SAMPLING OF THE RINSING WATER 12 : 22 : 15 ... - Soluscope
SAMPLING OF THE RINSING WATER 12 : 22 : 15 ... - Soluscope
SAMPLING OF THE RINSING WATER 12 : 22 : 15 ... - Soluscope
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>SAMPLING</strong> <strong>OF</strong> <strong>THE</strong> <strong>RINSING</strong> <strong>WATER</strong><br />
PURPOSE <strong>OF</strong> <strong>THE</strong> CONTROL<br />
D04139 - Instructions to sample rinsing water<br />
This procedure is intended to assure the quality of the water during the final rinsing of your endoscopes<br />
at the end of your <strong>Soluscope</strong> ENT treatment. It also allows to check the presence of a possible<br />
contamination of the machine internal circuits.<br />
SYN<strong>THE</strong>SIS <strong>OF</strong> <strong>THE</strong> PROCEDURE<br />
1<br />
Lid seal is cleaned with a disinfecting surface spray.<br />
2 Self-disinfection cycle 6 is launched.<br />
3 Access to bowl filling menu.<br />
4 Water is sampled from the bottom of the bowl in the sterile container.<br />
5 Bowl is drained and <strong>Soluscope</strong> ENT is reset by user.<br />
EQUIPMENT<br />
• A container with a capacity of about 200 cc sterile with shutting.<br />
• A pair of sterile gloves.<br />
• One syringe.<br />
• One SL-OTE<br />
STAGE 1<br />
1 Open the lid.<br />
2 Connect SL-OTE and launch a cycle 10 (inflation of the lid seal).<br />
3 Press MAIN once the lid is inflated.<br />
4 Use a disinfecting surface spray to disinfect the lid seal of your<br />
<strong>Soluscope</strong> ENT.<br />
5 Deflate the lid seal by opening the Luer-Lock on the side of your<br />
<strong>Soluscope</strong> ENT beneath the 0.2 μm terminal filter.<br />
6 Close the Luer-Lock.<br />
7 Close the lid.<br />
STAGE 2<br />
Launch a cycle 6.<br />
STAGE 3<br />
At the end of a self-disinfection cycle, start at the initial menu :<br />
1 2 : 2 2 : 1 5 P r e s s F 3<br />
1 5 2 6 t o s t a r t<br />
Press F2<br />
This document can be downloaded from our website www.soluscope.com (USERS section) /2<br />
2<br />
5
2/2<br />
P a s s w o r d :<br />
Enter 200 then press<br />
1 2 : 2 2 : 1 5 P r e s s F 3<br />
1 5 2 6 t o s t a r t<br />
Press F2<br />
S e r v i c C o n f i g S a v e<br />
& S a m p l F 2 F 3<br />
Press F<br />
Press F2<br />
Press F<br />
Press F until the two stars disappear<br />
Press F3 during 0 seconds to cycle <strong>Soluscope</strong> ENT.<br />
STAGE 4 - <strong>SAMPLING</strong><br />
D04139 - Instructions to sample rinsing water<br />
1 Deflate the lid seal by opening the Luer-Lock on the side of your <strong>Soluscope</strong> ENT under the 0.2 μm<br />
terminal filter (see photo 5 page 1).<br />
2 Open the lid.<br />
3 Wear sterile gloves.<br />
S e r v i c i n g :<br />
1 = F 2 2 = F 3<br />
B o w l D o s / L T D r a w<br />
F 1 F 2 F 3<br />
F i l l D r a i n C y c l e<br />
✹ ✹ ✹ ✹ P 2 5 . 8 C<br />
F i l l D r a i n C y c l e<br />
✹ ✹ P 2 5 . 8 C<br />
4 Using a syringe, collect 200 cc of water from the bottom of the bowl and fill the sterile sampling<br />
container.<br />
STAGE 5 - RESETTING SOLUSCOPE ENT<br />
Once the sampling is finished, do not forget to drain the bowl (press F2 Drain).<br />
Press MAIN to come back to the main menu.<br />
This document can be downloaded from our website www.soluscope.com (USERS section)