10/24/11<br />

Health Physicist monitors own medical dose<br />

from radioiodine thyroid ablation procedure<br />

California Campus RSO Meeting<br />

University of California, Irvine Medical Center<br />

October 24, 2011<br />

Sander Perle<br />

Dosimetry Services Division<br />

1


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Agenda<br />

1. Present State of Passive Dosimetry<br />

2. Future State of Passive Dosimetry<br />

3 Direct Ion Storage (Instadose) Type Test Information<br />

4. I-131 thyroid study with instadose & TLD<br />

5. Future<br />

2<br />

Dosimetry Services<br />

Division


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

What dosimetry is appropriate for<br />

a specific monitoring requirement<br />

§ What is the purpose for wearing the dosimeter<br />

§ What are the key requirements<br />

§ How accurate and precise<br />

§ Ease of obtaining the dose data<br />

§ Timeliness of obtaining the data<br />

10/24/11<br />

Dosimetry Services<br />

Division<br />

3


§ Film<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Minimum Reportable Dose is<br />

0.10 mSv<br />

Non-tissue Equivalent<br />

Energy Dependent<br />

Angular Dependent<br />

Requires Return processing<br />

Present state of passive dosimetry<br />

§ TLD<br />

§ Other <strong>Technologies</strong> have<br />

good features as well, but<br />

still require a processing<br />

facility to read and report<br />

doses to end users<br />

RPL<br />

OSL<br />

CR39<br />

Minimum Reportable Dose is<br />

0.01 mSv<br />

Tissue Equivalent<br />

Energy Independent<br />

Angular Independent<br />

Requires Return processing<br />

Dosimetry Services<br />

Division<br />

4


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Future State of Passive Dosimetry<br />

§ An accredited and approved remote read that does not<br />

require returning for processing and is electronically<br />

analyzed by trained staff that meets or exceeds technical<br />

requirements from customers and regulatory entities<br />

Dosimetry Services<br />

Division<br />

5


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Direct Ion Storage (DIS) Technology<br />

§ Instadose utilizes the DIS<br />

detector which stands for<br />

Direct Ion Storage<br />

§ DIS Dosimeter:<br />

§ Non Volatile Analog<br />

Memory Cell surrounded<br />

by a Gas Filled Ion<br />

Chamber<br />

§ For photon radiation,<br />

initial interactions take<br />

place in the wall material<br />

and secondary electrons<br />

ionize the gas of the<br />

chamber<br />

Dosimetry Services<br />

Division<br />

6


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

DIS Detector Element<br />

Dosimetry Services<br />

Division<br />

7


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Direct Ion Storage Technology,<br />

Instadose, Read on a PC or Mac<br />

§ Instadose allows users to have the flexibility to view their radiation dose at any<br />

time from any computer with internet access. Readings via a PC or Mac are<br />

enabled by a USB compatible detector.<br />

§ When a user wishes to obtain a reading they simply log-in to their account, plugin<br />

instadose to a USB port and obtain their read. The accumulated dose stored<br />

on instadose is processed through a proprietary algorithm. Once complete a<br />

graphical representation of the current dose will load on the screen. Users can<br />

also view their cumulative dose level by clicking "View Cumulative Dose."<br />

Dosimetry Services<br />

Division<br />

8


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Technical Specs<br />

Minimum Reportable Dose (MRD) 0.01 mSv<br />

Lower Limit of Detection (LLD) 0.01 mSv<br />

Useful Dose Range 0.01 mSv – 5 Sv<br />

Energy Response Photon 5 keV – 6 MeV<br />

Accreditations/Approvals/Licenses NVLAP (100555), HSE, United<br />

Arab Emirates, Victoria and New<br />

South Wales, Australia, Nigeria<br />

9<br />

Dosimetry Services<br />

Division<br />

9


2<br />

1.8<br />

1.6<br />

1.4<br />

1.2<br />

1<br />

0.8<br />

0.6<br />

0.4<br />

0.2<br />

0<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Instadose Energy Response<br />

Instadose Energy Response<br />

NVLAP Proficiency Blind Test , ANSI N13.11-2009<br />

Energy (keV)<br />

Dosimetry Services<br />

Division<br />

10


1.200<br />

1.150<br />

1.100<br />

1.050<br />

1.000<br />

0.950<br />

0.900<br />

0.850<br />

0.800<br />

0.750<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Instadose Linearity<br />

Instadose Linearity<br />

1 10 100 1000 10000 100000<br />

Irradiated Dose (mRad)<br />

Instadose Linearity<br />

Dosimetry Services<br />

Division<br />

11


2.00<br />

1.80<br />

1.60<br />

1.40<br />

1.20<br />

1.00<br />

0.80<br />

0.60<br />

-100 -50 0 50 100<br />

Degrees<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

NS 30 (24 keV)<br />

0.60<br />

-100 -50 0 50 100<br />

10/24/11<br />

NS 80 (65 keV)<br />

1.40<br />

1.20<br />

1.00<br />

0.80<br />

Degrees<br />

Instadose Angularity<br />

Vertical<br />

Horizontal<br />

Vertical<br />

Horizontal<br />

NS 60 (48 keV)<br />

1.40<br />

1.20<br />

1.00<br />

0.80<br />

0.60<br />

-100 -50 0 50 100<br />

Degrees<br />

Vertical<br />

Horizontal<br />

Dosimetry Services<br />

Division<br />

12


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Low Dose Testing<br />

Dosimetry Services<br />

Division<br />

13


§ Dose Rate Independent<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Instadose Dose Rate Independent<br />

Dosimetry Services<br />

Division<br />

14


§ 1.1 µSv/hr to 100 µSv/hr<br />

Response<br />

1.20<br />

1.10<br />

1.00<br />

0.90<br />

0.80<br />

0.70<br />

0.60<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Low Dose Rate Test Data<br />

Low Dose Rate Test<br />

0 20 40 60 80 100 120<br />

Dose Rate (µSv/hour)<br />

Dosimetry Services<br />

Division<br />

15


§ Two tests<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

Temperature Testing<br />

§ Test 1 - 3 Groups (reference, +40C, -20C) were irradiated to 3 mSv<br />

§ Test 2 – 2 Groups (0.35 mSv and 0.05 mSv) at +40C<br />

Dose Value (mR)<br />

10/24/11<br />

350<br />

330<br />

310<br />

290<br />

270<br />

250<br />

Instadose Temperature Test<br />

299.3 301.7 299.2<br />

Reference Dose +40 C -20 C<br />

Group<br />

Dose Value (mR)<br />

40.0<br />

35.0<br />

30.0<br />

25.0<br />

20.0<br />

15.0<br />

10.0<br />

5.0<br />

0.0<br />

Instadose Low Dose<br />

Temperature Test<br />

36.9 36.0<br />

4.6<br />

Reference Dose 35 mR 5 mR<br />

Group<br />

Dosimetry Services<br />

Division<br />

16


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

I-131 Radioiodine Thyroid Uptake Study<br />

§ August 29, 2011, radioiodine thyroid ablation procedure<br />

§ Administered 26.2 mCi (1 GBq) Sodium Iodide @9AM<br />

§ Instadose and TLD badges worn on lanyard (facing thyroid)<br />

§ Instadose on neck was read approximately every hour during daytime<br />

hours<br />

§ Instadose on neck was read prior to going to sleep, placed on window sill<br />

approximately 1 ft. from couch<br />

§ Instadose read first thing in the morning and the process described<br />

above was repeated<br />

§ After the first week the Instadose was read approximately 2 or 3 times a<br />

day<br />

§ Instadose and TLD badges strategically located throughout house to<br />

monitor doses<br />

§ TLDs were processed on September 8, 2011<br />

10/24/11<br />

Dosimetry Services<br />

Division<br />

17


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

I-131 Radioiodine Thyroid Uptake Study<br />

Dosimetry Services<br />

Division<br />

18


Account: Sandy, Inc. (363)<br />

Location: Main<br />

EID Name<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Report Run Time: 10/22/2011 8:38 AM<br />

I-131 Radioiodine Thyroid Uptake Study<br />

HISTORY DETAIL REPORT<br />

Report Period: 08/29/2011 12:00 AM - 09/05/2011 12:00 AM<br />

Serial<br />

Number<br />

1 2 3 4 5<br />

6 7 8<br />

Accredited by the National Institute of Standards and Technology through NVLAP<br />

for the specific scope of accreditation under lab code 100555-0<br />

SEE LAST PAGE FOR COMPLETE REPORT DETAILS BY COLUMN NUMBER<br />

IT IS RECOMMENDED THAT YOU KEEP THIS REPORT FOR YOUR RECORDS<br />

P.O. BOX 16451<br />

IRVINE, CA 92623<br />

U.S./CANADA: 800.359.9686<br />

www.instadose.com<br />

Body<br />

Region<br />

Ship To:<br />

Read<br />

Date/Time<br />

113 Sienna Ridge<br />

Mission Viejo CA 92692<br />

United States<br />

Dose Reported in Units of mrem<br />

Deep<br />

(Whole Body)<br />

Number of Reads: 12 Total: 821 821 821<br />

Instadose, Two 74537 Torso 9/5/2011 8:33:00 PM 1 1 1<br />

74537 Torso 9/3/2011 2:28:00 PM 12 12 12<br />

74537 Torso 8/31/2011 7:43:00 PM * * *<br />

74537 Torso 8/31/2011 7:42:00 PM 3 3 3<br />

74537 Torso 8/31/2011 10:01:00 AM 18 18 18<br />

74537 Torso 8/30/2011 10:03:00 AM 30 30 30<br />

74537 Torso 8/29/2011 9:07:00 PM 50 50 50<br />

Number of Reads: 7 Total: 114 114 114<br />

Perle, Sandy 11801 Torso 9/5/2011 8:38:00 PM 66 66 66<br />

11801 Torso 9/5/2011 8:07:00 AM * * *<br />

11801 Torso 9/5/2011 8:05:00 AM 4 4 4<br />

11801 Torso 9/4/2011 8:57:00 PM 98 98 98<br />

11801 Torso 9/4/2011 9:12:00 AM * * *<br />

11801 Torso 9/3/2011 8:52:00 PM 64 64 64<br />

11801 Torso 9/3/2011 2:00:00 PM 63 63 63<br />

11801 Torso 9/3/2011 8:01:00 AM * * *<br />

11801 Torso 9/2/2011 9:20:00 PM 110 110 110<br />

11801 Torso 9/2/2011 1:01:00 PM 52 52 52<br />

Eye<br />

Dosimetry Services<br />

Division<br />

Page 4 of 7<br />

Shallow<br />

(Skin)<br />

19


ead date time mrem/h Read<br />

8/28/2011 17:50 0 2<br />

8/29/2011 9:58 3 56<br />

8/29/2011 11:05 139 156<br />

8/29/2011 12:02 157 149<br />

8/29/2011 13:02 133 133<br />

8/29/2011 14:01 153 150<br />

8/29/2011 15:02 139 141<br />

8/29/2011 16:00 133 129<br />

8/29/2011 17:03 122 129<br />

8/29/2011 18:00 96 91<br />

8/29/2011 19:00 94 94<br />

8/29/2011 20:03 99 104<br />

8/29/2011 21:02 90 88<br />

8/29/2011 22:13 98 116<br />

8/30/2011 6:54<br />

8/30/2011 7:57 96 101<br />

8/30/2011 9:00 105 111<br />

8/30/2011 10:01<br />

8/30/2011 11:09 97 109<br />

8/30/2011 12:00 79 68<br />

8/30/2011 13:16 64 80<br />

8/30/2011 15:51 142<br />

8/30/2011 19:03 136<br />

8/30/2011 21:53 58 163<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

I-131 Radioiodine Thyroid Uptake Study<br />

read date time mrem/h Read<br />

8/31/2011 9:57<br />

8/31/2011 13:01 53 162<br />

8/31/2011 13:59 50 49<br />

8/31/2011 18:01 45 181<br />

8/31/2011 20:50 21 26<br />

8/31/2011 22:02 23 28<br />

9/1/2011 14:10 55<br />

9/1/2011 21:28 19 138<br />

9/2/2011 8:12 2<br />

9/2/2011 13:01 11 52<br />

9/2/2011 21:20 13 110<br />

9/3/2011 8:02<br />

9/3/2011 14:00 11 63<br />

9/3/2011 20:52 9 64<br />

9/4/2011 9:13<br />

9/4/2011 21:58 8 98<br />

9/5/2011 8:07<br />

9/5/2011 20:38 5 66<br />

Dosimetry Services<br />

Division<br />

20


dose rate (mrem·hr -1 )<br />

160<br />

140<br />

120<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Instadose response on neck (lanyard)<br />

Dosimetry Services<br />

Division<br />

21


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

Calculated uptake<br />

DRest = Γ·Aest·d -2 DRest = Γ·Aest·d -2 (e -­‐μ x)<br />

ft ft cm (skin shielding)<br />

Date/ time elapse time Aest: (Aest = A0e -ʎeff∆t ) 1 0.25 1<br />

DRcalc Acalc = (DRcalc)(Γ·∙d -2 ) -1 Fraction<br />

8/28/2011 17:50 0.000<br />

8/29/2011 9:00 26.200<br />

8/29/2011 9:58 0.041 3 0.099 0.004<br />

8/29/2011 11:05 0.047 139 3.959 0.151<br />

8/29/2011 12:02 0.039 5.240 11.5 184.4 165.2 157 4.474 0.171<br />

8/29/2011 13:02 0.042 5.219 11.5 183.7 164.6 133 3.774 0.144<br />

8/29/2011 14:01 0.041 5.199 11.4 183.0 163.9 153 4.350 0.166<br />

8/29/2011 15:02 0.042 5.178 11.4 182.3 163.3 139 3.952 0.151<br />

8/29/2011 16:00 0.040 5.158 11.3 181.6 162.7 133 3.772 0.144<br />

8/29/2011 17:03 0.044 5.137 11.3 180.8 162.0 122 3.476 0.133<br />

8/29/2011 18:00 0.040 5.117 11.3 180.1 161.4 96 2.717 0.104<br />

8/29/2011 19:00 0.042 5.097 11.2 179.4 160.7 94 2.665 0.102<br />

8/29/2011 20:03 0.044 5.076 11.2 178.7 160.1 99 2.812 0.107<br />

8/29/2011 21:02 0.041 5.056 11.1 178.0 159.4 90 2.552 0.097<br />

8/29/2011 22:13 0.049 5.032 11.1 177.1 158.7 98 2.786 0.106<br />

8/30/2011 6:54 0.362 4.862 10.7 171.1 153.3<br />

8/30/2011 7:57 0.044 4.841 10.7 170.4 152.7 96 2.740 0.105<br />

8/30/2011 9:00 0.044 4.821 10.6 169.7 152.0 105 2.995 0.114<br />

8/30/2011 10:01 0.042 4.802 10.6 169.0 151.4<br />

8/30/2011 11:09 0.047 4.780 10.5 168.3 150.7 97 2.746 0.105<br />

8/30/2011 12:00 0.036 4.764 10.5 167.7 150.2 79 2.234 0.085<br />

8/30/2011 13:16 0.052 4.740 10.4 166.9 149.5 64 1.817 0.069<br />

8/30/2011 15:51 0.108 4.692 10.3 165.2 148.0<br />

8/30/2011 19:03 0.134 4.633 10.2 163.1 146.1<br />

8/30/2011 21:53 0.118 4.581 10.1 161.3 144.5 58 1.637 0.062<br />

8/31/2011 9:57 0.503 4.367 9.6 153.7 137.7<br />

8/31/2011 13:01 0.128 4.314 9.5 151.8 136.0 53 1.498 0.057<br />

8/31/2011 13:59 0.040 4.297 9.5 151.3 135.5 50 1.434 0.055<br />

8/31/2011 18:01 0.168 4.229 9.3 148.9 133.3 45 1.273 0.049<br />

8/31/2011 18:04 0.002 4.228 9.3 148.8 133.3<br />

8/31/2011 18:05 0.001 4.228 9.3 148.8 133.3<br />

8/31/2011 18:08 0.002 4.227 9.3 148.8 133.3<br />

8/31/2011 19:37 0.061 4.202 9.2 147.9 132.5<br />

8/31/2011 20:50 0.051 4.182 9.2 147.2 131.9 21 0.602 0.023<br />

8/31/2011 22:02 0.050 4.162 9.2 146.5 131.2 23 0.666 0.025<br />

9/1/2011 14:10 0.672 3.904 8.6 137.4 123.1<br />

9/1/2011 21:28 0.304 3.792 8.3 133.5 119.6 19 0.538 0.021<br />

9/2/2011 8:12 0.447 3.634 8.0 127.9 114.6<br />

9/2/2011 13:01 0.201 3.565 7.8 125.5 112.4 11 0.307 0.012<br />

9/2/2011 21:20 0.347 3.449 7.6 121.4 108.8 13 0.376 0.014<br />

9/3/2011 8:02 0.446 3.306 7.3 116.4 104.2<br />

9/3/2011 14:00 0.249 3.228 7.1 113.6 101.8 11 0.300 0.011<br />

9/3/2011 20:52 0.286 3.142 6.9 110.6 99.1 9 0.265 0.010<br />

9/4/2011 9:13 0.515 2.991 6.6 105.3 94.3<br />

9/4/2011 21:58 0.531 2.844 6.3 100.1 89.7 8 0.218 0.008<br />

9/5/2011 8:07 0.423 2.731 6.0 96.1 86.1<br />

9/5/2011 20:38 0.522 2.599 5.7 91.5 81.9 5 0.150 0.006<br />

10/24/11<br />

Dosimetry Services<br />

Division<br />

22


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Loca&on<br />

Instadose vs. TLD results<br />

Instadose Reading<br />

(mSv) TLD Reading (mSv)<br />

Collar 35.15 30.25<br />

Computer Room #1 0.51 0.71<br />

Computer Room #2 0.47 0.6<br />

Window Sill #1 1.11 0.96<br />

Window Sill #2 1.11 0.83<br />

Spouse 0.05 0<br />

Dosimetry Services<br />

Division<br />

23


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Future?<br />

Dosimetry Services<br />

Division<br />

24


<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />

Learn more at: www.mirion.com<br />

10/24/11<br />

Questions?<br />

Dosimetry Services<br />

Division<br />

25

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!