 93/68 EEC.<br />

Electromagnetic Compatibility Directive 89/336/EEC, 92/31/EEC, 93/68/EEC.<br />

The Masini Institute of Milan tested it and copy of the certification is available on request.<br />

Low voltage Directive 72/23/EEC, 93/68/EEC.<br />

IMPORTANT !!<br />

The <strong>cabinet</strong>s we produce, according to the above statements, are manufactured in compliance with the laws on safety<br />

into force.


Particularly, they are manufactured in compliance with Art. 6 Dlgs 626/94 and its modification (Italian law in compliance<br />

with European Directive EEC/89/391) on the DUTY OF DESIGNERS, OF MANUFACTURERS, OF SUPPLIERS AND<br />

INSTALLATORS.<br />

The <strong>cabinet</strong>s are NOT submitted to the directive EEC 93/42 on Medical Devices, as indicated in the definition of “Medical<br />

Devices” reported in art. 1 point 2a of the Directive mentioned.<br />

STERIL IS NOT responsible for damages to people and objects due to a non proper use of the <strong>cabinet</strong>, and due to a fail<br />

to comply with the user instructions and maintenance that are always supplied with the <strong>cabinet</strong>s.<br />

Technical Specification<br />

Working surface in stainless steel AISI 304L “2B” finishing with 15 mm lift (see<br />

picture).<br />

Side walls in toughened safety glass with 3 holes 19 mm diameter for the easy<br />

passage of external connection of fluids and/or cables without necessity of factory<br />

assembly.<br />

Lighting on working surface by means of 2 fluorescent lamps of 36W fitted in noncontaminated<br />

area for HELIOS 48 and 58W for HELIOS 72.<br />

Construction structure in steel epoxy-polyester powder coated, resistant to the most<br />

common industrial disinfectants.<br />

Ventilation system by means of n°1 motor blower for HELIOS 48 dedicated to the<br />

one-way <strong>airflow</strong> in the working chamber.<br />

HELIOS 72 is supplied with n° 2 motor blowers in parallel way system<br />

The motor blower is centrifugal type, direct driven motor with double aspiration and<br />

protection factor IP55.<br />

Air filtration by means of disposable high retention capacity pre-filters and H14<br />

HEPA filters with efficiency 99,995% MPPS according to EN 1822 (99,999% tested<br />

with DOP @ 0,3 µm).<br />

Protection grid, for HEPA filter made by anodized aluminium.<br />

Command and Controls<br />

Air ventilation control by means of automatic regulation of the revolution velocity of the motor blower. High-resolution<br />

flow rate volumetric device, directly interfaced to microprocessor, achieve the automatic regulation. The microprocessor<br />

guarantees the activity of the motor blower and controls the optimal function even in presence of effects caused by<br />

progressive clogging of the HEPA filters.<br />

Alarm device optical (red light signal visible from working position and display message) and acoustic (buzzer) type,<br />

activated, at real time, by the microprocessor, with indication of type of alarm activate showing on a wide alpha numeric<br />

display.<br />

Monitoring with alarms of <strong>laminar</strong> air flow velocity out of the pre-set range;<br />

Alpha numeric display showing:<br />

• Laminar air flow speed in m/s;<br />

• Date and time<br />

• Hours counters for <strong>cabinet</strong> life<br />

• UV Lamp run hours<br />

• HEPA filters run hours<br />

• Date of last HEPA filters change<br />

• Date of last service<br />

• Change language<br />

• Change password<br />

• Setting UV timer program<br />

• Date of last power failure<br />

Tests control<br />

In our Quality Control Dept., according to our procedure WI-MFG-006 included in the Quality manual of Steril<br />

Equipment Division. The Quality System of Steril Equipment Division is certified in compliance to EN ISO 9001 –2000<br />

with a Certificate n° CERT-00669-95-AQ-MIL-SINCERT/B released by “DNV – Det Norske Veritas”.<br />

On request, validation tests at site with IQ/OQ protocols are available.<br />

TD_HeliosC_08.doc<br />

Rev. January 08


Standard features<br />

• Automatic air velocity regulation<br />

• Fluorescent lights<br />

• Power socket 4 A UNEL IP 55<br />

• DOP/DOS 100% test inlet hose-barb port<br />

Optional features<br />

• Steel support stand epoxy polyester coated<br />

• Chest of 3 drawers on pivoting wheels in steel epoxy coated<br />

• 30W germicidal lamp with front screen, wavelength 253.7 nm (UV-C)<br />

• Gas manual tap manufactured in compliance with UNI/CIG regulations. Tap is provided with the automatic safety valve<br />

for interruption in case of electric blackout and/or air velocity out of pre-set range. Terminal connection for 6 mm<br />

diameter gas line.<br />

• Bunsen burner with automatic switch ignition<br />

• Bunsen burner with automatic foot switch ignition<br />

• Compressed air line with manual tap<br />

• Nitrogen line with manual tap<br />

• Vacuum line with manual tap<br />

Maintenance<br />

The replacement of filters can be done from the front of the <strong>cabinet</strong>. Any other maintenance activity (control of the<br />

electric board, replacement of UV lamp) can be done from the top.<br />

TD_HeliosC_08.doc<br />

Rev. January 08


Technical characteristics<br />

Model Unit Helios 48 C Helios 72 C<br />

Overall dimension (WxDxH) mm 1310x875x1284 1920x875x1284<br />

Working area dimension (WxDxH) mm 1270x537x814 1880x537x814<br />

Package dimension (WxDxH) mm 1430x980x1700 2010x980x1700<br />

Front opening height mm 814 814<br />

Net weight Kg 160 240<br />

System ventilation data<br />

Flow rate m 3 /h 1316 1975<br />

Heat emission at 25 °C (1)<br />

Kcal/h 345 525<br />

Noise level dB (A)


M2754-02.DGN<br />

TD_HeliosC_08.doc<br />

Rev. January 08<br />

A<br />

HELIOS 48<br />

HELIOS 72<br />

A LA R M<br />

A B<br />

1310<br />

1920<br />

Esc<br />

NETZSPANNUNG<br />

BRENNG AS<br />

6<br />

M O DUS O<br />

LI NK<br />

0<br />

7<br />

M O DUS I<br />

RECHTS<br />

1<br />

M O DUS I I<br />

AUF<br />

2<br />

8<br />

BELEUCHTUNG<br />

AB<br />

3<br />

9<br />

UV LAM PE<br />

ENTER<br />

4<br />

0 1 2 3 4 5 6 7 8 9<br />

NEBENSTECKER E s c ESC- M UTE<br />

5<br />

1430<br />

2010<br />

829 (useful)<br />

25<br />

1284<br />

785<br />

1700<br />

B<br />

870<br />

reggia<br />

537 (useful)<br />

FRAGILE<br />

980

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!