ENGLISH<br />
24<br />
3. Afterwards, rinse the container thoroughly<br />
<strong>and</strong> let it dry. Using “Neodisher AN“ as<br />
cleaner (manufactured by DR. WEIGERT),<br />
it is also possible to clean the system components<br />
in a specialised dishwasher if the<br />
manufacturer’s instructions are followed.<br />
4.1.5 Cleaning the hose accessories<br />
1. Dispose of all hoses <strong>and</strong> bacteria filters<br />
which are intended for once-only use.<br />
2. Rinse out the silicone hoses or immerse<br />
them in the customary disinfectant solution.<br />
3. Afterwards, rinse the silicone hoses out<br />
thoroughly <strong>and</strong> let them dry.<br />
It is recommended that the hoses be exchanged<br />
every four weeks, <strong>and</strong> for every<br />
new patient using the aspirator.<br />
4.2. Maintenance <strong>and</strong> servicing<br />
• If the ASSKEA S <strong>and</strong> M-<strong>Series</strong> aspirators<br />
are used in accordance with the user manual,<br />
they require no maintenance.<br />
• Each time they are used, carry out a visual<br />
check <strong>and</strong> a check of functioning first. This<br />
must include all the accessories of the relevant<br />
unit.<br />
• The opening up <strong>and</strong>, likewise, the repair<br />
of the units is a matter solely for ASSKEA<br />
GmbH or specialist staff authorised by AS-<br />
SKEA GmbH who follow the servicing instructions<br />
issued by the manufacturer <strong>and</strong><br />
take all necssary technicl <strong>and</strong> hygienic precautions.<br />
• The unit may be returned to ASSKEA for<br />
repair either direct or via the specialist distributor<br />
from whom it was obtained.<br />
• All accessories must be cleaned <strong>and</strong> disinfected<br />
before return. The unit itself must be<br />
treated with surface disinfectant.<br />
ASSKEA GmbH neither guarantees faultless<br />
4.3. Checking the ASSKEA M- <strong>and</strong> S-<br />
<strong>Series</strong> aspirators<br />
• The periodic inspections must always<br />
comply with the currently applicable<br />
regulations <strong>and</strong> directives<br />
(such as IEC62353, 2007/47/EC, etc.).<br />
• ASSKEA recommends that a safety<br />
check (ASSKEA code: STK) be carried<br />
out at least every 24 months.<br />
• It is the equipment operator’s sole<br />
responsibility to ensure the safety<br />
check is carried out.<br />
• ASSKEA GmbH offers partners <strong>and</strong><br />
customers rapid <strong>and</strong> professional<br />
reconditioning, also any checks <strong>and</strong><br />
inspections necessary.