10/24/11<br />
Health Physicist monitors own medical dose<br />
from radioiodine thyroid ablation procedure<br />
California Campus RSO Meeting<br />
University of California, Irvine Medical Center<br />
October 24, 2011<br />
Sander Perle<br />
Dosimetry Services Division<br />
1
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Agenda<br />
1. Present State of Passive Dosimetry<br />
2. Future State of Passive Dosimetry<br />
3 Direct Ion Storage (Instadose) Type Test Information<br />
4. I-131 thyroid study with instadose & TLD<br />
5. Future<br />
2<br />
Dosimetry Services<br />
Division
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
What dosimetry is appropriate for<br />
a specific monitoring requirement<br />
§ What is the purpose for wearing the dosimeter<br />
§ What are the key requirements<br />
§ How accurate and precise<br />
§ Ease of obtaining the dose data<br />
§ Timeliness of obtaining the data<br />
10/24/11<br />
Dosimetry Services<br />
Division<br />
3
§ Film<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Minimum Reportable Dose is<br />
0.10 mSv<br />
Non-tissue Equivalent<br />
Energy Dependent<br />
Angular Dependent<br />
Requires Return processing<br />
Present state of passive dosimetry<br />
§ TLD<br />
§ Other <strong>Technologies</strong> have<br />
good features as well, but<br />
still require a processing<br />
facility to read and report<br />
doses to end users<br />
RPL<br />
OSL<br />
CR39<br />
Minimum Reportable Dose is<br />
0.01 mSv<br />
Tissue Equivalent<br />
Energy Independent<br />
Angular Independent<br />
Requires Return processing<br />
Dosimetry Services<br />
Division<br />
4
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Future State of Passive Dosimetry<br />
§ An accredited and approved remote read that does not<br />
require returning for processing and is electronically<br />
analyzed by trained staff that meets or exceeds technical<br />
requirements from customers and regulatory entities<br />
Dosimetry Services<br />
Division<br />
5
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Direct Ion Storage (DIS) Technology<br />
§ Instadose utilizes the DIS<br />
detector which stands for<br />
Direct Ion Storage<br />
§ DIS Dosimeter:<br />
§ Non Volatile Analog<br />
Memory Cell surrounded<br />
by a Gas Filled Ion<br />
Chamber<br />
§ For photon radiation,<br />
initial interactions take<br />
place in the wall material<br />
and secondary electrons<br />
ionize the gas of the<br />
chamber<br />
Dosimetry Services<br />
Division<br />
6
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
DIS Detector Element<br />
Dosimetry Services<br />
Division<br />
7
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Direct Ion Storage Technology,<br />
Instadose, Read on a PC or Mac<br />
§ Instadose allows users to have the flexibility to view their radiation dose at any<br />
time from any computer with internet access. Readings via a PC or Mac are<br />
enabled by a USB compatible detector.<br />
§ When a user wishes to obtain a reading they simply log-in to their account, plugin<br />
instadose to a USB port and obtain their read. The accumulated dose stored<br />
on instadose is processed through a proprietary algorithm. Once complete a<br />
graphical representation of the current dose will load on the screen. Users can<br />
also view their cumulative dose level by clicking "View Cumulative Dose."<br />
Dosimetry Services<br />
Division<br />
8
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Technical Specs<br />
Minimum Reportable Dose (MRD) 0.01 mSv<br />
Lower Limit of Detection (LLD) 0.01 mSv<br />
Useful Dose Range 0.01 mSv – 5 Sv<br />
Energy Response Photon 5 keV – 6 MeV<br />
Accreditations/Approvals/Licenses NVLAP (100555), HSE, United<br />
Arab Emirates, Victoria and New<br />
South Wales, Australia, Nigeria<br />
9<br />
Dosimetry Services<br />
Division<br />
9
2<br />
1.8<br />
1.6<br />
1.4<br />
1.2<br />
1<br />
0.8<br />
0.6<br />
0.4<br />
0.2<br />
0<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Instadose Energy Response<br />
Instadose Energy Response<br />
NVLAP Proficiency Blind Test , ANSI N13.11-2009<br />
Energy (keV)<br />
Dosimetry Services<br />
Division<br />
10
1.200<br />
1.150<br />
1.100<br />
1.050<br />
1.000<br />
0.950<br />
0.900<br />
0.850<br />
0.800<br />
0.750<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Instadose Linearity<br />
Instadose Linearity<br />
1 10 100 1000 10000 100000<br />
Irradiated Dose (mRad)<br />
Instadose Linearity<br />
Dosimetry Services<br />
Division<br />
11
2.00<br />
1.80<br />
1.60<br />
1.40<br />
1.20<br />
1.00<br />
0.80<br />
0.60<br />
-100 -50 0 50 100<br />
Degrees<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
NS 30 (24 keV)<br />
0.60<br />
-100 -50 0 50 100<br />
10/24/11<br />
NS 80 (65 keV)<br />
1.40<br />
1.20<br />
1.00<br />
0.80<br />
Degrees<br />
Instadose Angularity<br />
Vertical<br />
Horizontal<br />
Vertical<br />
Horizontal<br />
NS 60 (48 keV)<br />
1.40<br />
1.20<br />
1.00<br />
0.80<br />
0.60<br />
-100 -50 0 50 100<br />
Degrees<br />
Vertical<br />
Horizontal<br />
Dosimetry Services<br />
Division<br />
12
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Low Dose Testing<br />
Dosimetry Services<br />
Division<br />
13
§ Dose Rate Independent<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Instadose Dose Rate Independent<br />
Dosimetry Services<br />
Division<br />
14
§ 1.1 µSv/hr to 100 µSv/hr<br />
Response<br />
1.20<br />
1.10<br />
1.00<br />
0.90<br />
0.80<br />
0.70<br />
0.60<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Low Dose Rate Test Data<br />
Low Dose Rate Test<br />
0 20 40 60 80 100 120<br />
Dose Rate (µSv/hour)<br />
Dosimetry Services<br />
Division<br />
15
§ Two tests<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
Temperature Testing<br />
§ Test 1 - 3 Groups (reference, +40C, -20C) were irradiated to 3 mSv<br />
§ Test 2 – 2 Groups (0.35 mSv and 0.05 mSv) at +40C<br />
Dose Value (mR)<br />
10/24/11<br />
350<br />
330<br />
310<br />
290<br />
270<br />
250<br />
Instadose Temperature Test<br />
299.3 301.7 299.2<br />
Reference Dose +40 C -20 C<br />
Group<br />
Dose Value (mR)<br />
40.0<br />
35.0<br />
30.0<br />
25.0<br />
20.0<br />
15.0<br />
10.0<br />
5.0<br />
0.0<br />
Instadose Low Dose<br />
Temperature Test<br />
36.9 36.0<br />
4.6<br />
Reference Dose 35 mR 5 mR<br />
Group<br />
Dosimetry Services<br />
Division<br />
16
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
I-131 Radioiodine Thyroid Uptake Study<br />
§ August 29, 2011, radioiodine thyroid ablation procedure<br />
§ Administered 26.2 mCi (1 GBq) Sodium Iodide @9AM<br />
§ Instadose and TLD badges worn on lanyard (facing thyroid)<br />
§ Instadose on neck was read approximately every hour during daytime<br />
hours<br />
§ Instadose on neck was read prior to going to sleep, placed on window sill<br />
approximately 1 ft. from couch<br />
§ Instadose read first thing in the morning and the process described<br />
above was repeated<br />
§ After the first week the Instadose was read approximately 2 or 3 times a<br />
day<br />
§ Instadose and TLD badges strategically located throughout house to<br />
monitor doses<br />
§ TLDs were processed on September 8, 2011<br />
10/24/11<br />
Dosimetry Services<br />
Division<br />
17
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
I-131 Radioiodine Thyroid Uptake Study<br />
Dosimetry Services<br />
Division<br />
18
Account: Sandy, Inc. (363)<br />
Location: Main<br />
EID Name<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Report Run Time: 10/22/2011 8:38 AM<br />
I-131 Radioiodine Thyroid Uptake Study<br />
HISTORY DETAIL REPORT<br />
Report Period: 08/29/2011 12:00 AM - 09/05/2011 12:00 AM<br />
Serial<br />
Number<br />
1 2 3 4 5<br />
6 7 8<br />
Accredited by the National Institute of Standards and Technology through NVLAP<br />
for the specific scope of accreditation under lab code 100555-0<br />
SEE LAST PAGE FOR COMPLETE REPORT DETAILS BY COLUMN NUMBER<br />
IT IS RECOMMENDED THAT YOU KEEP THIS REPORT FOR YOUR RECORDS<br />
P.O. BOX 16451<br />
IRVINE, CA 92623<br />
U.S./CANADA: 800.359.9686<br />
www.instadose.com<br />
Body<br />
Region<br />
Ship To:<br />
Read<br />
Date/Time<br />
113 Sienna Ridge<br />
Mission Viejo CA 92692<br />
United States<br />
Dose Reported in Units of mrem<br />
Deep<br />
(Whole Body)<br />
Number of Reads: 12 Total: 821 821 821<br />
Instadose, Two 74537 Torso 9/5/2011 8:33:00 PM 1 1 1<br />
74537 Torso 9/3/2011 2:28:00 PM 12 12 12<br />
74537 Torso 8/31/2011 7:43:00 PM * * *<br />
74537 Torso 8/31/2011 7:42:00 PM 3 3 3<br />
74537 Torso 8/31/2011 10:01:00 AM 18 18 18<br />
74537 Torso 8/30/2011 10:03:00 AM 30 30 30<br />
74537 Torso 8/29/2011 9:07:00 PM 50 50 50<br />
Number of Reads: 7 Total: 114 114 114<br />
Perle, Sandy 11801 Torso 9/5/2011 8:38:00 PM 66 66 66<br />
11801 Torso 9/5/2011 8:07:00 AM * * *<br />
11801 Torso 9/5/2011 8:05:00 AM 4 4 4<br />
11801 Torso 9/4/2011 8:57:00 PM 98 98 98<br />
11801 Torso 9/4/2011 9:12:00 AM * * *<br />
11801 Torso 9/3/2011 8:52:00 PM 64 64 64<br />
11801 Torso 9/3/2011 2:00:00 PM 63 63 63<br />
11801 Torso 9/3/2011 8:01:00 AM * * *<br />
11801 Torso 9/2/2011 9:20:00 PM 110 110 110<br />
11801 Torso 9/2/2011 1:01:00 PM 52 52 52<br />
Eye<br />
Dosimetry Services<br />
Division<br />
Page 4 of 7<br />
Shallow<br />
(Skin)<br />
19
ead date time mrem/h Read<br />
8/28/2011 17:50 0 2<br />
8/29/2011 9:58 3 56<br />
8/29/2011 11:05 139 156<br />
8/29/2011 12:02 157 149<br />
8/29/2011 13:02 133 133<br />
8/29/2011 14:01 153 150<br />
8/29/2011 15:02 139 141<br />
8/29/2011 16:00 133 129<br />
8/29/2011 17:03 122 129<br />
8/29/2011 18:00 96 91<br />
8/29/2011 19:00 94 94<br />
8/29/2011 20:03 99 104<br />
8/29/2011 21:02 90 88<br />
8/29/2011 22:13 98 116<br />
8/30/2011 6:54<br />
8/30/2011 7:57 96 101<br />
8/30/2011 9:00 105 111<br />
8/30/2011 10:01<br />
8/30/2011 11:09 97 109<br />
8/30/2011 12:00 79 68<br />
8/30/2011 13:16 64 80<br />
8/30/2011 15:51 142<br />
8/30/2011 19:03 136<br />
8/30/2011 21:53 58 163<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
I-131 Radioiodine Thyroid Uptake Study<br />
read date time mrem/h Read<br />
8/31/2011 9:57<br />
8/31/2011 13:01 53 162<br />
8/31/2011 13:59 50 49<br />
8/31/2011 18:01 45 181<br />
8/31/2011 20:50 21 26<br />
8/31/2011 22:02 23 28<br />
9/1/2011 14:10 55<br />
9/1/2011 21:28 19 138<br />
9/2/2011 8:12 2<br />
9/2/2011 13:01 11 52<br />
9/2/2011 21:20 13 110<br />
9/3/2011 8:02<br />
9/3/2011 14:00 11 63<br />
9/3/2011 20:52 9 64<br />
9/4/2011 9:13<br />
9/4/2011 21:58 8 98<br />
9/5/2011 8:07<br />
9/5/2011 20:38 5 66<br />
Dosimetry Services<br />
Division<br />
20
dose rate (mrem·hr -1 )<br />
160<br />
140<br />
120<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Instadose response on neck (lanyard)<br />
Dosimetry Services<br />
Division<br />
21
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
Calculated uptake<br />
DRest = Γ·Aest·d -2 DRest = Γ·Aest·d -2 (e -‐μ x)<br />
ft ft cm (skin shielding)<br />
Date/ time elapse time Aest: (Aest = A0e -ʎeff∆t ) 1 0.25 1<br />
DRcalc Acalc = (DRcalc)(Γ·∙d -2 ) -1 Fraction<br />
8/28/2011 17:50 0.000<br />
8/29/2011 9:00 26.200<br />
8/29/2011 9:58 0.041 3 0.099 0.004<br />
8/29/2011 11:05 0.047 139 3.959 0.151<br />
8/29/2011 12:02 0.039 5.240 11.5 184.4 165.2 157 4.474 0.171<br />
8/29/2011 13:02 0.042 5.219 11.5 183.7 164.6 133 3.774 0.144<br />
8/29/2011 14:01 0.041 5.199 11.4 183.0 163.9 153 4.350 0.166<br />
8/29/2011 15:02 0.042 5.178 11.4 182.3 163.3 139 3.952 0.151<br />
8/29/2011 16:00 0.040 5.158 11.3 181.6 162.7 133 3.772 0.144<br />
8/29/2011 17:03 0.044 5.137 11.3 180.8 162.0 122 3.476 0.133<br />
8/29/2011 18:00 0.040 5.117 11.3 180.1 161.4 96 2.717 0.104<br />
8/29/2011 19:00 0.042 5.097 11.2 179.4 160.7 94 2.665 0.102<br />
8/29/2011 20:03 0.044 5.076 11.2 178.7 160.1 99 2.812 0.107<br />
8/29/2011 21:02 0.041 5.056 11.1 178.0 159.4 90 2.552 0.097<br />
8/29/2011 22:13 0.049 5.032 11.1 177.1 158.7 98 2.786 0.106<br />
8/30/2011 6:54 0.362 4.862 10.7 171.1 153.3<br />
8/30/2011 7:57 0.044 4.841 10.7 170.4 152.7 96 2.740 0.105<br />
8/30/2011 9:00 0.044 4.821 10.6 169.7 152.0 105 2.995 0.114<br />
8/30/2011 10:01 0.042 4.802 10.6 169.0 151.4<br />
8/30/2011 11:09 0.047 4.780 10.5 168.3 150.7 97 2.746 0.105<br />
8/30/2011 12:00 0.036 4.764 10.5 167.7 150.2 79 2.234 0.085<br />
8/30/2011 13:16 0.052 4.740 10.4 166.9 149.5 64 1.817 0.069<br />
8/30/2011 15:51 0.108 4.692 10.3 165.2 148.0<br />
8/30/2011 19:03 0.134 4.633 10.2 163.1 146.1<br />
8/30/2011 21:53 0.118 4.581 10.1 161.3 144.5 58 1.637 0.062<br />
8/31/2011 9:57 0.503 4.367 9.6 153.7 137.7<br />
8/31/2011 13:01 0.128 4.314 9.5 151.8 136.0 53 1.498 0.057<br />
8/31/2011 13:59 0.040 4.297 9.5 151.3 135.5 50 1.434 0.055<br />
8/31/2011 18:01 0.168 4.229 9.3 148.9 133.3 45 1.273 0.049<br />
8/31/2011 18:04 0.002 4.228 9.3 148.8 133.3<br />
8/31/2011 18:05 0.001 4.228 9.3 148.8 133.3<br />
8/31/2011 18:08 0.002 4.227 9.3 148.8 133.3<br />
8/31/2011 19:37 0.061 4.202 9.2 147.9 132.5<br />
8/31/2011 20:50 0.051 4.182 9.2 147.2 131.9 21 0.602 0.023<br />
8/31/2011 22:02 0.050 4.162 9.2 146.5 131.2 23 0.666 0.025<br />
9/1/2011 14:10 0.672 3.904 8.6 137.4 123.1<br />
9/1/2011 21:28 0.304 3.792 8.3 133.5 119.6 19 0.538 0.021<br />
9/2/2011 8:12 0.447 3.634 8.0 127.9 114.6<br />
9/2/2011 13:01 0.201 3.565 7.8 125.5 112.4 11 0.307 0.012<br />
9/2/2011 21:20 0.347 3.449 7.6 121.4 108.8 13 0.376 0.014<br />
9/3/2011 8:02 0.446 3.306 7.3 116.4 104.2<br />
9/3/2011 14:00 0.249 3.228 7.1 113.6 101.8 11 0.300 0.011<br />
9/3/2011 20:52 0.286 3.142 6.9 110.6 99.1 9 0.265 0.010<br />
9/4/2011 9:13 0.515 2.991 6.6 105.3 94.3<br />
9/4/2011 21:58 0.531 2.844 6.3 100.1 89.7 8 0.218 0.008<br />
9/5/2011 8:07 0.423 2.731 6.0 96.1 86.1<br />
9/5/2011 20:38 0.522 2.599 5.7 91.5 81.9 5 0.150 0.006<br />
10/24/11<br />
Dosimetry Services<br />
Division<br />
22
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Loca&on<br />
Instadose vs. TLD results<br />
Instadose Reading<br />
(mSv) TLD Reading (mSv)<br />
Collar 35.15 30.25<br />
Computer Room #1 0.51 0.71<br />
Computer Room #2 0.47 0.6<br />
Window Sill #1 1.11 0.96<br />
Window Sill #2 1.11 0.83<br />
Spouse 0.05 0<br />
Dosimetry Services<br />
Division<br />
23
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Future?<br />
Dosimetry Services<br />
Division<br />
24
<strong>Mirion</strong> offers an array of solutions and services for managing radiological hazards.<br />
Learn more at: www.mirion.com<br />
10/24/11<br />
Questions?<br />